The Mazor X is indicated for precise positioning of surgical instruments or spinal implants during general spinal and brain surgery. It may be used in either open or minimally invasive or percutaneous procedures. Mazor X 3D imaging capabilities provide a processing and conversion of 2D fluoroscopic projections from standard C-Arms into volumetric 3D image. It is intended to be used whenever the clinician and/or patient benefits from generated 3D imaging of high contrast objects. The Mazor X navigation tracks the position of instruments, during spinal surgery, in relation to the surgical anatomy and identifies this position on diagnostic or intraoperative images of a patient.
Device Story
Mazor X System (Stealth Edition) integrates navigation with robotic guidance; tracks compatible surgical instruments relative to patient anatomy during spinal surgery. Inputs include 2D fluoroscopic projections from standard C-Arms; system processes/converts these into volumetric 3D images. Navigation feature continuously updates instrument position on patient images. Used in OR by surgeons; provides real-time visual feedback of instrument trajectory and position relative to anatomy. Enhances precision for spinal implant placement; supports open, minimally invasive, and percutaneous procedures. System operates with or without navigation feature; maintains previously cleared robotic guidance trajectory accuracy (< 1.5mm).
Clinical Evidence
Bench testing only. Validation included navigation accuracy testing under worst-case scenarios, surgical system and arm testing, CT-Fluro and Scan & Plan workflow accuracy tests, and cadaver validation. Software validated per IEC 62304. Electrical safety and EMC testing performed per AAMI/ANSI ES 60601-1 and IEC 60601-1-2.
Technological Characteristics
Robotic stereotaxic instrument with integrated navigation. Connectivity: C-Arm interface for 2D-to-3D image processing. Software: IEC 62304 compliant. Electrical: AAMI/ANSI ES 60601-1, IEC 60601-1-2. Accuracy: < 1.5mm trajectory error. Operates in CT-Fluro and Scan & Plan workflows.
Indications for Use
Indicated for precise positioning of surgical instruments or spinal implants during general spinal and brain surgery in open, minimally invasive, or percutaneous procedures.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
StealthStation System with the S8 Spine Software (K170011)
StealthStation System with the S8 Spine Software (K162309)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Mazor Robotics Ltd. % Ahava Stein Regulatory Affairs Consultant A. Stein - Regulatory Affairs Consulting Ltd. 20 Hata'as St. Kfar Saba, 44425 Il
November 2, 2018
Re: K182077
Trade/Device Name: Mazor X System (Mazor X Stealth Edition) Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO, HAW, LLZ Dated: July 23, 2018 Received: August 1, 2018
Dear Ahava Stein:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/7 description: The image shows a digital signature. The name "Jesse Muir -S" is displayed twice, once as the name and once as the signer. The date of the signature is November 2nd, 2018, and the time is 11:25:49 -04'00'.
by Jesse Muir -S
Date: 2018.11.02
11:25:49 -04'00'
For: Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182077
Device Name Mazor X System (Mazor X Stealth Edition)
# Indications for Use (Describe)
Indications for Use The Mazor X is indicated for precise positioning of spinal implants during general spinal and brain surgery. It may be used in open or minimally invasive or percutaneous procedures.
Mazor X 3D imaging capabilities provide a processing and conversion of 2D fluoroscopic projections from standard C-Arms into volumetric 3D image. It is intended to be used whenever the clinician and/or patient benefits from generated 3D imaging of high contrast objects.
The Mazor X navigation tracks the position of instruments, during spinal surgery, in relation to the surgical anatomy and identifies this position on diagnostic or intraoperative images of a patient.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# SUMMARY OF SAFETY AND EFFECTIVENESS
# K182077
(Premarket Notification [510(k)] Number)
