NELLCOR PURITAN BENNETT ACHIEVA VENTILATOR AND REPORT GENERATOR
K990177 · Nellcor Puritan Bennett, Inc. · CBK · Nov 18, 1999 · Anesthesiology
Device Facts
Record ID
K990177
Device Name
NELLCOR PURITAN BENNETT ACHIEVA VENTILATOR AND REPORT GENERATOR
Applicant
Nellcor Puritan Bennett, Inc.
Product Code
CBK · Anesthesiology
Decision Date
Nov 18, 1999
Decision
ST
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Achieva Ventilator is intended to provide ventilatory support for pediatric (Patients should weigh no less than 11 lbs. (5 kg) and adult patients who require positive pressure mechanical ventilation. The ventilator is for use in home, institutional, and portable settings. The Achieva is contraindicated for use with anesthetic gases. This device is intended to be used on the order, and under the supervision, of a physician. The Achieva Report Generator is a management and information tool, to aid in the care of patients requiring ventilatory support. The information displayed by the Report Generator software is intended to be used by healthcare professionals to supplement data obtained using established clinical procedures. The Achieva Report Generator allows users to download, view, and print ventilator performance data thave been stored in the ventilator's internal memory. Achieva Report Generator users can also capture and display a snapshot of real-time waveform data (pressure, peak flow, and volume).
Device Story
Piston-driven ventilator delivering controlled volumes, inspiratory times, breath rates, oxygen concentrations, and PEEP levels; operates in Assist/Control, SIMV, and Spontaneous modes. Models include pressure support (PS/PSx) and internal oxygen blending (PSx/PS/X). Internal modem in PSx/X models. Powered by AC or internal 24 VDC battery (approx. 4 hours). Used in home, institutional, and transport settings under physician supervision. Achieva Report Generator software allows healthcare professionals to download, display, and print stored ventilator performance data and real-time waveforms (pressure, peak flow, volume) via computer. Facilitates clinical management by providing archived data for patient care review.
Clinical Evidence
No clinical data. Safety and efficacy were established through comprehensive non-clinical testing, including software, electrical, mechanical, environmental, and EMC validation. The device demonstrated compliance with IEC 601-1, IEC 601-1-2, UL 2601-1, and partial compliance with ASTM F1100-90 and ASTM F1246-91.
Technological Characteristics
Piston-driven positive pressure ventilator. Energy source: AC or 24 VDC battery. Connectivity: Internal modem (select models) and data download capability to PC. Compliance: IEC 601-1, IEC 601-1-2, UL 2601-1, ASTM F1100-90, ASTM F1246-91. Software: Report generator for data management and waveform display.
Indications for Use
Indicated for pediatric (≥5 kg) and adult patients requiring positive pressure mechanical ventilation in home, institutional, or nonemergency transport settings. Contraindicated for use with anesthetic gases.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
Predicate Devices
Nellcor Puritan Bennett Model LP10 Volume Ventilator (K903010)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for Nellcor Puritan Bennett. The logo consists of a black square with a white swirl design on the left and the company name on the right. The company name is stacked vertically, with "NELLCOR" on top, "PURITAN" in the middle, and "BENNETT" on the bottom.
Jelicer Puntan Bennett Inc 800 Northwest Boulevard neapolis MN 554-1-2625
10 497 4979 Tol. Fron 612 694 3500 Direct 612 59-1 3600 Fax
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is K 990177 ____________________________________________________________________________________________________________________________________________________
| Submitter's Name: | Nellcor Puritan Bennett, Inc. (subsidiary of Mallinckrodt, Inc.) |
|----------------------|------------------------------------------------------------------|
| Submitter's Address: | 2800 Northwest Boulevard, Minneapolis, MN, 55441 |
| Contact Person: | Cheryl Rosenberg |
| Phone Number: | (612) 694-3638 |
| FAX Number: | (612) 694-3600 |
| Summary Date: | January 14, 1999 |
Device Trade Name: Achieva Ventilator and Report Generator
Device Classification Names:
"Continuous Ventilator" (21 CFR 868.5895, Classification Number 73 CBK)
## Predicate Devices:
The Achieva Ventilator is substantially equivalent to both the Nellcor Puritan Bennett Model LP10 Volume Ventilator (K903010) and the Hamilton Medical Amadeus Ventilator (K894316).
