LIFESHIELD VISION INFUSION SET WITH PRE-PIERCED RESEAL
K113683 · Hospira, Inc. · FPA · Jan 12, 2012 · General Hospital
Device Facts
Record ID
K113683
Device Name
LIFESHIELD VISION INFUSION SET WITH PRE-PIERCED RESEAL
Applicant
Hospira, Inc.
Product Code
FPA · General Hospital
Decision Date
Jan 12, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
LIFESHIELD™ VISION™ Infusion sets are intended for the delivery of fluids including but not limited to blood and blood products from a container into a patients vascular system.
Device Story
LIFESHIELD™ VISION™ infusion sets are disposable, single-patient-use fluid delivery systems. They consist of tubing and pre-pierced access ports (male adapter plug, Y-site, T-connector) used to deliver fluids or blood products from a container to a patient's vascular system. The device is used in clinical settings, operated by healthcare professionals. The modification includes a change in the access port housing material from acrylic multipolymer to polycarbonate, the addition of a septum holder, and compatibility with BD plastic blunt cannulas. The device functions via gravity or in conjunction with Hospira infusion pumps. It allows for secure, needle-free or needle-based access to the infusion line, facilitating fluid administration while maintaining system integrity.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included biocompatibility (cytotoxicity, sensitization, irritation, systemic toxicity, hemocompatibility), sterility, particulate, leakage, tensile strength, filter characteristics, and dimensional/connection conformance per ISO 594-2.
Technological Characteristics
Materials: Polycarbonate housing, white septum holder. Energy source: Gravity or infusion pump. Connectivity: None. Sterilization: Validated per ISO 11137-2. Dimensions: Conforms to ISO 594-2 (6% Luer taper). Access: Compatible with plastic blunt cannulas and needles ≤18 gauge.
Indications for Use
Indicated for the delivery of fluids, including blood and blood products, from a container into a patient's vascular system. Intended for use as gravity sets or with dedicated Hospira Infusion Pumps.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
K113683
## LIFESHIELD™ VISION™ Pre-Pierced Reseal Modification Special 510(k) Date: December 14, 2011
JAN 1 2 2012
! t
イ .
### Section 6: 510(k) SUMMARY
A summary of 510(k) safety and effectiveness information in accordance with the requirements of
21 CFR 807.92. LIFESHIELD" VISION" Infusion sets with Pre-pierced Reseal modi
| Submitter Information | | |
|---------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|
| Name | Hospira, Incorporated | |
| Address | D-393, Bldg. H2<br>275 North Field Drive<br>Lake Forest, IL. 60046 | |
| Phone number | (224) 212-5316 | |
| Fax number | 224-212-5401 | |
| Establishment<br>Registration Number | Owner/Operator #9063339 | |
| Name of contact person | Karen Keener | |
| Date prepared | December 14, 2011 | |
| Name of device | | |
| Trade or proprietary<br>name | LIFESHIELD™ VISION™ Infusion sets with Pre-pierced Reseal Access Port | |
| Common or usual name | Fluid Delivery Tubing | |
| Classification name | Infusion Sets | |
| Classification panel | Class II | |
| Regulation | 21-CFR Part 880.5440 | |
| Product Code(s) | 80-FPA | |
| Legally marketed device(s) to<br>which equivalence is claimed | K941214 LIFESHIELD™ Extension Set<br>K052722 LIFESHIELD™ Latex Free Microbore Extension Set<br>K101677 Hospira Infusion Blood Sets | 10/03/1994<br>11/02/2005<br>08/11/2010 |
| | The changes addressed in this submission include:<br>Material change of the pre-pierced housing access port from an acrylic multipolymer to polycarbonate Addition of a Septum Holder using a medical grade material Minor Dimensional change to the pre-pierced access ports Use of BD plastic blunt cannula for port access | |
| Device description | The LIFESHIELD™ VISION™ infusion sets, are intended for use as gravity sets or<br>with dedicated Hospira Infusion Pumps. These devices are obtainable in custom<br>lengths and component options according to facility needs and physician<br>preference, Hospira infusion sets are disposable devices for single patient use,<br>which incorporate various set configurations and components.<br>The components include Pre-pierced access ports (Pre-pierced Male Adapter Plug,<br>Pre-pierced Y-site, and Pre-pierced T-connector) which are the subject of this<br>submission, may be shared across Hospira set families may be used with sets for | |
