PLUM 360 INFUSION SYSTEM WITH HOSPIRA MEDNET, SMART CARD PLUG N PLAY CE 3.0 MODULE FOR PLUM A+ INFUSION SYSTEM UPGRADE
K141789 · Hospira, Inc. · FRN · Jan 16, 2015 · General Hospital
Device Facts
Record ID
K141789
Device Name
PLUM 360 INFUSION SYSTEM WITH HOSPIRA MEDNET, SMART CARD PLUG N PLAY CE 3.0 MODULE FOR PLUM A+ INFUSION SYSTEM UPGRADE
Applicant
Hospira, Inc.
Product Code
FRN · General Hospital
Decision Date
Jan 16, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Plum 360™ Infusion System with Hospira MedNet™ and Plum A+™ (360 Enabled™) Infusion System are indicated for use in parenteral, enteral and epidural therapies and the administration of whole blood and blood products.
Device Story
Volumetric infusion system; delivers parenteral, enteral, epidural therapies and blood products via dedicated administration sets. User interface includes LCD, numeric keypad, audible/visual indicators, nurse call interface, and drug library. Hospira MedNet™ software enables communication between pump, server, and facility systems. Operated by clinicians in healthcare settings. Enhancements include CPU software v15.0 (drug library, autoprogramming, alarm management), connectivity engine CE 3.0 (improved Wi-Fi/throughput), and ergonomic hardware (enclosure, tubing guides, ratcheting pole clamp). Smart Card Plug 'n' Play module allows Plum A+™ upgrades. Output provides controlled fluid delivery; assists clinical decision-making via drug library safety limits and automated programming. Benefits include improved infusion safety, connectivity, and workflow efficiency.
Clinical Evidence
No clinical data. Bench testing only. System verification and validation confirmed design inputs met. Human factors studies conducted to validate use-related error mitigations with passing results.
Indicated for patients requiring parenteral, enteral, or epidural therapies and administration of whole blood products.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
Plum A+™ Infusion System with Hospira Mednet™ (K042081)
Lifeshield Vision Infusion Set with Pre-pierced Reseal (K113683)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a human figure, represented by three overlapping profiles, with the head tilted upwards.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 16, 2015
Hospira, Incorporated Yuliva Matlin. M.S., M.B.A. Associate Director of Global Regulatory Affairs 275 North Field Drive Lake Forest, Illinois 60045
Re: K141789
Trade/Device Name: Plum 360™ Infusion System with Hospira Mednet™, Smart Card Plug 'N Play CE 3.0 Upgrade Module for Plum A+TM Infusion System, and Plum Administration Sets Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: Class II
Product Code: FRN, PHC, FPA Dated: December 15, 2014 Received: December 17, 2014
Dear Ms. Malin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# Tina Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K141789
Device Name
Plum 360 Infusion System with Hospira Mednet Smart Card Plug 'N Play CE 3.0 Upgrade Module for Plum A+ Infusion System Plum Administration Sets
Indications for Use (Describe)
Plum 360™ Infusion System with Hospira MedNet™ and Plum A+™ (360 Enabled™) Infusion System are indicated for use in parenteral, enteral and epidural therapies and the administration of whole blood products.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Hospira logo. The logo features the company name "Hospira" in a sans-serif font, with the "H" slightly larger than the other letters. Above the name is an abstract graphic consisting of three curved lines that converge at a point, creating a sense of movement or flow.
# 510(k) Summary of Safety and Effectiveness
## K141789
| Submitter Information | |
|---------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | Hospira, Inc. |
| Address | 275 N Field Drive, Lake Forest, IL 60045 USA |
| Phone Number | 224-212-4857 |
| Fax Number | 224-212-5402 |
| Establishment Registration Number | 3005579246 |
| Name of Contact Person | Yuliya Matlin, M.S., M.BA |
| Date Prepared | June 30, 2014 |
| Name of Device | |
| Trade or Proprietary Name | Plum 360TM Infusion System with Hospira MedNetTM |
| | Smart Card Plug'n' Play Module with CE 3.0 for Plum A+TM Infusion System |
| | PlumTM Administration Sets |
| Common or Usual Name | Infusion Pumps, Infusion safety management software &<br>Intravascular Administration Sets |
| Classification Name | Infusion Pumps, Infusion safety management software &<br>Intravascular Administration Sets |
| Classification Panel | 80 |
| Regulation(s) | 21 CFR 880.5725 & 21 CFR 880.5440 |
| Product Code(s) | FRN, PHC, FPA |
| Legally Marketed Device(s) To Which<br>Equivalence is Claimed | Infusion System:<br>Plum A+TM Infusion System with Hospira MednetTM (K042081)<br>Administration Sets:<br>Abbott Plum A+ Pump and Sets (K982159)<br>Lifeshield Latex-Free Microbore Extentsion Set, Model 14949 and<br>Others (K052722)<br>Hospira Enteral Feeding Set, Model 20640 (K061432)<br>Hospira Infusion Blood Sets (K101677)<br>Hospira Plum Infusion Set; Hospira Infusion Set with Yellow Stripe<br>Tubing (K103224) |
| | Infusion Sets (K103344) |
| | Lifeshield Vision Infusion Set with Pre-pierced Reseal (K113683) |
| Reason for 510(k) Submission | Introduction of the next generation of Plum™ Infusion System-Plum<br>360™ Infusion System with Hospira MedNet™ with enhanced<br>software, hardware and connectivity features.<br><br>Introduction of an optional Smart Card Plug 'n' Play module with<br>CE 3.0 for Plum A+TM Infusion System upgrade. |
| | Changes to secure lock component of Plum™ Administration Sets. |
