GE DATEX-OHMEDA HELIOX DELIVERY SYSTEM

K041524 · Datex-Ohmeda, Inc. · CBK · Sep 20, 2004 · Anesthesiology

Device Facts

Record IDK041524
Device NameGE DATEX-OHMEDA HELIOX DELIVERY SYSTEM
ApplicantDatex-Ohmeda, Inc.
Product CodeCBK · Anesthesiology
Decision DateSep 20, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5895
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The GE Datex-Ohmeda Aptaér Heliox delivery system is designed to deliver heliox from a source gas cylinder to spontaneously breathing patients via a facemask using pressure support. A built-in nebulizer, the Aerogen Aeroneb Pro (K021175) is provided for adding nebulized medication to the delivered heliox. The system is designed for facility use and should only be used under the orders of a clinician. The Aptaér Heliox Delivery System is not intended as a life support device and is not intended for intubated patients.

Device Story

System delivers heliox gas from source cylinder to spontaneously breathing patients via facemask using pressure support; includes integrated Aerogen Aeroneb Pro nebulizer for medication delivery. Designed for facility use; operated under clinician orders. Not for life support or intubated patients. Provides therapeutic gas delivery to assist breathing; nebulizer adds medication to gas stream. Benefits patient by facilitating heliox therapy in non-intubated settings.

Clinical Evidence

Bench testing only. No clinical data provided. Compliance with voluntary standards (UL 2601, ASTM F100, EN/IEC 60601-1, EN/IEC 60601-1-2, EN 475, CGA V-1, ISO 5145, EN 980) used to demonstrate safety and effectiveness.

Technological Characteristics

Continuous ventilator for heliox delivery. Includes integrated Aerogen Aeroneb Pro nebulizer. Complies with UL 2601, ASTM F100 (Critical Care Ventilators), EN/IEC 60601-1 (Electrical Safety), EN/IEC 60601-1-2 (EMC), EN 475 (Alarm Signals), CGA V-1/ISO 5145 (Gas Cylinders), and EN 980 (Symbols).

Indications for Use

Indicated for spontaneously breathing patients requiring heliox delivery via facemask using pressure support. Not for life support or intubated patients. Must be used in a facility under clinician orders.

Regulatory Classification

Identification

A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and features the department's name around the perimeter. In the center is a stylized image of an eagle or bird-like figure with three curved lines representing its wings or body. The logo is black and white. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 OCT 1 8 2004 Mr. Dan Kosednar Manager, Regulatory Planning & Submissions Datex-Ohmeda, Incorporated CARE Business Area P.O. Box 7550 Madison, Wisconsin 53707 Re: K041524 Trade/Device Name: GE Datex-Ohmeda Aptaér Heliox Delivery System Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK Dated: August 27, 2004 Received: August 30, 2004 Dear Mr. Kosednar: This letter corrects our substantially equivalent letter of June 17, 2004 regarding the trade name. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent [{for the indications for use stated in the enclosure)] to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ ## Page 2 - Mr. Kosednar Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Que Saliem ba Chiu Lin,(Ph.D.) Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ K041524 | Date: | June 4, 2004 | SEP 20 2004 | |-----------------|-----------------------------------------------------------------------------------------------------------------|-------------| | Subject: | 510(k) Summary of Safety and Effectiveness Information<br>for the GE Datex-Ohmeda Aptaér Heliox Delivery System | | | Proprietary: | GE Datex-Ohmeda Aptaér Heliox Delivery System | | | Common: | Ventilator, Continuous | | | Classification: | Anesthesiology, 73 CBK , 21 CFR 868.5895 | | The 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 1992. The GE Datex-Ohmeda Aptaér Heliox Delivery System is substantially equivalent to the following currently marketed device: Viasys Avea Ventilator - Class II - 21CFR868.5895, which has been the subject of a cleared 510(k) with FDA log number K022674. Puritan Bennett 7200 Series Ventilator -Class II - 21CFR868.5895, which has been the subject of several cleared 510(k)s, most recently with FDA log number K922705 The GE Datex-Ohmeda Aptaér Heliox delivery system is designed to deliver heliox from a source gas cylinder to spontaneously breathing patients via a facemask using pressure support. A built-in nebulizer, the Aerogen Aeroneb Pro (K021175) is provided for adding nebulized medication to the delivered heliox. The system is designed for facility use and should only be used under the orders of a clinician. The Aptaér Heliox Delivery System is not intended as a life support device and is not intended for intubated patients. The GE Datex- Aptaér Heliox Delivery System was designed to comply with the applicable portions of the following voluntary standards; - I . UL 2601 General requirements for Medical Electrical Equipment - 2. ASTM F100 Particular Requirements for Critical Care Ventilators - 3. EN/IEC 60601-1: General requirements for Medical Electrical Equipment - 4. EN/IEC 60601-1-2: 1998 Medical Electrical Equipment Electromagnetic Compatibility - 5. EN 475 Electrically Generated Alarm Signals - 6. CGA V-1 ad ISO 5145 Medical Gas Cylinders Threaded Cylinders - 7. EN 980 Graphical Symbols {3}------------------------------------------------ The GE Datex-Ohmeda Aptaér Heliox Delivery System and the currently marketed device are substantially equivalent in design concepts, technologies and materials. The GE Datex-Ohmeda Aptaer Heliox Delivery System has been validated through rigorous testing that, in part, supports the compliance of GE Datex-Ohmeda Aptaér Heliox Delivery System to the standards listed above. Dan Kosednar, RAC Contact: Manager, Regulatory Planning and Submissions {4}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K041524 Device Name: GE Datex-Ohmeda Aptaér Heliox Delivery System Indications For Use: The GE Datex-Ohmeda Aptaér Heliox delivery system is designed to deliver The GE Datex-Onmeda Aptacl Ticlion delivery Systems heliox from a source gas cylinder to spontaneously breathing patients on Anragen nellox from a source gas bylinder to open . A built-in nebulizer, the Aerogen seated facemask using procedro on possible nebulized medication to the Aeroneb Pro (KUZTT7J) is provided for adaing naturing we and should only be used under the orders of a clinician. The Aptaér Heliox Delivery System is not intended as a life support device and is not intended for intubated patients. Prescription Use _XXX_ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use _ (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) сихбаевот Page 1 of 1___ (Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices 510(k) Number
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...