K101677 · Hospira, Inc. · FPA · Aug 11, 2010 · General Hospital
Device Facts
Record ID
K101677
Device Name
HOSPIRA INFUSION BLOOD SETS
Applicant
Hospira, Inc.
Product Code
FPA · General Hospital
Decision Date
Aug 11, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Hospira infusion sets are intended for the delivery of fluids, including but not limited to, blood and blood products, from a container to a patient's vascular system.
Device Story
Hospira PVC Infusion Blood Sets facilitate fluid and blood product delivery from containers to patient vascular systems. The device consists of non-DEHP PVC tubing and components, including a manual blood pump with an integrated ball check valve. The pump allows manual control of fluid flow. The device is intended for clinical use by healthcare professionals. It functions as a passive delivery system; no electronic or automated processing is involved. The design incorporates minor dimensional and geometric changes from previous iterations to accommodate the non-DEHP material. The device ensures safe fluid administration through standard luer connections and integrated filtration (200-micron filter).
Clinical Evidence
No clinical data. Bench testing only. Performance testing confirmed compliance with ISO 10993 (biocompatibility), ISO 11137 (sterility), ISO 8536-4 (gravity set characteristics), and ISO 594-2 (luer connection compatibility).
Technological Characteristics
Non-DEHP PVC tubing and components; manual blood pump with ball check valve; 200-micron filter. Connectivity: None (mechanical). Sterilization: ISO 11137-2. Standards: ISO 10993 (biocompatibility), ISO 8536-4 (infusion equipment), ISO 594-2 (luer fittings).
Indications for Use
Indicated for the delivery of fluids, including blood and blood products, from a container into a patient's vascular system.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
Lifeshield Latex-Free Veniloop Connector with Prepierced Injection Site, Non-DEHP (K063239)
Lifeshield Latex-Free Plum Non-Vented Blood set (K030002)
HEMA Y-Type Transfusion Blood Filter Pump Set 100 Inch-SL with Prepierced Reseals (Latex-Free) (K780880)
Related Devices
K213846 — Q2 Blood Administration Sets · Quest Medical, Inc. · Jun 5, 2022
K143082 — IV Administration Sets with 200(micro)m Blood Filter · B.Braun Medical, Inc. · Jul 24, 2015
K210335 — Blood Administration Set · Baxter Healthcare Corporation · Jun 10, 2021
K150513 — Imed Technology Intravascular Administration Set, Imed Technology Extension Set · Imed Technology, Inc. · May 28, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
Hospira Infusion Blood Sets K101677
## 510(k) SUMMARY
- B
A summary of 510(k) safety and effectiveness information in accordation with the requirements of 21 CFR 807.92.
| Submitter Information [21 CFR 807.929(a)(1)] | |
|-------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | Hospira Incorporated |
| Address | D-389, Bldg H2, 375 N. Field Drive, Lake Forest, Illinois 60045 |
| Phone number | 224-212-5316 |
| Fax number | 224-212-5401 |
| Establishment Registration<br>Number | 9063339 |
| Name of contact person | Karen Keener |
| Date prepared | July 20, 2010 |
| Name of the device [21 CFR 807.92(a)(2)] | |
| Trade or proprietary name | Hospira PVC Infusion Blood Set |
| Common or usual name | Fluid delivery tubing |
| Classification name | Sets, Administration, Intravascular |
| Classification panel | General Hospital |
| Regulation | 880.5440 |
| Product Code(s) | FPA |
| Legally marketed device(s)<br>to which equivalence is<br>claimed [21 CFR<br>807.92(a)(3)] | Hospira Infusion Blood sets utilize components from various Hospira<br>Infusion set families; including but not limited to:<br>Lifeshield Latex-Free Veniloop Connector with Prepierced Injection<br>Site, Non-DEHP (K063239)<br>Lifeshield Latex-Free Plum Non-Vented Blood set (K030002)<br>HEMA Y-Type Transfusion Blood Filter Pump Set 100 Inch-SL with<br>Prepierced Reseals (Latex-Free) (K780880) |
| Reason for 510(k)<br>submission | To offer a non-DEHP PVC alternative Infusion set with a manual<br>blood pump. |
| Device description<br>[21 CFR 807.92(a)(4)] | The Hospira PVC Infusion Blood Sets with/without the Manual blood<br>pump consists of Non-DEHP PVC tubing and components previously<br>cleared in Hospira 510k's. The change addressed in this submission,<br>is the material used to construct the cylindrical blood pump. The<br>modified design incorporates a ball check valve previously used in<br>the bulb style manual blood pump. Other changes include minor<br>dimensional and geometric changes. |
