IPD Dental Implant Abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for single or multiple dental prosthetic restorations.
Device Story
IPD Dental Implant Abutments are titanium alloy components (healing, temporary, cementing, or Ti-base) that attach to endosseous dental implants. The system includes titanium screws (Ti, TiN-coated, or TPA) to secure the prosthesis. Used by dental clinicians in clinics, the device provides a foundation for prosthetic restorations. Ti-base abutments are designed via a digital dentistry workflow (intra-oral scanner, CAD software, milling machine) to support a zirconia superstructure, which is cemented to the Ti-base. The final assembly is screw-retained to the implant. The device benefits patients by restoring dental function and aesthetics in edentulous or partially edentulous arches.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including sterilization validation (ISO 17665-1), biocompatibility (ISO 10993-5/10), dimensional analysis, digital workflow validation, and static/dynamic fatigue testing (ISO 14801).
Technological Characteristics
Materials: Titanium alloy (ISO 5832-3), TiN coating, or anodized. Design: Healing, temporary, cementing, and Ti-base abutments. Connectivity: Digital library files for CAD/CAM integration. Sterilization: Non-sterile, intended for end-user steam sterilization. Energy: Mechanical. Form factor: Various diameters/platforms compatible with Brånemark, Straumann, and Tapered Screw-Vent systems.
Indications for Use
Indicated for patients undergoing oral implant surgery requiring support for single or multiple dental prosthetic restorations in the maxillary or mandibular arch. Compatible with specific Brånemark, Straumann Tissue Level, and Tapered Screw-Vent implant systems.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
10/21/2022
Implant Protesis Dental 2004 S.L Francesc Fumanal Regulatory Affairs Manager Cami del Mig. 71. Mataro, Barcelona 08302 SPAIN
Re: K222215
Trade/Device Name: IPD Dental Implant Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA, PNP Dated: July 19, 2022 Received: July 25, 2022
Dear Francesc Fumanal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Andrew Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic. Anesthesia. Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222215
Device Name IPD DENTAL IMPLANT ABUTMENTS
#### Indications for Use (Describe)
IPD Dental Implant Abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for single or multiple dental prosthetic restorations.
| Implant System Compatibility | Implant Diameter (mm) | Platform Diameter |
|------------------------------|-----------------------|-------------------|
| Brånemark | 3.5 | NP (3.5 mm) |
| | 3.75 / 4.0 | RP (4.1 mm) |
| | 5.0 | WP (5.1 mm) |
| Straumann® Tissue Level | 3.3 / 4.1 / 4.8 | RN (4.8 mm) |
| | 4.8 | WN (6.5 mm) |
| Tapered Screw-Vent® | 3.7 / 4.1 | 3.5 mm |
| | 4.7 | 4.5 mm |
| | 6.0 | 5.7 mm |
#### Compatible Implant Systems
The zirconia superstructures for use with the Ti Base (Interface) are only intended to be designed and manufactured according to digital dentistry workflow. The workflow system integrates multiple components of the digital dentistry workflow: scan files from intra-oral scanners, CAD software, ceramic material, milling machine and associated tooling and accessories.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|
| <span style="font-size: 12pt;"> <span style="font-family: Symbol;"> <span style="color: black;">☑</span> </span> </span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
#### I. SUBMITTER
#### IMPLANT PROTESIS DENTAL 2004, S.L
Camí del Mig, 71. 1º 2ª 08302 Mataró (Barcelona), Spain.
#### Contact Person:
Francesc Fumanal +34 93 672 110 748 ffumanal@ipd2004.com
Date prepared: October 21, 2022.
#### II. DEVICE
| Device name: | IPD DENTAL IMPLANT ABUTMENTS |
|----------------------|------------------------------------------------------|
| Common Name: | ENDOSSEOUS DENTAL IMPLANT ABUTMENT |
| Classification Name: | Endosseous Dental Implant Abutment (21 CFR 872.3630) |
| Regulatory Class: | Class II |
| Product Code(s): | Primary: NHA; Secondary: PNP. |
#### III. PREDICATE DEVICE(S):
| Primary Predicate: | K201860 Elos Accurate® Hybrid Base™ |
|--------------------|-------------------------------------|
| Reference Devices: | K170588 DESS Dental Smart Solutions |
| | K153521 IH Implant System |
| | K173908 DESS Dental Smart Solutions |
| Compatible Implant Systems | | |
|----------------------------|-------------------------|-----------------------|
| Clearance | Device | Manufacturer |
| K022562 | Brånemark | Nobel Biocare AB |
| K130222 | Straumann® Tissue Level | Institut Straumann AG |
| K112160 | Tapered Screw-Vent® | Zimmer Biomet Dental |
{4}------------------------------------------------
#### IV. DEVICE DESCRIPTION
IPD Dental Implant Abutments is a dental implant abutment system composed of dental abutments and screws intended to be placed into dental implants to provide support for dental prosthetic restorations.
