K060549 · Pulmonetic Systems, Inc. · CBK · May 4, 2006 · Anesthesiology
Device Facts
Record ID
K060549
Device Name
AUTOMOBILE LIGHTER POWER CORD
Applicant
Pulmonetic Systems, Inc.
Product Code
CBK · Anesthesiology
Decision Date
May 4, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The LTV ventilator is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. The ventilator is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician. Specifically, the ventilator is applicable for adult and pediatric patients weighing at least 5 kg (11 lbs.), who require the following types of ventilatory support: - Positive Pressure Ventilation, delivered invasively (via ET tube) or non-invasively (via mask). - Assist/Control, SIMV, or CPAP modes of ventilation. - Breath types including Volume, Pressure Control and Pressure Support, . The ventilator is suitable for use in institutional, home and transport settings.
Device Story
Automobile Lighter Power Cord accessory for LTV 1000 ventilator; enables device operation in automobiles via +12V cigarette lighter or auxiliary power outlets (minimum 20A capacity). Used in institutional, home, and transport settings by qualified personnel. Provides power to ventilator to maintain continuous or intermittent ventilatory support for adult and pediatric patients (≥5 kg).
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
External power cable; connects to +12V automobile cigarette lighter or auxiliary power outlets; requires minimum 20 amperes current; designed for LTV Series ventilators.
Indications for Use
Indicated for adult and pediatric patients weighing at least 5 kg (11 lbs) requiring continuous or intermittent mechanical ventilatory support via invasive (ET tube) or non-invasive (mask) positive pressure ventilation, including Assist/Control, SIMV, or CPAP modes.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
Predicate Devices
LP-6 Ventilator with Cigarette Lighter Cable Y-CG1230 (K905244)
Related Devices
K032226 — LTV-1000 VENTILATOR · Pulmonetic Systems, Inc. · Oct 16, 2003
K060647 — LTV 1200 VENTILATOR · Pulmonetic Systems, Inc. · May 25, 2006
K040790 — MODIFICATION TO LTV 1000 VENTILATOR/BREATHING CIRCUITS · Pulmonetic Systems, Inc. · Jun 3, 2004
K082724 — NEWPORT HTS50 VENTILATOR WITH DUAL PAC BATTERY SYSTEM · Newport Medical Instruments, Inc. · Dec 16, 2008
K253099 — LIFEPAK 15 AC Power Adapter (41577-000333) · Physio-Control, Inc. · Oct 14, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
K 060549
MAY 4 2006
## 1. 510(K) SUMMARY
#### 1.1 SUBMITTER
Pulmonetic Systems, Inc. 17400 Medina Road, Suite 100 Minneapolis, Minnesota 55447-1341
| Contact Person: | Robert C. Samec | |
|-----------------|-----------------|-----------|
| | (763) 398-8305 | Telephone |
| | (763) 398-8400 | Facsimile |
### 1.2 DEVICE / TRADE NAME
Trade Name: Automobile Lighter Power Cord
Common Name: Ventilator Accessory
Classification Name: Ventilator, Continuous (Respirator) 868.5895
#### 1.3 SUBMISSION DATE
Submission Date: March 1, 2006
#### 1.4 DESCRIPTION
The LTV 1000 ventilator is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. The ventilator is suitable for use in institutional, home and transport settings, and is applicable for adult and pediatric patients weighing at least 5 kg (11 lbs.), who require the following types of ventilatory support:
- Positive Pressure Ventilation, delivered invasively (via ET tube) or non-invasively (via . mask).
- Assist/Control, SIMV, or CPAP modes of ventilation. .
- Breath types including Volume, Pressure Control and Pressure Support. .
The addition intended to be cleared by this submission is:
- The addition of the Automobile Lighter Power Cord to be externally attached . to the LTV Series to power the LTV Ventilator while operating in an automobile. The power cord is designed to connect to +12V automobile cigarette lighter or auxiliary power outlets capable of supplying as least 20 amperes of current.
{1}------------------------------------------------
## 1.5 INTENDED USE
The LTV ventilator is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. The ventilator is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician. Specifically, the ventilator is applicable for adult and pediatric patients weighing at least 5 kg (11 lbs.), who require the following types of ventilatory support:
- Positive Pressure Ventilation, delivered invasively (via ET tube) or non-invasively . (via mask).
- Assist/Control, SIMV, or CPAP modes of ventilation. .
- Breath types including Volume, Pressure Control and Pressure Support, .
The ventilator is suitable for use in institutional, home and transport settings.
# EQUIVALENCE TO PREDICATE DEVICE(S) 1.6
The Automobile Lighter Power Cord is substantially equivalent to the following listed devices:
| Predicate Device | 510(k) Clearance | Manufacturer |
|-------------------------------------------------------------|------------------|-----------------------------------------|
| LP-6 Ventilator with<br>Cigarette Lighter<br>Cable Y-CG1230 | K905244 | Puritan Bennett<br>Pleasanton, CA 94588 |
{2}------------------------------------------------
| SUBSTANTIAL EQUIVALENCE SUMMARY TABLE |
|---------------------------------------|
.
| Predicate Device | Modified Device | Discussion of Differences and<br>Similarities |
|---------------------------------------------------------------------------|------------------------------------------------------------------------|-----------------------------------------------|
| LP-6 Ventilator with<br>Cigarette Lighter Cable<br>P/N Y-CG1230 (K905244) | LTV 1000 Ventilator with<br>Automobile Lighter Power Cord<br>P/N 11544 | Same intended use. |
1
:
:
,
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.
4 2006 MAY
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Robert C. Samec Vice President, Product Assurance Pulmonetic Systems, Incorporated 17400 Medina Road, Suite 100 Minneapolis, Minnesota 55447-1341
Re: K060549
Trade/Device Name: Ventilator, Continuous (Respirator) Regulation Number: 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK Dated: March 1, 2006 Received: March 2, 2006
Dear Mr. Samec:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Samec
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: Ventilator, Continuous (Respirator)
Indications for Use:
The LTV 1000 ventilator is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. The ventilator is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician. Specifically, the ventilator is applicable for adult and pediatric patients weighing at least 5 kg (11 lbs), who require the following types of ventilatory support:
- Positive Pressure Ventilation, delivered invasively (via ET tube) or non-invasively (via mask). .
- Assist/Control, SIMV or CPAP modes of ventilation. .
The ventilator is suitable for use in institutional, home, or transport settings.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Tim Sullivan
ion of Anesthesioloo eneral Hos Infection Control. D
**510(k) Number:**
Page **_ of _**/
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.