Zimmer Zfx Abutment for Zimmer 3.1mmD Implant System

K142572 · Zimmer Dental, Inc. · NHA · Jan 21, 2015 · Dental

Device Facts

Record IDK142572
Device NameZimmer Zfx Abutment for Zimmer 3.1mmD Implant System
ApplicantZimmer Dental, Inc.
Product CodeNHA · Dental
Decision DateJan 21, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3630
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Zimmer Zfx Abutment for Zimmer 3.1mmD Implant System, Titanium is used as a terminal or intermediate abutment for a cemented prosthesis. The abutment can be used for a single or multiple-unit restoration

Device Story

Patient-specific dental implant abutment; designed for Zimmer 3.1mmD implants. Fabricated at manufacturing facility based on patient-specific anatomical needs; replaces manual chair-side or laboratory modification of stock abutments. Composed of Titanium alloy (Ti6Al4V ELI); secured to implant via separate titanium alloy screw. Used by dentists/laboratory technicians to support cemented single or multi-tooth restorations. Provides customized emergence profiles and angles; improves upon traditional stock abutment workflows by pre-machining modifications.

Clinical Evidence

No clinical testing performed. Substantial equivalence supported by non-clinical bench testing, including fatigue and compression testing per FDA guidance for Root-form Dental Implants and Endosseous Dental Implant Abutments, and MRI safety/compatibility evaluation.

Technological Characteristics

Material: Titanium alloy (Ti6Al4V ELI). Interface: Internal Hex/Conical connection, friction fit. Dimensions: 2.9mm platform diameter; 2.9-9.0mm cuff width; 3.0-12.0mm height; 0-30° cone angle. Sold non-sterile; validated for end-user sterilization (SAL 10^-6).

Indications for Use

Indicated for patients requiring single or multiple-unit cemented dental restorations supported by Zimmer 3.1mmD dental implants.

