Anatase Spine Surgery Navigation System and Remex Spine Surgery Navigation System II
K251315 · Remex Medical Corp. · OLO · May 29, 2025 · Neurology
Device Facts
Record ID
K251315
Device Name
Anatase Spine Surgery Navigation System and Remex Spine Surgery Navigation System II
Applicant
Remex Medical Corp.
Product Code
OLO · Neurology
Decision Date
May 29, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
The Anatase Spine Surgery Navigation System and Remex Spine Surgery Navigation System II are indicated for precise positioning of surgical instruments or spinal implants during general spine surgery when reference to a rigid anatomical structure, such as the vertebra, can be identified relative to a patient's fluoroscopic or CT imagery. It is intended as a planning and intraoperative guidance system to enable open or percutaneous image guided surgery by means of registering intraoperative 2D fluoroscopic projections to pre-operative 3D CT imagery. Example procedures include: Posterior-approach spinal implant procedures, such as pedicle screw placement, with the lumbar region.
Device Story
Navigation system for spine surgery; enables precise positioning of surgical instruments/implants. Inputs: pre-operative 3D CT imagery and intraoperative 2D fluoroscopic projections. Operation: registers 2D fluoroscopic projections to 3D CT imagery to provide real-time guidance. Output: visual navigation display for surgeon. Used in OR by surgeons for open or percutaneous procedures. Benefits: improved accuracy for spinal implant placement. Includes new image calibrators (21F/31F) for compatibility with 21cm/31cm flat panel C-arms and auto-labeling function for image registration.
Clinical Evidence
Bench testing only. Verification and validation activities performed per FDA software guidance and IEC 62304. Risk assessment conducted per ISO 14971. Design verification confirmed outputs meet input requirements using test methods identical to predicate K233513. All tests passed.
Technological Characteristics
Stereotaxic instrument; includes image calibrators (21F/31F) for 21cm/31cm flat panel C-arms. Software-based image registration and navigation. Connectivity: integrates with C-arm imaging systems. Software: includes auto-labeling for CT/C-arm registration. Design controls and risk management per ISO 14971 and IEC 62304.
Indications for Use
Indicated for patients undergoing general spine surgery, specifically posterior-approach spinal implant procedures (e.g., pedicle screw placement) in the lumbar region, where rigid anatomical structures (vertebrae) can be identified via fluoroscopic or CT imagery.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
Predicate Devices
Remex Spine Surgery Navigation System II (K233513)
K031156 — NAVITRACK SYSTEM-FLUOROSPINE, MODEL 900.005 · Orthosoft, Inc. · May 9, 2003
K212245 — Spine and Trauma Navigation System, Spine & Trauma 3D Navigation, Instrument Selection, Fluoro 3D, Registration Software Fluoro 3D, Registration Software Paired Point, Registration Software Spine Surface Matching, Spine Planning, Elements Screw Planning Spine, Elements Spine Screw Planning · Brainlab AG · Apr 13, 2022
K252950 — Paradigm System · Proprio, Inc. · Dec 15, 2025
K230783 — Anatase Spine Surgery Navigation System · Remex Medical Corp. · Apr 21, 2023
K211616 — ExcelsiusHub · Globus Medical, Inc. · Aug 25, 2021
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 29, 2025
Remex Medical Corp.
Cheng-Hsiung Wang
Official Correspondent
4F, No.9, Jingke Road, Nantun District
Taichung, 408209
Taiwan
Re: K251315
Trade/Device Name: Anatase Spine Surgery Navigation System and Remex Spine Surgery Navigation System II
Regulation Number: 21 CFR 882.4560
Regulation Name: Stereotaxic Instrument
Regulatory Class: Class II
Product Code: OLO
Dated: April 29, 2025
Received: April 29, 2025
Dear Cheng-Hsiung Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251315 - Cheng-Hsiung Wang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251315 - Cheng-Hsiung Wang
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali -S
Shumaya Ali, M.P.H.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251315 | |
| Device Name Anatase Spine Surgery Navigation System and Remex Spine Surgery Navigation System II | |
| Indications for Use (Describe) The Anatase Spine Surgery Navigation System and Remex Spine Surgery Navigation System II are indicated for precise positioning of surgical instruments or spinal implants during general spine surgery when reference to a rigid anatomical structure, such as the vertebra, can be identified relative to a patient's fluoroscopic or CT imagery. It is intended as a planning and intraoperative guidance system to enable open or percutaneous image guided surgery by means of registering intraoperative 2D fluoroscopic projections to pre-operative 3D CT imagery. Example procedures include: Posterior-approach spinal implant procedures, such as pedicle screw placement, with the lumbar region. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
REMEX
Remex Medical Corporation 4F, No. 9, Jingke Road, Nantun District, Taichung City408209, Taiwan
510(k) Summary
| Submitted By | REMEX MEDICAL CORP.
