NOBELPROCERA ZI ABUTMENT THOMMEN PLATFORMS

K110171 · Nobel Biocare USA, LLC · NHA · Jun 15, 2011 · Dental

Device Facts

Record IDK110171
Device NameNOBELPROCERA ZI ABUTMENT THOMMEN PLATFORMS
ApplicantNobel Biocare USA, LLC
Product CodeNHA · Dental
Decision DateJun 15, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 872.3630
Device ClassClass 2

Intended Use

The NobelProcera Zi Abutments Thommen Platforms are premanufactured prosthetic components directly connected to endosseous dental implants and are intended for use as an aid in prosthetic rehabilitation. They are compatible with the Thommen Medical SPI 3.5, 4.0, 4.5, 5.0, and 6.0 implants.

Device Story

NobelProcera Zi Abutment Thommen Platforms are patient-specific, machined endosseous dental implant abutments; constructed from zirconium oxide. Device connects directly to Thommen Medical SPI dental implants (platforms 3.5, 4.0, 4.5, 5.0, 6.0) to provide support for prosthetic restorations (crowns, bridges, overdentures). Used in clinical dental settings by practitioners; abutment serves as interface between implant and final restoration. Benefit includes patient-specific fit for prosthetic rehabilitation.

Clinical Evidence

Bench testing only. Fatigue testing performed on worst-case samples in accordance with FDA guidance 'Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments'. No clinical data.

Technological Characteristics

Material: Zirconium oxide. Form factor: Patient-specific, machined endosseous dental implant abutment. Compatibility: Thommen Medical SPI implant platforms 3.5, 4.0, 4.5, 5.0, 6.0. Mechanical performance validated via fatigue testing.

Indications for Use

Indicated for patients requiring prosthetic rehabilitation via endosseous dental implants. Compatible with Thommen Medical SPI 3.5, 4.0, 4.5, 5.0, and 6.0 implant platforms.

Regulatory Classification

Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K110171 # JUN 15 2011 #### 510(k) Summary of Safety and Effectiveness 1.4 | Submitted by: | Herbert Crane<br>Director, Global Regulatory Affairs | |----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Address: | Nobel Biocare USA LLC<br>22715 Savi Ranch Parkway<br>Yorba Linda, CA 92887 | | Telephone: | (714) 282-5074 | | Facsimile: | (714) 998-9348 | | Date of Submission: | January 13, 2011 | | Classification Name: | Endosseous Dental Implant Abutment (21 CFR 872.3630) | | Trade or Proprietary<br>or Model Name: | NobelProcera Zi Abutment Thommen Platforms | | Legally Marketed Devices: | Nobel Biocare - NobelProcera Ti Abutment Thommen Platforms<br>(K101714)<br>Thommen Medical AG - SPI Easy 4.0 Abutment (K071457)<br>Thommen Medical AG - SPI Dental Implant Abutments (K031747) | #### Device Description Nobel Biocare's NobelProcera Zi Abutment Thommen Platforms is an endosseous dental implant abutment. The abutment attaches directly to Thommen dental implants and provides a platform for restoration. Nobel Biocare's NobelProcera Zi Abutments are designed and made individually to fit the individual requirements for each patient. The NobelProcera Zi Abutments are made entirely of zirconium oxide. The abutments fit the Thommen SPI implant systems platform 3.5, 4.0, 4.5, 5.0, 6.0. #### Indications for Use The NobelProcera Zi Abutments Thommen Platforms are premanufactured prosthetic components directly connected to endosseous dental implants and are intended for use as an aid in prosthetic rehabilitation. They are compatible with the Thommen Medical SPI 3.5, 4.0, 4.5, 5.0, and 6.0 implants. #### Technical Modifications of Predicate Device The NobelProcera Zi Abutment for the Thommen Platform differs from the predicates by using the material zirconium oxide. Nobel Biocare Special 510(k) Notification NobelProcera Zi Abutments Thommen Platforms January 2011 {1}------------------------------------------------ Summary of testing to demonstrate safety and effectiveness Non-clinical test data was used to support the decision of safety and effectiveness. Clinical testing was not necessary. Non-clinical testing consisted of analysis of platforms to identify worst-case test samples and performance of fatigue testing of the worst-case samples in accordance with the FDA guidance Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments. ### Conclusion The information provided in this submission demonstrates that the device is substantially_ equivalent to the predicate devices. Nobel Biocare Special 510(k) Notification NobelProcera Zi Abutments Thommen Platforms January 2011 {2}------------------------------------------------ ubstantial Equivalence Comparison to Predicate Devices | ATTRIBUTE | CANDIDATE | PREDICATE | PREDICATE | | |---------------------------|--------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Design/<br>Construction | NobelProcera Zi<br>Abutment<br>Thommen Platforms | NobelProcera Ti Abutment<br>Thommen Platforms<br>(K101714) | Thommen Medical AG<br>SPI Dental Implant Abutments<br>(K031747) | | | Anatomical<br>Site | Patient specific / machined | Patient specific / machined | Patient specific / machined | | | Platform<br>Compatibility | Oral Cavity | Oral Cavity | Oral Cavity | | | Device<br>Material | Thommen<br>SPI implant systems platform<br>3.5, 4.0, 4.5, 5.0, 6.0 | Thommen<br>SPI implant systems platform<br>3.5, 4.5, 5.0, 6.0 | Thommen<br>SPI implant systems platform<br>3.5, 4.5, 5.0, 6.0 | | | Indications for<br>Use | Zirconium Oxide | CP Titanium<br>Titanium/vanadium alloy<br>The NobelProcera Ti Abutments<br>Thommen Platforms are<br>premanufactured prosthetic<br>components directly connected to<br>endosseous dental implants and<br>are intended for use as an aid in<br>prosthetic rehabilitation. They are<br>compatible with the Thommen<br>Medical SPI 3.5, 4.0, 4.5, 5.0, and<br>6.0 implants. | CP Titanium<br>The Thommen SPI Easy 4.0<br>Abutment is intended to be used<br>in conjunction with SPI System<br>dental implants in the maxillary<br>and/or mandibular arch to<br>provide support for crowns,<br>bridges or overdentures. | CP Titanium<br>The Thommen SPI Dental Implant<br>Abutments are intended to be<br>used in conjunction with SPI<br>System dental implants in the<br>maxillary and/or mandibular arch<br>to provide support for crowns,<br>bridges or overdentures. | {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Nobel Biocare AB C/O Mr. Herbert Crane Director, Global Regulatory Affairs Nobel Biocare USA LLC 22715 Savi Ranch Parkway Yorba Linda, California 92887 JUN 1 5 2011 Re: K110171 Trade/Device Name: NobelProcera Zi Abutment Thommen Platforms Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: May 13, 2011 Received: May 16, 2011 Dear Mr. Crane: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ ### Page 2 - Mr. Crane Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, for Anthony Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health ## Enclosure {5}------------------------------------------------ # Indications for Use KIIO17 | 510(k) Number (if known): Device Name: NobelProcera Zi Abutment Thommen Platforms Indications For Use: The NobelProcera Zi Abutments Thommen Platforms are premanufactured prosthetic components directly connected to endosseous dental implants and are intended for use as an aid in prosthetic rehabilitation. They are compatible with the Thommen Medical SPI 3.5, 4.0, 4.5, 5.0, and 6.0 implants. × Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (Please do Not WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Robert S. Betz, DDS for Dr. Susan Runner (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices Page 1 of 1 ----- 510(k) Number: K110171 000020
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...