The Straumann® Variobase™ Abutment is a titanium base placed onto Straumann dental implants to provide support for customized prosthetic restorations. Straumann® Variobase™ Abutments are indicated for screw-retained single tooth or cement-retained single tooth and bridge restorations. All digitally designed copings and/or crowns for use with the Straumann® Variobase™ Abutment system are intended to be sent to Straumann for manufacture at a validated milling center.
Device Story
Straumann Variobase Abutments are two-piece, stock titanium-base (Ti-base) abutments for Straumann dental implants. The system consists of a titanium-aluminum-niobium alloy base and a customized coping/crown. Dental laboratory technicians design the coping/crown using CAD software or traditional burnout methods. Manufacturing occurs via validated Straumann milling, traditional laboratory pressing, or casting. The device is used in dental clinics and laboratories. The healthcare provider uses the abutment to support prosthetic restorations; the output is a functional, customized tooth replacement. Benefits include improved prosthetic fit and support for various restoration types.
Clinical Evidence
Bench testing only. Dynamic fatigue testing was conducted in accordance with the FDA guidance document 'Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments' to evaluate the performance of the abutment system with various coping/crown materials.
Technological Characteristics
Two-piece abutment system. Base: Titanium-aluminum-niobium alloy (ISO 5832-11). Coping/crown materials: Type 4 metals (ISO 22674), cobalt-chromium, noble metal alloys, IPS e.max Press/CAD ceramics. Manufacturing: Validated milling, traditional pressing, or casting. Connectivity: None (mechanical).
Indications for Use
Indicated for patients requiring screw-retained or cement-retained single tooth or bridge restorations supported by Straumann dental implants.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three human profiles facing to the right, connected by a flowing ribbon-like shape.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 9,2015
Straumann USA, LLC Jennifer M. Jackson M.S. Senior Regulatory Affairs Project Manager 60 Minuteman Road Andover, MA 01810
Re: K142890
> Trade/Device Name: Straumann Variobase Abutment NNC, Straumann Variobase Abutment RN, Straumann Variobase Abutment WN, Straumann Variobase Abutment NC, Straumann Variobase Abutment RC, IPS e.max CAD MO Coping, IPS e.max CAD LT Crown, IPS e.max CAD HT Crown, coron CoCr Single Unit Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: December 22, 2014 Received: December 23, 2014
Dear Ms. Jackson.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Susan Russo, DDS, MA
Erin I. Keith, M.S. Director, Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K142890
### Device Name
Strauman® Variobase™ Abutment NNC, Strauman® Variobase™ Abutment RN, Strauman® Variobase™ Abutment WN, Strauman® Variobase™ Abutment NC, Strauman® Variobase™ Abutment RC, IPS e.max® CAD MO Coping, IPS e.max® CAD LT Crown, IPS e.max® CAD HT Crown, coron CoCr Single Unit
### Indications for Use (Describe)
The Straumann® Variobase™ Abutment is a titanium base placed onto Straumann dental implants to provide support for customized prosthetic restorations. Strauman® Variobase™ Abutments are indicated for screw-retained single tooth or cement-retained single tooth and bridge restorations. All digitally designed copings and/or crowns for use with the Straumann® Variobase™ Abutment system are intended to be sent to Straumann for manufacture at a validated milling center.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Straumann® Variobase™ Abutments
510(k) Summary
#### 510(k) Summary 5
#### Submitter's Contact Information 5.1
Straumann USA, LLC (on behalf of Institut Straumann AG)
60 Minuteman Road
Andover, MA 01810
Phone Number: 1-978-747-2509
Fax Number: 1-978-747-0023
Contact Person: Jennifer M. Jackson, MS
Date of Submission: January 8, 2015
#### 5.2 Name of the Device
| Trade Name: | Straumann® Variobase™ Abutment NNC, Straumann® Variobase™ Abutment RN, Straumann® Variobase™ Abutment WN,<br>Straumann® Variobase™ Abutment NC, Straumann® Variobase™ Abutment RC, IPS e.max® CAD MO Coping, IPS e.max® CAD LT Crown, IPS e.max® CAD HT Crown, coron CoCr Single Unit |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Dental Implant Abutment |
