K190182 · DEKA Research & Development · FRN · May 6, 2019 · General Hospital
Device Facts
Record ID
K190182
Device Name
Unity Subcutaneous Delivery System for Remodulin
Applicant
DEKA Research & Development
Product Code
FRN · General Hospital
Decision Date
May 6, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Unity Subcutaneous Delivery System for Remodulin® (the Unity System) is intended for continuous subcutaneous delivery of Remodulin (treprostinil) injection for use in adults (greater than 22 years of age).
Device Story
The Unity System is a wearable infusion pump for continuous subcutaneous delivery of Remodulin (treprostinil) to treat pulmonary arterial hypertension (PAH). It consists of a reusable pump, a disposable single-use cassette with a user-filled drug reservoir, a remote interface, and accessories. The device uses a micro-dosing pump mechanism supplemented by acoustic volume sensor (AVS) feedback to monitor and ensure delivery accuracy. It is used in professional healthcare facilities and home environments by patients or caregivers. The healthcare provider programs the infusion rate; the pump delivers the drug subcutaneously. The AVS technology provides real-time volume delivery confirmation. The system includes various alarms (e.g., occlusion, cassette depleted, battery low) to alert the user. By providing precise, continuous subcutaneous delivery of Remodulin, the device aims to manage PAH symptoms effectively while allowing patient mobility.
Clinical Evidence
Clinical study in healthy volunteers (n=not specified, ages 18-55) assessed in vivo infusion delivery accuracy of 0.9% sodium chloride and device reliability. Primary endpoints: volumetric flow rate accuracy vs. programmed rate and device logs. Results: Majority of data within accuracy specifications; brief intervals outside limits were attributed to measurement uncertainty or air bubbles. No evidence of environmental noise interference. No safety concerns identified.
Technological Characteristics
Wearable infusion pump; micro-dosing mechanism with acoustic volume sensor (AVS) feedback. Materials: Cassette (polycarbonate, bromobutyl, SEBS, polyurethane); Pump (ABS, polycarbonate, aluminum). Power: Rechargeable Li-ion battery (IEC 62133). Dimensions: 6cm x 6cm x 2cm; 50g. Connectivity: Wireless (coexistence tested per ANSI C63.27). Ingress protection: IP58. Sterilization: Radiation (ISO 11137). Software: Major level of concern (IEC 62304).
Indications for Use
Indicated for continuous subcutaneous delivery of Remodulin (treprostinil) injection in adults >22 years of age.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
K240256 — Remunity System · Deka Research and Development · Jun 12, 2024
K202690 — Remunity Pump for Remodulin (treprostinil) Injection · Deka Research and Development · Dec 30, 2020
K191313 — Unity Subcutaneous Infusion System for Remodulin · DEKA Research & Development · Feb 21, 2020
K241736 — Remunity 2.0 Pump for Remodulin (treprostinil) Injection · Deka Research and Development Corp. · Jan 16, 2025
K243354 — Remunity System · Deka Research & Development Corp. · Nov 27, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym followed by the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a solid blue square and the agency name in a lighter blue. The text reads "FDA U.S. FOOD & DRUG ADMINISTRATION".
May 6, 2019 DEKA Research & Development Brian Carney Project Regulatory Affairs Manager 340 Commercial Street Manchester, New Hampshire 03101
Re: K190182
Trade/Device Name: Unity Subcutaneous Delivery System for Remodulin Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion pump Regulatory Class: Class II Product Code: FRN Dated: January 31, 2019 Received: February 1, 2019
### Dear Brian Carney:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Tina Kiang, Ph.D Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190182
#### Device Name
Unity Subcutaneous Delivery System for Remodulin®
Indications for Use (Describe)
The Unity Subcutaneous Delivery System for Remodulin® (the Unity System) is intended for continuous subcutaneous delivery of Remodulin (treprostinil) injection for use in adults (greater than 22 years of age).
Type of Use (Select one or both, as applicable)
| <div> <object>X</object> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|---------------------------------------------------------------------------------|
| <div> <object> </object> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY - K190182
#### Submitter Information
| 510(k) Sponsor: | DEKA Research & Development<br>340 Commercial Street<br>Manchester, NH 03101 |
|-----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Brian Carney<br>Project Regulatory Affairs Manager<br>DEKA Research & Development Corporation<br>Phone: (603) 669-5139<br>Fax: (603) 624-0573<br>bcarney@dekaresearch.com |
| Date Prepared: | May 3, 2019 |
#### Proposed Device
| Common/Usual Name: | Infusion Pump |
|-------------------------|---------------------------------------------------|
| Trade/Proprietary Name: | Unity Subcutaneous Delivery System for Remodulin® |
| Classification Name: | Infusion Pump |
| Device Classification: | 880.5725 |
| Product Code: | FRN |
| Class | II |
| Device Panel: | General Hospital Devices Branch |
### Predicate Device
CADD-MS3 Ambulatory Infusion Pump (Smiths Medical) K051568. No reference devices are used in this submission.
