MAERSK MEDICAL PURELINE COMFORT SUBCUTANEOUS INFUSION SET
K972135 · P/L Biomedical · FPA · Aug 26, 1997 · General Hospital
Device Facts
Record ID
K972135
Device Name
MAERSK MEDICAL PURELINE COMFORT SUBCUTANEOUS INFUSION SET
Applicant
P/L Biomedical
Product Code
FPA · General Hospital
Decision Date
Aug 26, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Maersk Medical Pureline™ Subcutaneous Infusion Sets with Indwelling Catheters continue to be intended for the infusion and/or injection of fluids into the body below the surface of the skin. The indwelling catheter can be inserted independently from the infusion catheter and can be accessed for injections through the injection port. The indwelling catheter can also be securely attached to the infusion catheter by means of a proprietary "click-lock" connector for the infusion of drugs subcutaneously. The infusion set can be detached from the indwelling catheter, and the catheter capped, to allow freedom from the infusion set and pump for showers, athletics or other activities.
Device Story
Device is a sterile, single-use subcutaneous infusion set with an indwelling Teflon catheter. Components include an indwelling catheter assembly (Teflon cannula, adhesive fixation tape, injection port, female click-lock connector) and an infusion set (co-extruded tube, female luer lock, male click-lock connector). Insertion is performed using a stainless steel insertion cannula. The system allows for independent catheter placement and subsequent connection to an infusion pump via the proprietary click-lock mechanism. The infusion set can be detached, allowing the patient to cap the catheter for activities like showering or athletics. The device is used in clinical or home settings to facilitate subcutaneous drug delivery.
Clinical Evidence
Bench testing only. Biocompatibility and component security testing were performed to demonstrate substantial equivalence to the unmodified predicate device.
Technological Characteristics
Materials include Teflon (cannula), stainless steel (insertion cannula), and proprietary plastic (connectors). Features a co-extruded tube, female luer lock, and proprietary click-lock connector. Sterile, non-pyrogenic, single-use. No electronic or software components.
Indications for Use
Indicated for the infusion and/or injection of fluids into the body below the surface of the skin.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
K012754 — EZ SET INFUSION SET · Animas Corp. · Aug 27, 2001
K153257 — BD FlowSmart Set, MiniMed Pro-Set · Becton, Dickinson and Company · Dec 11, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
# ·510(K) SUMMARY
K972/35
#### 9 0 Summary of Safety and Effectiveness
AJG 26 EST
- 9.1 Trade/Proprietary Name: Maersk Medical Pureline™ Subcutaneous Infusion Set
- Subcutaneous Infusion Set with indwelling catheter 9.2 Common/Usual Name:
- 9.3 Classification Name: Intravascular Administration Set
- 9.4 Substantial Equivalence: The modified Pureline™ Comfort™ Subcutaneous Infusion Sets are substantially equivalent to the currently marketed Pureline™ Comfort™ Subcutaneous Infusion Sets (K95018).
- 9.5 Device Description
(
The general description, operation, construction and use of the Maersk Medical Pureline™ Subcutaneous Infusion Sets With Indwelling Catheters has not changed as a result of the modifications. They remain sterile, non-pyrogenic, single use subcutaneous infusion sets designed to be used with commercially available infusion devices. The following discussion remains valid.
There are two basic components provided for each device. The first is a stand-alone subcutaneous indwelling catheter. This catheter is provided as an integral assembly with a Teflon cannula, adhesive backed fixation tape, an iniection port and the female portion of a proprietary plastic "click-lock" connector. The catheter assembly comes with a stainless steel insertion cannula. The insertion cannula is mounted to a male portion of the proprietary plastic "click-lock" connector. The insertion cannula is come to the user inserted through the injection port and the inner lumen of the Teflon catheter with the needle end protruding past the tip of the catheter. The male connector is locked to the female connector on the indwelling catheter. A needle protector is assembled over the Teflon catheter and the insertion cannula. A separate male portion of the proprietary connector without the insertion cannula is provided in the package. This component is used to attach to the female connector of the catheter after the indwelling Teflon cannula has been inserted and the steel insertion cannula has been withdrawn. The connector protects the indwelling catheter when the infusion set is not attached.
The second component of the Maersk Medical Subcutaneous infusion Set with Indwelling Catheter is the infusion set. The infusion set is comprised of a co-extruded tube with a female luer lock connector at the pump end and a stainless steel needle incorporated into the male portion of the proprietary plastic "click-lock" connector at the patient end. In order to maintain the sterility of the infusion path, a male luer cap covers the female luer lock connector and the male "click-lock" connector comes attached to a mating female connector. The sets come packaged in blister packs sealed with paper lid stock.
- 9.6 Intended Use
The Intended Use for the device has not changed. The Maersk Medical Pureline™ Subcutaneous Infusion Sets with Indwelling Catheters continue to be intended for the infusion and/or injection of fluids into the body below the surface of the skin. The indwelling catheter can be inserted
{1}------------------------------------------------
independently from the infusion catheter and can be accessed for injections through the injection port. The indwelling catheter can also be securely attached to the infusion catheter by means of a proprietary "click-lock" connector for the infusion of drugs subcutaneously. The infusion set can be detached from the indwelling catheter, and the catheter capped, to allow freedom from the infusion set and pump for showers, athletics or other activities.
### 9.7 Technological Characteristics
The technological characteristics of the deice have not been affected by these modifications.
#### 9.8 Performance Data
The results of the biocompatibility testing and component security testing showed the device is substantially equivalent to the unmodified device.
#### 9.9 Conclusion
Based on the design equivalency and the functional and safety testing, Maersk Medical has determined that the Maersk Medical Pureline™ Comfort™ Indwelling Subcutaneous Catheter and Infusion Sets are substantially equivalent to the devices currently marketed in the United States.
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. In the center of the seal, there is a stylized image of three abstract shapes that resemble birds in flight. The seal is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Lee Leichter P/L Biomedical 7690 Cameron Circle 33912 Fort Myers, Florida
AUG 26 1997
K972135 Re : Maersk Medical Pureline Comfort Subcutaneous Trade Name: Infusion Set Requlatory Class: II Product Code: FPA Dated: May 14, 1997 Received: June 6, 1997
Dear Mr. Leichter:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531
{3}------------------------------------------------
Page 2 - Mr. Leichter
through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fdag.gov/cdrh/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## 510(k) File Number:
Maersk Medical Pureline™ Comfort™ (Disetronic Tender™) Subcutaneous Device Name: Infusion Set
- The Subcutaneous Infusion Sets with Indwelling Catheters are indicated for the Indications For Use: infusion and/or injection of fluids into the body below the surface of the skin.
(Division Sign-Citi) Division of Dental, Int and General Hospital D 5 ' Olk) Number _
Prescription Use
(Per 21 CFR 801.109)
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.