4F., No.9 Jingke Rd., Nantun District, Taichung, TW 408209
Tel. +886-4-23595336
Fax.+886-4-23598875 |
| --- | --- |
| Contact Person | Chegn-Hsiung Wang
Kevin816@mail.intai.com.tw |
| Prepared by | Jessie Wang
jessiewang@mail.intai.com.tw |
| Date Prepared | April 29th, 2025 |
| Device Name | Anatase Spine Surgery Navigation System and Remex Spine Surgery Navigation System II |
| Classification Name | Stereotaxic instrument |
| Regulation Number | 882.4560 |
| Product Codes | OLO |
| Device Class | Class II |
| Predicate Devices Information | K233513, Remex Spine Surgery Navigation System II (primary) |
| Device Description | The subject devices, Image calibrator-21F and Image calibrator-31F are designed for use with Anatase Spine Surgery Navigation System (K230783) and Remex Spine Surgery Navigation System II (K233513). The subject devices are modified from previously cleared devices to be compatible with the 21 cm/ 31 cm flat panel C-arm for CT image position and calibration.
The software upgrade is due to these newly added Image calibrators and one added function of auto-labeling for CT/C-arm image registration. |
| Indications for Use | The Anatase Spine Surgery Navigation System and Remex Spine Surgery Navigation System II, are indicated for precise positioning of surgical instruments or spinal implants during general spinal surgery when reference to a rigid anatomical structure, such as the vertebra, can be identified relative to a patient's fluoroscopic or CT imagery. It is intended as a planning and intraoperative guidance system to enable open or percutaneous image guided surgery by means of registering intraoperative 2D fluoroscopic projections to pre-operative 3D CT imagery.
Example procedures include:
Posterior-approach spinal implant procedures, such as pedicle screw placement, within the lumbar region. |
| Technological Characteristics | The subject devices have the same intended use, same indications for use, same fundamental technology, and similar design as the predicate devices, Anatase Spine Surgery Navigation System (K230783) and Remex Spine Surgery Navigation System II (K233513).
The difference between the subject devices and the predicate devices is the upgrade of the system software version, and new Image calibrator for |
K251315 - Page 1 of 2
{5}
REMEX
Remex Medical Corporation 4F, No. 9, Jingke Road, Nantun District, Taichung City408209, Taiwan
| | corresponding compatible devices. However, these modifications share the same function and fundamental technology with the predicate devices. | |
| --- | --- | --- |
| Performance Data | Verification and validation activities have been completed to provide sufficient assurance that the subject device meets the performance requirements under its indications for use conditions. Below is a summary of all performance tests which should be carried out on the subject device to demonstrated that the subject device performs as safely and effectively as the predicate device. | |
| | Test | Description |
| | Software | Software is verified and validated in accordance with FDA guidance for the content of premarket submissions for software contained in medical devices and IEC 62304. |
| | Risk Assessment | The effectiveness of all risk control measures is verified in accordance with ISO 14971. |
| | Design Verification | The design output fulfills all design input requirements. The test methods are the same as the previously cleared 510(k), K233513. |
| | The intended use and fundamental technology of the subject device are identical to the predicate devices and could share the same testing methods as the predicate devices. All the results are passed. Therefore, REMEX believes the testing demonstrated that the subject devices are as safe and effective as the predicate devices. | |
| Conclusion | The Anatase Spine Surgery Navigation System and Remex Spine Surgery Navigation System II and the predicate devices have the same intended use and have the same or similar technological characteristics. Moreover, the information contained in this submission demonstrates that any differences in their technological characteristics do not raise new issues of safety or effectiveness. Hence, The Anatase Spine Surgery Navigation System and Remex Spine Surgery Navigation System II is substantially equivalent to the predicate devices. | |
K251315 - Page 2 of 2
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.