P.004 ABUTMENTS

K071919 · Straumann USA (On Behalf of Institut Straumann Ag) · NHA · Sep 26, 2007 · Dental

Device Facts

Record IDK071919
Device NameP.004 ABUTMENTS
ApplicantStraumann USA (On Behalf of Institut Straumann Ag)
Product CodeNHA · Dental
Decision DateSep 26, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3630
Device ClassClass 2

Indications for Use

Both the P.004 RC/NC Gold Abutment for Bars and RC/NC Titanium Abutment for Bars are intended to be placed into dental implants to provide support for prosthetic reconstructions such as crowns, bridges and overdentures.

Device Story

The P.004 RC/NC Gold and Titanium Abutments are components of the Straumann P.004 Dental Implant System. These abutments are placed into endosseous dental implants to serve as the interface between the implant and the final prosthetic reconstruction. They provide the necessary support for crowns, bridges, and overdentures in partially or fully edentulous patients. The device is used in a clinical dental setting by a licensed dentist or oral surgeon. The abutments are mechanical components; they do not involve electronic processing, software, or algorithms. By providing a stable connection to the dental implant, they enable the restoration of dental function and aesthetics for the patient.

Clinical Evidence

Bench testing only.

Technological Characteristics

The abutments are manufactured from gold and titanium alloys. They function as mechanical interfaces for endosseous dental implants. The design is consistent with existing dental abutment standards for the Straumann P.004 system.

Indications for Use

Indicated for partially or fully edentulous patients requiring prosthetic reconstructions, including crowns, bridges, and overdentures, supported by endosseous dental implants.

Regulatory Classification

Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 222 % 2007 # Section I 510(k) Summary #### Applicant's Name and Address 1. Straumann US (on behalf of Institut Straumann AG) 60 Minuteman Rd. Andover, MA 01810 Telephone Number: 800-448-8168, ext 2507 Fax Number: 978-747-0023 Contact Person: Paige Sweeney Manager, Regulatory Affairs Date of Submission: July 11, 2007 #### 2. Name of the Device | Trade Name: | P.004 RC/NC Gold Abutment for Bars<br>P.004 RC/NC Titanium Abutment for Bars | |----------------------|------------------------------------------------------------------------------| | Common Name: | Abutment, Dental, Endosseous implants | | Classification Name: | Abutment, Dental, Endosseous implants | | Regulation Number: | §872.3630 | ### es Legally Marketed Device to which Equivalence is Claimed (Predicate Device) Straumann P.004 Dental Implants, K062129 RN synOcta® Gold Coping, K990342 RN synOcta® Titanium Coping, K990342 RN Abutment for Bars, K894844 #### 4. Description of the Device The Straumann P.004 Dental Implant System is an integrated system of endosseous dental implants, which are designed to support prosthetic devices for partially or fully edentulous patients. The system consists of a variety of dental implants, abutments and surgical and prosthetic parts and instruments. {1}------------------------------------------------ #### 5. Intended Use of the Device Both the P.004 RC/NC Gold Abutment for Bars and RC/NC Titanium Abutment for Bars are intended to be placed into dental implants to provide support for prosthetic reconstructions such as crowns, bridges and overdentures. #### Technological Characteristics 6. The proposed abutments are substantially equivalent to the currently marketed devices. The intended use is identical to the predicate devices. The proposed abutment has the same material composition, basic design and fundamental operating principles to the currently marketed devices. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized image of an eagle or bird-like figure with its wings spread, suggesting protection and care. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SEP 2 6 2007 Ms. Paige Sweeney Manager, Regulatory Affairs Straumann USA 60 Minuteman Road Andover, Massachusetts 01810 Re: K071919 Trade/Device Name: P.004 RC/NC Gold and Titanium Abutment for Bars Regulation Number: 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: July 11, 2007 Received: July 12, 2007 Dear Ms. Sweeney: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket . approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Ms. Sweeney Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Smute y. Mchael Oms. Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ K071919 ## Indications for Use Statement P.004 RC/NC Gold and Titanium Abutment for Bars Device Name: Indications for Use: Both the P.004 RC/NC Gold Abutment for Bars and RC/NC Titanium Abutment for Bars are intended to be placed into dental implants to provide support for prosthetic reconstructions such as crowns, bridges and overdentures. × Prescription Use AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Page 1 of 1 Concurrence of CDRH, Office of Device Evaluation (ODE) ![Signature](signature.png) | Division Sign-Off | |-----------------------------------------------| | Division of Anesthesiology, General Hospital, | | Infection Control, Dental Devices | | 510(k) Number: | K071919 | |----------------|---------| |----------------|---------| 510(k) Submission: P.004 RC/NC Gold and Titanium Abutments for Bars July 11, 2007 Proprietary and Confidential
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...