VITA ENAMIC Implant Solutions (IS)

K153645 · Vita Zahnfabrik H.Rauter GmbH Co. · NHA · Jun 30, 2016 · Dental

Device Facts

Record IDK153645
Device NameVITA ENAMIC Implant Solutions (IS)
ApplicantVita Zahnfabrik H.Rauter GmbH Co.
Product CodeNHA · Dental
Decision DateJun 30, 2016
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 872.3630
Device ClassClass 2
AttributesTherapeutic

Intended Use

VITA ENAMIC Implant Solutions is indicated for use as a component of a two-piece abutment system, consisting of a tibase component, and a mesostructure or abutment crown. VITA ENAMIC Implant Solutions is intended for use in partially or fully edentulous mandible or maxillae for fabrication of permanent, anterior or posterior, single unit CAD/CAM abutments. VITA ENAMIC Implant Solutions is indicated for use in conjunction with the CAD/CAM component of the Sirona Dental CAD/CAM system and the following Tibase abutments and implant bodies:

Device Story

VITA ENAMIC Implant Solutions (IS) is a hybrid ceramic CAD/CAM block (~86% feldspar ceramic, ~14% polymer) used to fabricate dental implant abutments or abutment crowns. The device is supplied as a block with a mandrel attachment and a pre-drilled hole. In a clinical setting, a dental professional uses Sirona CAD/CAM systems (inLab 15.0+, CEREC 4.4+) to mill the block into a custom mesostructure or abutment crown. This milled component is then bonded to a compatible titanium (Ti) base. The final two-piece assembly is secured to the patient's dental implant via a screw. The device allows for restorative customization in edentulous patients. The hybrid material is designed to mimic natural tooth properties, providing a restorative solution that integrates with existing implant systems.

Clinical Evidence

Bench testing only. No human clinical testing performed. Evidence includes dynamic fatigue testing per ISO 14801:2007, and compliance with ISO 6872:2008 (ceramic materials) and ISO 10477:2004 (polymer-based crown/bridge materials). Biocompatibility assessment performed per ISO 10993-1:2009 and ISO 7405:2008, leveraging data from reference device VITA ENAMIC.

Technological Characteristics

Hybrid ceramic block (~86% feldspar ceramic, ~14% polymer). Dimensions: 12x14x18 mm and 18x16x18 mm. Features pre-drilled hole for TiBase interface and mandrel for CAD/CAM milling. Energy source: mechanical milling. Connectivity: compatible with Sirona CAD/CAM systems. Sterilization: not specified (restorative material).

Indications for Use

Indicated for partially or fully edentulous patients requiring permanent, anterior or posterior, single-unit CAD/CAM dental implant abutments or abutment crowns, used in conjunction with Sirona TiBase systems.

