K960957 · Sims Deltec, Inc. · FRN · Jun 4, 1996 · General Hospital
Device Facts
Record ID
K960957
Device Name
EXTERNAL POWER SOURCE (EPS) SYSTEM, POWER PACK
Applicant
Sims Deltec, Inc.
Product Code
FRN · General Hospital
Decision Date
Jun 4, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Indications for Use
The External Power Source (EPS) System is designed to provide an alternative source of power for the following Deltec Pumps: CADD-1*, CADD-PCA*, CADD-PLUS*, CADD-TPN*, and CADD-Prizm™ ambulatory infusion pumps.
Device Story
External Power Source (EPS) System provides rechargeable power for CADD ambulatory infusion pumps; replaces standard 9V batteries. System includes rechargeable power pack, AC adaptor, battery adaptor, notched battery door, and nickel-metal hydride (NiMH) battery pack. Operates within 2°C to 40°C; weight approximately 570g. Includes low battery alarm and ready indicator lights. Used in clinical or home settings by patients or clinicians to ensure continuous pump operation during infusion therapy.
Clinical Evidence
No clinical data. Functional testing conducted per 'Rechargeable Battery Preliminary Guidance for Data to be Submitted to the Food and Drug Administration in Support of Premarket Notification Applications' (January 1994). Device performed as intended and within specifications.
Technological Characteristics
Rechargeable power pack using nickel-metal hydride (NiMH) battery chemistry. Includes AC adaptor, battery adaptor, and notched battery door. Weight: ~570g. Operating temperature: 2°C to 40°C. Features low battery alarm and ready indicator lights. System is an external power accessory for ambulatory infusion pumps.
Indications for Use
Indicated for use as an alternative power source for Deltec CADD-1, CADD-PCA, CADD-PLUS, CADD-TPN, and CADD-Prizm ambulatory infusion pumps.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
External Power Source (EPS) System, Power Pack (nickel cadmium, NiCad)
Related Devices
K971234 — ALEXANDER MANUFACTURING CO. RECHARGEABLE BATTERY PART NUMBER GL825 · Alexander Mfg. Co. · May 27, 1997
K972779 — ALEXANDER MANUFACTURING CO. RECHARGEABLE BATTERY PART NUMBER MS3030-P · Alexander Mfg. Co. · Oct 20, 1997
K971865 — ALEXANDER MANUFACTURING CO. RECHARGEABLE BATTERY PART NUMBER M12/500-3P · Alexander Mfg. Co. · Jul 3, 1997
K012259 — MEDFUSION 2001 & 2010, MODEL N7215IWC · National Custom Ent., Inc. · Sep 12, 2001
{0}
K960957
510(k) Summary of Safety and Effectiveness
Date Prepared: March 4, 1996
JUN - 4 1996
## I. GENERAL INFORMATION
Applicant's name and address:
SIMS Deltec, Inc.
1265 Grey Fox Road
St. Paul, MN 55112
Contact:
David H. Short
Manager, Regulatory Affairs
(612)628-7214
## II. PRODUCT INFORMATION
Common/Usual Name: Battery Pack
Classification Name: Unknown
Proprietary Name: External Power Source (EPS) System, Alternate Power Pack (nickel-metal hydride)
## III. EQUIVALENCE DEVICE COMPARISON
SIMS Deltec's External Power Source (EPS) System, Power Pack (nickel cadmium, NiCad)
## IV. DEVICE DESCRIPTION
The External Power Source (EPS) System is comprised of a rechargeable power pack, AC adaptor, battery adaptor, notched battery door, and replacement battery pack for the power pack. The EPS System was developed as an alternate power source to 9 volt batteries for various CADD* pumps Deltec manufactures. The EPS System is designed for use with CADD-1* pump, CADD-PLUS* pump, CADD-PCA* (Models 5800 and 5800R) pumps, and the CADD-Prizm™ pump. An alternate rechargeable power pack with replaceable nickel-metal hydride (NiNH) batteries has been developed and the subject of this 510(k) Notification.
Page 1 of 3
{1}
510(k) Summary of Safety and Effectiveness
Date Prepared: March 4, 1996
## IV. DEVICE COMPARISON
Provided in Table 1 is a feature comparison of the EPS System Power Pack with NiMH batteries and the EPS System Power Pack with NiCad batteries.
Table 1. Comparison Chart
| | External Power Source (EPS) System
Alternate Power Pack | External Power Source (EPS) System
Power Pack |
| --- | --- | --- |
| Manufacturer | SIMS Deltec, Inc. | SIMS Deltec, Inc. |
| Power Pack Feature |
| Battery Chemistry | nickel-metal hydride | nickel cadmium |
| Rechargeable Power Pack | YES | YES |
| Replaceable Battery Pack | YES | YES |
| Low Battery Alarm | YES | YES |
| Ready Indicator Lights | YES | NO |
| Operating Temperature | 2°C to 40°C | 2°C to 40°C |
| Weight | ≈570 g | ≈570 g |
## V. INTENDED USE
The External Power Source (EPS) System is designed to provide an alternative source of power for the following Deltec Pumps: CADD-1*, CADD-PCA*, CADD-PLUS*, CADD-TPN*, and CADD-Prizm™ ambulatory infusion pumps.
## VI. SUMMARY OF STUDIES
### A. Functional Testing
Function testing was conducted using the alternate power pack of the External Power Source (EPS) System in accordance with the Rechargeable Battery Preliminary Guidance for Data to be Submitted to the Food and Drug Administration in Support of Premarket Notification Applications (January 1994). Based upon the testing that was conducted, the power pack with NiMH batteries performed as intended and within specification.
### B. Clinical Studies
Clinical studies were not deemed necessary.
Page 2 of 3
{2}
510(k) Summary of Safety and Effectiveness
Date Prepared: March 4, 1996
## VII. CONCLUSION
Based upon the information provided above, the alternate power pack developed for use in the External Power Source (EPS) System is suitable for its intended use, i.e., as an alternative source of power for the following Deltec Pumps: CADD-1*, CADD-PCA*, CADD-PLUS*, CADD-TPN*, and CADD-Prizm™ ambulatory infusion pumps.
Page 3 of 3
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.