ALARIS PC UNIT, MODELS 8000 AND 8015

K091308 · Cardinal Health 303,Inc. · FRN · Jul 9, 2009 · General Hospital

Device Facts

Record IDK091308
Device NameALARIS PC UNIT, MODELS 8000 AND 8015
ApplicantCardinal Health 303,Inc.
Product CodeFRN · General Hospital
Decision DateJul 9, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5725
Device ClassClass 2

Indications for Use

The Alaris® PC Unit is the main user interface unit and power supply of the Alaris® System, a modular system to be used with Alaris® System modules (aka Medley™ System modules) intended for use in today's growing professional healthcare environment for facilities that utilize infusion and/or monitoring devices. The specific intended use for each Alaris® System module is specified in its respective submission.

Device Story

Alaris® PC Unit acts as central user interface and power supply for modular Alaris® System (Medley™) infusion and monitoring modules; used in professional healthcare environments. Device modification includes upgraded processor, increased memory, and color screen. Operates as host for various infusion/monitoring modules; provides power and UI for clinician interaction. Output displayed on screen for clinician monitoring and control of infusion/monitoring parameters. Benefits include improved processing speed and visual interface for clinical staff managing patient therapy.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Modular infusion/monitoring system component. Hardware includes upgraded processor, increased memory, and color display. Functions as power supply and user interface for Alaris® System modules. Fundamental scientific technology remains consistent with predicate device.

Indications for Use

Indicated for use in professional healthcare facilities utilizing infusion and/or monitoring devices. Intended to serve as the primary user interface and power supply for modular Alaris® System infusion and monitoring components.

Regulatory Classification

Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ CARDINAL HEALTH KC09/308 SPECIAL 510/k ALARIS® PC UNIT ## ATTACHMENT D ## SUMMARY OF SAFETY AND EFFECTIVENESS #### Cardinal Health 303, Inc. Alaris® PC Unit Models 8000 and 8015 JUL - 9 2009 | SUBMITTER'S NAME: | Cardinal Health 303, Inc. | |-------------------|-----------------------------------------------------| | | 10020 Pacific Mesa Blvd. | | | San Diego, CA 92121 | | | (858) 617-5925 | | | (858) 617-5977 FAX | | CONTACT PERSON: | Michelle J. Badal, RAC | | | Vice President, Regulatory Affairs | | DATE PREPARED: | May 1, 2009 | | DEVICE NAME: | Proprietary Name: Alaris® PC Unit | | | Common Name: Infusion Pump | | | Classification Name: Pump, Infusion, FRN (880.5725) | | PREDICATE DEVICE: | Alaris® PC Unit (K051641, October 20, 2005) | ## DEVICE DESCRIPTION The Alaris® System is a modular system that consists of a point-of-care unit (PC Unit) that provides the main user interface and power supply for the associated infusion and monitoring modules. The enhanced PC Unit will include a faster processor, increased memory, and a color screen. This update is only for the Alaris® PC Unit and does not require any change to the associated modules, systems, or accessories of the Alaris® System. {1}------------------------------------------------ ### SUMMARY OF SAFETY AND EFFECTIVENESS Page 2 of 2 #### SUBSTANTIAL EQUIVALENCE With the exception of the device modification presented in this submission, the Alaris® PC Unit is essentially the same as the originally submitted predicate device. The intended use, principles of operation, fundamental scientific technology, method of manufacture, and application are essentially the same. #### INTENDED USE The intended use of this device has not changed from the original submissions in terms of content or intent: The Alaris® PC Unit is the main user interface unit and power supply of the Alaris® System, a modular system to be used with Alaris® System modules (aka Medley " System modules) intended for use in today's growing professional healthcare environment for facilities that utilize infusion and/or monitoring devices. The specific intended use for each Alaris® System module is specified in its respective submission ## TECHNOLOGICAL CHARACTERISTICS A comparison of the technological characteristics of the Alaris® PC Unit and the predicate devices has been performed. The results of this comparison demonstrate that the modified Alaris® PC Unit is equivalent in technological characteristics to the unmodified device and that the fundamental scientific technology of the predicate device has not been altered. . {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle with its wings spread, symbolizing protection and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The seal is simple and monochromatic. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUL - 9 2009 Ms. Michelle J. Badal Vice President Regulatory Affairs Cardinal Health 303, Incorporated 10020 Pacific Mesa Boulevard San Diego, California 92121 Re: K091308 Trade/Device Name: Alaris® PC Unit Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: FRN Dated: June 26, 2009 Received: July 1, 2009 Dear Ms. Badal: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ #### Page 2- Ms. Badal Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrl/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. Anthony D. Watson Susan Runner, D.D.S., M.A. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Attachment B INDICATIONS FOR USE K091308 510(k) Number (if known): Device Name: Alaris® PC Unit #### Indications for Use: The Alaris® PC Unit is the main user interface unit and power supply of the Alaris® System, a modular system to be used with Alaris® System modules (aka Medley™ System modules) intended for use in today's growing professional healthcare environment for facilities that utilize infusion and/or monitoring devices. The specific intended use for each Alario® System module is specified in its respective submission. X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: K091308 00069
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...