Dentis s-Clean Abutment Mini

K220440 · Dentis Co., Ltd. · NHA · Jun 16, 2022 · Dental

Device Facts

Record IDK220440
Device NameDentis s-Clean Abutment Mini
ApplicantDentis Co., Ltd.
Product CodeNHA · Dental
Decision DateJun 16, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3630
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Dentis s-Clean Abutment Mini is an endosseous dental implant that is indicated for surgical placement in the upper and lower jaw arches, to provide a root form means for single or multiple-units prosthetic appliance attachment to restore a patient's chewing function. Implants can be placed with a conventional two stage surgical process with an option for transmucosal healing or they can be placed in a single stage surgical process for immediate loading when good primary stability has been achieved and with appropriate occlusal loading.

Device Story

The Dentis s-Clean Abutment Mini is a dental implant abutment system used to connect prosthetic appliances to endosseous dental implants. The device is used by dental clinicians in a clinical setting. It acts as an interface between the implant fixture and the final prosthesis, supporting chewing function. The system includes various abutment types (O-Ring, TiN-coated Sub-Octa, and CCM UCLA) that are screw-retained to the implant fixture. The clinician selects the appropriate abutment based on the clinical requirement (e.g., overdenture retention or fixed restoration). The device is provided non-sterile for end-user sterilization. By providing a stable connection to the implant, the device enables the restoration of dental function for the patient.

Clinical Evidence

No clinical data. Substantial equivalence is supported by non-clinical bench testing, including biocompatibility (ISO 10993-1), end-user sterilization validation (ANSI/AAMI ST79, ISO 17665, ISO 11737, ISO 11138), and MRI safety assessment based on scientific rationale and published literature.

Technological Characteristics

Materials: Ti-6Al-4V ELI (ASTM F136) and Cobalt-28Chromium-6Molybdenum Alloy (ASTM F1537). Features TiN coating on specific models. Connection: Screw-retained. Dimensions: Various diameters (4.0mm–4.8mm) and lengths. Sterilization: End-user sterilization required. No angled abutments; fatigue testing not required per FDA guidance.

Indications for Use

Indicated for patients requiring single or multiple-unit prosthetic restoration in the upper or lower jaw arches. Suitable for conventional two-stage surgical processes with transmucosal healing or single-stage processes with immediate loading, provided primary stability and appropriate occlusal loading are achieved.

