ExactechGPS® System
K233299 · Blue Ortho · OLO · Jan 26, 2024 · Neurology
Device Facts
| Record ID | K233299 |
| Device Name | ExactechGPS® System |
| Applicant | Blue Ortho |
| Product Code | OLO · Neurology |
| Decision Date | Jan 26, 2024 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 882.4560 |
| Device Class | Class 2 |
Intended Use
The ExactechGPS® System is intended for use during stereotaxic surgery to aid the surgeon in locating anatomical structures and aligning the endoprostheses with the anatomical structures. TKA Pro software application is specifically indicated for Total Knee Arthroplasty.
Device Story
ExactechGPS is an image-guided surgical navigation system used in orthopedic arthroplasty. It utilizes hardware trackers to capture intraoperative data regarding patient anatomy and surgical instrumentation. This data is transmitted to a hardware station, which computes and displays real-time information—including distances, angles, and prosthetic component placement—to assist surgeons in identifying and characterizing bone cuts. The system is operated by surgeons in the OR to improve the accuracy of endoprosthesis alignment. The device provides visual feedback to the surgeon, facilitating clinical decision-making during bone preparation and implant positioning, potentially improving surgical outcomes.
Clinical Evidence
Bench testing only. Performance verified through general function tests, electrical safety (IEC 60601-1), EMC compatibility (IEC 60601-1-2), photobiological safety (EN 62471), and sterilization validation (ISO 11135-1:2014).
Technological Characteristics
Stereotaxic navigation system comprising an optical localizer (camera), tablet, and hardware station. Connectivity via integrated Wi-Fi. Sterilization of drapes per ISO 11135-1:2014. Compliant with IEC 60601-1 and IEC 60601-1-2 standards.
Indications for Use
Indicated for patients undergoing stereotaxic surgery, specifically Total Knee Arthroplasty (TKA) when using the TKA Pro software application, to assist surgeons in anatomical localization and endoprosthesis alignment.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
Predicate Devices
- ExactechGPS System (K213877)
Related Devices
- K213877 — ExactechGPS® System · Blue Ortho · May 2, 2022
- K050752 — AESCULAP ORTHOPILOT 2 THA V 2.0 · Aesculap, Inc. · Apr 27, 2005
- K152764 — EXACTECH GPS · Blue Ortho · Jul 13, 2016
- K153396 — Aesculap OrthoPilot Next Generation · Aesculap Implant Systems, LLC · May 16, 2016
- K012937 — KNEE MODULE FOR THE STEALTHSTATION SYSTEM · Medtronic Surgical Navigation Technologies · Jan 25, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
January 26, 2024
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Blue Ortho Matthieu Coic QA RA Director 22 Chemin du Vieux Chene Meylan, 38240 France
#### Re: K233299
Trade/Device Name: ExactechGPS® System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instruments Regulatory Class: Class II Product Code: OLO Dated: December 29, 2023 Received: December 29, 2023
#### Dear Matthieu Coic:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Shumaya Ali -S
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K233299
Device Name ExactechGPS® System
Indications for Use (Describe)
The ExactechGPS® System is intended for use during stereotaxic surgery to aid the surgeon in locating anatomical structures and aligning the endoprostheses with the anatomical structures.
TKA Pro software application is specifically indicated for Total Knee Arthroplasty.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial;"> <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span></span> |
|--------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial;"> <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span></span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for BlueOrtho, a company specializing in augmented orthopedics. The logo features a stylized blue "C" shape on the left, followed by the text "BlueOrtho" in a clean, sans-serif font. Below "BlueOrtho" is the tagline "AUGMENTED ORTHOPAEDICS" in smaller, lighter font.
2 Chemin du Vieux Chêne 38240 Meylan France
#### ExactechGPS® System 510(k) Summary of Safety and Effectiveness
#### I. SUBMISSION DATE
September 29, 2023.
#### II. SUBMITTER:
BLUE ORTHO 22 Chemin du Vieux Chêne 38240 Meylan France Phone: +33 (0)4 58 00 35 25 Contact person: Matthieu COIC - Mail: matthieu.coic@blue-ortho.com
## III. US LOCAL AGENT
Exactech, Inc. 2320 NW 66th Ct. Gainesville, FL. 32653 Phone: 352-377-1140
## IV. INFORMATION ON DEVICES TO WHICH SUBSTANTIAL EQUIVALENCE IS CLAIMED
510(k) Number Trade or Proprietary Model Name Manufacturer ExactechGPS System #K213877 Blue Ortho This predicate has not been subject to a design-related recall.