## Device Description
The Unity Subcutaneous Delivery System for Remodulin® (Unity System) is a wearable infusion pump designed to deliver Remodulin® for the treatment of pulmonary arterial hypertension (PAH). It is intended for continuous subcutaneous delivery of FDA-approved Remodulin® (treprostimil) (hereinafter referred to as 'Remodulin (treprostinily), NDA 021272. The Unity System consists of several components: a wearable pump assembly, a remote interface, a filling and priming aid, and accessories (e.g., rechargeable batteries, battery charging cable, power adapter). A commercially available subcutaneous infusion set is connected to the pump assembly via a standard Luer connector for the delivery of Remodulin from the Unity System to the patient.
The pump assembly is composed of a reusable pump and a disposable single-use cassette with a user-filled drug reservoir, which infuses Remodulin subcutaneously into the patient based on an individualized programmed rate. Each disposable cassette may be used for up to 72 hours after attachment to the pump. The Unity System utilizes a micro-dosing pump mechanism supplemented with acoustic volume sensor feedback to ensure delivery accuracy.
The device is prescription use only.
{4}------------------------------------------------
# Indications for Use
The Unity Subcutaneous Infusion System (the Unity System) is intended for continuous subcutaneous delivery of Remodulin® (treprostinil) Injection for use in adults (greater than 22 years of age).
# Substantial Equivalence Discussion
## Intended Use Comparison
The table below includes a comparison of the intended use between the new device and those of the predicate device:
| Characteristic | Predicate<br>CADD-MS 3 Ambulatory<br>Infusion Pump<br>K051568 | Proposed<br>Unity Subcutaneous Deliver System<br>for Remodulin®<br>K190182 | |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The CADD-MS 3 Ambulatory<br>Infusion Pump is designed for<br>subcutaneous, intravenous, epidural<br>and intrathecal infusion of<br>medication. The Smiths Medical<br>MD, Inc, 3-ml Cartridge Reservoir<br>is designed for use with the CADD-<br>MS 3 for delivering medication | The Unity Subcutaneous Delivery System<br>for Remodulin® (the Unity System) is<br>intended for continuous subcutaneous<br>delivery of Remodulin (treprostinil)<br>injection for use in adults (greater than 22<br>years of age). | |
| Prescription Only or<br>Over the Counter | Prescription Only | Prescription Only | |
| Intended Population | Not Specified | Adults (>22 years of age) | |
| Environment of Use | In hospital or out-patient settings | In professional healthcare facility<br>and home healthcare environments | |
| Characteristic | Predicate Device<br>K051568 | Subject Device<br>K190182 | Discussion of Differences |
| Mechanism of<br>action | Microprocessor based<br>electronic syringe pump | Microprocessor controlled<br>Micro- dosing pump<br>mechanism supplemented<br>with acoustic volume sensor<br>(AVS) feedback for<br>monitoring delivery accuracy. | AVS technology provides<br>accuracy confirmation of the<br>volume of fluid delivered. This<br>was supported with clinical and<br>non-clinical testing. |
| Infusion<br>Accuracy | ±3% (nominal). | ±6% | Compliant with Remodulin<br>USPI an infusion accuracy of<br>±6% satisfies the Remodulin<br>delivery accuracy requirement<br>for administration by<br>subcutaneous injection |
| Maximum<br>Infusion pressure | 23 psi (159 kPa) | <16.4 psi (<113 kPa) | The maximum infusion pressure<br>is less that the predicate;<br>however, the infusion accuracy<br>was verified through<br>performance testing and<br>validated to be appropriate for<br>the intended use. |
| Programmable<br>Flow rate ranges | 0.000 mL/hr to 1.000 mL/hr<br>with increments of .002<br>mL/hr | 16 µl/hr to 225 µl/hr with<br>increments of 1 µl/hr | Infusion accuracy was<br>maintained within specification<br>across flow rates that span the<br>programmable flow rates of the<br>device through bench testing<br>and is appropriate based on the<br>intended use. |
| Time to occlusion<br>alarm | Maximum time to occlusion<br>alarm:<br>12 min. to 15 min. at 124<br>µl/hr (intermediate rate)<br>9.5 hr. to 19 hr. at 2 µl/hr<br>(minimum rate) | Maximum time to occlusion<br>alarm:<br><12 min. at rates ≥ 100 µl/hr<br>within 8 hr, at rates < 100<br>µl/hr | Time to occlusion is similar to<br>the predicate. Differences were<br>evaluated through a risk based<br>approach to be appropriate for<br>the intended use. |
| Post-occlusion<br>bolus | At minimum rate:<br>approximately 40 µl | <40 µl at all rates. | The range for the Unity System<br>is tailored for the delivery of<br>Remodulin. These differences<br>do not raise different questions |
