REAL INTELLIGENCE CORI

K221224 · Blue Belt Technologies, Inc. · OLO · Aug 19, 2022 · Neurology

Device Facts

Record IDK221224
Device NameREAL INTELLIGENCE CORI
ApplicantBlue Belt Technologies, Inc.
Product CodeOLO · Neurology
Decision DateAug 19, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.4560
Device ClassClass 2

Intended Use

Real Intelligence CORI (CORI) is intended to assist the surgeon in providing software-defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures.

Device Story

Robotic-assisted orthopedic surgical navigation and burring system; uses passive infrared tracking camera to map bone landmarks and implant geometry. Surgeon defines bone surface model and surgical plan; system controls robotic bur cutting engagement via exposure control (retracting bur into guard) and speed control (regulating tool control unit) based on proximity to planned surface. Used in hospital/surgical suite by trained medical professionals. Includes CORI Knee TENSIONER, an electromechanical accessory for quantitative assessment of soft tissue laxity during gap planning. Output provides spatial boundaries and navigation data to assist surgeon in bone preparation and implant placement. Benefits include improved accuracy in bone resection and ligament balancing.

Clinical Evidence

Bench testing only. Included biocompatibility (ISO 10993-1), electrical safety/EMC (IEC 60601-1, IEC 60601-1-2), software verification (IEC 62304), and usability engineering validation (IEC 62366-1). Testing confirmed design inputs met and performance is equivalent to predicates.

Technological Characteristics

Robotic-assisted navigation system using passive infrared tracking. Includes electromechanical TENSIONER for force feedback. Software version 1.7. Complies with IEC 60601-1, IEC 60601-1-2, IEC 62304, and IEC 62366-1. Materials evaluated per ISO 10993-1.

Indications for Use

Indicated for surgical procedures where stereotactic surgery is appropriate and reference to rigid anatomical bony structures can be determined, including unicondylar knee replacement (UKR), total knee arthroplasty (TKA), and total hip arthroplasty (THA).