Regulatory Classification

Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 21, 2015 Zimmer Dental, Inc. Ms. Christina Boydston Regulatory Affairs Manager 1900 Aston Avenue Carlsbad, California 92008 Re: K142572 Trade/Device Name: Zimmer Zfx Abutment for Zimmer 3.1mmD Implant System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous dental implant abutment Regulatory Class: II Product Code: NHA Dated: December 19, 2014 Received: December 22, 2014 Dear Ms. Boydston: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Tina Kiang -S for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health {2}------------------------------------------------ ![zimmer dental logo](zimmer_dental_logo.png) **Indications for Use** | 510(k) Number (if known): | K142572 | |---------------------------|---------| |---------------------------|---------| Device Name: Zimmer Zfx Abutment for Zimmer 3.1mmD Implant System Indications For Use: The Zimmer Zfx Abutment for Zimmer 3.1mmD Implant System, Titanium is used as a terminal or intermediate abutment for a cemented prosthesis. The abutment can be used for a single or multiple-unit restoration × Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) {3}------------------------------------------------ Zimmer Dental 1900 Aston Avenue Carlsbad, CA 92008 760.929.4300 (ph) 760.431.7811 (fax) # Traditional 510(k) PRE-MARKET NOTIFICATION 510(k) # 510(k) SUMMARY (21CFR807.92(a)) #### 1. Submitter's Information: | Name: | Zimmer Dental Inc. | |----------------|--------------------| | Address: | 1900 Aston Ave. | | | Carlsbad, CA 92008 | | Phone: | 760-929-4300 | | Contact: | Christina Boydston | | Date Prepared: | September 10, 2014 | - 2. Device Name: Trade Name: Regulation Number: Classification Code: Device Classification Name: Zimmer Zfx Abutment for Zimmer 3.1mmD Implant System 872.3630 NHA Abutment, Implant, Dental, Endosseous - 3. Predicate Device(s): Predicate Device No. 1 Trade Name: 510(k) Number: Regulation Number: Classification Code: Device Classification Name: Zimmer Patient-Specific Abutment, Internal Hex, Titanium K071439 872.3630 NHA Abutment, Implant, Dental, Endosseous ## 4. Device Description: The Zimmer Zfx Abutment for Zimmer 3.1mmD Implant System is designed for use with Zimmer 3.1mmD Dental Implants to support single or multi tooth restorations. The new abutment will be offered in a 2.9mm implant platform. The Zimmer Zfx Abutment for Zimmer 3.1mmD Implant System is a patient specific dental implant abutment. The purpose of a Patient-Specific abutment is to satisfy customer needs that are otherwise difficult to meet with off-the-shelf abutments. They can be manufactured in multiple sizes, shapes, and angles within the limits established in this submission. They frequently incorporate the modifications typically done at a dental laboratory or "chair-side" by a dentist. {4}------------------------------------------------ Traditional methodologies require the customer (dentist/laboratory technician) to begin with a "stock" abutment and use manual subtractive techniques to remove material from this original "stock" design. However, a Patient-Specific abutment will incorporate these same modifications desired by the customer (dentist/laboratory technician) at the time of fabrication at the manufacturing facility. The engineering drawings list ranges in areas (attributes) of the abutment that may be modified depending upon patient-specific needs. The abutment is composed of Titanium alloy (Ti6Al4V ELI), and secured to the implant with a separate Titanium alloy screw for retention. #### న్. Indications for Use: The Zimmer Zfx Abutment for Zimmer 3.1mmD Implant System. Titanium is used as a terminal or intermediate abutment for a cemented prosthesis. The abutment can be used for a single or multiple-unit restoration #### 6. Device Comparison: The new abutment device is substantially equivalent to the predicate relative to material, manufacturing process and general design features. It is fabricated from Titanium alloy. The new abutment will be affixed to the implant by a retaining screw, the same manner as the predicate. | Feature | New Device<br>Zimmer Zfx Abutment for Zimmer<br>2.9mmD Platform | Predicate #1<br>Zimmer Patient-Specific<br>Abutment, Internal Hex, Titanium | |-------------------------|-----------------------------------------------------------------|-----------------------------------------------------------------------------| | Material | Titanium 6Al-4V ELI | Titanium 6Al-4V ELI | | Implant<br>Interface | Internal Hex, Friction fit<br>Conical Connection, Friction fit | Internal Hex, Friction-Fit | | Emergence | Contoured/curved depending on<br>anatomy | Contoured/curved depending on<br>anatomy | | Margin | Pre-machined | Pre-machined | | Platform<br>Diameter | 2.9mm | 3.5, 4.5, 5.7mm | | Cuff Width/<br>Diameter | 2.9mm-9.0mm | 3.5mm-8.0mm | | Minimum<br>Height | 3.0 MIN Cone<br>3.0-12.0 Overall | 3.0-11.5 Cone<br>3.5-12.0 Overall | | Cone Angle | 0-30° | 0-30° | #### 6. Technological Characteristics {5}------------------------------------------------ | Retaining<br>Screw | Cat No. CUAS | Cat No. MHLAS | |--------------------|--------------|---------------| |--------------------|--------------|---------------| ### 8. Non-Clinical Testing: Non-clinical test data was used to support the decision of substantial Non-clinical testing consisted of performance of fatigue and equivalence. compression testing in accordance with the FDA guidance Class II Special Controls Guidance Document: Root-form Dental Implants and Endosseous Dental Implant Abutments. The testing indicates that the new device is strong enough to withstand the anticipated forces and demonstrated improvements over the predicate device(s). In addition, the Zimmer Zfx Abutment for Zimmer 3.1mmD Implant System will be sold non-sterile and will be sterilized by the end user. The sterilization procedures listed in the Instruction For Use were validated to provide a minimum sterility assurance level of 10-9. Additionally, Zimmer Dental implant systems were evaluated for interactions with magnetic fields during Magnetic Resonance Imaging (MRI) in accordance with the FDA Guidance: Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment. This was done to determine that the presence of the abutment poses no additional restrictions on MRI beyond those that would otherwise occur for the patient. #### Clinical Testing 9. No clinical testing was performed. Non-clinical testing was used to support the decision of safety and effectiveness. ## 10. Conclusion Based on our analysis, the device is substantially equivalent to the predicate.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...