Regulatory Classification

Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. Dentis Co., Ltd. % April Lee Consultant Withus Group Inc 106 Superior Irvine, California 92620 Re: K220440 Trade/Device Name: Dentis s-Clean Abutment Mini Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: May 14, 2022 Received: May 17, 2022 Dear April Lee: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Andrew I Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K220440 Device Name Dentis s-Clean Abutment Mini #### Indications for Use (Describe) The Dentis s-Clean Abutment Mini is an endosseous dental implant that is indicated for surgical placement in the upper and lower jaw arches, to provide a root form means for single or multiple-units prosthetic appliance attachment to restore a patient's chewing function. Implants can be placed with a conventional two stage surgical process with an option for transmucosal healing or they can be placed in a single stage surgical process for immediate loading when good primary stability has been achieved and with appropriate occlusal loading. | Type of Use (Select one or both, as applicable) | | |---------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------| | <div> <span style="font-family: DejaVu Sans, sans-serif;">✓</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span style="font-family: DejaVu Sans, sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | 2 Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### 510(k) Summary #### Submitter Dentis Co., Ltd. Gyu Ri Kim 99, Seongseoseo-ro, Dalseo-gu Daegu, 42718 Korea Email: kgr1026@dentis.co.kr Tel. +82-53-589-3541 Fax. +82-53-289-7922 ## Official Correspondent Withus Group Inc. April Lee 106 Superior, Irvine, CA 92620 USA Email: withus6664@gmail.com Phone: 1-909-274-9971 Fax: 1-909-460-8122 ### Device Information - . Trade Name: Dentis s-Clean Abutment Mini - . Common Name: Endosseous dental implant abutment - Classification Name: Endosseous dental implant abutment ● - . Product Code: NHA - Panel: Dental - Regulation Number: 872.3630 - Device Class: Class II - Date prepared: 06/15/2022 ● ### Predicate Devices: Primary Predicate K171027, Dentis Dental Implant System by Dentis Co., Ltd Reference Device K161244, s-Clean OneQ-SL Narrow Implant System by Dentis Co., Ltd. K210080, Dentis s-Clean s-Line Mini by Dentis Co., Ltd. #### Indication for Use: The Dentis s-Clean Abutment Mini is an endosseous dental implant that is indicated for surgical placement in the upper and lower jaw arches, to provide a root form means for single or multiple-units prosthetic appliance attachment to restore a patient's chewing function. Implants can be placed with a conventional two stage surgical process with an option for transmucosal healing or they can be placed in a single stage surgical process for immediate loading when good primary stability has been achieved and with appropriate occlusal loading. #### Device Description: Dentis s-Clean Abutment Mini is composed of Abutments are made of Ti-6AI-4V ELI according to ASTM F136 and Cobalt-28Chromium-6Molybdenum Alloy ASTM F1537. The s-Clean O-Ring Abutment s-Line and s-Clean TiN Half Coating Sub-Octa Abutment is screw assembled design but s-Clean CCM UCLA Abutment s-Line is screw separation design. The s-Clean TiN Half Coating Sub-Octa Abutment is always to be combined with a casting abutment (e.g., Gold Cylinder) or burn-out plastic sleeve for single-unit loading. The minimum post height of s {4}------------------------------------------------ Clean TiN Half Coating Sub-Octa Abutment when combined with casting abutment (e.g., Gold Cylinder) or burn-out plastic sleeve for single-unit loading is 4mm. The minimum post height of s-Clean CCM UCLA Abutment s-Line for single-unit loading is 4mm. The Subject device is compatible with implants cleared in K210080. Compatible fixture diameters are Ø5.8. 6.8 and Ø7.8 and lengths are 7.5. 