NNC CEMENTABLE ABUTMENT STRAIGHT,15 DEGREE ANGLE, TYPE A ,15 DEGREE ANGLE, TYPE B,NNC BASAL SCREW CEMENTABLE ABUTMENTS

K113283 · Straumann USA · NHA · Dec 28, 2011 · Dental

Device Facts

Record IDK113283
Device NameNNC CEMENTABLE ABUTMENT STRAIGHT,15 DEGREE ANGLE, TYPE A ,15 DEGREE ANGLE, TYPE B,NNC BASAL SCREW CEMENTABLE ABUTMENTS
ApplicantStraumann USA
Product CodeNHA · Dental
Decision DateDec 28, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 872.3630
Device ClassClass 2
AttributesTherapeutic

Intended Use

Abutments are used in connection with the prosthetic restoration of Straumann dental implants. Abutments are intended to be placed into dental implants to provide support for prosthetic reconstructions such as crowns and bridges. Straumann Narrow Neck CrossFit (NNC) Cementable Abutments are indicated for cement-retained single tooth and bridge restorations.

Device Story

Straumann NNC Cementable Abutments are permanent dental components for 3.3mm diameter tissue-level implants; used to support prosthetic crowns and bridges. Device consists of titanium grade 4 abutment and titanium alloy basal screw; available in straight or 15° angled configurations. Operated by dental professionals in clinical settings; placed into dental implants to provide stable base for cement-retained restorations. Benefits include restoration of tooth function and aesthetics for patients with missing teeth. Performance verified via fatigue testing per FDA guidance.

Clinical Evidence

Bench testing only. Fatigue testing conducted in accordance with FDA guidance document 'Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments' to verify performance and structural integrity.

Technological Characteristics

Materials: Titanium Grade 4 (abutment), Titanium Alloy (basal screw). Configuration: Straight and 15° angled. Connection: NNC tissue level implant interface (3.3mm diameter). Principle: Mechanical fixation for cement-retained dental prosthetics. Sterilization: Not specified.

Indications for Use

Indicated for cement-retained single tooth and bridge restorations in patients requiring prosthetic reconstruction on Straumann Narrow Neck CrossFit (NNC) tissue level dental implants (3.3mm diameter).

Regulatory Classification

Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ### Narrow Neck CrossFit (NNC) Cementable Abutments Special 510(k) Section 5: 510(k) Summary DEC 2 8 2011 #### 1. Applicant's Name and Address Straumann US (on behalf of Institut Straumann AG) 60 Minuteman Rd. Andover, MA 01810 Telephone Number: 800-448-8168, ext 2513 Fax Number: 978-747-0023 Contact Person: Elaine Alan Senior Regulatory Affairs Specialist #### 2. Date of Submission: November 4, 2011 #### 3. Name of the Device | Trade Name: | Straumann Narrow Neck CrossFit (NNC)<br>Cementable Abutments | |----------------------|--------------------------------------------------------------| | Common Name: | NNC Cementable Abutments | | Classification Name: | Endosseous Dental Implant Abutments | | Regulation Number: | §872.3630 | ### 4. Legally Marketed Devices to which Equivalence is Claimed (Predicate Device) Straumann Narrow Neck Connection (NNC) Gold Abutment for crowns, K111357 Straumann Narrow Connection (NC) Cementable Abutments, K080286 Straumann Narrow Neck (NN) CARES Titanium Abutment, K082545 #### 5. Description of the Device Straumann Narrow Neck CrossFit (NNC) Cementable Abutments are permanent abutments intended for placement onto the Straumann Narrow Neck CrossFit (NNC) tissue level Implants with the diameter of 3.3mm. The abutments are made of Titanium Grade 4 with a corresponding basal screw made of Titanium Alloy. The abutments are available in straight and 15° angled configurations. #### 6. Intended Use of the Device Response Letter: K113283 December 9, 2011 Straumann NNC Cementable Abutments Straumann, USA Page 7 of 14 {1}------------------------------------------------ # < 113283 ### Narrow Neck CrossFit (NNC) Cementable Abutments Special 510(k) Section 5: 510(k) Summary Abutments are used in connection with the prosthetic restoration of Straumann dental implants. Abutments are intended to be placed into dental implants to provide support for prosthetic reconstructions such as crowns and bridges. Straumann Narrow Neck CrossFit (NNC) Cementable Abutments are indicated for cement-retained single tooth and bridge restorations. #### 7. Technological Characteristics The proposed device is substantially equivalent to the currently marketed device. They share the same indication for use, prosthetic platform, implant/abutment connection, and fundamental operating principles. #### Performance Testing 8. Verification and validation testing were performed to ensure that the Straumann Narrow Neck CrossFit (NNC) Cementable Abutments function as intended and that the modifications do not impact the performance of the device. Testing included: 1. Performance Testing - i. Fatigue Testing in accordance to FDA guidance document "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments." #### 9. Conclusion The results from the testing conducted demonstrated that the Sraumann Narrow Neck CrossFit (NNC) Cementable Abutments function as intended and met the pre-determined acceptance criteria. December 9, 2011 {2}------------------------------------------------ K II 3283 ### Narrow Neck CrossFit (NNC) Cementable Abutments Special 510(k) Section 5: 510(k) Summary The Straumann Narrow Neck CrossFit (NNC) Cementable Abutments is a validated system. The results of the performance bench testing and risk analysis indicate that the Straumann Narrow Neck CrossFit (NNC) Cementable Abutments are substantially equivalent to the named predicate device and is safe and effective for its intended use. Response Letter: K113283 Straumann NNC Cementable Abutments December 9, 2011 Straumann, USA Page 9 of 14 {3}------------------------------------------------ Image /page/3/Picture/1 description: The image is a circular seal or logo. The seal contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the edge. In the center of the seal is a stylized image of an eagle with its wings spread, facing to the right. The eagle is composed of thick, curved lines, giving it a modern and abstract appearance. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 DEC 2 8 2011 Ms. Elaine Alan Senior Regulatory Affairs Specialist Straumann USA 60 Minuteteman Road Andover, Massachusetts 01810 Re: K113283 Trade/Device Name: Straumann Narrow Neck CrossFit (NNC) Cementable Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment . Regulatory Class: II Product Code: NHA Dated: December 9, 2011 Received: December 12, 2011 Dear Ms. Alan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device mendments, or to devices that have been reclassified in accordance with the provisions of Amendinens, or to av roometic Act (Act) that do not require approval of a premarket the rederal F 00d, Drag, and Coou may, therefore, market the device, subject to the general approvisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, legallents for annual vegainst misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your (1 will 1), it thay be basjon the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ ### Page 2 -- Ms. Alan Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing (1 CFR Partice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to Sempilation of fication" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem /default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.goy/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Nh for Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health ### Enclosure {5}------------------------------------------------ ## Indications for Use 510(k) Number (if known): Device Name: Straumann Narrow Neck CrossFit (NNC) Cementable Abutments Indications for Use: Abutments are used in connection with the prosthetic restoration of Straumann dental implants. Abutments are intended to be placed into dental implants to provide support for prosthetic reconstructions such as crowns and bridges. Narrow Neck CrossFit Cementable Abutments are indicated for cement-retained single tooth and bridge restorations. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-the-Counter Use (21 CFR 801 Subpart C) PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of DCRH, Office of Device Evaluation (ODE) Michael (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: Special 510(k) Submission: NNC Cementable Abutments November 4, 2011 113283 Page 1 of 1 Page 15 of 75 Straumann USA
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...