TK Intravascular Administration Set

K191640 · Anhui Tiankang Medical Technology Co., Ltd. · FPA · Aug 26, 2020 · General Hospital

Device Facts

Record IDK191640
Device NameTK Intravascular Administration Set
ApplicantAnhui Tiankang Medical Technology Co., Ltd.
Product CodeFPA · General Hospital
Decision DateAug 26, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5440
Device ClassClass 2
AttributesTherapeutic

Indications for Use

TK Intravascular Administration Set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein.

Device Story

Gravity-fed single-use intravascular administration set; delivers fluids from container to patient vein via catheter or needle. Components include spike, drip chamber, micron/fluid filters, tubing, clamps, flow regulator, Y-site, and luer connectors. Used in clinical settings; operated by healthcare professionals. Device facilitates controlled fluid infusion. Benefits patient by providing reliable, sterile pathway for intravenous therapy.

Clinical Evidence

Bench testing only. Biocompatibility testing performed per ISO 10993-1 (hemocompatibility, cytotoxicity, irritation, sensitization, systemic toxicity, pyrogenicity). Performance testing conducted per ISO 8536-4, ISO 80369-7/20, ISO 11607-1/2, ISO 11135, ISO 10993-7/12, ASTM F1980, ASTM F1929, ASTM F88, and USP 788.

Technological Characteristics

Materials: PVC, ABS, LDPE, PTFE, POM, MABS, Makrolon, HDPE. Gravity-fed infusion set. Components: spike, drip chamber, filters, tubing, clamps, flow regulator, Y-site, luer connectors. Sterilization: Ethylene Oxide (EO) per ISO 11135. Connectivity: None. Software: None.

Indications for Use

Indicated for patients requiring fluid administration from a container to the vascular system via a needle or catheter inserted into a vein.

Regulatory Classification

Identification

An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.