| Classification Name: | Abutment, Implant, Dental, Endosseous |
| Regulation Number: | §872.3630 |
#### Predicate Device(s) 5.3
- K120822, Straumann CARES Variobase Abutment NNC, RN, WN, NC, RC . (Institut Straumann AG)
- K120053, IPS e.max® Press Abutment Solutions (Ivoclar Vivadent AG) ●
- K132209, IPS e.max® CAD Abutment Solutions (Ivoclar Vivadent AG) ●
- K132219, Straumann Variobase Abutments ●
{5}------------------------------------------------
Straumann® Variobase™ Abutments
510(k) Summary
## 5.4 Device Description
The Straumann® Variobase™ Abutments are pre-manufactured (stock) abutments, sometimes referred to as "Ti-bases". Straumann® Variobase" Abutments are available to fit Straumann® dental implant platforms NNC (Narrow Neck CrossFit®), RN (Regular Neck), WN (Wide Neck), NC (Narrow CrossFit®), and RC (Regular CrossFit®). A dental laboratory technician would design the corresponding coping and/or crown (the second component of the Variobase two-piece abutment) and/or prosthetic restoration in the dental laboratory via their preferred workflow for pressing, or milling using either a burnout coping or STL model for open CAD software. The coping and/or crown would be manufactured via traditional laboratory methods for pressing or casting, or via validated Straumann milling.
The purpose of this submission is to add coping and/or crown materials to the Straumann® Variobase™ Abutment portfolio for the previously cleared validated Straumann milling workflow and to expand the design/manufacturing workflow to include in-lab pressing and casting techniques and materials.
## 5.5 Intended Use
The Straumann® Variobase™ Abutment is a titanium base placed onto Straumann dental implants to provide support for customized prosthetic restorations. Straumann® Variobase" Abutments are indicated for screw-retained single tooth or cement-retained single tooth and bridge restorations.
## 5.6 Indications for Use
The Straumann® Variobase™ Abutment is a titanium base placed onto Straumann dental implants to provide support for customized prosthetic restorations. Straumann® Variobase"" Abutments are indicated for screw-retained single tooth or cement-retained single tooth and bridge restorations. All digitally designed copings and/or crowns for use with the Straumann® Variobase"" Abutment system are intended to be sent to Straumann for manufacture at a validated milling center.
{6}------------------------------------------------
Straumann® Variobase™ Abutments
510(k) Summary
## 5.7 Technological Characteristics
Straumann® Variobase™ Abutments are two-piece abutments consisting of a premanufactured (stock) abutment made from a titanium-aluminum-niobium alloy and a coping and/or crown which is designed in the dental laboratory by a dental technician and manufactured via traditional in-lab methods of pressing or via validated Straumann milling.
The Ti-base components of the Straumann® Variobase™ Abutments are identical to the Ti-base components of the Straumann predicates (K120822 and K132219). The CAD and press materials which may be used to manufacture the coping/crown component of the Straumann Variobase Abutments are identical to the identified predicate devices and include:
| Casting: | Type 4 metals (ISO 22674) |
|-----------|------------------------------------------------|
| | Base metal alloy (e.g., cobalt-chromium (CoCr) |
| | Noble metal alloy (e.g., gold alloy) |
| Pressing: | IPS e.max® Press Ceramic (K120053) |
| CAD: | IPS e.max® CAD Ceramic (K132209) |
| | coron® |
#### 5.8 Performance Testing
The material used in the manufacture of Straumann® Variobase™ Abutments is a titanium-aluminum-niobium alloy which meets the requirements of ISO 5832-11. Bench testing was performed with each coping/crown material to evaluate the performance of the proposed Straumann® Variobase™ Abutments. Dynamic fatigue tests were conducted in according to the FDA guidance document "Guidance for Industry and FDA Staff – Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments".
{7}------------------------------------------------
Straumann® Variobase™ Abutments
510(k) Summary
## 5.9 Conclusion
The documentation submitted in this premarket notification demonstrates that the Straumann® Variobase™ Abutments are substantially equivalent to the predicate devices and do not pose new issues of safety and effectiveness when used as labeled.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.