9.5, 11.4 and 11.5mm. Compatible fixture, K210080, has dual screw-joint (structured with Upper and Lower screw joint) and a hex anti-rotation design connection. As explained in K210080, there are two-screw-joint features, 1.6M screw (Lower part) or 2.0M Screw (Upper Part) for diverse screw connection with other abutments. s-Clean O-Ring Abutment s-Line and s-Clean TiN Half Coating Sub-Octa Abutment have Upper M2.0 Screw and s-Clean CCM UCLA Abutment s-Line has Lower M1.6 screw. When dividing screw assemble design, screw assembled one-piece abutment is connected with upper screw and screw separated twopiece abutment is connected with lower screw. | No. | Device Name | Diameter X Total Length | Angulation | |-----|--------------------------------------------|----------------------------------------------------------------|------------| | 1 | s-Clean O-Ring Abutment s-Line | $\varnothing$ 4.5 X 10.71 and 12.71mm | 0° | | 2 | s-Clean TiN Half Coating Sub-Octa Abutment | $\varnothing$ 4.8 X 8.16, 8.66, 9.66, 10.66, 11.66 and 12.66mm | 0° | | 3 | s-Clean CCM UCLA Abutment s-Line | $\varnothing$ 4.0 X 1, 3mm (Gingival Height) | 0° | The dimensions of abutments are as following: #### The Abutments have below featured: | Name | Uses | Surface | Connection | |-----------------------------------------------|-----------------------------------------------------------------------------------------------------|-------------|--------------| | s-Clean O-Ring Abutment<br>s-Line | It is used to retain an overdenture prosthetics. | N/A | Screw retain | | s-Clean TiN Half Coating<br>Sub-Octa Abutment | This Abutment is connected with fixture and it<br>supports prosthesis which restore tooth function. | TiN Coating | Screw retain | | s-Clean CCM UCLA Abutment<br>s-Line | This Abutment is connected with fixture and it<br>supports prosthesis which restore tooth function. | N/A | Screw retain | ### Materials: - . s-Clean O-Ring Abutment s-Line and s-Clean TiN Half Coating Sub-Octa Abutment is fabricated from Ti-6A1-4V ELI (Conforming to ASTM Standard F136). - S-Clean CCM UCLA Abutment s-Line is fabricated from Cobalt-28Chromium-6Molybdenum . Alloy (Conforming to ASTM F1537) - CCM UCLA s-Line Abutment Screw is fabricated from Ti-6A1-4V ELI (Conforming to ASTM Standard F136). {5}------------------------------------------------ ## Summaries of Technology Characteristics The subject device is substantially equivalent to the current cleared devices. They are substantially equivalent in intended use, material and product spec as diameter and length. Comparison demonstrating Substantial Equivalence follows: ## 1. s-Clean O-Ring Abutment s-Line | | Subject Device | Primary Predicate | |------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Applicant | Dentis Co., Ltd. | Dentis Co., Ltd. | | Trade Name | Dentis s-Clean Abutment Mini | Dentis Dental Implant System | | 510(k) No. | | K171027 | | Classification<br>Name | Endosseous Dental Implant | Endosseous Dental Implant | | Product Code | NHA | DZE, NHA | | Description | Image: Dentis s-Clean Abutment Mini | Image: Dentis Dental Implant System | | Material | Ti-6Al-4V ELI (ASTM F136) | Ti-6Al-4V ELI (ASTM F136) | | Diameter (mm) | $\text{\O}4.5$ | $\text{\O}3.4$ and $\text{\O}4.5$ | | Length (mm) | 10.71 and 12.71 | 10.1, 11.6 and 13.6 | | Sterile | End User Sterilization | End User Sterilization | | Indications For<br>Use | The Dentis s-Clean Abutment Mini is an endosseous<br>dental implant that is indicated for surgical<br>placement in the upper and lower jaw arches, to<br>provide a root form means for single or multiple-<br>units prosthetic appliance attachment to restore a<br>patient's chewing function. Implants can be placed<br>with a conventional two stage surgical process with<br>an option for transmucosal healing or they can be<br>placed in a single stage surgical process for<br>immediate loading when good primary stability has<br>been achieved and with appropriate occlusal loading. | The Dentis Dental Implant System is an<br>endosseous dental implant that is indicated for<br>surgical placement in the upper and lower jaw<br>arches, to provide a root form means for single or<br>multiple-units prosthetic appliance attachment to<br>restore a patient's chewing function. Implants<br>can be placed with a conventional two stage<br>surgical process with an option for transmucosal<br>healing or they can be placed in a single stage<br>surgical process for immediate loading when<br>good primary stability has been achieved and<br>with appropriate occlusal loading. | | Substantial<br>Equivalence<br>Comparison | The subject device has same indications for Use, fundamental scientific technology, principle of<br>operation, technology, functions, and materials and has similar design with the primary predicate.