| Device Description | The Plum 360TM Infusion System with Hospira MedNet™ remains a<br>volumentric Infusion System intended for use in parenteral, enteral<br>and epidural therapies and the administration of whole blood and<br>blood products. Dedicated administration sets are used for delivery of<br>infusion therapy to the patient.<br><br>The device has a user interface consisting of LCD display, front<br>panel numeric keypads and keys, audible and visual indicators,<br>audible indicator volume control, nurse call interface, network<br>interface connectors, the drug library, and software that provides pre-<br>defined display screens for therapy programming.<br><br>The continued use of the Hospira MedNet™ Software allows<br>communication between the Infusion Pump, Hospira MedNet™<br>server and the facility's communication systems. |
| Intended Use/Indication For Use | Plum 360™ Infusion System with Hospira Mednet™ and Plum<br>A+TM (360 EnabledTM) Infusion System are indicated for use in<br>parenteral, enteral and epidural therapies and the administration of<br>whole blood and blood products. |
| Summary of the Technological Characteristics of the Device Compared to the Predicate Device | |
| | Plum 360TM Infusion System with Hospira MedNet™ employs the same fundamental scientific technology, principles of<br>operation and intended/indications for use. In addition, the subject device includes the following notable characteristics:<br>• Latest CPU software version 15.0 featuring drug library enhancements, autoprogramming and alarm management<br>enhancments.<br>• Enhanced connectivity engine CE 3.0 with imporved Wi-Fi Performance, network connectivity and throughput to<br>Hospira MedNet™<br>• Enhanced hardware including egronomic enclosure, proximal tubing guides, ratcheting pole clamp and keypad.<br>• Enhancements to dedicated administration sets to include a modified secure lock.<br>• An optional Smart Card Plug 'n' Play with CE 3.0 Module allows Plum A+TM Infusion System upgrades to the |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Hospira logo. The logo features the company name "Hospira" in a simple, sans-serif font. Above the name is an abstract graphic consisting of three curved lines that converge at the bottom, creating a stylized, arch-like shape.
### Performance Data
### Summary of Non-Clinical Tests Conducted For Determination of Substantial Equivalence
Performance Test Summary - Subject Device
configuration.
System verification and validation activities for subject devices confirmed that the system meets and design inputs. All the testing met the acceptance criteria.
Risk management activities are incorporated in to the design and development process and safety assurance cases has been generated to demonstrate the safety of the subject devcies.
Risk management as well as verification activities incorporate the principles of applicable FDA guidance such as "Total Product Life Cycle: infusion Pump –Premarket Notification [510(k)] Submissions (DRAFT GUIDANCE)"
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Hospira logo. The logo consists of the word "Hospira" in a sans-serif font, with a stylized graphic above it. The graphic is composed of three curved lines that converge at the top, resembling a stylized arch or a simplified representation of a medical device. The logo is in black and white.
### Plum 360™ Infusion system with Hospira MedNet™/ Smart Card Plug 'n' Play CE 3.0 Module Traditional 510(k) Section 5: 510(k) Summary
issued in April 23, 2010 and "General Principles of Software Validation; Final Guidance for Industry and FDA Staff," issued January 11, 2002.
Human Factors studies have been conducted to validate the effectiveness of use related error mitigations.
| Characteristics | Standard/Test/FDA Guidance | Results Summary |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|
| Electrical and Mechanical<br>Safety | ANSI/AAMI ES60601-1 Medical Electrical Equipment - Part<br>1: General Requirements for Basic Safety and Essential<br>Performance, 2005 / (R)2012 / C1:2009 / (R)2012, including<br>Amendment 1:2012 | PASS |
| Electromagnetic<br>Compatibility | IEC 60601-1-2 Medical electrical equipment - Part 1-2:<br>General requirements for basic safety and essential<br>performance - Collateral standard: Electromagnetic<br>compatibility - Requirements and tests, Edition 3:2007-03 | PASS |
| Alarms System | IEC 60601-1-8 Collateral Standard: General requirements,<br>tests and guidance for alarm systems in medical electrical<br>equipment and medical electrical systems, Edition 2.0 2006-10 | PASS |
| Infusion Pump Basic Safety<br>and Essential Performance | IEC 60601-2-24 Medical Electrical Equipment - Part-24:<br>Particular Requirements for the Basic Safety and Essential<br>Performance of Infusion Pump and Controllers, Edition 2.0<br>2012-10 | PASS |
| Biocompatibility | ISO 10993-1 Biological evaluation of medical devices - Part 1:<br>Evaluation and testing within a risk management process,<br>Fourth edition 2009-10-15 | PASS |
| Summary of Clinical Tests Conducted for Determination of Substantial Equivalence and/or of Clinical Information | | |
| Clinical evaluation is not required for this submission to support substantial equivalence. Human factors studies have been<br>conducted on subject devcies demonstrating passing results. | | |
| Conclusion | | |
| Hospira, Inc. considers Plum 360™ Infusion System Hospira MedNet™ and Plum A+™ (360 Enabled™) Infusion System<br>with Hospira MedNet™ to be as safe and effective as its predicate devices. Therefore, Plum 360™ Infusion System<br>compatible with Hospira MedNet™ and Plum A+™ (360 Enabled™) Infusion System with Hospira MedNet™ are<br>substantially equivalent to predicate devices. | | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.