Regulatory Classification

Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ August 19, 2022 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. Blue Belt Technologies, Inc. Corrine Herlinger Principal Regulatory Affairs Specialist 2905 Northwest Blvd., Ste. 40 Plymouth, Minnesota 55441 Re: K221224 Trade/Device Name: Real Intelligence Cori (Cori) Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: July 21, 2022 Received: July 21, 2022 Dear Corrine Herlinger: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For: Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K221224 Device Name REAL INTELLIGENCE CORI (CORI) Indications for Use (Describe) CORI is indicated for use in surgical procedures, in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be determined. These procedures include unicondylar knee replacement (UKR), total knee arthroplasty (TKA), and total hip atthroplasty (THA). | Type of Use (Select one or both, as applicable) | | |--------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------| | <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Blue Belt Technologies. The logo consists of a blue swirl design on the left, followed by the words "BLUE BELT" in large, blue, sans-serif font. Below "BLUE BELT" is the word "TECHNOLOGIES" in a smaller, gray, sans-serif font. The overall design is clean and modern. 2905 Northwest Boulevard, Suite 40, Plymouth, Minnesota 55441 USA Tel 763.452.4950 Fax 763.452.4675 www.bluebelttech.com # 510(k) Summary | 510(k) Owner | Blue Belt Technologies, Inc.<br>2905 Northwest Blvd Ste. 40<br>Plymouth, MN 55441 USA<br>Tel: (763) 452-4950<br>Fax: (763) 452-4675 | |----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person | Corrine Herlinger<br>Principal Regulatory Affairs Specialist<br>Tel: (412) 683-3844 x 4128<br>Email: corrine.herlinger@smith-nephew.com | | Date of Submission | April 26, 2022 | | Classification Reference | 21 CFR 882.4560 | | Product Code | OLO | | Supported Codes | HSX, JWH, MBH | | Common/Usual Name | Orthopedic Stereotaxic Instrument | | Trade/Proprietary Name | REAL INTELLIGENCE® CORI® (CORI) | | Primary Predicate Device | REAL INTELLIGENCE® CORI® (K220255) | | Secondary Predicate Device | OMNIBotics Knee System (K200888) | | Reason for Submission | This Traditional 510(k) submission supports an<br>update to CORI to allow the system to be used with<br>the CORI Knee TENSIONER (TENSIONER) and the<br>associated software upgrade, CORI 1.7. | K221224 - Page 1 of 8 {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Blue Belt Technologies. The logo consists of a blue swirl design on the left, followed by the words "BLUE BELT" in large, blue, sans-serif font. Below "BLUE BELT" is the word "TECHNOLOGIES" in a smaller, gray, sans-serif font. The overall design is clean and modern. ### Intended Use Real Intelligence CORI (CORI) is intended to assist the surgeon in providing software-defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures. ## Indications for Use CORI is indicated for use in surgical procedures, in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be determined. These procedures include unicondylar knee replacement (UKR), total knee arthroplasty (TKA), and total hip arthroplasty (THA). ## Device Description The subject of this Traditional 510(k) is REAL INTELLIGENCE CORI (CORI), a robotic-assisted orthopedic surgical navigation and burring system. CORI uses established technologies of navigation via a passive infrared tracking camera. Based on intraoperatively-defined bone landmarks and known geometry of the surgical implant, the system aids the surgeon in establishing a bone surface model for the target surgery and planning the surgical implant location. For knee applications, CORI then aids the surgeon in executing the surgical plan by controlling the cutting engagement of the surgical bur. CORI knee application software controls the cutting engagement of the surgical bur based on its proximity to the planned target surface. The cutting control is achieved with two modes: - . Exposure control adjusts the bur's exposure with respect to a guard. If the surgeon encroaches on a portion of bone that is not to be cut, the robotic system retracts the bur inside the guard, disabling cutting. - Speed control regulates the signal going to the tool control unit itself and limits the speed of the drill if the target surface is approached. Alternatively, the surgeon can disable both controls and operate the robotic drill as a standard navigated surgical drill. K221224 - Page 2 of 8 {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Blue Belt Technologies. The logo consists of a blue swirl design on the left, followed by the words "BLUE BELT" in large, bold, blue letters. Below "BLUE BELT" is the word "TECHNOLOGIES" in smaller, gray letters. ## Currently Supported Knee Implants The following Smith+Nephew knee implants are supported on CORI: ### Table 1: Currently Supported Smith+ Nephew Knee Implants | Implant Model Name | 510(k) Number | Classification<br>Product Code | |--------------------------------------------|---------------------------|--------------------------------| | STRIDE Unicondylar Knee | K123380 | HSX | | ZUK Select Knee System | K160738 | HSX | | JOURNEY II Unicompartmental Knee System | K191211 | HSX | | JOURNEY UNI | K102069 | HSX | | JOURNEY II CR | K121443 | JWH | | JOURNEY II BCS | K111711 | JWH | | JOURNEY II XR | K141471, K152726 | JWH | | LEGION CR/PS | K951987, K962557, K093746 | JWH | | LEGION Porous CR Femoral Components | K073325, K091543 | MBH | | LEGION Porous CR Narrow Femoral Components | K210566 | MBH | | LEGION Porous Tibia | K100897 | MBH | | Porous Tibia Baseplate | K211221 | MBH | | GENESIS II CR/PS | K951987, K962557 | JWH | | ANTHEM | K142807 | JWH | ## Discussion of Similarities and Differences This Traditional 510(k) submission supports an update to the CORI System to allow the system to be used with the CORI Knee TENSIONER (TENSIONER) and the associated software upgrade, CORI 1.7. The TENSIONER is an electromechanical accessory designed to provide feedback to CORI with respect to the force that the user applies to the joint during gap assessment. It is used by the surgeon in the gap planning stages of a total knee replacement procedure to provide a quantitative assessment of soft tissue laxity. The modifications made to CORI do not impact the system's intended use, indications for use, or fundamental scientific technology. Blue Belt Technologies believes that CORI is subject to premarket notification requirements under Section 510(k) of the Federal Food, Drug and Cosmetic Act ("FFDCA" or "the Act") and is substantially equivalent to the previously cleared REAL INTELLIGENCE CORI system, K220255, and the OMNIBotics Knee System, K200888. K221224 - Page 3 of 8 {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for Blue Belt Technologies. The logo consists of a blue swirl design on the left, followed by the words "BLUE BELT" in large, bold, blue letters. Below "BLUE BELT" is the word "TECHNOLOGIES" in smaller, gray letters. The logo is clean and modern, with a focus on the company name. ### Table 2: Predicate Device | Manufacturer | Description | Submission<br>Number | Clearance Date | |------------------------------|------------------------|----------------------|----------------| | Blue Belt Technologies, Inc. | REAL INTELLIGENCE CORI | K220255 | 03/29/2022 | | Corin Ltd. | OMNIBotics Knee System | K200888 | 06/27/2020 | ### Table 3: Summary of Technological Similarities with Predicate | Devices | Subject Device<br>CORI | Primary Predicate<br>CORI<br>K220255 | Secondary Predicate<br>OMNIBotics Knee System<br>K200888 | |-------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended use and<br>Indications for Use<br>[Same] | Same as primary<br>predicate. | REAL INTELLIGENCE CORI (CORI)<br>is intended to assist the surgeon<br>in providing software-defined<br>spatial boundaries for<br>orientation and reference<br>information to anatomical<br>structures during orthopedic<br>procedures. | The OMNIBotics Knee<br>System is indicated for<br>stereotaxic surgery to aid the<br>surgeon in locating<br>anatomical structures and<br>aligning endoprostheses with<br>these anatomical structures<br>during Total Knee<br>Arthroplasty. | | | | CORI is indicated for use in<br>surgical knee procedures in<br>which the use of stereotactic<br>surgery may be appropriate, and<br>where reference to rigid<br>anatomical bony structures can<br>be determined. These<br>procedures include unicondylar<br>knee replacement, total knee<br>arthroplasty, and total hip<br>arthroplasty. | The BalanceBot™ is indicated<br>as a tool for adjustment of<br>soft tissue and the femoral<br>implant to reduce instability<br>from flexion gap asymmetry.<br>The OMNIBotics Knee<br>System supports OMNI Apex<br>Knee™ implants and CORIN<br>Unity Knee™ implants. | | Devices | Subject Device<br>CORI | Primary Predicate<br>CORI<br>K220255 | Secondary Predicate<br>OMNIBotics Knee System<br>K200888 | | Technological<br>Characteristics –<br>Operating Principle<br>[Same] | Same as primary<br>predicate. | For knee applications, CORI uses<br>established technologies to<br>prepare bone for attachment of<br>UKR and TKA implant<br>components. In the case of a<br>total knee arthroplasty, the<br>bone surface may also be<br>prepared to receive the femoral<br>and tibial cutting guides with<br>final bone surface for receiving<br>the implant prepared using a<br>standard surgical saw.<br><br>CORI uses intraoperative data<br>collection (image-free or non-CT<br>data generation) to create a<br>model of the patient's femur<br>and/or tibia, dependent on the<br>procedure being performed, and<br>allows the surgeon to prepare a<br>surgical plan.<br><br>The system uses predefined<br>boundaries generated during<br>the planning process to control<br>the motion of the surgical bur<br>and limit the amount of bone<br>removed in order to shape the<br>condyles or tibial plateau in<br>preparation for placement of<br>the surgical implant.<br><br>Bur cutting is controlled either<br>by retracting the bur in a guard,<br>or by controlling the speed of<br>the bur as the target surface is<br>approached.<br><br>To support the hip application,<br>CORI uses a Virtual Machine<br>with hypervisor to enable the<br>Windows-based HIP7 software<br>to run on CORI. | N/A | | Devices | Subject Device<br>CORI | Primary Predicate<br>CORI<br>K220255 | Secondary Predicate<br>OMNIBotics Knee System<br>K200888 | | Technological<br>Characteristics -<br>Tensioning<br>Instrument<br>[Substantially<br>Equivalent] | The TENSIONER is an<br>electromechanical<br>tensioning<br>instrument with<br>reverse-action pliers<br>to distract the joint. | Consists of a Gap Planning stage<br>to assist users with (manual)<br>ligament balancing. | The BalanceBot is an electro-<br>mechanical tensioning<br>instrument with integrated<br>force sensors and position<br>actuators. | | Technological<br>Characteristics -<br>Gap Balancing<br>[Substantially<br>Equivalent] | The TENSIONER is<br>used during the Gap<br>Planning stage of the<br>CORI Total Knee<br>Arthroplasty<br>application and<br>provides feedback to<br>CORI with respect to<br>the force that the<br>user applies to the<br>joint to assist with<br>ligament balancing. | Consists of a Gap Planning stage<br>to assist users with (manual)<br>ligament balancing. | BalanceBot assists the<br>surgeon with ligament<br>balancing during total knee<br>arthroplasty. | | Knee Implant<br>Product Codes<br>Supported<br>[Same] | Same as primary<br>predicate. | HSX, JWH, MBH | N/A | | Environment of Use<br>[Same] | Same as primary<br>predicate. | CORI is intended to be used by<br>trained medical professionals in<br>a hospital or clinical setting<br>equivalent to an orthopedic<br>surgery suite. | N/A | {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for Blue Belt Technologies. The logo features a blue swirl design on the left, followed by the words "BLUE BELT" in large, bold, blue letters. Below "BLUE BELT" is the word "TECHNOLOGIES" in smaller, gray letters. 