LRT ATTACHMENT

K112333 · Bcf Certification, Inc. · NHA · Jan 30, 2012 · Dental

Device Facts

Record IDK112333
Device NameLRT ATTACHMENT
ApplicantBcf Certification, Inc.
Product CodeNHA · Dental
Decision DateJan 30, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3630
Device ClassClass 2

Indications for Use

The LRT Attachment System is indicated for providing a mean of attaching a removable prosthetic device to an endosseous implant body.

Device Story

LRT Attachment System provides mechanical retention for removable dental prostheses on endosseous implants. Device consists of cone-shaped ball abutments that mate with reciprocal covers in denture bases. Unlike predicate snap-on systems relying on metal expansion/constriction, LRT features a mechanical lock-release system activated by digital pressure via identified buttons. Used in dental clinics; operated by dentists or assistants to facilitate manipulation for patients with severe impairments. System improves ease of handling and prosthetic stability.

Clinical Evidence

Clinical evidence includes 18 years of clinical testing to evaluate device integration and handling. In vivo testing confirmed the functionality of the lock/release mechanism. Biocompatibility was assessed via in-vitro cytotoxicity screening assays in accordance with USP 87.

Technological Characteristics

Mechanical dental attachment system. Features cone-shaped ball abutments and a digital pressure-activated lock/release mechanism. Biocompatibility verified per USP 87 standards.

Indications for Use

Indicated for patients requiring attachment of a removable prosthetic device to an endosseous implant body.

Regulatory Classification

Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K112333 JAN 3 0 2012 ## 510(k) Summary | 1. | Name/Address of Submitter: | Innovation TAP inc.<br>1441, rue du Collège<br>Suite 4<br>Montréal, Québec<br>H4L 2L8 CANADA | |----|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------| | 2. | Contact Person: | Emmanuel Montini, Ph.D., Adm.A<br>Co-President, BCF Certification inc.<br>Phone: (514) 927-4635<br>Fax: (514) 397-8515<br>Email: emontini@bcf.ca | | 3. | Date Summary Prepared: | July 15th, 2011 | | 4. | Classification: | 21 CFR 872.3630 | | 5. | Common Name: | Accessory to endosseous dental implan<br>abutment | | 5. | Device Name: | LRT Attachment System | | 6. | Product Code: | NHA | ### 5. Predicate Devices: | LRT<br>System | Attachment | Locator implant anchor K994257<br>Ball Attachment system K955372<br>O-ring Prosthetic Attachment K001638 | |---------------|------------|----------------------------------------------------------------------------------------------------------| |---------------|------------|----------------------------------------------------------------------------------------------------------| ### Device Description: 6. Alike the Locator implant, the O-ring Prosthetic Attachment and the Nobel Ball Attachment systems, the LRT Attachment System provides a mean for attaching a removable prosthetic device to an endosseous implant body with the use of a mechanical based retention, hence eliminating permanent structures that prevent easy access to implant abutments. The LRT Attachment System requires the use of two ball abutments to receive the The presence of additional implant abutments, although not attachment. mandatory, contributes to the support and stability of the prosthesis. These abutments are cone shaped and mate with a reciprocal cover in the denture base. All predicate devices are snap-on devices. The retentive female part is embedded in the denture part and snaps on another part (e.g. a ball abutment). The retentive {1}------------------------------------------------ action relies on the expansion and constriction of the metal female counterpart. The LRT Attachment System provides for an innovative solution by featuring a lock-release retention system activated by digital pressure. This mechanical lock/release action offers ease of manipulation. One of the key elements of the LRT Attachemt System is the easily identified "buttons" of the lock/release mechanism which allow an assistant to manipulate the denture for the severely impaired person. #### 7. Indication for Use: The LRT Attachment System is indicated for providing a mean of attaching a removable prosthetic device to an endosseous implant body. - Brief Description of Clinical and Non-clinical Testing: The activation of the 8. mechanism to provide the lock/release function was tested in vivo to insure the ease of handling. Over 18 years of clinical testing were performed to fine tune the device and its integration to dental prostheses. In these clinical tests, the LRT Attachment System was found substantially equivalent to the cited predicates. Furthermore, Cytotoxicity in-vitro screening assays in accordance with USP 87 were successfully performed to assess the biocompatibility of the material. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. Inside the circle is an abstract image of an eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Mr. Louis-Paul Marin Consultant BCF Certification Incorporation 1100 Rene-Levesque Boulevard, West 25th Floor Montreal CANADA H4L 2L8 Re: K112333 Trade/Device Name: LRT Attachment System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: January 19, 2012 Received: January 23, 2012 Dear Mr. Marin: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. JAN 3 0 2012 {3}------------------------------------------------ ### Page 2 - Mr. Marin Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to Compilation ification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the 1 samall Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, es fer Anthony Diktson Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health ### Enclosure {4}------------------------------------------------ KII 2333 Indication for Use 510(k) Number (if known): # Device Name: LRT ATTACHMENT SYSTEM Indication for Use: The LRT Attachment System is indicated for providing a mean of attaching a removable prosthetic device to an endosseous implant body. ## Concurrence of CDRH Office of Device Evaluation Prescription Use (per 21CFR 801.109) : ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ 2291689.1 OR Over-the-counter Use Susan Gunner 510(k) Number: (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 1-1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...