## V. PROPOSED DEVICE DESCRIPTION
Trade or Proprietary or Model Name(s): ExactechGPS® System Common Name: Surgical navigation system Classification Name: Orthopedic Stereotaxic Instrument (21 CFR 882.4560, product code OLO) Classification: Class II Device Description: The ExactechGPS System proposed in this submission is a modification of the ExactechGPS System cleared per 510(k) #K213877.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for BlueOrtho Augmented Orthopaedics. The logo features a blue square with a white "C" shape inside on the left. To the right of the square, the words "BlueOrtho" are written in a blue sans-serif font. Below "BlueOrtho", the words "AUGMENTED ORTHOPAEDICS" are written in a smaller, lighter blue font.
? Chemin du Vieux Chêne 8240 Mevlan -rance
The ExactechGPS System is an image-quided surgery, or navigation, system intended to be used during orthopedic surgical procedures to intraoperatively assist surgeons during arthroplasty. The ExactechGPS System enables surgeons to acquire intraoperative data by computing and displaying information such as distances, angles, and placement of prosthetic components in order to identify and characterize bone cuts necessary to achieve surgical goals.
The ExactechGPS System works with ExactechGPS hardware trackers that communicate intraoperative data to the ExactechGPS hardware station to provide surgeons with real-time information on the positions of patient anatomical structures and instrumentation used to prepare bone during stereotaxic surgery.
The proposed modifications do not change the ExactechGPS System general intended use, general design features, or basic fundamental scientific technology.
#### VI. INDICATIONS FOR USE
The Exactech GPS is intended for use during stereotaxic surgery to aid the surgeon in locating anatomical structures and aligning the endoprostheses with the anatomical structures.
TKA Pro software application is specifically indicated for Total Knee Arthroplasty.
#### OF TECHNOLOGICAL CHARACTERISTICS WITH THE VII. COMPARISON PREDICATE DEVICE
This submission proposes the following hardware modifications to the ExactechGPS system:
- A new optical localizer (or camera), to avoid obsolescence of sub-component . and implement minor design improvements.
- A new Tablet component, to avoid obsolescence of previous Tablet. This new . Tablet also implies minor design changes to the mechanical frame and other subcomponents of the station.
- . The wireless feature is now provided by the Tablet itself, not anymore using a Wi-Fi dongle
The ExactechGPS system cleared per 510(k) #K213877 is compatible with the ExactechGPS TKA Pro clinical software applications.
The proposed ExactechGPS system that includes changes listed hereabove is compatible with the same ExactechGPS TKA Pro clinical software application.
Based on test bench testing, the proposed changes do not alter significantly the performances of the devices and the way it is used:
- The proposed ExactechGPS system have the same general features and . dimensions, with the same technology and performances.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for BlueOrtho. The logo consists of a blue square with a white "C" inside, followed by the words "BlueOrtho" in blue. Below the words "BlueOrtho" is the phrase "AUGMENTED ORTHOPAEDICS" in a smaller font, also in blue. The logo is simple and modern, and the use of blue gives it a sense of trust and reliability.
- . The redesigned electronic card only resolves obsolescence issues with no impact on user, global performances and safety.
- . The use of the native Wi-Fi feature of the tablet only allow to retrieve the external additional dongle used with the predicate device.
## VIII. PERFORMANCE DATA
Testing information demonstrating safety and effectiveness of the ExactechGPS System in the intended environment of use is supported by testing that was conducted in-house and by accredited testing laboratories.
This submission includes or references the following non-clinical testing:
- . General Functions tests to ensure design is fully verified and validated.
- Electrical safety testing per IEC 60601-1 demonstrated the proposed . ExactechGPS System is compliant with the requirements of the standard.
- EMC Compatibility testing per IEC 60601-1-2 demonstrated the proposed ● ExactechGPS System is compliant with the requirements of the standard.
- . Analysis of photobiological safety per EN 62471 demonstrated the proposed ExactechGPS System is compliant with the requirements of the standard.
- Sterilization validation per ISO 11135-1:2014 demonstrated the proposed . ExactechGPS Sterile Drape is compliant with the requirements of the standard.
#### IX. SUBSTANTIAL EQUIVALENCE CONCLUSION
A comparison of specific features included in this submission demonstrates the proposed ExactechGPS System is substantially equivalent to the cited predicate cleared per #K213877. The devices share identical intended use, identical general design features and basic fundamental scientific technology.