STRAUMANN P.004 CEMENTABLE ABUTMENTS, TEMPORARY COPINGS AND PROTECTIVE CAPS
K072071 · Institut Straumann AG · NHA · Aug 15, 2007 · Dental
Device Facts
| Record ID | K072071 |
| Device Name | STRAUMANN P.004 CEMENTABLE ABUTMENTS, TEMPORARY COPINGS AND PROTECTIVE CAPS |
| Applicant | Institut Straumann AG |
| Product Code | NHA · Dental |
| Decision Date | Aug 15, 2007 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 872.3630 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
Abutments are placed into dental implants to provide support for prosthetic restorations such as crowns, bridges and overdentures, Abutments can be used in single tooth replacements and multiple tooth restorations. Temporary Copings are intended to serve as a base for temporary restorations for up to 28 days. Protective Caps are intended to protect the outer configuration of the abutment and to maintain and condition the contours of the soft tissue during the healing phase for up to 6 months.
Device Story
Straumann P.004 system comprises endosseous dental implants, abutments, temporary copings, and protective caps. Abutments connect to implants to support prosthetic restorations (crowns, bridges, overdentures) in partially or fully edentulous patients. Temporary copings provide a base for temporary restorations; protective caps maintain soft tissue contours during healing. Used in dental clinics by dental professionals. Mechanical components; no electronic or software elements.
Clinical Evidence
Bench testing only.
Technological Characteristics
Endosseous dental implant abutments, temporary copings, and protective caps. Mechanical design; no energy source or software. Materials and design are identical to previously cleared predicate devices.
Indications for Use
Indicated for partially or fully edentulous patients requiring dental implant-supported prosthetic restorations (crowns, bridges, overdentures). Temporary copings indicated for temporary restorations (up to 28 days). Protective caps indicated for soft tissue contouring/protection during healing (up to 6 months).
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Predicate Devices
- P.004 RC Anatomic Abutments (K062129)
- Temporary Copings (K041070)
- ITI Protective Healing Caps (K962023)
Related Devices
- K080239 — P.004 ABUTMENTS, P.004 TEMPORARY COPINGS, P.004 PROTECTIVE CAPS, P.004 PERMANENT COPINGS · Straumann Manufacturing, Inc. · Apr 30, 2008
- K080286 — CEMENTABLE ABUNTMENTS; TEMPORARY COPINGS; PROTECTIVE CAPS · Straumann USA · Mar 7, 2008
- K091701 — STRAUMANN MODIFIED DENTAL ABUTMENT · Institut Straumann AG · Dec 22, 2009
- K071919 — P.004 ABUTMENTS · Straumann USA (On Behalf of Institut Straumann Ag) · Sep 26, 2007
- K073553 — CAMLOG IMPLANT SYSTEM ABUTMENTS · Altatec GmbH · Mar 5, 2008
Submission Summary (Full Text)
{0}------------------------------------------------
K072071
Section H 510(k) SUMMARY AUG 15 2007
## 1. Applicant's Name and Address
Straumann US (on behalf of Institut Straumann AG) 60 Minuteman Rd. Andover, MA 01810 Telephone Number: 978-747-2513 978-747-0023 Fax Number: Contact Person: Elaine Alan Regulatory Affairs Specialist Date of Submission: July 27. 2007
#### 2. Name of the Device
| Trade Name: | P.004 RC Cementable Abutments |
|----------------------|---------------------------------------|
| Common Name: | Abutment, Dental, Endosseous implants |
| Classification Name: | Abutment, Dental, Endosseous implants |
| Regulation Number: | 21 CFR 872.3630 |
# 3. Legally Marketed Device to which Equivalence is Claimed (Predicate Device)
P.004 RC Anatomic Abutments, K062129 Temporary Copings, K041070 ITI Protective Healing Caps, K962023
#### 4. Description of the Device
The Straumann P.004 Dental Implant System is an integrated system of endosseous dental implants, which are designed to support prosthetic devices for partially or fully edentulous patients. The system consists of a variety of dental implants, abutments and surqical and prosthetic parts and instruments.
Abutments are placed into dental implants to provide support for prosthetic restorations, temporary copings support a temporary restoration out of occlusion and protective caps protect the abutment during the healing phase.
#### 5. Intended Use of the Device
Abutments are placed into dental implants to provide support for prosthetic restorations such as crowns, bridges and overdentures, Abutments can be used in single tooth replacements and multiple tooth restorations. Temporary Copings are intended to serve as a base for temporary restorations for up to 28 days. Protective Caps are intended to protect the outer configuration of the abutment and to maintain and
{1}------------------------------------------------
2 of 2
K072071
condition the contours of the soft tissue during the healing phase for up to 6 months.
### ર. Technological Characteristics
The modified Anatomic Abutments, Temporary Copings and Protective Caps are substantially equivalent to the currently cleared devices. The intended use is identical to the predicate devices. The proposed devices have the same material composition, basic design and fundamental operating principles to the currently cleared devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with three stripes forming its wing, symbolizing service and protection. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 1 5 2007
Institut Straumann AG C/O Ms. Elaine Alan Regulatory Affairs Specialist Straumann USA 60 Minuteman Road Andover, Massachusetts 01810
Re: K072071
Trade/Device Name: P.004 Cementable Abutments, Temporary Copings and Protective Caps Regulation Number: 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: July 27, 2007 Received: July 30, 2007
Dear Ms. Alan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Alan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chris Liu, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K072071
# INDICATIONS FOR USE STATEMENT
Device Name: P.004 Cementable Abutments
Indications for Use:
Abutments are intended to be placed into dental implants to provide support for prosthetic reconstructions such as crowns, bridges and overdentures.
Device Name: P.004 Protective Caps
Indications for Use:
Protective Caps are intended to protect the outer configuration of the abutment and to maintain and condition the contours of the soft tissue during the healing phase for up to 6 months.
Device Name: P.004 Temporary Copings
Indications for Use:
Temporary Copings are intended to serve as a base for temporary restorations for up to 28 days.
× Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Susan Evans
vision of Anesthesiology, General Hospital, Intection Control, Dental Devices
| 510(k) Submission: RC Cementable Abutments | |
|--------------------------------------------|-------|
| 510(k) Number: | K1207 |
| July 27, 2007 | |
| Straumann US | |
| Page 5 | |
Page 5