SMITH & NEPHEW PIGALILEO SCREW TARGETING SYSTEM, VERSION 1.1

K092497 · Smith & Nephew, Inc. · OLO · Sep 11, 2009 · Neurology

Device Facts

Record IDK092497
Device NameSMITH & NEPHEW PIGALILEO SCREW TARGETING SYSTEM, VERSION 1.1
ApplicantSmith & Nephew, Inc.
Product CodeOLO · Neurology
Decision DateSep 11, 2009
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 882.4560
Device ClassClass 2

Intended Use

The Smith & Nephew PiGalileo Screw Targeting System is intended to be an intraoperative image guided localization system. It is a computer assisted orthopedic surgery tool to aid the surgeon with drill positioning for screws during intramedullary nail implantation. It provides information to the surgeon that is used to place surgical instruments during surgery utilizing intraoperatively obtained electromagnetic tracking data. The Smith & Nephew PiGalileo Screw Targeting System V 1.1 is indicated for long bone fractures treated with intramedullary nails in which the use of stereotactic surgery may be appropriate.

Device Story

Computer-assisted orthopedic surgery system; utilizes electromagnetic tracking to assist surgeons in locating/positioning screws during intramedullary nail implantation. System inputs: electromagnetic signals from sensor-embedded instruments placed within controlled, varying magnetic fields. Operation: induced voltages in sensor coils calculated to determine 3D virtual position of instruments; avoids line-of-sight constraints of optical systems. Output: real-time 3D spatial visualization of surgical instruments. Used in OR by orthopedic surgeons; provides guidance for drill positioning. Benefits: improved accuracy in screw placement during trauma surgery; reduced reliance on direct line-of-sight.

Clinical Evidence

No clinical data provided; device relies on bench testing and substantial equivalence to the predicate system.

Technological Characteristics

Electromagnetic tracking system; utilizes sensor-embedded instruments and magnetic field generators. 3D spatial measurement via induced voltage calculation. Class II stereotaxic instrument. Software-controlled.

Indications for Use

Indicated for patients with long bone fractures requiring intramedullary nail fixation where stereotactic surgery is appropriate.

Regulatory Classification

Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K042497 f. lot 1 # SEP 1 1 2009 ### Smith & Nephew, Inc. Summary of Safety and Effectiveness PiGalileo Screw Targeting System V1.1 Contact Person and Address Date of Summary: 08/12/2009 Regina Holmes Regulatory Affairs Specialist Smith & Nephew Orthopaedics 1450 Brooks Road Memphis, TN 38116 (901) 399-6538 Name of Device: Smith & Nephew PiGalileo Screw Targeting System V1.1 Common Name: Computer Assisted Surgery System #### Device Description The PiGalileo Screw Targeting System is a computer controlled electromagnetic tracking system. It assists the surgeon in locating and positioning screws in an intramedullary nail implant during orthopedic trauma surgery. The link between the sterile surgical area (patient) and the instrument system is provided through an electromagnetic tracking system. Electromagnetic spatial measurement systems determine the location of instruments that are embedded with sensor coils. When the sensor-embedded instrument is placed inside controlled, varying magnetic fields, voltages are induced in the sensor coils. These induced voltages are used by the measurement system to calculate a 3D virtual position of the instrument. Because the magnetic fields are of a low field strength and can safely pass through human tissue, location measurement of an object is possible without the line-of-sight constraints of an optical spatial measurement system that requires a camera. #### Device Classification 21 CFR 882.4560 Stereotaxic Instrument - Class II #### Indications for Use The Smith & Nephew PiGalileo Screw Targeting System is intended to be an intraoperative image guided localization system. It is a computer assisted orthopedic surgery tool to aid the surgeon with drill positioning for screws during intramedullary nail implantation. It provides information to the surgeon that is used to place surgical instruments during surgery utilizing intraoperatively obtained electromagnetic tracking data. The Smith & Nephew PiGalileo Screw Targeting System V 1.1 is indicated for long bone fractures treated with intramedullary nails in which the use of stereotactic surgery may be appropriate. #### Substantial Equivalence Information The overall design of the PiGalifeo Screw Targeting System V1.1 is substantially equivalent to the previously cleared device listed below: | Manufacturer | Description | 510(k) | Clearance Date | |----------------------|---------------------------------------|---------|----------------| | Smith & Nephew, Inc. | PiGalileo Screw Targeting System V1.0 | K090420 | 10/29/08 | {1}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol. SEP 1 1 2009 Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002 Smith & Nephew % Ms. Regina Holmes Regulatory Affairs Specialist 1450 East Brooks Road Memphis, Tennessee 38116 Re: K092497 Trade/Device Name: PiGalileo Screw Targeting System V1.1 Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instruments Regulatory Class: II Product Code: OLO Dated: August 12, 2009 Received: August 14, 2009 Dear Ms. Holmes: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {2}------------------------------------------------ ## Page 2 – Ms. Regina Holmes If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.html for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ### Indications for Use 510(k) Number (if known): Device Name: PiGalileo Screw Targeting System V1.1 Indications for Use: The Smith & Nephew PiGalileo Screw Targeting System is intended to be an intraoperative image guided localization system. It is a computer assisted orthopedic surgery tool to aid the surgeon with drill positioning for screws during intramedullary nail implantation. It provides information to the surgeon that is used to place surgical instruments during surgery utilizing intraoperatively obtained electromagnetic tracking data. The Smith & Nephew PiGalileo Screw Targeting System V 1.1 is indicated for long bone fractures treated with intramedullary nails in which the use of stereotactic surgery may be appropriate. Prescription Use x (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Nifre Odu for mxm (Division Sign-Off) (Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number. Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...