Abutments provide basis for single or multiple tooth prosthetic restorations. They are available in a variety of connection types to enable compatibility with commercially available dental implants systems.
IPD Dental Implant Abutments submission includes the following categories of dental abutment designs:
- Healing abutments; -
- -Temporary abutments;
- Cementing titanium abutments; -
- -Titanium base (interface) abutments;
The system also includes the corresponding Titanium Screws intended to attach the prosthesis to the dental implant. Specifically:
- Ti Screw: Used during restoration fabrication.
- TiN Screw: Used in finished restorations, with TiN coating.
- TPA Screw: Used in finished angulated restorations, with TiN coating.
All subject abutments and screws are made of titanium alloy conforming to ISO 5832-3 "Implants for surgery - Metallic materials - Part 3: Wrought titanium 6-aluminium 4-vanadium alloy".
IPD dental implant abutments and screws are compatible with the following commercially available dental implant systems:
{5}------------------------------------------------
Traditional 510(k) - K222215
| Compatible<br>Implant<br>System | Type of<br>connection | Implant<br>Diameter<br>(mm) | Platform<br>Diameter | Device<br>category | Healing<br>Abutment | Cementing<br>/<br>Temporary<br>Abutments | Ti Base<br>(Interface)<br>Abutment |
|---------------------------------|-----------------------|-----------------------------|-------------------------------------------|----------------------------------------|----------------------------------|--------------------------------------------------------|-------------------------------------------------|
| | | | | Material<br>IPD<br>Abutment<br>Systems | Titanium<br>alloy, ISO<br>5832-3 | Titanium<br>alloy,<br>ISO 5832-3,<br>Temp.<br>Anodized | Titanium<br>alloy,<br>ISO 5832-3,<br>TiN coated |
| Brånemark | External | 3.5<br>3.75 / 4.0<br>5.0 | NP (3.5 mm)<br>RP (4.1 mm)<br>WP (5.1 mm) | AA | 3.5<br>4.1<br>5.1 | 3.5<br>4.1<br>5.1 | 3.5<br>4.1<br>5.1 |
| Straumann®<br>Tissue<br>Level | Internal | 3.3/4.1/4.8<br>4.8 | RN (4.8 mm)<br>WN (6.5 mm) | DA | 4.8 | 4.8<br>6.5 | 4.8<br>6.5 |
| Tapered<br>Screw-<br>Vent® | Internal | 3.7 / 4.1<br>4.7<br>6.0 | 3.5 mm<br>4.5 mm<br>5.7 mm | FA | 3.5<br>4.5 | 3.5<br>4.5<br>5.7 | 3.5<br>4.5<br>5.7 |
Table 5.1. Summary of IPD abutments categories with compatibilized OEM Implant Systems.
The abovementioned Compatible Implant Systems received FDA-clearance with the following 510(k) concurrence numbers: Brânemark K022562 (aka Various Brånemark System Dental Implant Products); Straumann® Tissue Level K130222 (aka ITI Synocta Meso Abutments), and Tapered Screw-Vent K113756 (aka Tapered Screw-Vent® X Implant).
Ti Base (Interface) abutments are used as an interface between the dental implant and the zirconia superstructure. They are attached (screw-retained) to the implant and cemented to the zirconia superstructure.
The Ti Base (Interface) is a two-piece abutment composed of the Ti Base (Interface) as the bottomhalf and the zirconia superstructure as the top-half. It consists of a pre-manufactured prosthetic component in Titanium alloy per ISO 5832-3, as well as the supporting digital library file for FDAcleared design software (e.g., 3Shape Abutment Designer™ Software cleared under K151455) which enables the design of a patient-specific superstructure by the laboratory/clinician and which will be manufactured in FDA-cleared Zirconia (e.g., DD Bio Z, K142987) according to digital dentistry workflow at the point of care or at a dental laboratory.
The design and fabrication of the zirconia superstructure for Ti Bases (Interfaces) will be conducted using a digital dentistry workflow requiring the use of the following equipment:
Scanner: 3D Scanner D850 Design Software: 3Shape Abutment Designer Software, K151455. Zirconia Material: DD Bio Z, K142987 Milling machine: Brand: Dental Concept System Model: DC1 Milling System Cement: Multilink® Automix, K123397
{6}------------------------------------------------
Ti Base (Interface) abutments design parameters for the zirconia superstructure are defined as follows:
| Minimum wall thickness: | 0.43 mm |
|---------------------------------------------------|---------|
| Minimum post height for single-unit restorations: | 4.75 mm |
| Maximum gingival height: | 6.0 mm |
| Maximum angulation (Occlusal channel) | 20° |
All CAD/CAM superstructures are for straight abutments only.