Special Controls

*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around a symbol. To the right of the seal, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on the top line and "ADMINISTRATION" on the bottom line. August 26, 2020 Anhui Tiankang Medical Technology Co., Ltd. Bai Baodong RA Manager No.228 Weiyi Road, Economic Development Zone Tianchang, 239300 Anhui China Re: K191640 Trade/Device Name: TK Intravascular Administration Set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: FPA Dated: July 28, 2020 Received: July 28, 2020 Dear Bai Baodong: We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Payal Patel Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ {3}------------------------------------------------ No.228 Weiyi Road ,Economic Development Zone ,Tianchang City, Anhui, China [TK Intravascular Administration Set] ## 510(K) Summary: K191640 ### Date prepared: 25.08.2020 ### 1. Submitter Name and Address: | Owner Name: | Anhui Tiankang Medical Technology Co., Ltd. | |-----------------|------------------------------------------------------------------------------| | Address: | No.228 Weiyi Road .Economic Development Zone<br>Tianchang City, Anhui, China | | Contactor Name: | Bai Baodong, RA Manager | | TEL: | +86-550-7309187 | | E-mail: | tkquality@126.com | Manufacturer Nạme: Anhui Tiankang Medical Technology Co., Ltd | Address: | No.228 Weiyi Road, Economic Development Zone<br>Tianchang City, Anhui, China | |----------|------------------------------------------------------------------------------| |----------|------------------------------------------------------------------------------| ## US Agent: | US Agent: | James H. Liao | |-----------|-----------------------------------------------------| | Address: | 6775 Verde Ridge Rd., Rancho Palos Verdes, CA 90275 | | TEL: | (310) 375-8169 | | Email: | James@Sino2us.com | # 2. Submission Devices Information: Trade/Proprietary Name: TK Intravascular Administration Set Common Name: Intravascular administration set {4}------------------------------------------------ No.228 Weiyi Road ,Economic Development Zone ,Tianchang City, Anhui, China [TK Intravascular Administration Set] Regulation: 21CFR 880.5440 Regulation name: Set, Administration, Intravascular. Class: II Product Codes: FPA Submission Type: 510(K) ### 3. Predicate Device Information: Trade Name: U&U Intravascular Administration Set 510(K) Number: K151151 Intravascular administration set Common Name: Regulation: 21CFR 880.5440 Regulation name: Set, Administration, Intravascular. Class: II Product Codes: FPA ### 4. Device Description: #### TK Intravascular Administration Set The TK Intravascular Administration Set is a gravity single use device sterilized with Ethylene Oxide gas. It is utilized to administer fluids from a container to a patient's vascular system through a catheter or needle inserted into a vein. The device includes the following components: Protective cap of spike, Spike, Micron filter, Air inlet, Drip chamber, Fluid filter, Tubing, Slide clamp, Pinch clamp, Flow regulator, Roller, Male luer lock, Stopcock (without needleless luer access valve), 4.0mm Connector, Y-site (access by needle), Rotating luer lock, Back check valve, Luer lock cap. The device is available in multiple configurations listed below. | Ref | Description | |-----|-------------| |-----|-------------| {5}------------------------------------------------ No.228 Weiyi Road ,Economic Development Zone ,Tianchang City, Anhui, China [TK Intravascular Administration Set] | Number | | |----------|----------------------------------| | TKIV-F05 | Intravascular Administration Set | | TKIV-F09 | Intravascular Administration Set | | TKIV-F10 | Intravascular Administration Set | | TKIV-F11 | Intravascular Administration Set | | TKIV-F12 | Intravascular Administration Set | | TKIV-F13 | Intravascular Administration Set | | TKIV-F14 | Intravascular Administration Set | | TKIV-F15 | Intravascular Administration Set | ### 5. Indications for use: TK Intravascular Administration Set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. ## 6. Comparison of technological characteristics with the predicate: TK Intravascular Administration Set is constructed of high grade extruded Polyvinyl Chloride (PVC). The primary components of TK Intravascular Administration Set are manufactured to identical or similar specifications of the predicate device listed above. The intended use, basic design, function and materials used are identical or similar to the predicate device. The differences between the subject device and the predicate device are the y-site in the subject device is a septum while the predicate utilized a needless luer access valve and the material for the stopcock is made of Makrolon for the subject device while the predicate is made of ABS HDPE. These differences in technology do not affect the safety and effectiveness of the subject device because the evaluation were performed using industry consensus standards and evaluated per the performance requirements of the device. Therefore, the different technological characteristics do not raise new questions of safety and effectiveness and is substantially equivalent to the predicate device. {6}------------------------------------------------ No.228 Weiyi Road ,Economic Development Zone ,Tianchang City, Anhui, China [TK Intravascular Administration Set] | Element of Comparison | SUBJECT DEVICE<br>K191640 | PREDICATE DEVICE<br>K151151 | Comment | |------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------| | Indications for Use | The TK Intravascular<br>Administration Set is a<br>device used to administer<br>fluids from a container to a<br>patient's vascular system<br>through a needle or catheter<br>inserted into a vein. | The U&U Intravascular<br>Administration Set is a<br>device used to administer<br>fluids from a container to a<br>patient's vascular system<br>through a needle or catheter<br>inserted into a vein. | Same | | Mode of Fluid Delivery | Gravity | Gravity | Same | | Y-site | Septum | Luer access valve | Different - see<br>discussion above | | Tubing | The tubing shall be<br>transparent or sufficiently<br>translucent so that the<br>interface of air and water<br>during the passage of air<br>bubbles can be observed<br>with normal or corrected<br>vision. | The tubing shall be<br>transparent or sufficiently<br>translucent so that the<br>interface of air and water<br>during the passage of air<br>bubbles can be observed<br>with normal or corrected<br>vision. | Same | # Intravascular Administration Set Comparison Table {7}------------------------------------------------ No.228 Weiyi Road ,Economic Development Zone ,Tianchang City, Anhui, China [TK Intravascular Administration Set] | Materials | | | | |-------------------------|-----------------------------------------------|-----------------------------------------------|------------------------------------------------| | Tubing | PVC | PVC | | | Drip chamber | PVC | PVC | | | Spike | ABS | ABS | | | Protective cap of spike | LDPE | LDPE | | | Micron filter | PTFE | PTFE | | | Air inlet | PVC | PVC | | | Fluid filter | ABS | ABS | Different - see<br>discussion above | | Slide clamp | ABS | ABS | | | Pinch clamp | POM | POM | Biocompatibility<br>testing per ISO<br>10993-1 | | Flow regulator | ABS | ABS | | | Roller Clamp | ABS | ABS | | | Male luer lock | MABS | MABS | | | Stopcock | MAKROLON | ABS HDPE | | | 4.0mm Connector | MABS | MABS | | | Y-site | MABS | MABS | | | Rotating luer lock | MABS | MABS | | | Back check valve | MABS | MABS | | | Luer lock cap | HDPE | HDPE | | | Pyrogenicity | <0.5 EU | <0.5 EU | Same | | | | | | | | Method: by EO | Method: by EO | | | | Sterility inspection is | Sterility inspection is | | | Sterility | based on the methods | based on the methods | Same | | | stipulated in USP<71>, | stipulated in USP<71>, | | | | and the results are in line | and the results are in line | | | | with requirements of | with requirements of | | | | USP<71>: No microbial<br>growth was observed. | USP<71>: No microbial<br>growth was observed. | | | Shelf life | 3 years | Unknown | Evaluated per<br>ASTM testing | | Labeling | Meet the requirements of<br>21 CFR Part 801 | Meet the requirements of<br>21 CFR Part 801 | Same | {8}------------------------------------------------ No.228 Weiyi Road ,Economic Development Zone ,Tianchang City, Anhui, China [TK Intravascular Administration Set] ## 7. Non Clinical Performance Testing - A. Biocompatibility Testing (ISO 10993-1:2018): - 1. Hemocompatibility (ISO 10993-4:2017) - 2. Cytotoxicity (ISO 10993-5:2017) - 3. Irritation (ISO 10993-10:2017) - 4. Sensitization (ISO 10993-10:2017) - 5. Acute Systemic Toxicity (ISO 10993-11:2017) - 6. Bacterial Endotoxins (ISO 10993-11:2017) - 7. Pyrogenicity (ISO 10993-11:2017) - 8. Subacute Systemic Toxicity (ISO 10993-11:2017)/ Subchronic Systemic Toxicity Study (ISO 10993-11:2017) - B. Other Non-clinical Test Conducted Using the Following Guidance and Standards: - FDA Guidance July 2018: Intravascular Administration Sets Premarket Notification Submissions [510(k)] - ISO 8536-4: 2010, Infusion equipment for medical use - Part 4: Infusion sets for single use, gravity feed [Including: Amendment 1(2013)] {9}------------------------------------------------ ## 510(k) Submission No.228 Weiyi Road ,Economic Development Zone ,Tianchang City, Anhui, China [TK Intravascular Administration Set] - ISO 80369-7: 2016, Small-bore connectors for liquids and gases in healthcare applications -- Part 7: Connectors for intravascular or hypodermic applications - ISO 80369-20: 2015, Small-bore connectors for liquids and gases in healthcare applications - Part 20: Common test methods - ISO 11607-1: 2016, Packaging for terminally sterilized medical device -- Part 1: Requirements for materials, sterile barrier systems and packaging systems - ISO 11607-2: 2016, Packaging for terminally sterilized medical devices -- Part 2: Validation requirements for forming, sealing and assembly processes - ISO 11135: 2014, Sterilization of health care products-Ethylene oxide-Requirements for the development, validation and routine control of a sterilization process for medical devices - . ISO10993-7: 2008, Biological evaluation of medical devices -- Part 7: Ethylene oxide sterilization residuals - ISO10993-12: 2012, Biological evaluation of medical devices-Part 12: Sample preparation and reference materials - ASTM F1980- 2016, Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices - ASTM F1929- 2015, Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration - ASTM F88- 2015, Standard Test Method for Seal Strength of Flexible Barrier Materials ● - USP 788- Particulate matter for Injection # 8. Conclusions: The Indications for Use, technological characteristics, and performance testing demonstrate that the subject device is substantially equivalent as the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...