K120799 · Yukon Medical, LLC · FPA · Mar 30, 2012 · General Hospital
Device Facts
Record ID
K120799
Device Name
ARISURE NEUTRAL VALVE
Applicant
Yukon Medical, LLC
Product Code
FPA · General Hospital
Decision Date
Mar 30, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The Arisure™ Neutral Valve is a single use, sterile, non-pyrogenic device intended for use as an accessory to intravascular administration sets for the administration or withdrawal of fluids from a patient through a cannula placed in the vein or artery. The Arisure™ Neutral Valve may be used with low-pressure power injectors up to 300 psi.
Device Story
Arisure™ Neutral Valve is a sterile, single-use, luer-activated needle-free connector; used as an accessory to intravascular administration sets. Device consists of polycarbonate housings, retention ring, and pre-slit silicone piston lubricated with silicone oil. Principle of operation: male luer activation deforms the internal pre-slit silicone piston, opening a fluid pathway for administration or withdrawal of fluids; piston returns to original shape upon disconnection. Used in clinical settings by healthcare providers. Device facilitates fluid access to venous or arterial cannulas; supports low-pressure power injection up to 300 psi. Hemocompatibility testing confirms low hemolysis rates during blood withdrawal, comparable to 21GA needles. Benefits include needle-free access and secure fluid management.
Clinical Evidence
Bench testing only. Evaluated microbial ingress, liquid/vacuum leakage, flow rate, tensile strength, and disconnection bolus. Hemocompatibility testing compared blood withdrawal via Arisure™ Neutral Valve versus 21GA needle; both demonstrated hemolytic percentages below 2% (ASTM F 756-00), confirming non-hemolytic properties.
Technological Characteristics
Sterile, single-use, luer-activated needle-free connector. Materials: Polycarbonate (housings, retention ring), silicone rubber (piston), fluorosilicone (lubricant). Principle: Mechanical deformation of pre-slit elastomeric sleeve over internal post. Dimensions/form factor: Standard luer-compatible. Connectivity: Standalone mechanical interface. Sterilization: Not specified. No software/algorithm.
Indications for Use
Indicated for use as an accessory to intravascular administration sets for the administration or withdrawal of fluids from a patient through a cannula placed in the vein or artery. Compatible with low-pressure power injectors up to 300 psi.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
K213004 — Needle Free Connector · Medcaptain Life Science Co., Ltd. · May 10, 2022
K083723 — A6 LUER ACCESS DEVICE · B.Braun Medical, Inc. · Jun 5, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
510(K) Notification
Yukon Medical, LLC
Arisure™ Nestral Valve
# 5.0 510(k) Summary
MAR 3 0 2012
ﺮ ":"
-
#### 5.1. Submitter
| Submitter's Name<br>and Address | Todd Korogi, President | |
|---------------------------------|--------------------------------------------------|--|
| | Yukon Medical, LLC | |
| | 4819-400 Emperor Blvd. | |
| | Durham, NC 27703 | |
| | Phone: 919-595-8250 | |
| | Email: tkorogi@yukonmedical.com | |
| Contact Person | Carl Dupper, Director of Engineering and Quality | |
| | Yukon Medical | |
| | 4819-400 Emperor Blvd. | |
| | Durham, NC 27703 | |
| | Phone: 919-595-8257 | |
| | Email: cdupper@yukonmedical.com | |
| Prepared By | James Fentress, Product Development Engineer | |
| | Gilero Biomedical | |
| | 4022-300 Stirrup Creek Dr. | |
| | Durham, NC 27703 | |
| | Phone: 919-595-8236 | |
| | Email: jfentress@gilero.com | |
| Date Prepared | 01 November 2011 | |
#### 5.2. Device Identification
| Trade Name | Arisure™ Neutral Valve |
|----------------|--------------------------------------------------------------------------------------|
| Common Name | Needle-free access device, needle-free connector, luer<br>access device |
| Classification | Intravascular Administration Set<br>Product Code - FPA<br>Regulation 21 CFR 880.5440 |
#### 5.3. Predicate Devices
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
| Device | Manufacturer | 510(k) |
|--------------------------------------------------------------|-------------------|---------|
| B3300<br>MicroCLAVE®<br>Connector; C1000<br>CLAVE® Connector | ICU Medical, Inc. | K970855 |
{1}------------------------------------------------
#### 5.4. Device Description
The Arisure™ Neutral Valve is a sterile, single-use, swab-able, normally closed, luer-activated, valved connector. The Arisure™ Neutral Valve consists of five primary components: lower housing, upper housing, retention ring, pre-slit silicone piston, and lubricating silicone oil. The lower housing, made of polycarbonate, consists of a male luer and a long feature that provides support for the silicone piston to move upon. The pre-slit silicone piston attaches to the lower housing. A series of round protrusions on the inside of the silicone piston provides sliding seals over the long feature on the lower housing. These sliding seals allow fluid to travel through the silicone piston when the piston is accessed with a male luer. The retention ring, made of polycarbonate, fits over the silicone piston and provides extra rigidity. The silicone piston and top housing together comprise the female end of the needle-free valve. The top housing, made of polycarbonate, fits over the top of the piston and is permanently affixed to the bottom housing. When pressed downward by a male luer, the slit in the piston opens, and the sliding seal on the inside of the piston opens a fluid pathway through the device. Silicone oil is used to lubricate the silicone piston. The Arisure™ Neutral Valve does not contain any natural rubber latex.