The laboratory designed superstructure is attached to Ti Base (Interface) by the use of an FDA-cleared cement (i.e. Multilink® Automix, K123397). The resulting final prosthetic restoration is screwed to the dental implant. All subject abutments are single-use and provided non-sterile. Final restoration (which includes the corresponding screw) is intended to be sterilized at the dental clinic before it is placed in the patient.
### V. INDICATIONS FOR USE
IPD Dental Implant Abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for single or multiple dental prosthetic restorations.
| Implant System Compatibility | Implant Diameter (mm) | Platform Diameter |
|------------------------------|-----------------------|-------------------|
| Brånemark | 3.5 | NP (3.5 mm) |
| Brånemark | 3.75 / 4.0 | RP (4.1 mm) |
| Brånemark | 5.0 | WP (5.1 mm) |
| Straumann® Tissue Level | 3.3 / 4.1 / 4.8 | RN (4.8 mm) |
| Straumann® Tissue Level | 4.8 | WN (6.5 mm) |
| Tapered Screw-Vent® | 3.7 / 4.1 | 3.5 mm |
| Tapered Screw-Vent® | 4.7 | 4.5 mm |
| Tapered Screw-Vent® | 6.0 | 5.7 mm |
#### Compatible Implant Systems
The zirconia superstructures for use with the Ti Base (Interface) are only intended to be designed and manufactured according to digital dentistry workflow system integrates multiple components of the digital dentistry workflow: scan files from intra-oral scanners, CAD software, CAM software, ceramic material, milling machine and associated tooling and accessories.
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The subject device is substantially equivalent in indications and design principles to the primary and reference predicate devices. Comparative tables of indications for use and relevant technological characteristics have been provided in the following pages.
{7}------------------------------------------------
Table 5.2. Indications for Use Statements.
| Indications for Use Statements | | | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|--------------------------------|
| Subject device | IPD Dental Implant Abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for single or multiple dental prosthetic restorations. | | |
| IPD Dental Implant Abutments (Implant Protesis Dental 2004 SL) | Compatible Implant Systems | | |
| | Implant System Compatibility | Implant Diameter (mm)<br>Platform Diameter | |
| | Brånemark | 3.5<br>3.75 / 4.0<br>5.0<br>NP (3.5 mm)<br>RP (4.1 mm)<br>WP (5.1 mm) | |
| | Straumann® Tissue Level | 3.3 / 4.1 / 4.8<br>4.8<br>RN (4.8 mm)<br>WN (6.5 mm) | |
| | Tapered Screw-Vent® | 3.7 / 4.1<br>4.7<br>6.0<br>3.5 mm<br>4.5 mm<br>5.7 mm | |
| The zirconia superstructures for use with the Ti Base (Interface) are only intended to be designed and manufactured according to digital dentistry workflow. The workflow system integrates multiple components of the digital dentistry workflow: scan files from intra-oral scanners, CAD software, CAM software, ceramic material, milling machine and associated tooling and accessories. | | | |
| Primary Predicate Device | | | |
| K201860<br>Elos Accurate® Hybrid Base™ | The Elos Accurate® Hybrid Base™ is intended for attaching to dental implants in order to provide basis for single or multiple tooth prosthetic restorations. The Hyblid Base™ is used as an interface between a dental implant and a zirconia superstructure and will be attached to the implant using a prosthetic screw and attached to the zirconia superstructure by cementing. The Elos Accurate® Hybrid Base™ is compatible with the implant systems listed in table 1: | | |
| | Table 1. | | |
| | Implant Platform compatibility | Platform diameter [mm]<br>Implant Body diameter [mm] | |
