K-SHIELD ADVANTAGE PORT ACCESS INFUSION SET (PAIS)
K123344 · Kawasumi Laboratories, Inc. · FPA · Mar 7, 2013 · General Hospital
Device Facts
Record ID
K123344
Device Name
K-SHIELD ADVANTAGE PORT ACCESS INFUSION SET (PAIS)
Applicant
Kawasumi Laboratories, Inc.
Product Code
FPA · General Hospital
Decision Date
Mar 7, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The devices are used to administer solutions to a surgically implanted port.
Device Story
K-Shield Advantage Port Access Infusion Set (PAIS) is a disposable, sterile, single-use intravascular administration set. Device utilizes a non-coring Huber needle to access surgically implanted ports for solution administration. Features include optional needle injection site or needleless access connector (NAC), tubing (micro or standard bore), and an optional Anti-Needlestick Protector (ANSP) that covers the needle tip upon removal to prevent accidental needle stick injuries and blood-borne pathogen exposure. Used in hospitals and healthcare facilities by clinicians. Device design facilitates safe port access and reduces occupational risk for healthcare providers.
Clinical Evidence
Bench testing only. Device performance verified against design requirements and specifications. Compliance with EN ISO 8536-4:2007 (Infusion equipment for medical use) and biocompatibility evaluation per ISO 10993-1:2009 confirmed safety and effectiveness.
Technological Characteristics
Materials: polypropylene, polyvinylchloride, stainless steel. Components: non-coring Huber needle, wing, tubing, clamp, female conical fitting, locking cap. Sterilization: Ethylene oxide per EN ISO 11135-1:2007. Biocompatibility: ISO 10993-1:2009. Configuration: Single-use, disposable, gravity feed. No software or electronic components.
Indications for Use
Indicated for patients requiring administration of solutions via a surgically implanted port in a hospital or healthcare facility setting.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
Kawasumi Port Access Infusion Set with Anti Needle Stick Protector (K060580)
Submission Summary (Full Text)
{0}------------------------------------------------
K123344
MAR 0 7 2013
# 5. 510(k) Summary
#### Contact information
Submitter:
Mr. Takeshi Yuasa General Manager, Overseas Regulatory Affairs Kawasumi Laboratories, Inc. Shinagawa Intercity Tower B, 9th Floor 2-15-2, Konan, Minato-ku, Tokyo 108-6109, Japan PHONE: 81-3-5769-2664
Contact: Mr. Kazuhiko Hashimoto Kawasumi Laboratories America, Inc. 4723 Oak Fair Blvd. Tampa, Florida 33610 PHONE: 813-630-5554 FAX: 813-630-5033 E-mail: KHashimoto@kawasumiamerica.com
Date of preparation: 10/29/2012
Device Name: K-Shield Advantage Port Access Infusion Set
Proprietary name: K-Shield Advantage Port Access Infusion Set
Common name: Port Access Products (PAIS series)
Classification Name: set, administration, intravascular
Product Code: FPA
#### Predicate Information
The proposed K-Shield Advantage Port Access Infusion Set (PAIS) is substantially equivalent in both function and use to the predicate device, Kawasumi Port Access Infusion Set with Anti Needle Stick Protector (K060580 cleared on 06/06/2006). The predicate is a legally marketed device with the same product code as the proposed device.
{1}------------------------------------------------
# Device Description
The K-Shield Advantage Port Access Infusion Set (PAIS) is a safety port access device used to administer solutions to a surgically implanted port. This device is designed utilizing a non-coring Huber needle to access the implanted port. The K-Shield Advantage PAIS has an integral safety device intended to protect against accidental needle stick injuries and infection caused by blood borne pathogens. Upon removal from the port, the Anti- Needlestick Protector (ANSP) covers the needle tip protecting against accidental needle stick injuries.
The devices are disposable ethylene oxide sterilized medical devices which are constructed from non-coring needle (Huber needle), wing, tubing (either micro bore or standard bore), clamp, female conical fitting and locking cap. The device has optional injection site (needle injection, needleless access connector (NAC), or no injection site) and/or anti-needlestick protector (ANSP).
| Tubing | Anti Needle Stick Protector | Injection Site |
|---------------|-----------------------------|-------------------------------------------------------------------------|
| Micro bore | With | Needle injection site<br>NAC (Needleless Access<br>Connector<br>without |
| | Without | Needle injection site<br>NAC (Needleless Access<br>Connector<br>without |
| Standard bore | With | Needle injection site<br>NAC (Needleless Access<br>Connector<br>without |
| | Without | Needle injection site<br>NAC (Needleless Access<br>Connector<br>without |
# Table 1. Table of Possible Device Configurations:
### Device Characteristics:
The devices are single use, disposable ethylene oxide sterilized medical devices. Sterilization validation was conducted per EN ISO 11135-1:2007 Sterilization of health care products - Ethylene oxide - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices. The devices are considered blood path, indirect contact per section 5.2.2(a) of ISO 10993-1: 2009 Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process.
{2}------------------------------------------------
# Environment of Use:
The PAIS is intended for use at healthcare facilities or in hospitals.
# Materials of Use
General type of material used: polypropylene, polyvinylchloride, stainless steel. The materials are evaluated per ISO 10993-1: 2009 Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process and have been determined to be biocompatible.
#### Intended Use
The devices are used to administer solutions to a surgically implanted port.
The indications are equivalent to the predicate device.
#### Technological Characteristics
The device is technologically equivalent to the predicate device. Both devices are safety port access devices used to administer solutions to a surgically implanted port. The devices come in similar sizes and have similar components. Both devices are sterilized under ethylene oxide-gas sterilization.
#### Performance Data
Bench testing was performed and confirms that the device meets design requirements and specifications. The devices comply with the International standard EN ISO 8536-4: 2007-Infustion equipment for medical use -Part 4: Infusion sets for single use, gravity feed mainly, to be used primarily in a hospital / healthcare facility setting. The devices were evaluated to be safe and effective as a medical device based on design verification test results.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
### March 7, 2013
Kawasumi Laboratories. Incorporated C/O Ms. Christina Henza Regulatory Affairs Specialist Regulatory Compliance Associates, Incorporated 7401 104th Avenue, Suite 160 KENOSHA WI 53142
Re: K123344
Trade/Device Name: K-Shield Advantage Port Access Infusion Set (PAIS) Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: December 11, 2012 Received: January 28, 2013
Dear Ms. Henza:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any-Federal statutes and regulations administered by other-Federal-agencies. You-must-comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Image /page/4/Picture/7 description: The image contains a signature with the word "for" at the top left. The signature is illegible, but it appears to contain the words "with" and "urohity" or a similar word. The signature is written in black ink on a white background, and there is some background noise or texture visible in the image.
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
510(k) Number (if known): Pending->
Device Name: K-Shield Advantage Port Access Infusion Set (PAIS)
Indications For Use: The devices are used to administer solutions to a surgically implanted port.
× Prescription Use _ (Part 21 CFR 801 Subpart D)
:
AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Sajjad H. Syed
Digitally signed by Sajjad H. Syed
DN; C=US, o=U.S. Government, ou=HHS,
ou=FDA au People, cn=Sajjad H. Syed,
0.9.2342.19200300.100.1.1-2000601742
Date: 2013.03.05 13:23:33-05'00'
(Division Sign-Off)
Division of Anesthesiology, General Hospital
Infection Control, Dental Devices
| 510(k) Number: | K123344 |
|----------------|---------|
|----------------|---------|
Page 1 of __1_
7
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
510(k) Number:__
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.