SIGNATURE EDITION INFUSION SYSTEM, MODELS 710X, 720X

K070268 · Cardinal Health 303,Inc. · FRN · Feb 23, 2007 · General Hospital

Device Facts

Record IDK070268
Device NameSIGNATURE EDITION INFUSION SYSTEM, MODELS 710X, 720X
ApplicantCardinal Health 303,Inc.
Product CodeFRN · General Hospital
Decision DateFeb 23, 2007
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 880.5725
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Signature Edition® Infusion System is intended for use in today's growing professional healthcare environment including healthcare facilities, home care, and medical transport that utilize infusion systems for the delivery of fluids, medications, blood and blood products. The Signature Edition® Infusion System is indicated for continuous or intermittent delivery through clinically acceptable routes of administration such as intravenous (IV), intra-arterial (IA), subcutaneous, epidural, enteral, or irrigation of fluid spaces.

Device Story

Electrical volumetric infusion pump; controls flow rate/monitors delivery of drugs, fluids, blood products. Used in professional healthcare, home care, medical transport. Operated by clinicians/trained personnel. Provides continuous or intermittent infusion. Benefits patient by enabling precise administration of cardiovascular drugs, chemotherapy, anesthesia, and blood transfusions. Device modification subject of this 510(k).

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Electrical volumetric infusion pump. Fundamental scientific technology and materials unchanged from predicate. Standalone device.

Indications for Use

Indicated for continuous or intermittent delivery of fluids, medications, blood, and blood products via IV, IA, subcutaneous, epidural, enteral, or irrigation routes in healthcare facilities, home care, and medical transport settings.

Regulatory Classification

Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ CARDINAL HEALTH, ALARIS PRODUCTS K070268 # ATTACHMENT D FEB 2 3 2007 # SUMMARY OF SAFETY AND EFFECTIVENESS Cardinal Health Alaris Products #### Signature Edition® Infusion System Models 710X and 720X # SUBMITTER'S NAME: Cardinal Health 303, Inc. dba Cardinal Health, Alaris Products 10221 Wateridge Circle San Diego, CA 92121-2772 (858) 458-7830 (858) 458-6114 FAX CONTACT PERSON: Stacy L. Lewis Principal Regulatory Affairs Specialist DATE PREPARED: January 26, 2007 Proprietary Name: DEVICE NAME: Signature Edition® Infusion System Common Name: Infusion Pump Classification Name: Pump, Infusion, FRN (880.5725) PREDICATE DEVICE: Signature Edition® Infusion System (K931549), October 12, 1993 # DEVICE DESCRIPTION The subject of this Special 510(k) submission is the Signature Edition® Infusion System Models 710X and 720X. This device is essentially the same as the originally submitted predicate device, the Signature Edition® Infusion System (submitted as the Models EZ 1 and EZ 2 Infusion Pumps, K931549, October 12, 1993), with the exception of the device {1}------------------------------------------------ # SUMMARY OF SAFETY AND EFFECTIVENESS Page 2 of 3 modification provided in this submission. Other prior submissions include K032147 (Signature Edition Infusion System), which covered minor aggregate changes to the Signature Edition Infusion System and included all Signature Edition models (700X, 710X, 720X, 713X and 723X). The same device modification provided in this submission was also submitted for the Signature Edition® Gold Infusion System, Models 713X and 723X (K063288) and was cleared to market on November 28, 2006. The electrical volumetric pumps are used to control the rate or monitor the flow of solution or medication for delivery of drugs, fluids, and blood products. Infusion pumps have been proven to be useful in applications such as continuous epidural anesthesia, administration of IV cardiovascular drugs, chemotherapy, and blood transfusions. #### SUBSTANTIAL EQUIVALENCE With the exception of the device modification presented in this submission, the Signature Edition® Infusion System is essentially the same as the originally submitted predicate device. The intended use, principles of operation, fundamental scientific technology, method of manufacture, and application are essentially the same. #### INTENDED USE The intended use of this device has not changed from the original submissions in terms of content or intent: The Signature Edition® Infusion System is intended for use in today's growing professional healthcare environment including healthcare facilities, home care, and medical transport that utilize infusion systems for the delivery of fluids, medications, blood and blood products. 070268 00171 {2}------------------------------------------------ # SUMMARY OF SAFETY AND EFFECTIVENESS Page 3 of 3 The Signature Edition® Infusion System is indicated for continuous or intermittent delivery through clinically acceptable routes of administration such as intravenous (IV), intra-arterial (IA), subcutaneous, epidural, enteral, or irrigation of fluid spaces. ### TECHNOLOGICAL CHARACTERISTICS A comparison of the technological characteristics of the Signature Edition® Infusion System and the predicate devices has been performed. The results of this comparison demonstrate that the modified Signature Edition® Infusion System is equivalent in technological characteristics to the unmodified device and that the fundamental scientific technology of the predicate device has not been altered. 070268 {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" are arranged in a circular pattern around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 #### FEB 2 3 2007 Ms. Stacey L. Lewis Principal Regulatory Affairs Specialist Cardinal Health 303, Incorporated 10221 Wateridge Circle San Diego, California 92121-2772 Re: K070268 Trade/Device Name: Signature Edition® Infusion System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: FRN Dated: January 26, 2007 Received: January 29, 2007 Dear Ms. Lewis: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Ms. Lewis Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Attachment B INDICATIONS FOR USE 510(k) Number: K 070268 (To Be Assigned By FDA) Device Trade Names: Signature Edition® Infusion System #### Indications For Use: The Signature Edition® Infusion System (Models 710X and 720X) is intended for use in today's growing professional healthcare environment including healthcare facilities, home care, and medical transport that utilize infusion systems for the delivery of fluids, medications, blood and blood products. The Signature Edition® Infusion System is indicated for continuous or intermittent delivery through clinically acceptable routes of administration such as intravenous (IV), intra-arterial (IA), subcutaneous, epidural, enteral, or irrigation of fluid spaces. Prescription Use (Per 21 CFR 801.109) Over-The-Counter Use PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) OR Concurrence of CDRH, Office of Device Evaluation (ODE) Anthony V. Watson Anesthesiology, General Hospital, Anesthesia, Medical Services K970268 65166
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...