| | Nobel Replace NP | 3.5<br>3.5 | |
| | Nobel Replace RP | 4.3<br>4.3 | |
| Nobel Replace WP | 5 | 5 | |
| Nobel Replace 6.0 | 6 | 6 | |
| Nobel CC 3.0 | 3 | 3 | |
| Nobel CC NP | 3.5 | 3.5 & 3.75 | |
| Nobel CC RP | 3.9 | 4.3 & 5 | |
| Nobel CC WP | 5.1 | 5.5 | |
| Straumann Bone Level NC | 3.3 | 3.3 | |
| Straumann Bone Level RC | 4.1 & 4.8 | 4.1 & 4.8 | |
| Astra Tech 3.0 | 3 | 3 | |
| Astra Tech 3.5/4.0 | 3.5 & 4 | 3.5 & 4 | |
| Astra Tech 4.5/5.0 | 4.5 & 5 | 4.5 & 5 | |
| Astra Tech EV 3.0 | 3 | 3 | |
| Astra Tech EV 3.6 | 3.6 | 3.6 | |
| Astra Tech EV 4.2 | 4.2 | 3.6 & 4.2 | |
| Astra Tech EV 4.8 | 4.8 | 4.2 & 4.8 | |
| Astra Tech EV 5.4 | 5.4 | 5.4 | |
| Brånemark NP | 3.5 | 3.3 | |
| Brånemark RP | 4.1 | 3.75, 4 & 5 | |
| Brånemark WP | 5.1 | 5 & 6 | |
| The zirconia superstructures for use with the Elos Accurate® Hybrid Base™ are only intended to be designed and manufactured according to digital dentistry workflow. The workflow system integrates multiple components of the digital dentistry workflow: scan files from Intra-Oral Scanners, CAD software, CAM software, ceramic material, milling machine and associated tooling and accessories. | | | |
| Reference Devices | | | |
| K170588<br>DESS Dental Smart Solutions (Terrats Medical) | DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.<br>All digitally designed custom abutments for use with Ti Base abutments or Pre-milled (Blank) abutments are to be sent to a Terrats Medical validated milling center for manufacture. | | |
| Compatible Implant Systems | | | |
| | Compatible Implant System | Implant Body Diameter (mm) | Implant Platform Diameter (mm) |
| | 3i Certain® | 3.25, 4.0, 5.0 | 3.4, 4.1, 5.0 |
| | 3i OSSEOTITE® | 3.25, 3.75, 4.0, 5.0 | 3.4, 4.1, 5.0 |
| | OsseoSpeed™ | 3.5, 4.0, 5.0 | 3.5/4.0, 4.5/5.0 |
| | FRIADENT XIVE | 3.4, 3.8, 4.5 | 3.4, 3.8, 4.5 |
| | NobelActive® | 3.5, 4.3, 5.0 | NP, RP |
| | NobelReplace Conical | 3.5, 4.3, 5.0 | NP, RP |
| | Nobel Replace Trilobe | 3.5, 4.3, 5.0 | NP, RP, WP |
| | Brånemark | 3.5, 3.75/4.0, 5.0 | NP, RP, WP |
| | Straumann® Bone Level | 3.3, 4.1, 4.8 | NC, RC |
| | Straumann® Tissue Level | 3.3, 4.1, 4.8 | RN, WN |
| | Tapered Screw-Vent® | 3.7, 4.1, 4.7, 6.0 | 3.5, 4.5, 5.7 |
| K153521<br>IH Implant System<br>(Sewonmedix) | IH Implant System is device made of titanium and titanium alloy indicated for in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained or screw retained restorations and terminal or interminal abutment support for fixed bridgework. IH Implant System is for single and two stage surgical procedures. It is intended for delayed loading. | | |
| K173908<br>DESS Dental Smart Solutions<br>(Terrats Medical) | DESS Dental Smart Solutions are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.<br><br>All digitally designed custom abutments for use with Aurum™ Abutment or Pre-milled Blank are to be sent to a Terrats Medical validated milling center for manufacture. | | |
| | Compatible Implant Systems | | |
| | Implant System Compatibility | Implant Body | Implant Platform |
| | 3i Certain® | 3.25, 4.0, 5.0 | 3.4, 4.1, 5.0 |
| | 3i OSSEOTITE® | 3.25, 3.75, 4.0, 5.0 | 3.4, 4.1, 5.0 |
| | OsseoSpeed™ | 3.5, 4.0, 5.0 | 3.5/4.0, 4.5/5.0 |
| | FRIADENT XIVE | 3.4, 3.8, 4.5 | 3.4, 3.8, 4.5 |
| | NobelActive® | 3.5, 4.3, 5.0 | NP, RP |
| | NobelReplace® Conical | 3.5, 4.3, 5.0 | NP, RP |
| | NobelReplace® Trilobe | 3.5, 4.3, 5.0 | NP, RP, WP |
| | Branemark | 3.5, 3.75/4.0, 5.0 | NP, RP, WP |
| | Straumann® Bone Level | 3.3, 4.1, 4.8 | NC, RC |
| | Straumann® Tissue Level | 3.3, 4.1, 4.8 | RN, WN |
| | Tapered Screw-Vent®…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.