K093094 · Cardinal Health 207, Inc. · CBK · May 11, 2010 · Anesthesiology
Device Facts
Record ID
K093094
Device Name
VELA VENTILATOR
Applicant
Cardinal Health 207, Inc.
Product Code
CBK · Anesthesiology
Decision Date
May 11, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Intended Use
The VELA Ventilator is intended to provide continuous or intermittent respiratory support in an institutional health care environment (e.g. hospitals) for individuals who require mechanical ventilation. The ventilator is a restricted medical device intended for use only by qualified trained personnel under the direction of a physician. It may be used on adult and pediatric patients. Specifically, the ventilator is applicable for adult and pediatric patients weighting at least 5 kg (11lb), who require the following types of ventilation support, as prescribed by an attending physician. Positive pressure ventilation Assist / control, SIMV, or CPAP modes of ventilation The ventilator is suitable for use in institutional and transport settings. It is not intended of use in an emergency medical transport ventilator or for home care applications.
Device Story
VELA Ventilator is a self-contained, servo-controlled, software-driven ventilator for adult and pediatric patients (≥5 kg). Device provides positive pressure ventilation (Assist/Control, SIMV, CPAP) in institutional and transport settings. Features include a flat-panel color LCD, touch screen, membrane keys, and dial for parameter adjustment. Precision gas delivery turbine with active inhalation/exhalation control. Optional mainstream CO2 monitoring via external CAPNOSTAT 5 sensor. Software performs real-time charting, digital monitoring, and automated calculations including f/Vt (rapid shallow breathing index) and circuit resistance. Operated by qualified personnel under physician direction. Output allows clinicians to monitor ventilation effectiveness and adjust therapy settings. Benefits include flexible, easy-to-use respiratory support with integrated monitoring capabilities.
Clinical Evidence
Bench testing only. Performance testing verified that the VELA Ventilator meets its performance requirements and is substantially equivalent to legally marketed predicate devices.
Technological Characteristics
Servo-controlled, software-driven ventilator. Features precision gas delivery turbine, flat-panel color LCD, touch screen interface. Connectivity includes LEMO connector for external CO2 sensor (SELV). Software-based monitoring of f/Vt and circuit resistance. Designed for institutional use with liquid-resistant casing.
Indications for Use
Indicated for adult and pediatric patients weighing at least 5 kg (11 lb) requiring continuous or intermittent mechanical ventilatory support (Positive pressure, Assist/Control, SIMV, or CPAP) in institutional or transport settings. Contraindicated for emergency medical transport or homecare use.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
K013642 — AVEA · Bird Products Corp. · May 22, 2002
K073069 — MODIFICATION TO AVEA VENTILATOR · Viasys Respiratory Care, Inc. · Jan 25, 2008
K162738 — Puritan Bennett 980 Ventilator System · Covidien · Oct 27, 2017
K181170 — Trilogy Evo Universal · Respironics, Inc. · Jul 18, 2019
K140575 — ENGSTROM CARESTATION, ENGSTROM PRO · Datex-Ohmeda, Inc. · Feb 3, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
K093094
MAY 1 1 2010
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
| Manufacturing<br>Site: | Cardinal Health 211<br>1100 Bird Center Drive<br>Palm Springs, CA 92262 |
|-------------------------------------------|-------------------------------------------------------------------------|
| Contact: | Farokh Etemadieh (714) 919-3249 (phone); (714) 283-8420 (fax) |
| Summary Date | June 5,-2009 |
| Device Trade<br>Name: | VELA Ventilator |
| Device<br>Common/Classific<br>ation Name: | Classification name: 868.5895 Continuous Ventilator, 73 CBK |
| Establishment<br>Registration<br>Number | 202171 |
| Device Class: | Class II |
| Classification<br>Panel: | Anesthesiology |
| Predicate Device: | The predicate devices are: |
| | • Maquet Servo-I K022132<br>• LTV 1200: K060647 |
- Cardinal Health AVEA Ventilator: K073069 .
{1}------------------------------------------------
Device Description:
A VELA with the CO2 monitoring option is intended to monitor the effectiveness of ventilation. It may be used on adult and pediatric patients. It is suitable for use in a hospital critical care environment, and is a restricted medical device intend for use by qualified trained personnel under the direction of a physician.
