The ET Abutments are indicated for use with ET Dental Implants to provide support to prosthetic restoration such as crowns, bridges and overdentures in partially or fully edentulous patients.
Device Story
The ET Abutment System consists of various dental abutments (angled, cylinder, goldcast, multi, O-ring, etc.) designed to connect to ET Dental Implants. These components provide support for prosthetic restorations like crowns, bridges, and overdentures. The devices are manufactured from medical-grade titanium alloy (ASTM F136), CP titanium, or polymers (POM, NBR, Polycarbonate). They are supplied non-sterile and are intended for steam sterilization by the user. The system is used in a clinical dental setting by a dentist or dental professional. The abutments are placed onto the dental implant to serve as the interface for the final prosthesis. By providing a stable connection between the implant and the restoration, the system enables the functional and aesthetic replacement of missing teeth, benefiting patients by restoring oral function.
Clinical Evidence
No clinical performance data was submitted. Substantial equivalence is supported by non-clinical performance testing, including fatigue testing in accordance with ISO 14801 and MRI compatibility assessments based on scientific rationale and published literature.
Technological Characteristics
Materials: Titanium alloy (ASTM F136), CP Titanium, POM, NBR, Polycarbonate. Sensing/Actuation: None (mechanical interface). Energy Source: None. Connectivity: None. Sterilization: Moist heat (ISO 17665-1). Form factor: Various lengths/diameters for dental abutments.
Indications for Use
Indicated for partially or fully edentulous patients requiring prosthetic restoration support (crowns, bridges, overdentures) via ET Dental Implants.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Hiossen Inc. Peter Lee QA/RA Manager 85 Ben Fairless Drive Fairless Hills, Pennsylvania 19030
Re: K222636
Trade/Device Name: ET Abutment System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: March 30, 2023 Received: April 3, 2023
# Dear Peter Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
# Sincerely. Andrew I. Steen -S
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known) K222636
Device Name
# ET Abutment System
Indications for Use (Describe)
The ET Abutments are indicated for use with ET Dental Implants to provide support to prosthetic restoration such as crowns, bridges and overdentures in partially or fully edentulous patients.
Type of use (Select one or both, as applicable)
区 Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 807 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Service (301) 443-6740 EF
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Hiossen Implant. The logo has the letters "HI" in red, with a curved line above the letters. The word "OSSEN" is in gray, and the word "IMPLANT" is in a smaller font below the word "OSSEN". There is a registered trademark symbol to the right of the word "OSSEN".
85 Ben Fairless Drive s Hills, PA 19030 888-768-0001 www.hiossen.com
# Section 6 510(k) Summary - 12 PAGES
| 6.1 Submitter Information:<br>Submitted by: | Hiossen, Inc.<br>85 Ben Fairless Drive<br>Fairless Hills, PA 19030 |
|-------------------------------------------------------------------|--------------------------------------------------------------------|
| Contact Person:<br>Telephone Number:<br>Fax Number: | Peter Lee<br>267-759-7031<br>267-759-7031 |
| Date Prepared: | April 28, 2023 |
| 6.2 Device Name:<br>• Proprietary Name:<br>• Classification Name: | ET Abutment System<br>Implant, Endosseous, Root |
| • Proprietary Name: | ET Abutment System |
|------------------------|--------------------------------|
| • Classification Name: | Implant, Endosseous, Root-form |
| • CFR Number: | 872.3630 |
| • Device Class: | Class II |
| • Product Code: | NHA |
# 6.3 Predicate Devices:
| Primary | 510(k) | Manufacturer(s) |
|----------------------------------|---------|--------------------------|
| Osstem Abutment System | K182091 | Osstem Implant Co., Ltd. |
| Reference | 510(k) | Manufacturer(s) |
| Osstem Implant System - Abutment | K161689 | Osstem Implant Co., Ltd. |
| ET US SS Prosthetic System | K160670 | Osstem Implant Co., Ltd. |
| Hiossen Prosthetic System | K140507 | Osstem Implant Co., Ltd. |
| ET/SS Implant System | K120847 | Osstem Implant Co., Ltd. |
| Prosthetic System | K110308 | Osstem Implant Co., Ltd. |
| HS/HG Prosthetic System | K100245 | Osstem Implant Co., Ltd. |
| HU/HS/HG Prosthetic System | K081575 | Osstem Implant Co., Ltd. |
| US/SS/GS System | K073247 | Osstem Implant Co., Ltd. |
| US System | K062030 | Osstem Implant Co., Ltd. |
# 6.4 Description of Device:
The ET Abutment System are indicated for use with ET dental implants to provide support to prosthetic restoration such as crowns, bridges and overdentures in partially or fully edentulous patients. Manufactured from medical grade titanium alloy and delivered non-sterilized.