| Characteristic | Predicate Device<br>K051568 | Subject Device<br>K190182 | Discussion of Differences |
| | at intermediate rate:<br>approximately 210 µl | | of safety and effectiveness and<br>have been verified through<br>performance testing. |
| Alarms & Alerts | • Battery depleted<br>• Battery low<br>• Blockage detected<br>• Cartridge empty<br>• Cartridge removed<br>• Cartridge very low<br>• Depletes soon<br>• Edit not saved<br>• Key stuck<br>• Program defaulted<br>• Pump stopped<br>• Site change reminder<br>• System fault | • Battery depleted<br>• Battery Low (pump)<br>• Battery Low (remote)<br>• Cassette Depleted<br>• Cassette Problem<br>• Cassette Removed<br>• Depletes Soon<br>• Pump Error<br>• Pump Failure<br>• Occlusion<br>• Delivery Stopped<br>• Basal Not Started<br>• Idle<br>• Software Version Error<br>• Tech<br>• Excessive Noise<br>• No Communication<br>• Message Timeout<br>• Pairing Failed<br>• Walkaway | Alarms were validated based on<br>the intended use of the device<br>and the unique technological<br>characteristics. |
| Device Service<br>Life | Not stated | 3 years | The device was evaluated<br>through performance testing to<br>meet the essential performance<br>throughout the service life. |
| Dimensions &<br>Weight | 3.2 in x 1.8 in x 0.95 in<br>Approximately 3.2 oz<br>including battery and<br>cartridge | 6 cm x 6 cm x 2 cm<br>(2.4 in x 2.4 in x 0.4 in)<br>50 g (1.76 oz) | The devices are similar in size<br>and intended to be ambulatory. |
| Materials | Cartridge: Polypropylene<br>(plunger and other materials<br>unknown) | Cassette fluid path:<br>Polycarbonate, Bromobutyl,<br>SEBS, polyurethane<br><br>Pump: ABS, Polycarbonate,<br>Aluminum<br><br>Cartridge: Polycarbonate,<br>Acrylic, polyurethane | The appropriate<br>biocompatibility and toxicology<br>testing has been conducted on<br>all materials found in the fluid<br>path. Non-clinical testing<br>supported the differences in<br>materials. The differences in the<br>materials do not raise different<br>questions of safety and<br>effectiveness. |
| Characteristic | Predicate Device<br>K051568 | Subject Device<br>K190182 | Discussion of Differences |
| Environment of Use | In hospital or out-patient settings | In professional healthcare facility and home healthcare environments | Same |
| Ingress protection | IPX8 when pump's labels and outer shell are intact | IP58 when connected to the reservoir | The Unity System provides protection against particle ingress in addition to moisture protection. This is appropriate based on the intended use of the device. |
| Power source | 1 AAA alkaline battery | Rechargeable Lithium-Ion Battery | The differences were evaluated through performance testing; Lithium-Ion Battery complies with IEC 62133. |
| Storage Conditions | Temperature:<br>-4°F to 140°F (-20°C to 60°C)<br>Non-condensing humidity:<br>up to 90%<br>Pressure: 700 hPa to 1060 hPa | Temperature:<br>-13°F to 158°F (-25°C to 70°C)<br>Non-condensing humidity:<br>up to 90%.<br>Pressure: 500 hPa to 1060 hPa | The differences in storage conditions were evaluated through performance testing. |
| Operating Conditions | Temperature:<br>35.6°F to 104°F (2°C to 40°C)<br>Non-condensing humidity:<br>up to 90%<br>Pressure: 700 hPa to 1060 hPa | Temperature:<br>41°F to 104°F (5°C to 40°C)<br>Non-condensing humidity:<br>up to 90%<br>Pressure: 700 hPa to 1060 hPa | The operating conditions are similar. The differences in operating conditions were evaluated through performance testing. |
| Accessories | | | |
| Remote user feedback | Audible, vibratory | Audible, vibratory | Same |
| Administration Set | Unomedical Comfort Infusion Set | Medtronic Quick-set Infusion Set Medtronic Silhouette and Infusion Set Smiths Medical Cleo 90 Infusion Set | Similar; compatibility testing was completed to verify use with the designated infusion sets |
| Cartridge | Smiths Medical, Inc. 3 ml Medication Cartridge, | Remodulin Unity cassettes, 3 ml, User filled | Both cartridges can be user filled and are appropriate for the design of the infusion pump. |
## Discussions of differences in Indications for Use statement
The indications for use statement for the Unity System is not identical to the predicate is indicated for four different delivery types, including subcutaneous delivery, whereas the Unity System limits delivery to subcutaneous infusion. Additionally, the predicate device does not specify a particular drug or set of drugs. The Unity System is labeled only for use with Remodulin. The Unity System's is a specific subset of the predicate device's indications for use and is therefore not a new intended use. Additionally, compatibility with Remodulin specifically has been verified through performance testing.