| Intended use of the device<br>[21 CFR 807.92(a)(5)] | Hospira infusion sets are intended for the delivery of fluids, including<br>but not limited to, blood and blood products, from a container to a<br>patient's vascular system. |
{1}------------------------------------------------
#### Hospira Infusion Blood Sets K101677
Attachment 2
| Performance Data [21 CFR 807.92(b)] | | | |
|---------------------------------------------------------------------------------------------------------------|----------------------|-----------------------------------------------------------------------------------------------------------|--------------------|
| Summary of non-clinical tests conducted for determination of substantial equivalence<br>[21 CFR 807.92(b)(1)] | | | |
| Performance Test Summary-New Device | | | |
| Characteristic | Standard/Test Method | Standard / Test Title | Device Performance |
| Biocompatibility | ISO 10993-5: 2009 | Cytotoxicity | Pass |
| Biocompatibility | ISO 10993-10: 2002 | Sensitization | Pass |
| Biocompatibility | ISO 10993-10: 2002 | Irritation /<br>Intracutaneous<br>Reactivity | Pass |
| Biocompatibility | ISO 10993-11:2006 | Systemic Toxicity<br>(Acute) | Pass |
| Biocompatibility | ISO 10993-4:2002 | Hemocompatibility | Pass |
| SAL 10-6 | ISO 11137-2:2006 | Sterility | Pass |
| Packaging | ISO 1135-4:2004 | Transfusion<br>Equipment for<br>Medical Use | Pass |
| Particulate | | | |
| Bond Strength | | | |
| Piercing function | | | |
| Flow characteristics | | | |
| Specific Required<br>Physical Characteristics | | | |
| Gravity Set<br>Characteristics | 8536-4:2004 | Infusion<br>Equipment for<br>Medical Use; Part<br>4. | Pass |
| Dimensional<br>Conformance<br>and<br>Connection compatibility | ISO 594-2 | Conical Fittings<br>with a 6% (Luer)<br>Taper for syringes,<br>needles, and<br>certain other<br>equipment | Pass |
Summary Discussion of Bench Performance Data
The Hospira Blood administration sets with 200-micron filter and Non-DEHP PVC manual blood pump passed all specified test requirements.
The validation and verification testing confirmed product meets user needs and design inputs for an administration set containing a manual blood pump.
Testing also confirmed physical attributes and device performance meet requirements of the standards listed in the 'Performance test summary" above. These standards address sterility, biocompatibility, particulate, leakage, tensile strength, filter characteristics, drip tube performance, flow control, and simulated actual conditions of use.
{2}------------------------------------------------
# K101677
#### Hospira Infusion Blood Sets K101677
#### Attachment 2
Summary of clinical tests conducted for determination of substantial equivalence or of clinical information
[21 CFR 807.92(b)(2)]
Not required. Bench testing can adequately address the safety and efficacy of this product.
Statement of Safety and Efficacy
[21 CFR 807.92]
The Hospira Infusion Blood Sets with/without Manual Blood Pump meet the functional claims, and intended use as described in the product labeling.
The safety and effectiveness are equivalent to the predicate Hospira Administration Blood Sets with Manual Blood Pump. The claim for substantial equivalence is supported by the information provided in the 510(k) submission.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993=0002
Ms. Karen Keener Global Regulatory Associate Hospira, Incorporated 375 North Field Drive Lake Forest, Illinois 60045
Re: K101677
Trade/Device Name: Hospira Infusion Blood Sets Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: July 29, 2010 Received: July 30, 2010
AUG11 2010
Dear Ms. Keener:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2- Ms. Keener
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance,
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours
Susan Dunne
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology; General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Section 5:
# 510(k) Number (if known) K IO| 6 7 7
Device Name: Hospira Infusion Blood Sets
Indications for Use: Hospira Administration sets are intended for the delivery of fluids including but not limited to blood and blood products from a container into a patients vascular system.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over –The- Counter Use {Part 21 CFR *)& Subpart C)
(Please do not write below this line – continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K101677
Section 5: Indications for use Page 1 of 1
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.