MEDFUSION MODEL 2001 SYRINGE INFUSION PUMP
Device Facts
| Record ID | K955231 |
|---|---|
| Device Name | MEDFUSION MODEL 2001 SYRINGE INFUSION PUMP |
| Applicant | Medex, Inc. |
| Product Code | FRN · General Hospital |
| Decision Date | Jul 18, 1996 |
| Decision | ST |
| Submission Type | Traditional |
| Regulation | 21 CFR 880.5725 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
The intended medical application of Medex Medfusion 2000 Series syringe pumps is to produce controlled movement of the plunger of a syringe to inject a set amount of therapeutic fluid into a patient within a hospital setting at a set rate and at set times.
Device Story
Medfusion 2001, 2010, & 2010i syringe pumps; hospital-use infusion devices. Function: controlled plunger movement for precise therapeutic fluid delivery. Modification: replacement of microprocessor/memory chip with drop-in microprocessor-EPROM daughter board; associated software memory address updates. Operation: identical to predicate models. Output: controlled fluid infusion. Benefit: consistent, timed medication delivery.
Clinical Evidence
Bench testing only; no clinical data provided.
Technological Characteristics
Syringe pump; plastic exterior; no fluid/drug contacting surfaces. Energy source: same as predicate. Microprocessor/EPROM daughter board hardware. Software: modified memory addresses to accommodate hardware change.
Indications for Use
Indicated for patients requiring controlled infusion of therapeutic fluids via syringe pump in a hospital setting.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
- Medex Medfusion 2001 (previous model)
- Medex Medfusion 2010 (previous model)
- Medex Medfusion 2010i (previous model)
Related Devices
- K031185 — ABBOTT PLUM A+ INFUSION PUMP WITH A HPL/RS PLUG-AND-PLAY MODULE · Abbott Laboratories · Apr 30, 2003
- K033425 — AMBIT INTERMITTENT AMBULATORY INFUSION PUMP · Sorenson Medical, Inc. · Nov 6, 2003
- K033325 — AMBIT CONTINUOUS AMBULATORY INFUSION PUMP · Sorenson Medical, Inc. · Nov 6, 2003
- K982640 — MEDEX 3000 SERIES SYRINGE INFUSION PUMP · Medex, Inc. · Apr 6, 1999
- K041191 — COLLEAGUE VOLUMETRIC INFUSION PUMPS · Baxter Healthcare Corp · Jun 25, 2004