Regulatory Classification

Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of three human profiles facing right, stacked on top of each other to create a sense of depth and unity. The profiles are rendered in a dark color, contrasting with the white background. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 30, 2016 Vita Zahnfabrik H. Rauter GmbH Co. % Ms. Nevine Erian Director, Regulatory Affairs & Compliance (consultant) Vita North America, Inc. 22705 Savi Ranch Parkway, Suite 100 Yorba Linda. California 92887 Re: K153645 Trade/Device Name: VITA ENAMIC® Implant Solutions (IS) Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: May 27, 2016 Received: June 3, 2016 Dear Ms. Erian: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Tina Kiang -S for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use #### 510(k) Number (if known) K153645 Device Name VITA ENAMIC Implant Solutions (IS) #### Indications for Use VITA ENAMIC Implant Solutions is indicated for use as a component of a two-piece abutment system, consisting of a tibase component, and a mesostructure or abutment crown. VITA ENAMIC Implant Solutions is intended for use in partially or fully edentulous mandible or maxillae for fabrication of permanent, anterior or posterior, single unit CAD/CAM abutments. VITA ENAMIC Implant Solutions is indicated for use in conjunction with the CAD/CAM component of the Sirona Dental CAD/CAM system and the following Tibase abutments and implant bodies: | Table 1 - Compatible Sirona Ti Base/Implant System and VITA ENAMIC IS Interface | | | | | | | |---------------------------------------------------------------------------------|-------------------|------------------|-----------------------------|-------------------------|-----------------|---------------------| | Sirona<br>TiBase Type | Abutment<br>Screw | Sirona<br>Part # | VITA ENAMIC<br>IS Interface | Implant<br>Manufacturer | Implant System | Implant<br>Diameter | | NBRS 3.5 | M1.8 | 6282474 | L | Nobel Biocare | Replace® NP | 3.5 mm | | NBRS 4.3 | M2 | 6282482 | L | Nobel Biocare | Replace® RP | 4.3 mm | | NBRS 5.0 | M2 | 6282490 | L | Nobel Biocare | Replace® WP | 5.0 mm | | NBRS 6.0 | M2 | 6282508 | L | Nobel Biocare | Replace® 6.0 | 6.0 mm | | NBB 3.4 | M1.6 | 6282516 | L | Nobel Biocare | Brånemark® | 3.3 mm | | NBB 4.1 | M2 | 6282524 | L | Nobel Biocare | Brånemark® | 3.75 / 4.0 mm | | NB A 4.5 | M1.6 | 6308188 | L | Nobel Biocare | Nobel Active NP | 3.5 mm | | NB A 5.0 | M2 | 6308253 | L | Nobel Biocare | Nobel Active RP | 4.3 / 5.0 mm | | SSO 3.5 | M1.8 | 6284231 | L | Straumann® | Tissue level NN | 3.5 mm | | SSO 4.8 | M2 | 6284249 | L | Straumann® | Tissue Level RN | 4.8 mm | | SSO 6.5 | M2 | 6284256 | L | Straumann® | Tissue Level WN | 6.5 mm | | S BL 3.3 | M1.6 | 6308154 | L | Straumann® | Bone Level NC | 3.3mm | | S BL 4.1 | M1.6 | 6308337 | L | Straumann® | Bone Level NC | 4.1 / 4.8 mm | | ATOS 3.5/4.0 | M1.6 | 6282532 | L | Astra Tech | OsseoSpeed™ | 3.5 S/4.0 S mm | | ATOS 4.5/5.0 | M1.6 | 6282540 | L | Astra Tech | OsseoSpeed™ | 4.5 / 5.0 mm | | FX 3.4 | M1.6 | 6282433 | S | Friadent | Frialit®Xive® | 3.4 mm | | FX 3.8 | M1.6 | 6282441 | S | Friadent | Frialit®Xive® | 3.8 mm | | FX 4.5 | M1.6 | 6282458 | L | Friadent | Frialit®Xive® | 4.5 mm | | FX 5.5 | M1.6 | 6282466 | L | Friadent | Frialit®Xive® | 5.5 mm | | BO 3.4 | M2 | 6282557 | L | Biomet 3i | Ex. hex | 3.4 mm | | BO 4.1 | M2 | 6282565 | L | Biomet 3i | Ex. hex | 4.1 mm | | BO 5.0 | M2 | 6282573 | L | Biomet 3i | Ex. hex | 5.0 mm | | B C 3.4 | M1.6 | 6308048 | S | Biomet 3i | Certain® | 3.5 mm | | B C 4.1 | M1.6 | 6308097 | L | Biomet 3i | Certain® | 4.5 mm | | B C 5.0 | M1.6 | 6308121 | L | Biomet 3i | Certain® | 5.7 mm | {3}------------------------------------------------ | Table 1 – Compatible Sirona Ti Base/Implant System