The ET Abutments are available in various lengths and diameters; configurations are listed in the table below.
| EK DENTAL ABUTMENTS | Diameter (mm) | Length (mm) |
|--------------------------------|---------------|-------------|
| ET Angled | 4.0 ~ 6.0 | 13.0, 15.0 |
| ET Convertible Angled Cylinder | 4.0, 4.8, 6.0 | 7.5 |
| ET Goldcast | 4.0, 4.5 | 13.0 ~ 16.0 |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Hiossen Implant. The word "Hiossen" is written in a stylized font, with the "H" in red and the rest of the letters in gray. The word "Implant" is written in smaller gray letters below the word "Hiossen". The logo is simple and modern, and it is likely used to promote the company's dental implant products.
5 Ben Fairless Drive A 19030 www.hiossen.com
| ET Multi Angled | 4.80 | 9.00 ~ 13.00 |
|-------------------------------------|------------------|------------------|
| ET Multi Combination Cylinder | 5.00 | 7.30 |
| ET O-ring | 3.50, 4.60 | 1.15, 1.50 |
| ET Retainer, Retainer Cap | 5.00 | 4.40 |
| ET Port | 3.70 | 7.00 ~ 13.00 |
| ET Male Cap | 5.50 | 2.25 |
| ET Males | 4.75 | 1.80 |
| ET Stud | 3.50 | 9.00 ~ 15.00 |
| ET Temporary | 4.00, 4.50 | 13.00 ~ 16.00 |
| ET Transfer | 4.00 ~ 7.00 | 8.00 ~ 15.00 |
| ET Freeform ST | 4.00 ~ 7.00 | 14.00, 15.00 |
| ET Rigid | 4.00 ~ 7.00 | 4.00, 5.50, 7.00 |
| ET Rigid Protect Cap | 4.00 ~ 7.00 | 4.00, 5.50, 7.00 |
| ET Esthetic Low Temporary Cylinder | 4.80, 5.30 | 12.00 |
| ET Multi | 4.80 | 9.00 ~ 13.00 |
| ET Esthetic Low Gold Cylinder | 4.80 | 12.00 |
| ET Convertible Combination Cylinder | 4.00, 4.80, 6.00 | 7.00 |
| ET Convertible Temporary Cylinder | 4.00, 4.80, 6.00 | 9.80 |
| ET Convertible | 4.00, 4.80, 6.00 | 6.00 ~ 11.00 |
| ET Convertible GoldCast Cylinder | 4.00, 4.80, 6.00 | 11.75, 12.15 |
| ET Convertible Plastic Cylinder | 4.00, 4.80, 6.00 | 12.00 |
| ET Convertible Protect Cap | 4.00, 4.80, 6.00 | 2.90, 3.50 |
| ET Esthetic Low Plastic Cylinder | 4.80 | 12.00 |
| ET Esthetic Low Healing Cap | 4.80 | 4.60 |
The ET Abutment System are similar to other commercially available products based on the intended use, the technology used, the claims, the material composition employed and performance characteristics.
# 6.5 Indication for Use:
The ET Abutment System are indicated for use with ET dental implants to provide support to prosthetic restoration such as crowns, bridges and overdentures in partially or fully edentulous patients.
# 6.6 Substantial Equivalence:
#### 6.6.1 ET Abutment System
The information and date provided in this submission established the ET Abutment System are substantially equivalent to the primary predicate devices listed below. Proposed devices are the same devices Osstem (FEI 3003394081) received clearance listed as predicate and reference devices.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Hiossen Implant. The word "Hiossen" is written in a stylized font, with the "H" and the curved line above it in red, and the rest of the word in gray. Below the word "Hiossen" is the word "IMPLANT" in smaller, gray letters. The logo is simple and modern, and it is likely used to promote the company's dental implant products.