## Device Description:
The Achieva Ventilator is a piston driven device which can deliver a wide range of volumes, inspiratory times, breath rates, oxygen concentrations, and PEEP levels. The Achieva has three operating modes (Assist/Control, SIMV, and Spontaneous) and six ventilation modes.
The Achieva Ventilator is available in four different models. The Achieva PSx and Achieva PS have pressure support capability. The Achieva PSx, Achieva PS, and Achieva X have an internal oxygen blender. The Achieva PSx and Achieva X have an internal modem.
The Achieva Ventilator is compatible with previously cleared Nellcor Puritan Bennett patient circuits and accessories. When AC power is unavailable, the Achieva can operate
{1}------------------------------------------------
from an internal 24 VDC battery for approximately 4 hours (normal load) or an optional 24 VDC external battery for approximately 20 hours (normal load).
The Achieva Ventilator is also capable of downloading data to the Achieva Report Generator for purposes of displaying and archiving ventilator data with a computer.
# Indications For Use:
The Achieva Ventilator is intended to provide ventilatory support for pediatric {Patients should weigh no less than 11 lbs. (5 kg)} and adult patients who require positive pressure mechanical ventilation. The ventilator is for use in home, institutional, and nonemergency transport settings.
The Achieva Ventilator is contraindicated for use with anesthetic gases.
The Achieva Report Generator software is intended to be used by healthcare professionals to supplement data obtained using established clinical procedures.
# Nonclinical Performance:
The performance of the Achieva and its interface with the Achieva Report Generator was comprehensively tested (including software, electrical, mechanical, environmental and EMC). All functions as defined in the published specifications were completely validated.
The Achieva complies with the following standards (partial compliance where noted):
- IEC 601-1 (Medical Electrical Equipment Part 1: General Requirements for Safety) .
- IEC 601-1-2 (Medical Electrical Equipment Part 1: General Requirements for ● Safety. 2. Collateral Standard: Electromagnetic Compatibility - Requirements and Tests)
- . UL 2601-1 (Medical Electrical Equipment, Part 1: General Requirements for Safety)
- . ASTM F1100-90 (Standard Specification for Ventilators Intended for Use in Critical Care) (partial compliance)
- ASTM F1246-91 (Standard Specification for Electrically Powered Home Care . Ventilators, Part 1-Positive-Pressure Ventilators and Ventilator Circuits) (partial compliance)
# Clinical Performance:
Clinical testing was not performed on the device. Safety and efficacy were established through non-clinical testing.
## Conclusions from Nonclinical Tests:
The Achieva performs as intended according to its performance specification. The Achieva is substantially equivalent to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three wing-like shapes, representing health, services, and people. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle symbol.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Nov 1 8 1999
Ms. Cheryl Rosenberg Senior Requlatory Affairs Project Manager Mallinckrodt Inc. 2800 Northwest Boulevard Minneapolis, MN 55441-2625
K990177 Re: Nellcor Puritan Bennett Achieva Ventilator and Report Generator Regulatory Class: II (two) Product Code: CBK Dated: August 25, 1999 Received: August 26, 1999
Dear Ms. Rosenberg:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Druq Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
{3}------------------------------------------------
Page 2 - Ms. Cheryl Rosenberg
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
foams A Weckodran fer,
Celia M. Witten, Ph.D., M.D. Acting Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications For Use Statement
Achieva Ventilator and Report Generator Device Name:
Indications For Use:
The Achieva Ventilator is intended to provide ventilatory support for pediatric (Patients should weigh no less than 11 lbs. (5 kg) and adult patients who require positive pressure mechanical ventilation. The ventilator is for use in home, institutional, and portable settings. The Achieva is contraindicated for use with anesthetic gases. This device is intended to be used on the order, and under the supervision, of a physician.
The Achieva Report Generator is a management and information tool, to aid in the care of patients requiring ventilatory support. The information displayed by the Report Generator software is intended to be used by healthcare professionals to supplement data obtained using established clinical procedures. The Achieva Report Generator allows users to download, view, and print ventilator performance data thave been stored in the ventilator's internal memory. Achieva Report Generator users can also capture and display a snapshot of real-time waveform data (pressure, peak flow, and volume).
#### Concurence of CDRH, Office of Device Evaluation (ODE)
Prescription Use X OR Over-The-Counter Use __
510(k) number: K990177
ie am SA Weedon
reurnloqical Devices 5, (k) Number
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.