and VITA ENAMIC IS Interface | | | | | | | |---------------------------------------------------------------------------------|-------------------|------------------|-----------------------------|-------------------------|---------------------|---------------------| | Sirona<br>TiBase Type | Abutment<br>Screw | Sirona<br>Part # | VITA ENAMIC<br>IS Interface | Implant<br>Manufacturer | Implant System | Implant<br>Diameter | | ZTSV 3.5 | M1.8 | 6282581 | L | Zimmer | Tapered screw-Vent® | 3.5 mm | | ZTSV 4.5 | M1.8 | 6282599 | L | Zimmer | Tapered screw-Vent® | 4.5 mm | | ZTSV 5.7 | M1.8 | 6282607 | L | Zimmer | Tapered screw-Vent® | 5.7 mm | Prescription Use _____X (21 CFR Part 801 Subpart D) AND/OR Over-The-Counter Use _ (21 CFR Part 807 Subpart C) # (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON SEPARATE PAGE IF NEEDED) # Concurrence of CDRH, Office of Device Evaluation (ODE) {4}------------------------------------------------ # K153645 510(k) Summary | Submitter | Vita Zahnfabrik H.Rauter GmbH Co.<br>Spitelgasse 3<br>Bad Sackingen, D-79713<br>Germany<br>Establishment Reg. No. 1000625496 | |------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact | Bernd Walker<br>Head of Regulatory Affairs and Quality Systems<br>Phone (+49) 7761/562-361<br>Fax (+49) 7761/562-384 | | Official Correspondent | Nevine Erian<br>Director, Regulatory Affairs & Compliance (Consultant)<br>VITA North America, Inc.<br>22705 Savi Ranch Parkway, Suite 100<br>Yorba Linda, CA 92887<br>Establishment Reg. No. 2082832<br>Phone (949) 370-7155<br>Fax (714) 221-6759<br>E-mail: nerian@vident.com | | Date Prepared | June 30, 2016 | | Trade/Device Name<br>■ | VITA ENAMIC® Implant Solutions (IS) | | ■<br>Common Name | Abutment, implant, dental, endosseous | - Classification Name Endosseous Dental Implant Abutment - Regulation Number 21 CFR 872.3630 - . Product Code NHA #### Predicate Devices inCoris ZI meso (Sirona Dental) – K111421 – Primary Predicate VITA ENAMIC® (VITA Zahnfabrik GmbH) – K122269 IPS e.max® CAD Abutment Solutions (Ivoclar Vivadent) – K132209 {5}------------------------------------------------ ### Device Description VITA ENAMIC Implant Solutions (IS) is a hybrid ceramic CAD/CAM block, consisting of ~86% (by weight) feldspar ceramic and ~14% polymer. The VITA ENAMIC Implant Solutions is milled then bonded to a titanium (Ti) base for a two-element solution with mesostructure or a single-element solution with abutment crown. The compatible Ti bases to VITA ENAMIC IS block interface size and CAD/CAM system are as follows: | Ti Base: | Sirona TiBase (K111421) | |-----------------|---------------------------------------------------------------------------------| | CAD/CAM System: | Sirona software – inlab 15.0 and above<br>Sirona software – CEREC 4.4 and above | | Titanium Bases: | Compatible titanium bases are shown in Table 1 | | Table 1 - Compatible Sirona Ti Base/Implant System and VITA ENAMIC IS Interface | | | | | | | |---------------------------------------------------------------------------------|-------------------|------------------|-----------------------------|-------------------------|---------------------|---------------------| | Sirona<br>TiBase Type | Abutment<br>Screw | Sirona<br>Part # | VITA ENAMIC<br>IS Interface | Implant<br>Manufacturer | Implant System | Implant<br>Diameter | | NBRS 3.5 | M1.8 | 6282474 | L | Nobel Biocare | Replace® NP | 3.5 mm | | NBRS 4.3 | M2 | 6282482 | L | Nobel Biocare | Replace® RP | 4.3 mm | | NBRS 5.0 | M2 | 6282490 | L | Nobel Biocare | Replace® WP | 5.0 mm | | NBRS 6.0 | M2 | 6282508 | L | Nobel Biocare | Replace® 6.0 | 6.0 mm | | NBB 3.4 | M1.6 | 6282516 | L | Nobel Biocare | Brånemark® | 3.3 mm | | NBB 4.1 | M2 | 6282524 | L | Nobel Biocare | Brånemark® | 3.75 / 4.0 mm | | NB A 4.5 | M1.6 | 6308188 | L | Nobel Biocare | Nobel Active NP | 3.5 mm | | NB A 5.0 | M2 | 6308253 | L | Nobel Biocare | Nobel Active RP | 4.3 / 5.0 mm | | SSO 3.5 | M1.8 | 6284231 | L | Straumann® | Tissue level NN | 3.5 mm | | SSO 4.8 | M2 | 6284249 | L | Straumann® | Tissue Level RN | 4.8 mm | | SSO 6.5 | M2 | 6284256 | L | Straumann® | Tissue Level WN | 6.5 mm | | S BL 3.3 | M1.6 | 6308154 | L | Straumann® | Bone Level NC | 3.3mm | | S BL 4.1 | M1.6 | 6308337 | L | Straumann® | Bone Level NC | 4.1 / 4.8 mm | | ATOS 3.5/4.0 | M1.6 | 6282532 | L | Astra Tech | OsseoSpeed™ | 3.5 S/4.0 S mm | | ATOS 4.5/5.0 | M1.6 | 6282540 | L | Astra Tech | OsseoSpeed™ | 4.5 / 5.0 mm | | FX 3.4 | M1.6 | 6282433 | S | Friadent | Frialit®Xive® | 3.4 mm | | FX 3.8 | M1.6 | 6282441 | S | Friadent | Frialit®Xive® | 3.8 mm | | FX 4.5 | M1.6 | 6282458 | L | Friadent | Frialit®Xive® | 4.5 mm | | FX 5.5 | M1.6 | 6282466 | L | Friadent | Frialit®Xive® | 5.5 mm | | BO 3.4 | M2 | 6282557 | L | Biomet 3i | Ex. hex | 3.4 mm | | BO 4.1 | M2 | 6282565 | L | Biomet 3i | Ex. hex | 4.1 mm | | BO 5.0 | M2 | 6282573 | L | Biomet 3i | Ex. hex | 5.0 mm | | B C 3.4 | M1.6 | 6308048 | S | Biomet 3i | Certain® | 3.5 mm | | B C 4.1 | M1.6 | 6308097 | L | Biomet 3i | Certain® | 4.5 mm | | B C 5.0 | M1.6 | 6308121 | L | Biomet 3i | Certain® | 5.7 mm | | ZTSV 3.5 | M1.8 | 6282581 | L | Zimmer | Tapered screw-Vent® | 3.5 mm | | ZTSV 4.5 | M1.8 | 6282599 | L | Zimmer | Tapered screw-Vent® | 4.5 mm | | ZTSV 5.7 | M1.8 | 6282607 | L | Zimmer | Tapered screw-Vent® | 5.7 mm | {6}------------------------------------------------ ## Statement of Intended Use VITA ENAMIC Implant Solutions is indicated for use as a component of a two-piece abutment system, consisting of a tibase component, and a mesostructure or abutment crown. VITA ENAMIC Implant Solutions is intended for use in partially or fully edentulous mandible or maxillae for fabrication of permanent, anterior or posterior, single unit CAD/CAM abutments. VITA ENAMIC Implant Solutions is indicated for use in conjunction with the CAD/CAM component of the Sirona Dental CAD/CAM system and the following Tibase abutments and implant bodies: | Table 1 - Compatible Sirona Ti Base/Implant System and VITA ENAMIC IS Interface | | | | | | | |---------------------------------------------------------------------------------|-------------------|------------------|-----------------------------|-------------------------|---------------------|---------------------| | Sirona<br>TiBase Type | Abutment<br>Screw | Sirona<br>Part # | VITA ENAMIC<br>IS Interface | Implant<br>Manufacturer | Implant System | Implant<br>Diameter | | NBRS 3.5 | M1.8 | 6282474 | L | Nobel Biocare | Replace® NP | 3.5 mm | | NBRS 4.3 | M2 | 6282482 | L | Nobel Biocare | Replace® RP | 4.3 mm | | NBRS 5.0 | M2 | 6282490 | L | Nobel Biocare | Replace® WP | 5.0 mm | | NBRS 6.0 | M2 | 6282508 | L | Nobel Biocare | Replace® 6.0 | 6.0 mm | | NBB 3.4 | M1.6 | 6282516 | L | Nobel Biocare | Brånemark® | 3.3 mm | | NBB 4.1 | M2 | 6282524 | L | Nobel Biocare | Brånemark® | 3.75 / 4.0 mm | | NB A 4.5 | M1.6 | 6308188 | L | Nobel Biocare | Nobel Active NP | 3.5 mm | | NB A 5.0 | M2 | 6308253 | L | Nobel Biocare | Nobel Active RP | 4.3 / 5.0 mm | | SSO 3.5 | M1.8 | 6284231 | L | Straumann® | Tissue level NN | 3.5 mm | | SSO 4.8 | M2 | 6284249 | L | Straumann® | Tissue Level RN | 4.8 mm | | SSO 6.5 | M2 | 6284256 | L | Straumann® | Tissue Level WN | 6.5 mm | | S BL 3.3 | M1.6 | 6308154 | L | Straumann® | Bone Level NC | 3.3mm | | S BL 4.1 | M1.6 | 6308337 | L | Straumann® | Bone Level NC | 4.1 / 4.8 mm | | ATOS 3.5/4.0 | M1.6 | 6282532 | L | Astra Tech | OsseoSpeed™ | 3.5 S/4.0 S mm | | ATOS 4.5/5.0 | M1.6 | 6282540 | L | Astra Tech | OsseoSpeed™ | 4.5 / 5.0 mm | | FX 3.4 | M1.6 | 6282433 | S | Friadent | Frialit®Xive® | 3.4 mm | | FX 3.8 | M1.6 | 6282441 | S | Friadent | Frialit®Xive® | 3.8 mm | | FX 4.5 | M1.6 | 6282458 | L | Friadent | Frialit®Xive® | 4.5 mm | | FX 5.5 | M1.6 | 6282466 | L | Friadent | Frialit®Xive® | 5.5 mm | | BO 3.4 | M2 | 6282557 | L | Biomet 3i | Ex. hex | 3.4 mm | | BO 4.1 | M2 | 6282565 | L | Biomet 3i | Ex. hex | 4.1 mm | | BO 5.0 | M2 | 6282573 | L | Biomet 3i | Ex. hex | 5.0 mm | | B C 3.4 | M1.6 | 6308048 | S | Biomet 3i | Certain® | 3.5 mm | | B C 4.1 | M1.6 | 6308097 | L | Biomet 3i | Certain® | 4.5 mm | | B C 5.0 | M1.6 | 6308121 | L | Biomet 3i | Certain® | 5.7 mm | | ZTSV 3.5 | M1.8 | 6282581 | L | Zimmer | Tapered screw-Vent® | 3.5 mm | | ZTSV 4.5 | M1.8 | 6282599 | L | Zimmer | Tapered screw-Vent® | 4.5 mm | | ZTSV 5.7 | M1.8 | 6282607 | L | Zimmer | Tapered screw-Vent® | 5.7 mm | {7}------------------------------------------------ ### Material Composition VITA ENAMIC IS is identical in composition to VITA ENAMIC (K122269). #### Technological Characteristics VITA ENAMIC IS is offered in a block form, with a mandrel attachment, to permit securing it into a CAD/CAM machine for milling into its final form. The VITA ENAMIC IS blocks have a pre-drilled hole compatible with Sirona TiBase (K111421) to allow the mesostructure or abutment crown bonded to a titanium base to be connected to the implant. VITA ENAMIC IS blocks are identical to VITA ENAMIC blocks (K122269) in geometry, with the exception of a pre-drilled hole. Because the subject device and its predicates inCoris Zl meso, IPS e.max CAD Abutment Solutions do not share similar chemical or physical properties, a reference device, VITA Enamic, was introduced by the sponsor. The differences between VITA ENAMIC IS and the predicate devices do not render the subject device not substantially equivalent for the following reasons: - ı VITA ENAMIC IS has non-identical, but similar indications as the inCoris Zl meso and IPS e.max CAD Abutment Solutions, and meets the same biocompatibility, ISO 6872:2008 and ISO 14801:2007 test criteria. While the verbiage of the indications for use statements differs, the intended use for each is similar. All three are restorative blocks intended to serve as milled components of dental implant abutments and abutment crowns, and all three share implant compatibilities with the same Sirona TiBase devices. - . VITA ENAMIC IS has the identical chemical composition and physical properties as VITA ENAMIC. As such, the data supports the substantial equivalence of the subject and predicate devices. #### Non-Clinical Performance Testing VITA ENAMIC IS meets the applicable requirements of the following FDA recognized standards: - . ISO 6872:2008 – Dentistry – Ceramic materials - . ISO 10477:2004(E) - Dentistry - Polymer-based crown and bridge materials standard. {8}------------------------------------------------ - I ISO 14801:2007 - Dentistry - Implants - Dynamic fatigue test for endosseous dental implants Bench test results allowed us to conclude that VITA ENAMIC IS is well suited for its intended use. ## Biocompatibility A biocompatibility assessment was performed on VITA ENAMIC IS in accordance with ISO 10933-1:2009 – Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process, and ISO 7405:2008 – Dentistry – Evaluation of Biocompatibility of Medical Devices Used in Dentistry. Testing from reference device VITA ENAMIC was leveraged to support biocompatibility. This assessment supports that VITA ENAMIC IS is biocompatible and concludes that the device is compatible for its intended use. # Clinical Performance Data Not applicable. No human clinical testing was performed to support the substantial equivalence of VITA ENAMIC IS. #### Substantial Equivalence #### Material VITA ENAMIC IS has the identical chemical composition as the predicate device, VITA ENAMIC (K122269). # Physical Properties VITA ENAMIC IS has identical