| Device | Proposed Device<br>ET Abutment System | Primary Predicate Devices<br>Osstem Abutment System |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Hiossen, Inc. | Osstem Co., Ltd. |
| 510(K) No. | New device | K182091 |
| Intended use | The ET Abutments are indicated for use<br>with ET Dental Implants to provide support<br>to prosthetic restoration such as crowns,<br>bridges and overdentures in partially or fully<br>edentulous patients. | Osstem Abutment System is intended for<br>use with a dental implant to provide support<br>for prosthetic restorations such as crowns,<br>bridges, or overdentures.<br>• Link Abutment for CEREC |
| | | The Link Abutment for CEREC is titanium<br>alloy abutments placed onto OSSTEM<br>dental implants to provide support for<br>customized prosthetic restorations. Link<br>Abutment for CEREC is indicated for screw-<br>retained single tooth or cement-retained<br>single tooth and bridge restorations. All<br>digitally designed copings and/or crowns for<br>use with the Link abutment for CEREC is to<br>be scanned using Sirona CEREC AC or<br>CEREC AF or CEREC AI, designed using<br>Sirona inLab software (Version 3.65) or<br>Sirona CEREC Software (Version 4.2) and<br>manufactured using a Sirona CEREC or<br>inLab MC X or MC XL milling unit. CAD/CAM<br>manufacturing/milling occurs at dental<br>laboratories per the design limitations of the<br>Sirona CEREC. |
| Sterilization | • Delivered non-sterilized<br>• Steam sterilized by user | • Delivered non-sterilized<br>• Steam sterilized by user |
| Packaging | • Housed in Tyvek-lidded blister tray<br>• Placed in a tamper-evident outer<br>package. | • Housed in Tyvek-lidded blister tray<br>• Placed in a tamper-evident outer<br>package. |
| ET ANGLED | | |
| Design | | |
| Surface | Machine surface | Machine surface |
| Material | • Titanium alloy Ti-6Al-4V (ASTM F136) | • Titanium alloy Ti-6Al-4V (ASTM F136) |
| Diameters(mm) | 4.0 ~ 6.0 | 4.0 ~ 6.0 |
| Heights(mm) | 8.0 | 8.0 |
| Angulation | 17° | 17° |
| ET CONVERTIBLE ANGLED CYLINDER | | |
| Design | | |
| Surface | Machine surface | Machine surface |
| Material | • Titanium CP Grade 3 | • Titanium CP Grade 3 |
| Diameters(mm) | 4.0, 5.0 | 4.0, 5.0, 6.0 |
| Height(mm) | 8.0 | 8.0 |
| ET MULTI ANGLED | | |
| Design | Image: ET MULTI ANGLED | Image: ET MULTI ANGLED |
| Surface | Machine surface | Machine surface |
| Material | • Titanium alloy Ti-6Al-4V (ASTM F136) | • Titanium alloy Ti-6Al-4V (ASTM F136) |
| Diameters(mm) | 4.8 | 4.8 |
| G/H(mm) | $2.5 ~ 5.0$ | $2.5 ~ 5.0$ |
| Angulation | $17°, 30°$ | $17°, 30°$ |
| ET MULTI COMBO CYLINDER | | |
| Design | Image: ET MULTI COMBO CYLINDER | Image: ET MULTI COMBO CYLINDER |
| Surface | Machine surface | Machine surface |
| Material | • Titanium CP Grade 3 | • Titanium CP Grade 3 |
| Diameters(mm) | 5.0 | 5.0 |
| Height(mm) | 7.3 | 7.3 |
| ET PORT | | |
| Design | Image: ET PORT | Image: ET PORT |
| Surface | Machine surface | Machine surface |
| Material | • Titanium alloy Ti-6Al-4V (ASTM F136) | • Titanium alloy Ti-6Al-4V (ASTM F136) |
| Diameters(mm) | 3.7 | 3.7 |
| G/H(mm) | $1.0~5.0$ | $1.0~5.0$ |
| ET Male Cap | | |
| Design | Image: ET Male Cap | Image: ET Male Cap |
| Surface | Machine surface | Machine surface |
| Material | • Titanium alloy Ti-6Al-4V (ASTM F136) | • Titanium alloy Ti-6Al-4V (ASTM F136) |
| Diameters(mm) | 5.5 | 5.5 |
| Height(mm) | 2.25 | 2.25 |
| ET Port Males | | |
| Design | Image: ET Port Males | Image: ET Port Males |
| Material | • Nylon | • Nylon |
| Diameters(mm) | 4.73 | 4.73 |
| Height(mm) | 1.80 | 1.80 |
| ET Port Extended Males | | |
| Design | Image: ET Port Extended Males | Image: ET Port Extended Males |
| Material | • Nylon | • Nylon |
| Diameters(mm) | 4.75 | 4.75 |
| Height(mm) | 1.80 | 1.80 |
| ET STUD | | |
| Design | Image: ET STUD | Image: ET STUD |
| Surface | Machine surface | Machine surface |
| Material | • Titanium alloy Ti-6Al-4V (ASTM F136) | • Titanium alloy Ti-6Al-4V (ASTM F136) |
| | | |
| Diameters(mm) | 3.5 | 3.5 |
| G/H(mm) | 1.0 ~ 6.0 | 1.0 ~ 6.0 |
| ET O-RING | | |
| Design | Image: ET O-RING Design | Image: ET O-RING Design |
| Material | • Acrylonitrile & Butadiene Polymer (NBR) | • Acrylonitrile & Butadiene Polymer (NBR) |
| Diameter | 3.50, 4.60 | 3.50, 4.60…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.