{1}------------------------------------------------
| Section 6: | 510(k) SUMMARY | |
|---------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|
| Intended use of the device | Infusion sets are intended for the delivery of fluids including but not limited to<br>blood and blood products from a container into a patient's vascular system. gravity or other pump platforms. | |
| Summary of the technological characteristics of the device compared to the predicate device | | |
| Characteristic | Proposed Device | Predicate |
| Intended Use | Same | Same |
| Set Functionality | Same | Same |
| Visual characteristics | Colorless polycarbonate of similar size | Green Acrylic Multipolymer |
| | white septum holder | Does not contain septum holder. |
| Bonding processes | Same | Same |
| Biocompatibility | Same | Same |
| Principle of Operation | Same | Same |
| Port Access | Plastic Blunt Cannula and needles<br>≤18guage | Metal Cannula and needles |
| PERFORMANCE DATA | | |
| SUMMARY OF NON-CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE* | | |
| Performance Test Summary-New Device | | |
| Characteristic | Standard/Test Method | Standard / Test Title<br>Device Performance |
| Biocompatibility | ISO 10993-5: 2009 | Cytotoxicity<br>Pass |
| Biocompatibility | ISO 10993-10: 2010 | Sensitization<br>Pass |
| Biocompatibility | ISO 10993-10: 2010 | Irritation / Intracutaneous<br>Reactivity<br>Pass |
| Biocompatibility | ISO 10993-11:2006 | Systemic Toxicity (Acute)<br>Pass |
| Biocompatibility | ISO 10993-4:2006 | Hemocompatibility<br>Pass |
| SAL 10-6 | ISO 11137-2:2006 | Sterility<br>Pass |
| Dimensional conformance<br>and<br>Connection compatibility | ISO 594-2 1998 | Conical fittings with a 6%<br>(LUER) taper for syringes,<br>needles and certain other<br>equipment<br>Pass |
Summary discussion of Bench Performance Data
The LIFESHIELD" VISION" Infusion sets with Pre-pierced reseal access ports have passed all specified test requirements.
The validation and verification testing have confirmed these devices meet user needs and design inputs for an infusion set.
Testing also confirmed physical attributes and device performance meet requirements of the standards listed in the "Performance Test Summary-New Devices" table above. These starility, biocompatibility, particulate, leakage, tensile strength, and filter characteristics.
{2}------------------------------------------------
### Section 6: 510(k) SUMMARY
### CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL DATA
Statement of Safety and Efficacy:
The LIFESHIELD" VISION" Infusion sets with pre-pierced reseal access port meet the functional claims, and intended use as described in the product labeling. The safety and effectiveness, are substantially equivalent to the predicate Hospira Infusion sets cleared in K941214 LIFESHIELD " Extension Set, K052722 LIFESHIELD" Latex Free Microbore Extension Set and K101677 Hospira Infusion Blood Sets.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three stripes forming its wing and body. The eagle is enclosed in a circle of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JAN 1 2 2012
Ms. Karen Keener Senior Specialist Hospira, Incorporated 375 N. Field Drive Lake Forest, Illinois 60045
Re: K113683
Trade/Device Name: LIFESHIELD™ VISION™ Infusion sets Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: December 14, 2011 Received: December 15, 2011
Dear Ms. Keener:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
## Page 2 - Ms. Keener
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony D. Watson
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Section 5: Indications for Use
510(k) Number (unknown at this time) & 113683
## Device Name: LIFESHIELD™ VISION™ Infusion sets
### Indications for Use:
LIFESHIELD™ VISION™ Infusion sets are intended for the delivery of fluids including but not limited to blood and blood products from a container into a patients vascular system.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (Part 21 CFR 807 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
Khd C. A
(Division Sign-Off) 1/9/12
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K113683
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.