QBY · Positive Airway Pressure System

Anesthesiology · 21 CFR 868.5273 · Class 2

Overview

Product CodeQBY
Device NamePositive Airway Pressure System
Regulation21 CFR 868.5273
Device ClassClass 2
Review PanelAnesthesiology

Identification

A positive airway pressure delivery system is a prescription noninvasive ventilatory device that delivers expiratory positive airway pressure for patients suffering from obstructive sleep apnea. The system also provides positive airway pressure during incipient apnea. The system may include a dedicated flow generator and a patient interface.

Classification Rationale

Class II (special controls). The special controls for this device are:

Special Controls

In combination with the general controls of the FD&C Act, the positive airway pressure delivery system is subject to the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible. (2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including the following: (i) Waveform testing must simulate breathing conditions and evaluate pressure and airflow response over a range and combination of high and low breath rates and tidal volumes. (ii) Use life testing must demonstrate adequate device performance over the labeled use life of the device. (iii) Device integrity testing must demonstrate that the device can withstand typical forces expected during use. (iv) Carbon dioxide rebreathing testing must be performed. (v) System flow rate, maximum expiratory pressure, inhalation pressure, and intra-mask static pressure testing must be performed. (vi) Air bolus testing must demonstrate that the device can withstand worst-case scenario air pressures. (vii) Maximum limited pressure testing of the flow generator in single fault condition must be performed. (viii) Maximum output temperature testing of delivered gas, if humidified, must be performed. (3) Performance data must validate reprocessing instructions for any reusable components of the device. (4) Performance data must demonstrate the electrical, thermal, and mechanical safety and the electromagnetic compatibility of the device. (5) Software verification, validation, and hazard analysis must be performed. (6) Labeling must include the following: (i) Therapy pressure range; (ii) Use life and replacement schedule for all components; (iii) Cleaning instructions; and (iv) Instructions for assembly and connection of device components.

Recent Cleared Devices (2 of 2)

RecordDevice NameApplicantDecision DateDecision
K200698Somnera SystemFresca Medical, Inc.Jun 29, 2020SESE
DEN170089Curve Positive Airway Pressure SystemFresca MedicalJun 5, 2018DENG

Top Applicants

Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...