SpineAR SNAP
K213034 · Surgical Theater, Inc. · SBF · Sep 29, 2022 · Neurology
Device Facts
| Record ID | K213034 |
| Device Name | SpineAR SNAP |
| Applicant | Surgical Theater, Inc. |
| Product Code | SBF · Neurology |
| Decision Date | Sep 29, 2022 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 882.4560 |
| Device Class | Class 2 |
| Attributes | Software as a Medical Device |
Intended Use
SpineAR SNAP is intended for use or pre-operative surgical planning on-screen and in a virtual environment, and intraoperative surgical planning and visualization on-screen and in an augment using the HoloLens2 and Magic Leap 1 AR headset displays with validated navigation systems as identified in the device labeling. SpineAR SNAP is indicated for spinal stereotaxic surgery, and where reference to a rigid anatomical structure, such as the spine, can be identified relative to images of the anatomy. SpineAR is intended for use in spinal implant procedures, such as Pedicle Screw Placement, in the lumbar and thoracic regions with the Magic Leap 1 AR headset, and in the lumbar region with the HoloLens2 AR headset. The virtual display should not be relied upon solely for absolute positional information and should always be used in conjunction with the displayed 2D stereotaxic information.
Device Story
Software-based system for pre-operative planning and intra-operative visualization in spinal surgery. Inputs: 2D medical images (CT, MR, XA) and real-time tool position/orientation data from 3rd-party navigation systems (e.g., Medtronic StealthStation S8). Transforms 2D images into interactive 3D scenes; overlays guidance objects (directional markers, virtual screws) and navigation data. Outputs: 3D stereoscopic visualization on HD touch screen, VR headset (pre-op), or AR wireless headset (intra-op). Used in OR by surgeons. Provides visual guidance (dotted line, orientation line, ILS) with color-coded cues to assist instrument alignment. Does not perform tracking itself; relies on connected navigation system. Benefits: enhanced spatial awareness and guidance during spinal implant procedures.
Clinical Evidence
Bench testing only. Navigation accuracy testing compared post-surgical CT scans to pre-surgical plans, demonstrating mean positional error of 1.02 mm and mean trajectory error of 1.79°, within the predicate's <2.0 mm and <2° limits. Human factors and usability validation confirmed safe task performance in simulated OR environments. Latency testing confirmed projection delays <250 ms (avg 195 ms). EMC and wireless coexistence testing performed per IEC 60601-1-2, AAMI TIR69, and ANSI IEEE C63.27.
Technological Characteristics
Software-based system running on COTS high-performance desktop PC. Interfaces with 3rd-party navigation systems for tracking. Displays: HD touch screen, Microsoft HoloLens2, Magic Leap 1. Guidance: Dotted line, orientation line, ILS. Connectivity: Networked/wireless. Standards: IEC 60601-1-2:2014+A1:2020 (EMC), AAMI TIR69:2017/(R)2020, ANSI IEEE C63.27-2017 (wireless).
Indications for Use
Indicated for spinal stereotaxic surgery in patients requiring spinal implant procedures (e.g., Pedicle Screw Placement). Lumbar and thoracic regions supported with Magic Leap 1; lumbar region supported with HoloLens2.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
Predicate Devices
- StealthStation™ S8 Spine Software v1.3.0 (K201189)
Reference Devices
- Surgical Navigation Advanced Platform (SNAP) (K160584)
Related Devices
- K243623 — SpineAR SNAP (SyncAR Spine) · Surgical Theater, Inc. · Dec 24, 2024
- K252054 — SpineAR SNAP (SyncAR Spine) · Surgical Theater, Inc. · Sep 29, 2025
- K212245 — Spine and Trauma Navigation System, Spine & Trauma 3D Navigation, Instrument Selection, Fluoro 3D, Registration Software Fluoro 3D, Registration Software Paired Point, Registration Software Spine Surface Matching, Spine Planning, Elements Screw Planning Spine, Elements Spine Screw Planning · Brainlab AG · Apr 13, 2022
- K233513 — Remex Spine Surgery Navigation System II · Remex Medical Corp. · May 1, 2024
- K250879 — Paradigm System · Proprio, Inc. · Jun 4, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".
September 29, 2022
Surgical Theater, Inc. Kevin Murrock Sr. Director of Quality and Regulatory 30559 Pinetree Road. Suite 206 Pepper Pike. Ohio 44124
Re: K213034
Trade/Device Name: SpineAR SNAP Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO, LLZ Dated: August 26, 2022 Received: August 30, 2022
Dear Kevin Murrock:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumava Ali. M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213034
Device Name SpineAR SNAP
#### Indications for Use (Describe)
SpineAR SNAP is intended for use or pre-operative surgical planning on-screen and in a virtual environment, and intraoperative surgical planning and visualization on-screen and in an augment using the HoloLens2 and Magic Leap 1 AR headset displays with validated navigation systems as identified in the device labeling.
SpineAR SNAP is indicated for spinal stereotaxic surgery, and where reference to a rigid anatomical structure, such as the spine, can be identified relative to images of the anatomy. SpineAR is intended for use in spinal implant procedures, such as Pedicle Screw Placement, in the lumbar and thoracic regions with the Magic Leap 1 AR headset, and in the lumbar region with the HoloLens2 AR headset.
The virtual display should not be relied upon solely for absolute positional information and should always be used in conjunction with the displayed 2D stereotaxic information.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Surgical Theater. The logo consists of a stylized letter "S" in red and orange gradient, followed by the words "SURGICAL" in red and "THEATER" in orange. The word "THEATER" has a registered trademark symbol next to it.
SpineAR SNAP
## 510(k) SUMMARY
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR 8807.92.
Date Prepared: 27-Sep-2022
## Manufacturer/Submitter:
Surgical Theater, Inc. 30559 Pinetree Road, Suite 206 Pepper Pike, Ohio 44124 Phone: (216) 496-7884 Fax: (216) 916-3806
Establishment Registration Number: 3010197287
## Contact Person:
Kevin M. Murrock Sr. Director of Quality and Regulatory Surgical Theater, Inc. 30559 Pinetree Road, Suite 206 Pepper Pike, Ohio 44124 Phone: (330) 472-6520 Email: kmurrock@surgicaltheater.com
## Name of Device
- Trade Name: SpineAR SNAP ● Common Name: Augmented Reality System Classification Name: Orthopedic Stereotaxic Instrument ●
- Regulation Number: 21 CFR 882.4560 ●
- Product Code: OLO
- Regulatory Classification: ● II
- Device Panel: Orthopedic
## Predicate Device
Device Name: StealthStation™ S8 Spine Software v1.3.0 Manufacturer: Medtronic Navigation, Inc. 510(k) Number: K201189
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Surgical Theater, with the words "Surgical Theater" in red and gold. Below the logo is the text "510(k) Premarket Notification K213034" in bold font. The text is underlined with a double line.
SpineAR SNAP
# Reference Device
Device Name: Surgical Navigation Advanced Platform (SNAP) Manufacturer: Surgical Theater, Inc. 510(k) Number: K160584
## Indications for Use:
SpineAR SNAP is intended for use for pre-operative surgical planning on-screen and in a virtual environment, and intra-operative surgical planning and visualization on-screen and in an augmented environment using the HoloLens2 and Magic Leap 1 AR headset displays with validated navigation systems as identified in the device labeling.
SpineAR SNAP is indicated for spinal stereotaxic surgery, and where reference to a rigid anatomical structure, such as the spine, can be identified relative to images of the anatomy. SpineAR is intended for use in spinal implant procedures, such as Pedicle Screw Placement, in the lumbar and thoracic regions with the Magic Leap 1 AR headset, and in the lumbar region with the HoloLens2 AR headset.
The virtual display should not be relied upon solely for absolute positional information and should always be used in conjunction with the displayed 2D stereotaxic information.
## Device Description:
The SpineAR SNAP does not require any custom hardware and is a software-based device that runs on a high-performance desktop PC assembled using "commercial off-the-shelf" components that meet minimum performance requirements.
The SpineAR SNAP software transforms 2D medical images into a dynamic interactive 3D scene with multiple point of views for viewing on a high-definition (HD) touch screen monitor. The surgeon prepares a pre-operative plan for stereotaxic spine surgery by inserting guidance objects such as directional markers and virtual screws into the 3D scene. Surgical planning tools and functions are available on-screen and when using a virtual reality (VR) headset. The use of a VR headset for preoperative surgical planning further increases the surgeon's immersion level in the 3D scene by providing a 3D stereoscopic display of the same 3D scene displayed on the touch screen monitor.
By interfacing to a 3rd party navigation system such as a Medtronic StealthStation S8, the SpineAR SNAP extracts the navigation data (i.e. tool position and orientation) and presents the navigation data into the advanced interactive, high quality 3D image, with multiple point of views on a high-definition (HD) touch screen monitor. Once connected, the surgeon can then execute the plan through the intraoperative use of the SpineAR SNAP's enhanced visualization and guidance tools.
The SpineAR SNAP supports three (3) guidance options from which the surgeon selects the level of guidance that will be shown in the 3D scene. The guidance options are dotted line (indicates deviation distance), orientation line (indicates both distance and angular deviation), and ILS (indicates both distance and angular deviation using crosshairs). Visual color-coded cues indicate alignment of the tracker tip to the guidance object (e.g. green = aligned).
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Surgical Theater. The logo consists of a stylized letter "S" in red and orange gradient, followed by the words "SURGICAL" in red and "THEATER" in yellow. A registered trademark symbol is next to the word "THEATER". There are two black lines underneath the text.
The 3D scene with guidance tools can also be streamed into an AR wireless headset (Magic Leap 1 or HoloLens2) worn by the surgeon during surgery. The 3D scene and guidance shown within the AR headset is projected above the patient and does not obstruct the surgeons view of the surgical space.
## Summary of Technological Characteristics
The SpineAR SNAP is similar in its technological features to its predicate device, the StealthStation S8 Spine Software v1.3.0 (K201189). Both systems are indicated for spinal stereotaxic surgery, and where reference to a rigid anatomical structure, such as the spine, can be identified relative to images of the anatomy.
The SpineAR SNAP does not include any of the physical stereotaxic instrumentation, including camera system, and relies on the positional information supplied by the connected navigation system (e.g. Medtronic StealthStation S8 system with StealthStation Spine Software v1.3.0) to track the position of the surgical instruments relative to the surgical anatomy.
The SpineAR SNAP includes an AR headset that displays the 2D images and 3D scene including navigation and enhanced guidance information in an AR wireless headset (e.g. Microsoft HoloLens2) worn by the surgeon. The predicate device displays 2D and 3D images including navigation information with minimal guidance on a system monitor, and does not support use of AR headset.
| Feature | Subject Device: | Predicate Device: | Reference Device: |
|-------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | SpineAR SNAP | StealthStation S8 Spine<br>Software v1.3.0 | Surgical Navigation<br>Advanced Platform (SNAP) |
| Indications for Use | SpineAR SNAP is intended<br>for use for pre-operative<br>surgical planning on-screen<br>and in a virtual environment,<br>and intra-operative surgical<br>planning and visualization<br>on-screen and in an<br>augmented environment<br>using the HoloLens2 and<br>Magic Leap 1 AR headset<br>displays with validated<br>navigation systems as<br>identified in the device<br>labeling.<br>SpineAR SNAP is indicated<br>for spinal stereotaxic<br>surgery, and where reference<br>to a rigid anatomical<br>structure, such as the spine,<br>can be identified relative to<br>images of the anatomy.<br>SpineAR is intended for use<br>in spinal implant procedures,<br>such as Pedicle Screw | The StealthStation™ System,<br>with StealthStation™ Spine<br>Software, is intended as an aid<br>for precisely locating<br>anatomical structures in either<br>open or percutaneous<br>neurosurgical and orthopedic<br>procedures. Their use is<br>indicated for any medical<br>condition in which the use of<br>stereotaxic surgery may be<br>appropriate, and where<br>reference to a rigid anatomical<br>structure, such as the spine or<br>pelvis, can be identified<br>relative to images of the<br>anatomy.<br>This can include the following<br>spinal implant procedures,<br>such as:<br>o Pedicle Screw Placement<br>o Iliosacral Screw Placement<br>o Interbody Device Placement | The SNAP is intended for use<br>as a software interface and<br>image segmentation system<br>for the transfer of imaging<br>information from a medical<br>scanner such as a CT or MR<br>scanner to an output file.<br>It is also intended for use in<br>simulating and evaluating<br>surgical treatment options both<br>pre-operatively and intra-<br>operatively with validated<br>systems as identified in the<br>device labeling. |
| Feature | Subject Device:<br>SpineAR SNAP | Predicate Device:<br>StealthStation S8 Spine<br>Software v1.3.0 | Reference Device:<br>Surgical Navigation<br>Advanced Platform (SNAP) |
| | Placement, in the lumbar and<br>thoracic regions with the<br>Magic Leap 1 AR headset,<br>and in the lumbar region<br>with the HoloLens2 AR<br>headset.<br><br>The virtual display should<br>not be relied upon solely for<br>absolute positional<br>information and should<br>always be used in<br>conjunction with the<br>displayed 2D stereotaxic<br>information. | | |
| Main system<br>components | Cart with computer,<br>touchscreen monitor,<br>UPS and input devices. Software application AR Headset VR Headset | Platform including cart,<br>computer, monitor and<br>tracking cameras Software application Reflective markers -<br>Spheres Accessories (Instrument<br>adaptors, referencing<br>system) | Cart with computer,<br>touchscreen monitor and<br>input devices. Software application VR Headset |
| System Accuracy<br>Requirements | Provided by the connected<br>3rd party navigation system<br>as the subject device does not<br>include a camera system or<br>tracking arrays.<br><br>Maintains the system<br>accuracy of the connected<br>navigation system.<br><br>3D positional accuracy with<br>a mean positional error of<br>$\le$ 2.0 mm and mean<br>trajectory error of $\le$ 2<br>degrees. | Under representative worst-<br>case Configuration, the<br>StealthStation S8 Spine<br>software v1.3.0, has<br>demonstrated performance in<br>3D positional accuracy with a<br>mean positional error of $\le$ 2.0<br>mm and mean trajectory error<br>of $\le$ 2 degrees.<br><br>Mean Accuracy Values<br>(StealthAiR Spine):<br>Positional Error - 1.01 mm<br>Trajectory Error - 0.37<br>degrees<br><br>Mean Accuracy Values<br>(Overlapping Slices):<br>Positional Error - 0.51 mm<br>Trajectory Error -0.41<br>degrees | Provided by the connected 3rd<br>party navigation system as the<br>subject device does not<br>include a camera system or<br>tracking arrays. |
| Imaging Modalities | CT, MR and XA | X-Ray Based Imaging | CT, MR and XA |
| Registration Features | All registration is performed<br>by the connected 3rd party<br>navigation system. | PointMerge Registration SurfaceMerge<br>Registration | All registration is performed<br>by the connected 3rd party<br>navigation system. |
| Feature | Subject Device:<br>SpineAR SNAP | Predicate Device:<br>StealthStation S8 Spine<br>Software v1.3.0 | Reference Device:<br>Surgical Navigation<br>Advanced Platform (SNAP) |
| Planning Features | Plan Entry and Target<br>Selection 3D Model Building Directional Markers | Plan Entry and Target<br>Selection 3D Model Building Deformity Planning | Plan Entry and Target<br>Selection 3D Model Building |
| Medical Device<br>Interfaces | The system does not interface<br>directly with the imaging<br>modality device. | O-arm Imaging System Ziehm Vision FD Vario<br>3D C-Arm ISO-C 3D C-Arm Ziehm Vision RFD 3D C-<br>arm Stealth-Midas MR8 Orbic 3D C-Arm | The system does not interface<br>directly with the imaging<br>modality device. |
| Localization<br>Technology | The subject device does not<br>include a camera system or<br>tracking arrays, but instead<br>relies on the stereotaxic<br>guidance provided from a<br>connected 3rd party<br>navigation system to track<br>patient anatomy and surgical<br>tools. | Optical (infra-red) | The subject device does not<br>include a camera system or<br>tracking arrays, but instead<br>relies on the stereotaxic<br>guidance provided from a<br>connected 3rd party navigation<br>system to track patient<br>anatomy and surgical tools. |
| Connection to 3rd<br>party surgical<br>navigation system to<br>track patient anatomy<br>and surgical tools. | Yes. Medtronic<br>StealthStation S8 Spine<br>Software v1.3.0 | No | Medtronic StealthStation S8<br>Cranial Version 1.3.0-49 and<br>Brainlab Cranial Version 2.1 |
| Capability of creating<br>3D models of patient<br>data from 2D scan<br>slices. | Yes | Yes | Yes |
| Virtual Reality (VR)<br>Headset Display for<br>Pre-Operative Use | Yes | No | Yes |
| Augmented Reality<br>(AR) Headset<br>Display for Intra-<br>operative Use | Yes | No | No |
| Intra-operative<br>visualization and<br>guidance tools shown<br>on a wireless AR<br>headset display. | Yes | No | No |
| Feature | Subject Device:<br>SpineAR SNAP | Predicate Device:<br>StealthStation S8 Spine<br>Software v1.3.0 | Reference Device:<br>Surgical Navigation<br>Advanced Platform (SNAP) |
| Visual Guidance<br>Options with color-<br>coded cues. | 1) Dotted Line<br>2) Orientation Line<br>3) ILS | No | No |
The following table compares the key features of the subject and predicate devices:
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Surgical Theater. The logo consists of a stylized letter "S" in red and orange gradient, followed by the words "SURGICAL" in red and "THEATER" in yellow. A black line is present under the logo.
SpineAR SNAP
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Surgical Theater. The logo consists of a stylized letter "S" in red and orange gradient, followed by the words "SURGICAL" in red and "THEATER" in yellow. A black line is present at the bottom of the logo.
SpineAR SNAP
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Surgical Theater. The logo consists of a stylized letter "S" in a gradient of red and orange, followed by the words "SURGICAL" in red and "THEATER" in orange. A black line is present at the bottom of the logo.
The differences between the subject and predicate device do not raise any new questions regarding safety and effectiveness. Based on the information provided in this 510(k) submission. the SpineAR SNAP is considered substantially equivalent to the predicate device in terms of fundamental scientific technology.
## Performance Data
The following non-clinical performance testing was conducted to evaluate the subject device to ensure the SpineAR SNAP meets its intended use and performance requirements.
- Software Verification and Validation was performed to test the software requirements specifications for the SpineAR SNAP.
- Human Factors and Usability Validation demonstrated that the intended users of a SpineAR SNAP can safely and effectively perform tasks for the intended uses in the expected use environments, i.e. intraoperative use of the AR headsets. Validation took place under conditions of simulated use with users providing feedback on the usability of the software for performing defined tasks.
- Navigation Accuracy testing was conducted to demonstrate that SpineAR does not negatively affect the navigation accuracy of the connected Medtronic StealthStation 8 surgical navigation system. The final placement of each screw in a spine model was assessed via a post-surgical CT scan and compared to the pre-surgical plan to calculate the absolute displacement of the screw tip and absolute angular error of the screw. The summary of accuracy test results below demonstrate performance in 3D positional and trajectory accuracy that is within the system accuracy of the StealthStation S8 (mean positional error < 2.0 mm and mean trajectory error of < 2°):
- Mean Positional/Displacement Error: 1.02 mm O
- Max Positional/Displacement Error: 2.80 mm O
- Mean Trajectory/Angular Error: 1.79° O
- o Max Trajectory/Angular Error: 3.00°
- Verification of virtual screw library by verifying the length and diameter (e.g. 40 mm x 5.5 mm) of virtual screws shown in the 3D model accurately represent the real screws used during surgery, and screws are accurately positioned at the tip of the tracked tool.
- Performance of the Headset display was demonstrated by verifying the following elements: Field of View (FOV), resolution, luminance, transmittance, distortion, contrast ratio, temporal, display noise and motion-to-photon latency.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Surgical Theater. The logo consists of a stylized letter "S" in orange and yellow gradient, followed by the words "SURGICAL" in red and "THEATER" in yellow. The words are stacked on top of each other, with "SURGICAL" on top and "THEATER" on the bottom. There is a trademark symbol next to the word "THEATER".
- Projection latency testing was performed to assess the time delay between movement of a navigated handheld instrument and its display in the scene shown in the Magic Leap 1 and Microsoft HoloLens2 AR headsets. Based on experience with other types of simulators and feedback from surgeons during usability testing, a time delay requirement of < 250 ms was established. The summary of latency test results below demonstrates the projection delay for both AR headsets is acceptable:
- o Average Delay: 195 ms for HoloLens2 and 195 ms for Magic Leap 1
- o Worst Case Delay: 231 ms for HoloLens2 and 231 ms for Magic Leap 1
- Electromagnetic Compatibility (EMC) was tested in accordance with IEC 60601-1-● 2:2014+A1:2020 for all system components including both the Magic Leap 1 and Microsoft HoloLens2 AR headsets.
- Wireless coexistence was evaluated and tested per AAMI TIR69: 2017/(R)2020 and ANSI IEEE C63.27-2017.
Performance data demonstrated the SpineAR SNAP functions as intended without raising new safety or effectiveness issues, and does not impact accuracy or clinical workflow. The test results support a finding that the SpineAR SNAP is substantially equivalent to the predicate StealthStation S8 Spine Software v1.3.0.
## Substantial Equivalence Conclusion
The SpineAR SNAP is substantially equivalent to the StealthStation™ S8 Spine Software v1.3.0. The SpineAR SNAP has the same intended use and similar indications, technological characteristics, and principles of operation as the predicate device. Any differences between the subject and predicate device do not raise any new concerns regarding safety and effectiveness.