<br>The difference between the subject and primary predicate is the Dimensions of the device and the<br>compatible implant body. The subject device's diameter( $\text{\O}2.88$ ) with length (9.11mm) is smaller size<br>than the primary predicate but doesn't affect device's fundamental functions and safety; therefore, it is<br>substantial equivalent. | | {6}------------------------------------------------ | | Subject Device | Primary Predicate | | Subject Device | Primary Predicate Device | Reference Device | |------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Applicant | Dentis Co., Ltd. | Dentis Co., Ltd. | Applicant | Dentis Co., Ltd. | Dentis Co., Ltd. | Dentis Co., Ltd. | | Trade Name | Dentis s-Clean Abutment Mini | Dentis Dental Implant System | Trade Name | Dentis s-Clean Abutment Mini | Dentis Dental Implant System | s-Clean OneQ-SL Narrow Implant<br>System | | 510(k) No. | - | K171027 | 510(k) No. | - | K171027 | K161244 | | Classification<br>Name | Endosseous Dental Implant | Endosseous Dental Implant | Classification<br>Name | Endosseous Dental Implant | Endosseous Dental Implant | Endosseous Dental Implant | | Product Code | NHA | DZE, NHA | Product Code | NHA | DZE, NHA | DZE, NHA | | Description | Image: Dentis s-Clean Abutment Mini | Image: Dentis Dental Implant System | Class | Class II | Class II | Class II | | Material | Ti-6Al-4V ELI (ASTM F136) | Ti-6Al-4V ELI (ASTM F136) | Description | | | | | Diameter (mm) | Ø4.8 | Ø4.8 | Material | Cobalt-28Chromium-6Molybdenum<br>Alloy<br>(ASTM F1537) | Cobalt-28Chromium-<br>6Molybdenum Alloy<br>(ASTM F1537) | Cobalt-28Chromium-6Molybdenum<br>Alloy<br>(ASTM F1537) | | Length (mm) | 8.16, 8.66, 9.66, 10.66, 11.66 and 12.66 | 9.15, 9.65, 10.65, 11.65, 12.65 and 13.65 | Diameter (mm) | $Ø4.0$ | $Ø4.5$ | $Ø4.0$ | | Surface Treatment | TiN Coating | TiN Coating | Gingival Height | 1mm and 3mm | 1mm and 3mm | 1mm | | Sterile | End User Sterilization | End User Sterilization | Sterile | End User Sterilization | End User Sterilization | End User Sterilization | | Indications For<br>Use | The Dentis s-Clean Abutment Mini is an endosseous<br>dental implant that is indicated for surgical placement<br>in the upper and lower jaw arches, to provide a root<br>form means for single or multiple-units prosthetic<br>appliance attachment to restore a patient's chewing<br>function. Implants can be placed with a conventional<br>two stage surgical process with an option for<br>transmucosal healing or they can be placed in a single<br>stage surgical process for immediate loading when<br>good primary stability has been achieved and with<br>appropriate occlusal loading. | The Dentis Dental Implant System is an<br>endosseous dental implant that is indicated for<br>surgical placement in the upper and lower jaw<br>arches, to provide a root form means for single or<br>multiple-units prosthetic appliance attachment to<br>restore a patient's chewing function. Implants can<br>be placed with a conventional two stage surgical<br>process with an option for transmucosal healing or<br>they can be placed in a single stage surgical<br>process for immediate loading when good primary<br>stability has been achieved and with appropriate<br>occlusal loading. | Indications For<br>Use/<br>Intended Use | The Dentis s-Clean Abutment Mini<br>is an endosseous dental implant that<br>is indicated for surgical placement<br>in the upper and lower jaw arches,<br>to provide a root form means for<br>single or multiple-units prosthetic<br>appliance attachment to restore a<br>patient's chewing function.<br>Implants can be placed with a<br>conventional two stage surgical<br>process with an option for<br>transmucosal healing or they can be<br>placed in a single stage surgical<br>process for immediate loading when<br>good primary stability has been<br>achieved and with appropriate<br>occlusal loading. | The Dentis Dental Implant<br>System is an endosseous dental<br>implant that is indicated for<br>surgical placement in the upper<br>and lower jaw arches, to provide<br>a root form means for single or<br>multiple-units prosthetic<br>appliance attachment to restore<br>a patient's chewing function.<br>Implants can be placed with a<br>conventional two stage surgical<br>process with an option for<br>transmucosal healing or they<br>can be placed in a single stage<br>surgical process for immediate<br>loading when good primary<br>stability has been achieved and<br>with appropriate occlusal<br>loading. | The s-Clean OneQ-SL Narrow<br>Implant System is indicated for<br>use in partially or fully edentulous<br>mandibles and maxillae, in support<br>of single or multiple-unit<br>restorations including; cemented,<br>retained, screw retained, or<br>overdenture restorations, and<br>terminal or intermediate abutment<br>support for fixed bridgework. This<br>system is dedicated for one and<br>two stage surgical procedures and<br>not dedicated for immediate<br>loading. This system is intended<br>for delayed loading. Only, s-Clean<br>OneQ-SL Narrow fixture (3.0, 3.3,<br>3.7mm) is limited to replacement<br>of maxillary lateral incisors and<br>mandibular incisors. | | Substantial<br>Equivalence<br>Comparison | The subject device has same indications for Use, fundamental scientific technology, principle of operation,<br>technology, functions, and materials and has similar design with the primary predicate.