#### 5.5. Intended Use
The Arisure™ Neutral Valve is a single use, sterile, non-pyrogenic device intended for use as an accessory to intravascular administration sets for the administration or withdrawal of fluids from a patient through a cannula placed in the vein or artery. The Arisure™ Neutral Valve may be used with low-pressure power injectors up to 300 psi.
The indications statement for the Arisure™ Neutral Valve differs from the indications statement of the predicate device (presented in 510(k) K970855) in that it additionally states that the Arisure™ Neutral Valve can be used to withdraw fluid from a patient, rather than just administer fluid. This difference is not critical to the intended therapeutic use of the device, and the difference does not affect the safety and effectiveness of the device when used as indicated.
The primary concern with the withdrawal of fluids from a patient through an cannula placed in the vein or artery of a patient is damage that can occur to the withdrawn fluid sample; damage to blood. In addition to the traditional tests for material hemocompatibility presented in the biocompatibility section (section 15), additional testing was also conducted on the Arisure™ Neutral Valve to determine whether the withdrawal of blood using the Arisure™ Neutral Valve caused damage to the blood. Section 15 details this test, which shows that when compared to blood withdrawn using a 21GA needle, the Arisure™ Neutral Valve has less hemolysis. A general threshold for assessing the hemolytic property of a given material is that hemolytic percentages below 2% are considered nonhemolytic (ASTM F 756-00). Both the 21GA needle and the Arisure™ Neutral Valve had hemolytic percentages below 2%.
{2}------------------------------------------------
In addition to the hemolytic testing conducted on the Arisure™ Neutral Valve, the predicate CLAVE® device is also used for the withdrawal of fluid. ICU Medical Instructions M1-1053 Rev01 illustrates the use of the CLAVE® device for withdrawing blood from a patient.
#### 5.6. Predicate Device Comparison (Technological Characteristics)
A direct comparison of the technical and performance characteristics of the Arisure™ Neutral Valve and the technical and performance characteristics of the predicate device demonstrates equivalency. Minor differences in subject and predicate device characteristics do not introduce any new safety or efficacy concerns.
| Table 5.6: Device Technological Comparison | | |
|---------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject Device Arisure™ Neutral<br>Valve | Predicate Device MicroCLAVE®<br>B3300 Connector and CLAVE®<br>C1000 Connector<br>(510(k):K970855) | |
| Materials | | |
| Neutral Valve Materials:<br>Lower Housing (Internal Conduit) -<br>Polycarbonate | Materials per 510(k): K970855<br>Internal Conduit- Polycarbonate | |
| Pre-slit Silicone Piston: silicone rubber | Silicone Seal[Silicone Piston] - Silicone<br>Rubber | |
| Lubricant - Fluorosilicone | Lubricant - Fluorosilicone | |
| Upper Housing - Polycarbonate | [upper]Housing - Polyester | |
| Retention Ring- Polycarbonate | | |
| Principles of<br>Operation | The female Luer valve opens to permit the<br>introduction or withdrawal of fluids when<br>accessed by a male Luer tip. This access<br>deforms a compressible element that returns<br>to its original shape through the mechanical<br>properties of the deformable element. | The female Luer valve opens to permit the<br>introduction or withdrawal of fluids when<br>accessed by a male Luer tip. This access<br>deforms a compressible element that returns<br>to its original shape through the mechanical<br>properties of the deformable element. |
| Technology<br>and Design | When activated by a male Luer, a pre-slit<br>elastomeric sleeve advances over an internal<br>post, opening a fluid pathway that connects<br>the female and male ends of the device. | When activated by a male Luer, a pre-slit<br>elastomeric sleeve advances over an internal<br>post, opening a fluid pathway that connects<br>the female and male ends of the device. |
{3}------------------------------------------------
#### 5.7. Discussion of Non-clinical tests
The following non-clinical tests have been done in support of the substantial equivalence of the Arisure™ Neutral Valve to the Predicate CLAVE® devices:
- Microbial Ingress Testing .
- Biocompatibility Testing .
- Liquid Leak (open position) .
- Liquid Leak (closed position) .
- Vacuum Leak (open position) .
- Vacuum Leak (closed position) .
- Flow Rate .
- Tensile Strength .
- Disconnection Bolus .
Results from non-clinical testing indicate that the Arisure™ Neutral valve is as safe and effective as the predicate device. Testing conducted that reference the design verification and validation testing of the Arisure™ Neutral valve meet pre-determined acceptance criteria for the device. Testing conducted as a direct comparison of the Arisure™ Neutral valve and predicate device demonstrate that the equivalence of the devices.
#### 5.8. Conclusion
The Arisure™ Neutral Valve is equivalent to the predicate device and is safe and effective for its intended use.
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002
Yukon Medical, LLC C/O Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25TH Street N.W. Buffalo, Minnesota 55313
MAR 3 0 2012
Re: K120799
Trade/Device Name: Arisure™ Neutral Valve Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set. Regulatory Class: II Product Code: FPA Dated: March 15, 2012 Received: March 16, 2012
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
### Page 2 - Mr. Job
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony Dimon
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# 4.0 Indications for Use Statement
510(k) Number (if known): K120799
Device Name: Arisure™ Neutral Valve
Indications for Use:
The Arisure™ Neutral Valve is a single use, sterile, non-pyrogenic device intended for use as an accessory to intravascular administration sets for the administration or withdrawal of fluids from a patient through a cannula placed in the vein or artery. The Arisure™ Neutral Valve may be used with low-pressure power injectors up to 300 psi.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Ruld C. Chign 3/29/2012
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page 1 of 1
120799 510(k) Number:
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.