BIOMEDIX I.V. EXTENSION LINE W/ STOPCOCK MODEL B50-102/B51-102
K961928 · Biomedix, Inc. · FPA · Sep 16, 1996 · General Hospital
Device Facts
Record ID
K961928
Device Name
BIOMEDIX I.V. EXTENSION LINE W/ STOPCOCK MODEL B50-102/B51-102
Applicant
Biomedix, Inc.
Product Code
FPA · General Hospital
Decision Date
Sep 16, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This product would be used by Nurses and Paramedics in prehospital, emergency room and field situations.
Device Story
I.V. Extension with Stopcock facilitates intravenous fluid/drug administration; provides extra length between I.V. supply and patient; includes four-way stopcock for fluid source selection; features slip-type clamp to prevent backflow/dilution. Used by nurses and paramedics in prehospital, emergency room, and field settings. Available in two models: standard needle-type injection site or blunt cannula (needleless) injection site. Device connects to primary I.V. set to allow direct injection at patient catheter site.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
I.V. extension line assembly with four-way stopcock. Features include threaded and/or slip luer connections, 3.75 mL capacity, and 0.114" I.D. tubing. Available with needle-type or blunt cannula (needleless) injection sites. Includes slip-type tubing clamp. Construction utilizes commercially available components similar to existing I.V. sets.
Indications for Use
Indicated for use by nurses and paramedics in prehospital, emergency room, and field settings to extend intravenous lines and provide additional ports for the administration of fluids and drugs.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
Baxter PHARMASEAL K52 NOVEX Three-Way Stopcock with Extension Tube (K52)
Medex Extension Set
McGaw G0719 I.V. Extension Set
McGaw G0726 I.V. Extension Set
McGaw G0716 I.V. Extension Set
Related Devices
K991932 — INTRAVASCULAR ADMINISTRATION SET · The Kipp Group · Aug 20, 1999
K153293 — Extension Set · B.Braun Medical, Inc. · Apr 6, 2016
K171118 — Carefusion NeutraClear multi-fuse Extension Set with needle-free connector(s) · Cair Lgl · Aug 15, 2017
K092172 — EXEL I.V. EXTENSION SET · Exelint Intl. Co. · Jan 4, 2010
K980992 — IV ADMINISTRATION SETS WITH NEEDLELESS ACCESS DEVICES · Venetec Intl., Inc. · May 28, 1998
Submission Summary (Full Text)
{0}
BIOMEDIX
quality medical products
K961928
SEP 16 1996
TO: Food and Drug Administration
Center for Devices and Radiological Health (HFZ-401)
9200 Corporate Blvd.
Rockville, MD 20850
FROM: Daniel J. Dyar, President
DATE: May 17, 1996 (Revised June 11, 1996)
SUBJECT: 510 K Summary
TRADE NAME: I.V. Extension with Stopcock.
COMMON NAME: I.V. Extension
CLASSIFICATION NAME: [ 880.5440 ] Intravascular Administration Set
SUBSTANTIALLY
EQUIVALENT TO:
Biomedix, Inc. claims substantial equivalence regarding this design of the I.V. Extension Line Assembly with Stopcock as follows:
Products currently on the market include:
Baxter PHARMASEAL K52 NOVEX Three-Way Stopcock with Extension Tube.
( Sterile, For Single Use Only, Capacity 2.3 mL, Length 50.8 cm. Blue Protector covers Male Adapter. )
Medex Extension Set - 12" Extension with "Y" Injection Site.
McGaw - G0719 I.V. Extension Set, Female and Needle Adapter, Vol. 3 mL., Length 21"
G0726 I.V. Extension Set with (2) Injection Sites, Vol. 6 mL., Length 33"
G0716 I.V. Extension Set with (2) Injection Sites with check valve and
flashtube., Vol. 6 mL., Length 34"
PAGE 15
45 South Main Street P.O. Box 231 Spencer, Indiana 47460 800-627-2765 812-829-9667 FAX: 812-829-9650
{1}
510 K SUMMARY ( CONTINUED )
DESIGN COMPARISONS:
| COMPONENT | BIOMEDIX | BAXTER K52 | MEDEX EXTENSION SET |
| --- | --- | --- | --- |
| Stopcock | All 3 ports include both threaded and / or slip luer connections. | ( 1 ) Port has removeable needle injection site. ( 2 ) remaining ports have slip luer conn. only. | ( 1 ) Y - Port for Sharps Devices. |
| Capacity | 3.75 ml. | 2.3 ml. | N/A |
| Tubing | .114 +/- .002" I.D. | .088-.090" I.D. | N/A |
| Connections | Choice of Needle or Blunt Cannula Y-Site with Male Slip Luer Connector. | Male Slip Luer Connector Only. | Male Slip Luer Connector Only. |
The Biomedix design allows injection of solutions directly at the Patient I.V. Catheter. A slip-type clamp on the tubing portion of the I.V. Extension Line allows the line to be shut-off preventing backflow and/or dilution of administered fluids away from the patient.
DESCRIPTION:
The I.V. Extension Line Assembly with Stopcock is attached to the Primary I.V. to facilitate two advantages. One, is to add extra length between the I.V. Supply and the Patient. A second, is to provide additional supply ports for the administration of alternate solutions and special drugs that are administered intravenously. A four-way stopcock offers selection of the fluid types and sources. Biomedix is offering two models, one with a standard Needle-Type Injection Site at the Outlet and a second with a Blunt Cannula (Needleless) Type Injection Site at the outlet.
INTENDED USE:
This product would be used by Nurses and Paramedics in prehospital, emergency room and field situations.
PAGE 16
{2}
510 K SUMMARY ( CONTINUED )
## SUMMARY OF DEVICE COMPARISONS:
Basic construction of both Biomedix and Competitive Products of this type are of commercially available components. Biomedix already employs a process and construction similar to this in their existing Selec-3 Primary I.V. Set.
Sincerely,

Daniel J. Dyar
President Biomedix, Inc.
PAGE 16a
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.