<br>The difference between the subject and primary predicate is the lengths of the device and the compatible<br>implant body. The subject device has shorter size than the primary predicate but doesn't affect device's<br>fundamental functions and safety; therefore, it is substantial equivalent. | | Substantial<br>Equivalence<br>Comparison | The subject device has same indications for Use, fundamental scientific technology, principle of operation,<br>technology, functions, and materials and has similar design with the primary predicate.<br>The difference between the subject and primary predicate is the diameter of the device and the compatible<br>implant body. To support this discrepancy, K161244 was added; therefore, it is substantial equivalent. | | | # 2. s-Clean TiN Half Coating Sub-Octa Abutment {7}------------------------------------------------ # 3. s-Clean CCM UCLA Abutment s-Line {8}------------------------------------------------ ## Non-Clinical Testing Below tests were performed for predicate devices and leveraged for the subject device: - Biocompatibility testing according to ISO 10993-1:2009 for Ti-6A1-4V material abutments . referenced in K210080, for Cobalt-28Chrominum-6Molybdenum material abutment referenced in K161244 and for Ti-6Al-4V with TiN Coating abutments referenced in K171027 - End User Sterilization Validation Test Report according to ANSI/AAMI ST79, ISO 17665-1, ● ISO 17665-2. ISO 11737-1, ISO 11737-2, and ISO 11138-1 referenced in K111364 for Ti-6Al-4V material abutments non-coating and TiN-Coating. The results of the above tests have met the criteria of the standards and demonstrated the substantial equivalence with the predicate device. Non-clinical test data was conducted in accordance with FDA Guidance "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments". Biocompatibility testing for Ti-6AI-4V wth TiN Coating material abutments was conducted on the predicate device, K171027 and can be leveraged for the subject device because both devices have same material and manufacturing process which is substantially equivalent. Fatigue testing was not performed for the subject device according to the ISO 14801:2016 because this subject device system does not include any angled abutments. Based on the FDA Guidance "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments", the fatigue testing is not necessary if the system does not consist of angled abutments. Also, the surface modification for the subject device was not tested or provided because the surface modification information was provided in the predicate devices and can be leveraged for the subject device. End User Sterilization Validation Test Report for Cobalt-27Chrominum-6Molybdenum was conducted on the device that is not cleared by FDA. The test article is the largest surface area among the devices that have the same manufacturing process and material. Clinical testing was not necessary to establish substantial equivalency of the device. - MR Environment Condition - Non-clinical worst-case MRI review was performed to evaluate the Dentis s-Clean Abutment Mini in the MRI environment using scientific rationale and published literature (e.g., Woods, Terry O., Jana G. Delfino, and Sunder Rajan. "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices." Journal of Testing and Evaluation 49.2 (2019): 783-795), based on the entire system including all variations (all compatible implant bodies, dental abutments, and fixation screws) and material composition. Rationale addressed parameters per the FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetically induced displacement force and torque {9}------------------------------------------------ ### Conclusion Dentis s-Clean Abutment Mini constitutes a substantially equivalent medical device, meeting all the declared requirements of its intended use. This system has the same intended use and fundamental scientific technology as its predicate de…
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.