CoreLink Navigation Instruments
K212825 · Corelink, LLC · OLO · Oct 13, 2021 · Neurology
Device Facts
| Record ID | K212825 |
| Device Name | CoreLink Navigation Instruments |
| Applicant | Corelink, LLC |
| Product Code | OLO · Neurology |
| Decision Date | Oct 13, 2021 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 882.4560 |
| Device Class | Class 2 |
Intended Use
CoreLink Navigation Instruments are intended to be used in the preparation and placement of CoreLink screws (Tiger and Entasis) during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medtronic StealthStation® System S8 (V1.2.20), which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.
Device Story
CoreLink Navigation Instruments are non-sterile, reusable stainless steel manual surgical instruments; designed to interface with Medtronic StealthStation® System S8. Instruments assist surgeons in spinal surgery by providing navigation for precise placement of Tiger and Entasis screws. System utilizes CT/MR models, fluoroscopy, or digitized anatomical landmarks to track instrument position relative to rigid structures (vertebrae). Surgeon operates instruments in OR; navigation system provides real-time visual feedback on display. Output enables surgeon to accurately guide screw placement; potentially reducing surgical error and improving patient outcomes in spinal procedures.
Clinical Evidence
Bench testing only. Confirmatory positional accuracy testing was performed per ASTM F2554 using the Medtronic StealthStation® System S8 Version 1.2.0 (1.2.0-20) to verify that system modifications maintain adequate accuracy.
Technological Characteristics
Materials: Stainless steel. Form factor: Manual surgical instruments. Connectivity: Interfaces with Medtronic StealthStation® System S8. Standards: ASTM F2554 (positional accuracy). Sterilization: Reusable (non-sterile provided).
Indications for Use
Indicated for patients undergoing spinal surgery requiring preparation and placement of CoreLink screws (Tiger and Entasis) where stereotactic navigation is appropriate for locating anatomical structures in open or minimally invasive procedures.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
Predicate Devices
- CoreLink Navigation Instruments (K182731)
- Medtronic Navigated Reusable Instruments (K161210)
- Medtronic Navigated Reusable Instruments (K170679)
Related Devices
- K214059 — CoreLink Navigation Instruments · Corelink, LLC · Jan 19, 2022
- K230880 — CoreLink Navigation Instruments · Corelink, LLC · May 2, 2023
- K182731 — Corelink Navigation Instruments · Corelink, LLC · Dec 13, 2019
- K220973 — CoreLink Navigation Instruments · Corelink, LLC · Jun 1, 2022
- K182345 — Spineology Navigation Instruments · Spineology, Inc. · Sep 21, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 13, 2021
CoreLink, LLC % Nathan Wright Engineer & Regulatory Specialist Empirical Testing Corp. 4628 Northpark Drive Colorado Springs, Colorado 80918
Re: K212825
Trade/Device Name: CoreLink Navigation Instruments Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: September 2, 2021 Received: September 3, 2021
#### Dear Nathan Wright:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA
Page 2
has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES | Form Approved: OMB No. 0910-0120 |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|
| Food and Drug Administration | Expiration Date: 06/30/2020 |
| Indications for Use | See PRA Statement on last page. |
| 510(k) Number (if known) | K212825 |
| Device Name | CoreLink Navigation Instruments |
| Indications for Use (Describe) | |
| CoreLink Navigation Instruments are intended to be used in the preparation and placement of CoreLink screws (Tiger and Entasis) during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medtronic StealthStation® System S8 (V1.2.20), which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy. | |
| Type of Use (Select one or both, as applicable) | |
| <span style="text-decoration: underline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) <span style="text-decoration: underline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: | |
| Department of Health and Human Services | |
| Food and Drug Administration | |
| Office of Chief Information Officer | |
| Paperwork Reduction Act (PRA) Staff | |
| PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
| | |
FORM FDA 3881 (7/17)
{3}------------------------------------------------
# K212825 - 510(K) Summary
| Submitter's Name: | CoreLink, LLC |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Submitter's Address: | 2072 Fenton Logistics Park<br>St. Louis, Missouri 63026 |
| Submitter's Telephone: | 888-349-7808 |
| Contact Person: | Nathan Wright MS<br>Empirical Testing Corp.<br>719-351-0248<br>nwright@empiricaltech.com<br>Image: Empirical Testing Corp. Logo |
| Date Summary was Prepared: | September 2, 2021 |
| Trade or Proprietary Name: | CoreLink Navigation Instruments |
| Common or Usual Name: | Orthopedic Stereotaxic Instruments |
| Classification: | Class II per 21 CFR §882.4560 |
| Product Code: | OLO |
| Classification Panel: | Orthopedic |
### DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:
CoreLink Navigation Instruments are non-sterile, re-usable manual surgical instruments made from stainless steel. These instruments are designed to interface with the Medtronic StealthStation® Navigation System to assist surgeons in precisely locating anatomical structures.
The purpose of this submission is to clear the CoreLink Navigation Instruments for use with additional versions of the StealthStation software and to add instruments to the previously cleared CoreLink Navigation Instruments set.
### INDICATIONS FOR USE
CoreLink Navigation Instruments are intended to be used in the preparation and placement of CoreLink screws (Tiger and Entasis) during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medtronic StealthStation® System S8 (V1.2.0), which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.
# TECHNOLOGICAL CHARACTERISTICS
The subject and predicate devices have nearly identical technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. Specifically, the following characteristics are identical between the subject and predicates:
- Indications for use ●
- Materials of manufacture .
{4}------------------------------------------------
- Principles of operation ●
#### Table 5-1 Predicate Devices
| 510k Number | Trade or Proprietary or<br>Model Name | Manufacturer | Predicate<br>Type |
|----------------------|---------------------------------------------|----------------------------|-------------------|
| K182731 | CoreLink Navigation<br>Instruments | CoreLink, LLC | Primary |
| "K161210"; "K170679" | Medtronic Navigated<br>Reusable Instruments | Medtronic Sofamor<br>Danek | Additional |
# Performance Data
In support of this Special 510(k) submission, CoreLink, LLC has conducted confirmatory positional accuracy testing per ASTM F2554 with Medtronic StealthStation® System S8 Version 1.2.0 (1.2.0-20) to demonstrate that the system modifications provide adequate and substantially equivalent accuracy for their intended use.
## CONCLUSION
The overall technology characteristics and performance data lead to the conclusion that the CoreLink Navigation Instruments are substantially equivalent to the predicate device.