## Discussions of differences in intended population
The Unity system is indicated for adults (greater than 22 years of age) whereas the predicate does not specify an indicated age group. The age group for the subject device was validated in the Human Factors testing.
#### Discussions of differences in environment of use
The environment of use for the subject device is identical to the predicate device.
{5}------------------------------------------------
# Comparison of Technological Characteristics with the Predicate Device
Table 1 compares the characteristics of the predicate device and includes an assessment of differences between them and why the difference between the subject device and predicate device do not introduce new or different questions of safety and effectiveness.
### Table 1. Comparison of Predicate and Proposed Devices
{6}------------------------------------------------
{7}------------------------------------------------
# Non-Clinical/ Performance Testing:
The following performance data/non-clinical testing was provided in support of the substantial equivalence determination for the Unity System.
{8}------------------------------------------------
A safety assurance case was provided for the Unity Subcutaneous Delivery System for Remodulin, as recommended in the FDA guidance document, Infusion Pumps Total Product Life Cycle.
The stated goal of the safety assurance case is: "The Unity Remodulin Subcutaneous Infusion System is adequately safe for its intended use"
The assurance case defined the device system, including the indications for use, system definition, operational description, patient populations, and use environments. The supporting assurance arguments covered the following attributes:
- All hazards associated with the system have been identified and adequately addressed .
- Device reliability is adequate ●
- . The device design requirements are adequately verified and validated
The following specific evidence was included within the assurance case to demonstrate that the subject device is verified and validated for its intended use and to demonstrate substantial equivalence to the predicate devices:
| Software &<br>Cybersecurity | Software documentation is included according to FDA's Guidance for the<br>Content of Premarket Submissions for Software Contained in Medical Devices<br>for Major level of concern for the software embedded in the Unity System. Software validation was conducted according to IEC 62304 and the FDA<br>guidance document General Principles of Software Validation – Final Guidance<br>for Industry and FDA Staff. Cybersecurity risks were assessed and documentation is included based on the<br>FDA's Guidance of Premarket Submissions for Management of Cybersecurity in<br>Medical Device. AAMI ANSI IEC 62304:2006 Medical Device Software – Software Life Cycle<br>Process |
|-----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Electrical Safety | AAMI/ANSI ES 60601-1:2005/(R)2012 + A1:2012, C1:2009/(R)2012 +<br>A2:2010/(R) 2012 Medical electrical equipment – Part 1: General requirements<br>for basic safety and essential performance |
| EMC | IEC 60601-1-2:2014 Medical Electrical Equipment - Part 1: General<br>Requirements for Safety-2-Collateral Standard: Electromagnetic Compatibility-<br>Requirements & Tests |
| Wireless Coexistence<br>& RF Wireless testing | FDA Guidance Document "Radio Frequency Wireless Technology in Medical<br>Devices - Guidance for Industry and Food and Drug Administration Staff" ANSI C63.27-2017 American National Standard for Evaluation of Wireless<br>Coexistence |
| Accessory<br>compatibility | Verification of the pump essential performance was completed with labeled<br>administration sets |
| | |
| Device performance | ● The essential performance requirements of the device were verified through<br>performance testing in accordance with the intended use of the device and in<br>accordance with the FDA Guidance “Infusion Pumps Total Product Life Cycle” |
| Battery testing | ● Li-ion battery safety successfully tested per IEC 62133 |
| Human Factors | ● Human factors studies per the FDA Guidance Applying Human Factors and<br>Usability Engineering to Medical Devices (February 3, 2016). The human<br>factors studies were conducted with the intended user population, use<br>environment and use scenarios to simulate clinical conditions. Results of the<br>human factors testing demonstrate validation of the device per the intended use. |