2905 Northwest Boulevard, Suite 40, Plymouth, Minnesota 55441 USA Tel 763.452.4950 Fax 763.452.4675 www.bluebelttech.com K221224 - Page 5 of 8 ReshapingMobility™ {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the logo for Blue Belt Technologies. The logo consists of a stylized blue swirl on the left, followed by the words "BLUE BELT" in large, bold, blue letters. Below "BLUE BELT" is the word "TECHNOLOGIES" in smaller, gray letters. 2905 Northwest Boulevard, Suite 40, Plymouth, Minnesota 55441 USA Tel 763.452.4950 Fax 763.452.4675 www.bluebelttech.com K221224 - Page 6 of 8 ReshapingMobility™ {9}------------------------------------------------ Image /page/9/Picture/0 description: The image shows the logo for Blue Belt Technologies. The logo consists of a blue swirl design on the left, followed by the words "BLUE BELT" in large, bold, blue letters. Below "BLUE BELT" is the word "TECHNOLOGIES" in smaller, gray letters. ## Non-Clinical Testing (Bench) Design verification and validation testing demonstrated that CORI meets all design requirements and is as safe and effective as its primary predicate device (K220255) and secondary predicate device OMNIBotics Knee System (K200888). Comprehensive performance testing demonstrated that the system meets required design inputs. Performance data consisted of physical performance testing for all system components. Additionally, the following evidence was provided: - . Biocompatibility evaluation demonstrating that the system satisfies the requirements of ISO 10993-1 Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a risk management process. - . Safety and Electromagnetic Compatibility (EMC) testing demonstrating that the device complies with IEC 60601-1 Medical Electric Equipment – Part 1: General Requirements for Basic Safety and Essential Performance and IEC 60601-1-2 Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements and Tests. - . Software verification testing, including software integration and workflow testing, was completed. Software was developed in accordance with IEC 62304 Medical device software -Software life cycle processes, and this submission contains documentation per the requirements of FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. - . Usability Engineering Validation Testing demonstrating that representative users were able to safely and effectively use CORI and TENSIONER in a simulated use environment. Human factors and usability engineering processes were followed per IEC 62366-1:2015+A1:2020 Application of Usability Engineering to Medical Devices. All verification and validation testing concluded with acceptable results. Based on these test results, Blue Belt Technologies has concluded that all design inputs have been met and that the design verification and validation testing performed did not raise any new questions of safety or effectiveness. ## Conclusions The subject device, CORI, described in this submission has the same intended use and the same fundamental scientific technology as the primary predicate device, K220255. The primary difference between the subject device and CORI (K220255) is the update to allow the system to be used with TENSIONER and the associated software upgrade, CORI 1.7. The main functionality of the Total Knee Arthroplasty (TKA) software is the same as presented in K201022, with modifications to allow for communication with TENSIONER. The addition of TENSIONER does not change the system's method for gap calculation, which remains the same as the previously cleared devices. While the technological characteristics differ between manual and a tool-assisted gap-balancing measurement, the questions regarding accuracy and implementation of intraoperative gap-balancing K221224 - Page 7 of 8 {10}------------------------------------------------ Image /page/10/Picture/0 description: The image shows the logo for Blue Belt Technologies. The logo consists of a blue swirl design on the left, followed by the words "BLUE BELT" in large, bold, blue letters. Below "BLUE BELT" is the word "TECHNOLOGIES" in smaller, gray letters. apply to both the subject and primary predicate device. Additionally, usability testing demonstrated that users are able to successfully perform gap balancing using TENSIONER and the CORI system; therefore, the difference of the technological characteristics does not introduce new questions of safety or effectiveness. Blue Belt Technologies believes that FDA can find CORI to be substantially equivalent to the predicate devices, REAL INTELLIGENCE CORI system, K220255, and the OMNIBotics Knee System, K200888. The information presented in this 510(k) premarket notification demonstrates that CORI is as safe and effective as the primary predicate CORI system (K220255). K221224 - Page 8 of 8
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...