X12 STERILE REFLECTIVE MARKER SPHERE

K111443 · Pinsco Inc. Dba B & L Engineering · OLO · Sep 29, 2011 · Neurology

Device Facts

Record IDK111443
Device NameX12 STERILE REFLECTIVE MARKER SPHERE
ApplicantPinsco Inc. Dba B & L Engineering
Product CodeOLO · Neurology
Decision DateSep 29, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.4560
Device ClassClass 2

Intended Use

Passive Retro-Reflective Markers, disposable, used as an accessory to aid in the autoregistration and localization of rigid patient anatomical structures in either open or percutaneous image guided surgical procedures.

Device Story

Disposable passive retro-reflective marker spheres; used as accessories in image-guided surgery. Device consists of two hemispheres coated with retro-reflective foil, bonded on a central pin (M3 thread or snap hole). Used in OR settings by surgeons/clinical staff to aid auto-registration and localization of rigid anatomical structures. Reflectivity allows optical tracking systems to identify marker position in 3D space. Benefits include accurate anatomical localization during surgical procedures. Gamma sterilized; single-use.

Clinical Evidence

Bench testing only. Testing verified retro-reflectivity in polar and equatorial views, mechanical integrity via water soak, and surface durability via abrasion testing, all compared against predicate devices.

Technological Characteristics

Two hemispheres coated with retro-reflective foil, bonded on a central pin. Features M3 female thread or snap hole for mounting. Gamma sterilized. Passive optical tracking technology.

Indications for Use

Indicated for use as an accessory to aid in the auto-registration and localization of rigid patient anatomical structures during open or percutaneous image-guided surgical procedures.

Regulatory Classification

Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K111443 ## PREMARKET NOTIFICATION 510 (K) SUMMARY SEP 2 9 2011 | 1. | Submitters Name: | Dancho Jordanov | |----|-----------------------|-----------------------------------------------------------------------------------------| | 2. | Address: | X12 Co., Ltd.<br>J&L Center, fl.4, rm.403<br>46 Lubliana str.<br>1618 Sofia<br>Bulgaria | | 3. | Telephone No:<br>Fax: | +359 878 350 021 | | 4. | Date Prepared: | April 28, 2011 | | 5. | Trade Name: | X12 Sterile Reflective Marker Sphere | | 6. | Common Name: | Disposable Reflective Marker Sphere | | 7. | Classification Name: | Neurological Stereotaxic Instrument<br>Orthopedic Stereotaxic Instrument | X12's Sterile Reflective Marker Spheres are substantially 8. Predicate Devices: equivalent to those legally marketed by BrainLAB AG, Norhern Digital Inc. and ILUMARK GmbH (see table, below). This is based upon their intended use, design and materials of construction. | Name | Common Name | Product Code | 510(k) Number | Date | |------------------|-----------------------------------------|--------------|---------------|------------------| | BrainLAB AG | Disposable Reflective<br>Marker Spheres | HAW, OLO | K100038 | July 14, 2010 | | Northern Digital | NDI Passive Spheres | HAW | K033621 | January 27, 2004 | | ILUMARK GmbH | Navigation Marker | HAW, OLO | K103192 | February 7, 2011 | ### Device Description: 9. D X12's Sterile Reflective Marker Spheres consist of two hemispheres, coated with a retroreflective foil, bonded together on a central pin to form a sphere. The central pin can be threaded with an M3 female thread or have a snap hole to mate with a mounting pin. The retro-reflective foil is applied in such a way as to provide reflectivity from all angles that is equivalent to the {1}------------------------------------------------ predicate devices. The spheres are individually packaged in a special sterile package and are gamma sterilized. #### 10. Indications For Use: Passive Retro-Reflective Markers, disposable, used as an accessory to aid in the autoregistration and localization of rigid patient anatomical structures in either open or percutaneous image guided surgical procedures. #### l ] . Performance Testing: X12's Sterile Reflective Marker Spheres were tested to verify their retro-reflectivity in comparison to the predicate devices. Both the polar and equatorial views showed retroreflectivity that was equivalent to the predicate devices. A water soak test was conducted to show that the markers maintain there mechanical properties after being soaked in water. A surface abrasion test was performed which showed that the surface durability was substantially equivalent to the predicate devices. #### 12. Sterilization: X12's Sterile Reflective Marker Spheres are inserted (6 each), in special "made for sterilization" plastic shells which are sealed in a nitrogen environment in plastic sterilization packets. Each packet is inserted in a box and is gamma sterilized in a carton containing 50 boxes. Each box is printed with pertinent labeling which includes the date of manufacture, expiration date, manufacturer batch and sterilization lot numbers and the manufacturer's name and address, as well as instructions for use. #### 13. Conclusion: Based on the available 510 (K) summaries and statements and the information provided here, we conclude that X12's Sterile Reflective Marker Spheres are substantially equivalent to the predicate devices which are legally marketed in the United States under the FDA regulations. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows a logo with a circular border containing text, surrounding a stylized bird-like symbol. The text within the border is small and difficult to read, but it appears to follow the curve of the circle. The bird-like symbol consists of three curved lines that resemble the silhouette of a bird in flight. The logo is black and white and has a simple, clean design. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -W066-G609 Silver Spring, MIJ 20993-0002 Pinsco Inc. dba B & L Engineering % Mr. Lee A. Barnes 1901 Carnegie Avenue, Suite Q Santa Ana, California 92705 SEP 29 201 Re: K111443 · Trade/Device Name: X12 Sterile Reflective Marker Sphere Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Requlatory Class: Class II Product Code: OLO, HAW Dated: September 07, 2011 Received: September 12, 2011 Dear Mr. Barnes: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Insting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {3}------------------------------------------------ Page 2 - Mr. Lee A. Barnes comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. If vou desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.lda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/MedicalDevices/Safetv/ReportalProblem/default.htm for the CDRH3's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number: K111443 ## Device Name: X12 STERILE REFLECTIVE MARKER SPHERE Indications for Use: Used as an accessory to aid in the auto-registration and localization of rigid patient anatomical structures in either open or percutaneous image guided surgical procedures. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) | Concurrence of CDRH, Office of Device Evaluation (ODE) | | |--------------------------------------------------------|--| |--------------------------------------------------------|--| (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices . 510(k) Number K111443 Page 1 of 1
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%