The VELA Series Ventilator is a modified Tbird Ventilator and like its predecessor, is an easy to use, self-contained, servo-controlled, software-driven ventilator. It has a dynamic range of breathing gas delivery that provides for pediatric through adult patients. Its revolutionary user interface module provides maximum flexibility with simple operator interaction. It has a flat panel color LCD with real time charting and digital monitoring capabilities, a touch screen for easy interaction, membrane keys and a dial for changing settings and operating parameters. A precision gas delivery turbine with servo controlled active inhalation and exhalation improves performance over previous generations.
The VELA Series Ventilator may be configured as a conventional ventilator or non-invasive positive pressure ventilator (NPPV). It is easy to clean and its design does not allow liquids to pool on the casing, reducing the likelihood of fluid leakage into the body of the ventilator.
The VELA Series Ventilator base model comes with comprehensive features for the critical care environment. Optional features can be added at time of purchase or at a later date.
{2}------------------------------------------------
The Vela ventilator is intended to provide continuous or intermittent Intended Use: ventilatory support for the care of individuals who require mechanical verification. The ventilator is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician. Specifically, the ventilator is applicable for adult and pediatric patients weighing at least 5 kg (11lb), who require the following general types of ventilation support, as prescribed by attending physician:
> Positive pressure ventilation Assist/Control, SIMV, or CPAP modes of ventilation
The ventilator is suitable for use in institutional and transport settings. It is not intended for use as an emergency medical transport ventilator or for homecare applications.
Substantial Equivalence: Predicate Device Information
The predicate device(s) are:
- Maquet Servo-I: K022132 .
- LTV 1200: K060647 .
- Cardinal Health AVEA: K073069 .
- The modifications to VELA associated with this submittal as follows:
- An End Tidal CO2 monitoring option consisting of an external mainstream CO2 ● sensor module, the CAPNOSTAT 5, OEMed from Respironics and for added safety connected to the device using a LEMO connector at extra safety low voltage (SELV).
- The ventilator will be applicable for adult and pediatric patients weighing at least 5 kg . (11 Ib), who require continuous respiratory support, as prescribed by an attending physician.
- Monitoring of f/Vt and Circuit Resistance. . F/Vt monitoring by definition is the rapid shallow breathing index calculated by the ratio of the monitored spontaneous breath rate and monitored tidal volume. F/Vt is described at page 118 and Circuit Resistance at page 75 of the Vela operator manual L2854-101 Rev. B. and exhibit A.
- The setting range of the PSV Cycle changed from 5 30% to 5 70%. .
Summary of Testing and Validation:
In Summary Performance testing verified that the VELA Ventilator meets it's performance requirements and that this device is substantially equivalent to predicate medical devices currently legally marketed in the United States.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
### MAY 1 1 2010
Mr. Farokh Etemadieh Regulatory Affairs Consultant Cardinal Health 207, Incorporated 22745 Savi Ranch Parkway Yorba Linda, California 92887
Re: K093094
Trade/Device Name: Vela Ventilator Regulation Number: 21 CFR 868,5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK Dated: April 22, 2010 Received: April 27, 2010
Dear Mr. Etemadieh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2- Mr. Farokh Etemadieh
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 53) -542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOtfices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
h for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indication for Use
510(k) number (if known): K093094
Device Name: VELA Ventilator Indications for Use:
> The VELA Ventilator is intended to provide continuous or intermittent respiratory support in an institutional health care environment (e.g. hospitals) for individuals who require mechanical ventilation. The ventilator is a restricted medical device intended for use only by qualified trained personnel under the direction of a physician. It may be used on adult and pediatric patients. Specifically, the ventilator is applicable for adult and pediatric patients weighting at least 5 kg (11lb), who require the following types of ventilation support, as prescribed by an attending physician.
- Positive pressure ventilation
- Assist / control, SIMV, or CPAP modes of ventilation
The ventilator is suitable for use in institutional and transport settings. It is not intended of use in an emergency medical transport ventilator or for home care applications.
#### WARNING
The Vela ventilator is approved for institutional use only and should not be used to transport patients outside of the institutional setting.
Prescription Use V AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE OF NEEDED)
> Concurrence of CDRH Office of Device Evaluation ODE) Sheltons and
L. Schulten
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices . 510(k) Number: K093094 Page of (Posted
November 13, 2003)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.