physical properties as the predicate device, VITA ENAMIC (K122269). | Technical<br>Attributes | VITA<br>ENAMIC IS | inCoris Zl meso | IPS e.max CAD<br>Abutment Solutions | VITA<br>ENAMIC | |---------------------------|-------------------------------------------------------------|-------------------------------------------------------------|------------------------------------------------------------|----------------------------------| | Indication | Implant Abutment<br>and Abutment<br>Crown | Implant Abutment<br>and Abutment<br>Crown | Implant Abutment and<br>Abutment Crown | Crown & Veneers | | CAD/CAM<br>Material | Yes | Yes | Yes | Yes | | Physical<br>Configuration | Block with a pre-<br>drilled hole and<br>mandrel attachment | Block with a pre-<br>drilled hole and<br>mandrel attachment | Block with a pre-drilled<br>hole and mandrel<br>attachment | Block with mandrel<br>attachment | ### Technical Comparison of VITA ENAMIC IS to Predicate Devices {9}------------------------------------------------ | Technical<br>Attributes | VITA<br>ENAMIC IS | inCoris ZI meso | IPS e.max CAD<br>Abutment Solutions | VITA<br>ENAMIC | |-------------------------------------|----------------------------------------------------------------------------|----------------------------------------------------|-------------------------------------------------------------------------------------------|---------------------------------------------------------------------------| | Block Sizes | 12 x 14 x 18 mm and<br>18 x 16 x 18 mm<br>with Small or Large<br>hole size | One block size with<br>Small or Large hole<br>size | 12.4 x 14.5 x 18.0 mm<br>and<br>17.8 x 15.8 x 18.0 mm<br>with Small or Large hole<br>size | 12 x 14 x 18 mm and<br>8 x 10 x 15 mm<br>with Small or Large<br>hole size | | Shades &<br>Translucency | Different shades<br>Translucent &<br>High Translucent | Two shades | Different shades<br>Medium Opacity & Low<br>Translucency | Different shades<br>Translucent &<br>High Translucent | | Bonds to<br>Titanium Base | Sirona TiBase with<br>Small & Large hole | Sirona TiBase with<br>Small & Large hole | Sirona TiBase with Small<br>& Large hole | No | | Maximal<br>Abutment<br>Angulation | 20° | 20° | 20° | Not applicable | | Abutment<br>Fixation | Screwed | Screwed | Screwed | Not applicable | | Screw | Part of implant<br>system | Part of implant<br>system | Part of implant system | Not applicable | | Compatible<br>Titanium Base | Sirona TiBase<br>(K11142) | Sirona TiBase<br>(K11142) | Sirona TiBase<br>(K11142) | Not applicable | | 1 or 2 piece<br>Abutment | 2 piece — ENAMIC<br>IS to TiBase | 2 piece — Incoris ZI<br>to TiBase | 2 piece - IPS e.max<br>CAD to TiBase | Not applicable | | Mechanical<br>Testing | Dynamic Fatigue<br>Testing per ISO<br>14801 | Dynamic Fatigue<br>Testing per ISO<br>14801 | Dynamic Fatigue<br>Testing per ISO 14801 | Not applicable | | Flexural<br>Strength | 140 ± 10 MPa | > 900 MPa | 360 ± 60 MPa | 140 ± 10 MPa | | Fracture<br>Toughness | 1.5 MPa m0.5 | 5.9 Mpa m0.5 | 2.0 – 2.5 MPa m0.5 | 1.5 MPa m0.5 | | Young's<br>Modulus of<br>Elasticity | 30 GPa | Unknown | 95 ± 5 GPa | 30 GPa | # Conclusion Information provided by VITA North America, Inc. demonstrates that VITA ENAMIC Implant Solutions is substantially equivalent to the predicate devices. VITA ENAMIC IS shares the same indications and similar physical properties as the nCoris ZI meso and IPS e.max CAD Abutment Solutions. VITA ENAMIC IS has identical chemical composition and physical properties as VITA ENAMIC. - Comparisons of the indications for use, chemical composition, technological characteristics and physical properties of VITA ENAMIC IS to the predicate devices all support the substantial equivalence of VITA ENAMIC Implant Solutions to its predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...