# Submitter Information 1.
| Manufacturer Name and Address |
|-------------------------------|
| Mazor Robotics Ltd. |
| PO Box 3104, |
| 5 Shacham St., |
| Caesarea Park North 3088900, |
| Israel |
| Official Correspondent |
|-----------------------------------------------|
| Ahava Stein |
| A. Stein - Regulatory Affairs Consulting Ltd. |
| 20 Hata'as St. (Beit Hapaamon, Suite 102) |
| Kfar Saba 4442520, |
| Israel |
#### Date Prepared: July 24, 2018 2.
## 3. Device Name Mazor X
| Proprietary Name: | Mazor X System (Mazor X Stealth Edition) |
|--------------------------|------------------------------------------------------------------------------------------------|
| Common Name: | Combination of:<br>1. Stereotaxic instrument; and<br>2. System, Image Processing, Radiological |
| FDA Classification Name: | 21 CFR 882.4560; Stereotaxic instrument |
# FDA Classification: Class II, Product Code OLO, HAW and LLZ
# 4. Predicate Devices
The Mazor X is substantially equivalent to the following device:
| Manufacturer | Device | 510(k) | Date Cleared |
|---------------------|----------------------------|----------------|----------------|
| Mazor Robotics Ltd. | Mazor X System | K180307 | April 30, 2018 |
| Medtronic | StealthStation System with | K170011 | May 1, 2017 |
| Navigation Inc. | the S8 Spine Software | and<br>K162309 | March 31, 2017 |
{4}------------------------------------------------
## 5. Device Description
The Mazor X System integrates a new Navigation feature, which enables tracking compatible spine instruments. The previously cleared Mazor X System enables mechanical positioning of a tool or instrument and determining its orientation and trajectory. The new Navigation feature tracks the position of compatible surgical instruments in or on the patient anatomy during surgery and continuously updates the instrument position on the image of the patient's anatomy. The modified Mazor X System can operate with or without the Navigation feature.
## Indications for Use 6.
The Mazor X is indicated for precise positioning of surgical instruments or spinal implants during general spinal and brain surgery. It may be used in either open or minimally invasive or percutaneous procedures.
Mazor X 3D imaging capabilities provide a processing and conversion of 2D fluoroscopic projections from standard C-Arms into volumetric 3D image. It is intended to be used whenever the clinician and/or patient benefits from generated 3D imaging of high contrast objects.
The Mazor X navigation tracks the position of instruments, during spinal surgery, in relation to the surgical anatomy and identifies this position on diagnostic or intraoperative images of a patient.
### 7. Performance Standards
There are no performance standards under the Federal Food, Drug and Cosmetic Act, for the Mazor X.
# Performance Testing 8.
The following performance tests were conducted on the modified Mazor X System with the Navigation feature.
{5}------------------------------------------------
- Mazor X System Accuracy Workflow Validation Test (VV-07444)
The key performance testing to validate the modified Mazor X System navigation feature consisted of navigation accuracy testing. The navigation accuracy validation included testing the overall navigation accuracy under worst-case scenario navigation tool positioning. The overall accuracy under worst-case scenario navigation tool positioning calculated the position and trajectory errors.
Minor modifications were made to the released Mazor X System and accessories and therefore, the following validation tests were conducted:
- Mazor X System Surgical System Test (MCP0721-01)
- · Mazor X System Surgical Arm Test (MCP0695-01)
- · Mazor X System Accuracy (CT-Fluro, Scan & Plan) Tests) (CT-Fluro -MCP0881-01 & Scan & Plan - MCP0882-01)
- · Mazor X System Accessories Tests Summary (MCP0885-01)
- · Cadaver validation activities (GC17021, TP-GC17021)
The tests performed to validate the modifications to the previously cleared Mazor X System included testing the modified Surgical System and the modified Surgical Arm. The modified Mazor X System was also tested to ensure that the software changes did not affect the previously cleared robotic guidance trajectory accuracy (i.e.. < 1.5mm), in the CT-Fluro and Scan & Plan workflows, following the software modifications. Testing was also performed for the new and modified Mazor X System accessories.
The Mazor X System software with the navigation feature was underwent software validation according to FDA Guidelines for Software Validation and the following standard:
- IEC 62304 Medical Device Software: Software Life Cycle Processes (2015) .
The modified Mazor X System underwent the following testing according to international and FDA recognized standards:
- . AAMI/ANSI ES 60601-1:2005/(R):2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medical Electrical Equipment - Part 1:
{6}------------------------------------------------
General Requirements for Basix Safety and Essential Performance (IEC 60601-1:2005, Mod) 2012
- . IEC 60601-1-2:2014 Medical Electrical Equipment Part 1-2: General Requirements for Basic Safety and Essential Performance Collateral Standard: Electromagnetic Compatibility
### 9. Technological Characteristics Compared to Predicate Device
The intended use and technological characteristics of the modified Mazor X System are substantially equivalent to the intended use and technological characteristics of the previously cleared Mazor X System (K180307) and the FDA cleared, predicate StealthStation Surgical System with S8 Spine Software (K170011 & K162309). The performance of the modified Mazor X System is substantially equivalent to the performance of the FDA cleared, Mazor X System, as demonstrated by the software validation and functional performance testing. The functional performance testing also demonstrated that the navigation capabilities of the Mazor X System are substantially equivalent to the navigation capabilities of the predicate StealthStation Surgical System with S8 Spine Software (K170011 & K162309).
### 10. Conclusion
The performance testing and comparison to the predicate devices demonstrate that the Mazor X system is as safe, as effective and performs as well as the legally marketed Mazor X System (K180307) and the StealthStation Surgical System (K170011 & K162309) predicate devices. Therefore, the Mazor X system is substantially equivalent to the Mazor X System and the StealthStation Surgical System with S8 Spine Software.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.