K072588 · Respironics Inc., Sleep & Home Respiratory Group · CBK · Dec 18, 2007 · Anesthesiology
Device Facts
Record ID
K072588
Device Name
PERFORMAX SE TOTAL FACE MASK
Applicant
Respironics Inc., Sleep & Home Respiratory Group
Product Code
CBK · Anesthesiology
Decision Date
Dec 18, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Performax SE Total Face Mask is intended to provide a patient interface for application of noninvasive ventilation. The mask is to be used as an accessory to ventilators which have adequate alarms and safety systems for ventilator failure, and which are intended to administer CPAP or positive pressure ventilation for treatment of respiratory failure, respiratory insufficiency or obstructive sleep apnea. The mask is for multi-patient reuse on patients weighing > 30 kg), who are appropriate candidates for noninvasive ventilation, in the hospital/institutional environment only.
Device Story
The Respironics Performax SE Total Face Mask is a patient interface for noninvasive ventilation; it connects to a ventilator circuit to deliver CPAP or positive pressure ventilation. The device consists of a polycarbonate faceplate and a silicone cushion seal; it features headgear hooks for attachment. It is designed for multi-patient reuse in hospital or institutional environments. The mask is available in two sizes (small and large). It functions as an accessory to ventilators equipped with appropriate alarms and safety systems. The device helps treat respiratory failure, respiratory insufficiency, or obstructive sleep apnea by providing a secure seal for air delivery. Clinicians select the mask for patients weighing over 30 kg who are candidates for noninvasive ventilation.
Clinical Evidence
Bench testing only.
Technological Characteristics
Polycarbonate faceplate; silicone cushion seal; polycarbonate standard elbow interface. Two sizes available (small, large). Designed for multi-patient reuse. Non-powered, passive interface device.
Indications for Use
Indicated for patients > 30 kg requiring noninvasive ventilation for respiratory failure, respiratory insufficiency, or obstructive sleep apnea. For use in hospital/institutional settings only.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
K101129 — ESPRIT VENTILATOR WITH AF531 SE FULL FACE MASK · Respironics, Inc. · Sep 16, 2010
K243394 — AF531 Oro-Nasal SE Face Mask · Respironics, Inc. · Dec 30, 2024
K072592 — PERFORMAX TOTAL FACE MASK · Respironics Inc., Sleep & Home Respiratory Group · Dec 19, 2007
K150638 — AF541 SE Full Face Mask · Respironics, Inc. · Sep 18, 2015
K041362 — RESMED HOSPITAL FULL FACE MASK · Resmed, Ltd. · Jul 15, 2004
Submission Summary (Full Text)
{0}------------------------------------------------
Traditional 510(k) Tab 5 – 510(k) summary
K072588
Respironics Performax SE Total Face Mask
# TAB 5
| 510(K) SUMMARY | |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date of Submission | 12 September 2007 |
| Official Contact | Zita A. Yurko<br>Director, Regulatory Affairs<br>Respironics, Inc.<br>1001 Murry Ridge Lane<br>Murrysville, PA 15668<br>Zita.yurko@respironics.com<br>724-387-4120 t<br>724-882-4120 c<br>724-387-4216 f<br>DEC 1 8 2007 |
| Classification Reference | 21 CFR 868.5895; 21CFR 858.5905 |
| Product Code | CBK - Continuous Ventilator; BZD - non-continuous ventilator |
| Common/Usual Name | Ventilator, continuous, facility use; ventilator, non-continuous (respirator) |
| Proprietary Name | Respironics Performax SE Total Face Mask |
| Predicate Device(s) | Respironics Total Face Mask (K992969) - BZD<br>Image 3 SE Face Mask (K023135) - CBK |
| Reason for submission | new device |
## Substantial Equivalence
The Respironics Performax SE Total Face Mask has the following similarities to the previously cleared predicate device:
- Same intended use.
- Same operating principle. ப
- Same technology. D
- Same manufacturing process. ා
{1}------------------------------------------------
This premarket notification submission demonstrates that the Performax SE Total Face Mask is substantially equivalent to the Respironics Total Face Mask (K992969). Design modifications have been made to the Total Face Mask for this submission. These modifications are described here in. Based on the testing performed, none of the design modification affect the safety or effectiveness of the device.
The following changes have been made:
- 프 The change in the face plate design to contour the face
- I The change in the sealing cushion design
- r The addition of a standard elbow without exhalation similar to the standard elbow used on the Image 3 SE. This design is intended for use with a ventilator that has an integral safety valve.
- ı The addition of the claim for multi-patient use to be included.
### Intended Use
The Performax SE Total Face Mask is intended to provide a patient interface for application of noninvasive ventilation. The mask is to be used as an accessory to ventilators which have adequate alarms and safety systems for ventilator failure, and which are intended to administer CPAP or positive pressure ventilation for treatment of respiratory failure, respiratory insufficiency or obstructive sleep apnea.
The mask is for multi-patient reuse on patients weighing > 30 kg), who are appropriate candidates for noninvasive ventilation, in the hospital/institutional environment only.
#### Device Description
The Respironics Performax SE Total Face Mask consists of a polycarbonate faceplate with silicon cushion seal for the face. Like the device predicate, Image 3SE full face mask (K023135), the interface to the patient circuit is a polycarbonate standard elbow. The design of the standard elbow (SE) is the same as the SE used on the Image 3SE Full Face mask. The mask when used with the standard elbow has one integrated exhalation feature, which includes one port on the faceplate. Similar to the device predicate, Total Face Mask (K992969), the mask faceplate contains headgear hooks upon which the premium headgear is attached. The mask is available in two sizes - small and large.
{2}------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
The Respironics Performax SE Total Face Mask is intended for use with a patient circuit that is used to connect the device to the patient interface device (mask). A typical patient circuit consists of a six-foot disposable or reusable smooth lumen 22mm tubing, a method of venting exhaled gases.
. . . . . . . . . . . ........................................................................................................................................................
..... .. .....................................................................................................................................................................
(End of Tab.)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
: : :
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three heads, representing the department's focus on health, human services, and well-being. The eagle is surrounded by a circular border with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" written around it.
DEC 1 8 2007
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Zita Yurko Director, Regulatory Affairs Respironics Incorporated, Sleep & Home Respiratory Group 1001 Murry Ridge Lane Murrysville, Pennsylvania 15668
Re: K072588
Trade/Device Name: Respironics Performax SE Total Face Mask Regulation Number: 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK Dated: November 21, 2007 Received: November 23, 2007
Dear Ms. Yurko:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Yurko
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): ________________
:
Device Name: ____Respironics Performax SE Total Face Mask_____________________________________________________________________________________________________________________
The Performax SE Total Face Mask is intended to provide a patient interface for application of noninvasive ventilation. The mask is to be used as an accessory to ventilators which have adequate alarms and safety systems for ventilator failure, and which are intended to administer CPAP or positive pressure ventilation for treatment of respiratory failure, respiratory insufficiency or obstructive sleep apnea.
The mask is for multi-patient reuse on patients weighing > 30 kg, who are appropriate candidates for noninvasive ventilation, in the hospital/institutional environment only.
Prescription Use ____________________________________________________________________________________________________________________________________________________________
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Aut J. Wate
K072588
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches