IMNS MEDACTA NAVIGATION SYSTEM

K083872 · Medacta International S.A. · OLO · Jan 15, 2009 · Neurology

Device Facts

Record IDK083872
Device NameIMNS MEDACTA NAVIGATION SYSTEM
ApplicantMedacta International S.A.
Product CodeOLO · Neurology
Decision DateJan 15, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.4560
Device ClassClass 2
Attributes3rd-Party Reviewed

Intended Use

The iMNS Medacta Navigation System is intended to be used to support the surgeon during specific orthopedic surgical procedures by providing information on bone resections, instrument and implant positioning during joint replacement. The iMNS Medacta Navigation System provides computer assistance to the surgeon based on anatomical landmarks and other specific data obtained intra-operatively that are used to place surgical instruments. Examples of some surgical procedures include but are not limited to: Total Knee Replacement Minimally Invasive Total Knee Replacement

Device Story

iMNS Medacta Navigation System provides intraoperative computer-aided navigation for total knee replacement. System uses infrared camera to track passive markers attached to patient bones; monitors spatial location of anatomical landmarks (femur head, knee, ankle). Proprietary software (Evolis Global Femur First) processes tracking data to guide bone resections and implant positioning. Surgeon operates system via keyboard, foot pedal, or mouse; views real-time measurements on computer monitor. System provides guidance for navigated instruments; surgeon retains control over all surgical decisions. Benefits include improved accuracy of bone resections and implant placement. Used in OR by surgeons.

Clinical Evidence

No clinical data. Performance testing conducted via design verification to written protocols with pre-defined acceptance criteria and design validation in a simulated user setting by a surgeon.

Technological Characteristics

Stereotaxic optical tracking system; infrared camera with LED emitters; passive markers. Components: computer, monitor, interface devices (keyboard, foot pedal, mouse), manual reusable surgical instruments. Software: proprietary Medacta navigation software. Connectivity: standalone system. Sterilization: manual reusable instruments.

Indications for Use

Indicated for patients undergoing orthopedic surgical procedures, specifically Total Knee Replacement and Minimally Invasive Total Knee Replacement, requiring computer-assisted navigation for bone resection and implant positioning.

Regulatory Classification

Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K083872 Image /page/0/Picture/1 description: The image shows the logo for Medacta International. The logo consists of two triangles pointing upwards, followed by the word "Medacta" in a sans-serif font. Below the word "Medacta" is the word "International" in a smaller font, followed by a small square with a plus sign inside. ## 510(k) Summary - Manufacturer: MEDACTA International SA Strada Regina CH6874 Castel San Pietro Switzerland Phone (+41) 91 696 60 60 FAX (+41) 91 696 60 66 JAN 1 5 2009 - Contact Person: Ms. Natalie J. Kennel Consultant NJK & Associates, Inc. 13721 Via Tres Vista San Diego, CA 92129 USA Phone: (858) 705-0350 Fax: (858) 764-9739 email: NKennel@njkconsulting.com Date Prepared: Dec. 11. 2008 ## DEVICE INFORMATION Trade/Proprietary Name: iMNS Medacta Navigation System Common/Classification Name: > 21 CFR 882.4560 Class II Device Product Code: HAW K061362 PiGalileo™ Total Knee Replacement (TKR) Predicate Device: System Product Description: The iMNS Medacta Navigation System is a device for computer aided navigation of surgical instruments used in total knee replacement surgery. The system works on the common principle of stereotaxic technology in which passive markers are securely mounted on the patient's bones and an infrared camera is used to monitor the spatial location of those marker and and information is used to locate the anatomical landmarks such as centers of rotation of the femur head, knee and ankle intraoperatively. These measurements are displayed on a computer screen in real time. The {1}------------------------------------------------ instruments are then outfitted with the passive markers to improve the positioning of the cutting guides. The information from the system with the "navigated" instruments assists the surgeon in optimally conducting the bone resections and positioning of the orthopedic surgical implants. The surgeon maintains control of the surgery and makes any decisions required with regard to bone resections and implant positioning but the iMNS Medacta Navigation System provides real time support and information throughout the surgery. The iMNS Medacta Navigation System consists of the following key components: - . An acquisition system composed of two infrared cameras equipped with infrared light emitting diodes (LED) to track the position of the passive markers, - . A computer running the proprietary Medacta software and a monitor, - . Interface devices of a keyboard, foot pedal and optional mouse to control the system, and - Manual reusable surgical instruments. . The software application called Evolis Global Femur First is designed to work with Medacta's Evolis Total Knee System, cleared under K081023. The manual reusable surgical instruments include instruments specifically designed for navigated surgery and other standard surgical instruments needed to conduct total knee replacement. # Indications for Use: The iMNS Medacta Navigation System is intended to be used to support the surgeon during specific orthopedic surgical procedures by providing information on bone resections, instrument and implant positioning during joint replacement. The iMNS Medacta Navigation System provides computer assistance to the surgeon based on anatomical landmarks and other specific data obtained intra-operatively that are used to place surqical instruments. Examples of some surgical procedures include but are not limited to: Total Knee Replacement Minimally Invasive Total Knee Replacement # Performance Testing No performance standards applicable to this device have been adopted under Section 514 of the Food, Drug and Cosmetic Act. Performance testing of the iMNS Medacta Navigation System was conducted in accordance with design controls, international standards and FDA guidance documents. The iMNS Medacta Navigation System was tested as part of design verification to written protocols with pre-defined acceptance criteria. The {2}------------------------------------------------ testing met all acceptance criteria. Design validation was conducted on the iMNS Medacta Navigation System in a simulated user setting by a surgeon and demonstrated that the system meets user needs and intended uses. #### Comparison to Predicate Device The iMNS Medacta Navigation System is similar to or identical to the predicate device, the PiGalileo™ Total Knee Replacement System in the following aspects: Intended Use/Indications for Use, design features, operating principle, type of optical tracking system, graphical user interface. control switch, passive locators, navigated surgery process, sterility, and biocompatibility aspects. The main differences between the iMNS Medacta Navigation System and the predicate device are the type of computer, the lack of a motorized cutting guide, and the model of the optical tracking system. None of these differences raise any new issues of safety or effectiveness. ### Conclusion: The data and information provided in this submission support the conclusion that the iMNS Medacta Navigation System is substantially equivalent to its predicate device, Plus Orthopedics PiGalileo™ Total Knee Replacement System with respect to indications for use, operating principle, and technological characteristics. {3}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of an abstract eagle or bird-like figure with three curved lines forming its body and wings. To the left of the emblem, the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the emblem. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Medacta International. SA % Underwriters Laboratories, Inc. Ms. Casey Conry 1285 Walt Whitman Road Melville, New York 11747 JAN 1 5 2009 Re: K083872 Trade/Device Name: iMNS Medacta Navigation System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: II Product Code: OLO, HAW Dated: December 19, 2008 Received: December 29, 2008 Dear Ms. Conry: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ #### Page 2 - Ms. Casey Conry This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Mark N Mulkerson - Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ #### Indications for Use Statement 510(k) Number (if known): K083872 Device Name: iMNS Medacta Navigation System Indications for Use: The iMNS Medacta Navigation System is intended to be used to support the surgeon during specific orthopedic surgical procedures by providing information on bone resections, instrument and implant positioning during joint replacement. The iMNS Medacta Navigation System provides computer assistance to the surgeon based on anatomical landmarks and other specific data obtained intra-operatively that are used to place surgical instruments. Examples of some surgical procedures include but are not limited to: Total Knee Replacement Minimally Invasive Total Knee Replacement Prescription Use Over-The-Counter Use X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page (Division Sign-Off) Division of General, Restorative, and Neurological Devices 510(k) Number
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...