| Reprocessing,<br>Cleaning, Sterility | ● ISO 11137-1:2006 + A1:2013 Sterilization of health care products - Radiation -<br>Part 1: Requirements for development, validation and routine control of a<br>sterilization process for medical devices<br>● ISO 11607-1:2006 + A1:2014 Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems<br>● Validation per the FDA Guidance for Reprocessing Medical Devices in Health<br>Care Settings: Validation Methods and Labeling (March 17, 2015) confirmed<br>cleaning and disinfection instruction provided in instructions for use |
| Biocompatibility | ● The materials used for the Unity System comply with biocompatibility<br>requirements outlined in ISO 10993-1:2009 and the Guidance for Industry and<br>Food and Drug Administration Staff, Use of International Standard ISO 10993<br>Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within<br>a Risk Management Process and are considered to be biocompatible.<br>● Testing was conducted for cytotoxicity, sensitization, irritation, pyrogenicity,<br>and hemolysis. The acute systemic toxicity, subchronic systemic toxicity, and<br>genotoxicity endpoints were addressed through the extractables/leachables study<br>with a toxicological risk assessment.<br>● AAMI ST72: 2011 Bacterial Endotoxins – Test methods, routine monitoring, and<br>alternatives to batch testing<br>● USP 39 <788>:2016 Particulate Matter in Injections |
{9}------------------------------------------------
# Clinical Testing
A clinical study was conducted to support the substantial equivalence of the device and to clinically validate the Acoustic Volume Sensor (AVS) design. The Performance Assessment of The Unity Infusion System with 0.9% Sodium Chloride Injection in Healthy Volunteers study was conducted to establish evidence that the AVS measurement process consistently produces results within an acceptable, predetermined measurement uncertainty. All participants were healthy subjects between 18 and 55 years of age, inclusive, at the time of informed consent.
## Objective
The objective of this study was to assess the performance of the Unity System to deliver 0.9% sodium chloride (normal saline [NS]) by subcutaneous (SC) infusion to healthy volunteers by assessing the following:
{10}------------------------------------------------
- In vivo infusion delivery accuracy assessed by measured volumetric flow rate compared to programmed flow rate of the Unity System to demonstrate environmental noise of other elements of daily use interfered with the ability of the Unity System to deliver accurately when participants were challenged to loud sound exposures.
- . In vivo the Unity System reliability assessed by device history logs (flow rate, malfunctions, etc.) and subject diaries (environmental exposures)
#### Criteria for Evaluation
Pump Performance: In vivo infusion delivery accuracy was assessed by measured volumetric flow rate compared to programmed flow rate of the Unity System. In vivo Unity System reliability was assessed by device history logs (flow rate, malfunction, complications, etc.) and subject diaries (environmental exposures).
Safety: Per protocol, all adverse events (AEs), serious adverse events (SAEs), and unanticipated adverse device effects (UADEs) were to be documented from Baseline until the subject was either discontinued from the study or all End of Infusion study assessments were complete. All AEs associated with a device complication were to be investigated.
#### Conclusions
Pump Performance: The study successfully demonstrated that the majority of measured data fell within the accuracy specification. There were brief intervals (less than six how rate measured was outside the performance specification limit for under delivery (-6%). Each of these intervals were within the measurement uncertainty of the measure equipment used in the study or associated with the delivery of air bubbles, and all were within 25%. The study found no evidence of environmental noise or other elements of daily use interfered with the ability of the Unity System to deliver accurately.
Safety: No safety concerns related to use of the Unity System were identified during this study.
#### Conclusion
Differences between the intended use and technological characteristics of the subject device compared to the predicate do not raise different questions of safety and effectiveness. The device is supported by non-clinical testing, clinical testing, and risk management activities. The Unity Subcutaneous Delivery System for Remodulin is Substantially Equivalent (SE) to the CADD-MS 3 Ambulatory Infusion Pump, cleared under K051568.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.