The AF541 SE Full Face Mask is intended to provide a patient interface for application of noninvasive ventilation. The mask is to be used as an accessory to ventilators which have adequate alarms and safety systems for ventilation failure, and which are intended to administer CPAP or positive pressure ventilation for treatment of respiratory failure, respiratory insufficiency or obstructive sleep apnea. The mask is for single use in the hospital/institutional environment only. The mask is to be used on patients (>40lbs/20kg) who are appropriate candidates for noninvasive ventilation.
Device Story
Oral-nasal full face mask; provides patient interface for noninvasive ventilation; accessory to ventilators administering CPAP or positive pressure; used in hospital/institutional environments. Features interchangeable Over-the-Nose (OTN) and Under-the-Nose (UTN) cushions on a common frame; includes 4-point headgear and capstrap for oral access. Clinician-operated; facilitates respiratory support for patients with respiratory failure or obstructive sleep apnea. Benefits include patient-specific fit via multiple cushion sizes and configurations, ensuring effective pressure delivery and patient comfort during ventilation therapy.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical bench testing, including biocompatibility (muscle implantation, genotoxicity, intracutaneous injection, Kligman maximization, agar diffusion) and performance testing (pressure drop, leak, deadspace).
Technological Characteristics
Materials: Silicone (cushion), Polycarbonate (frame), Polypropylene (elbow), Nylon/Spandex/Polyurethane foam (headgear). Design: Oral-nasal mask with interchangeable OTN/UTN cushions. Standards: ISO 17510-2, ISO 10993-1, ISO 14971. Non-sterile, single-use. No software or electronic components.
Indications for Use
Indicated for patients >40lbs/20kg requiring noninvasive ventilation, CPAP, or positive pressure ventilation for respiratory failure, respiratory insufficiency, or obstructive sleep apnea in hospital/institutional settings.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 18, 2015
Respironics, Inc. Ms. Shaylee Masilunas Regulatory Affairs Engineer 1001 Murry Ridge Lane Murrysville, PA 15668
Re: K150638
Trade/Device Name: AF541 SE Full Face Mask Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: Class II Product Code: CBK Dated: August 14, 2015 Received: August 17, 2015
Dear Ms. Shaylee Masilunas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K150638
Device Name AF541 SE Full Face Mask
#### Indications for Use (Describe)
This Mask is intended to provide a patient interface for application of noninvasive ventilation. The mask is to be used as an accessory to ventilators which have adequate alams and safety systems for ventilation failure, and which are intended to administer CPAP or positive pressure ventilation for treatment of respiratory insufficiency or obstructive sleep apnea. The mask is for single use in the hospital/institutional environment only. The mask is to be used on patients (>401bs/20kg) who are appropriate candidates for noninvasive ventilation.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Page 1 of 1
PSC Publishing Services (Jul) 443-6740 EB
{3}------------------------------------------------
# 510(k) Summary
| Date | March 9, 2015 |
|---------------------------------------|---------------------------------------------------------------------|
| 510(k) Owner | Respironics, Inc.<br>1001 Murry Ridge Lane<br>Murrysville, PA 15668 |
| | (724) 387-7729<br>(724) 387-3999 (fax) |
| Official Contact | Shaylee Masilunas |
| | Regulatory Affairs Engineer |
| Establishment<br>Registration # | 2518422 |
| Proprietary Name | AF541 SE Full Face Mask |
| Common/Usual Name | Mask Accessory to a Continuous Ventilator |
| Classification Panel | Anesthesiology Devices |
| Classification<br>Reference | 21 CFR 868.5895 |
| Classification Name /<br>Product Code | CBK - Continuous Ventilator (Anesthesiology) |
| Predicate Device(s) | AF531 SE Full Face Mask (K101129) |
## Indication for Use
The AF541 SE Full Face Mask is intended to provide a patient interface for application of noninvasive ventilation. The mask is to be used as an accessory to ventilators which have adequate alarms and safety systems for ventilation failure, and which are intended to administer CPAP or positive pressure ventilation for treatment of respiratory failure, respiratory insufficiency or obstructive sleep apnea. The mask is for single use in the hospital/institutional environment only. The mask is to be used on patients (>40lbs/20kg) who are appropriate candidates for noninvasive ventilation.
{4}------------------------------------------------
### Device Description
The AF541 SE Full Face Mask is an oral-nasal full face mask that is available in two cushion configurations. The AF541 SE Full Face has two cushion configurations, an Over the Nose (OTN) cushion and an Under the Nose (UTN) cushion. The AF541 SE Full Face Mask will have interchangeable cushions that attach to a common frame. There will be four sizes available for the over the nose option (S, M, L, XL). The UTN option will have three sizes available (A, B, C). The AF541 SE will include a 4 point headgear and capstrap to allow for oral access with either headgear option.
### Similarities and Differences of the Subject Device Compared to the Predicate Devices
The AF541 SE Full Face Mask has the following similarities to the previously cleared predicate devices AF531 SE Full Face Mask (K101129):
- Similar intended use ●
- Same operating principle ●
- Similar design ●
- Similar materials
- Similar manufacturing process
The AF541 SE Full Face Mask has the following differences in the technological characteristics to the previously cleared predicate devices AF531 SE Full Face Mask (K101129):
- Patient population ●
- Patient usage type ●
- Interchangeable cushion design
- . Mask materials
#### Table 1: Comparison Table of Respironics AF541 EE Full Face Mask and predicate device
| Area | Predicate Device<br>AF531 with Standard Elbow (SE) –<br>Small Size (K101129) | Subject Device<br>AF541 SE Full Face Mask |
|-----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended<br>Use | The AF531 SE is intended to provide<br>a patient interface for application of<br>noninvasive ventilation. The mask is<br>to be used as an accessory to<br>ventilators which have adequate<br>alarms and safety systems for<br>ventilation failure, and which are<br>intended to administer CPAP or<br>positive pressure ventilation for<br>treatment of respiratory failure,<br>respiratory insufficiency or<br>obstructive sleep apnea. The mask is<br>for single use in the<br>hospital/institutional environment | This AF541 SE Full Face Mask is<br>intended to provide a patient interface<br>for application of noninvasive<br>ventilation. The mask is to be used as<br>an accessory to ventilators which have<br>adequate alarms and safety systems<br>for ventilation failure, and which are<br>intended to administer CPAP or positive<br>pressure ventilation for treatment of<br>respiratory failure, respiratory<br>insufficiency or obstructive sleep<br>apnea. The mask is for single use in<br>the hospital/institutional environment |
{5}------------------------------------------------
| Area | Predicate Device<br>AF531 with Standard Elbow (SE) -<br>Small Size (K101129) | Subject Device<br>AF541 SE Full Face Mask |
|--------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| | only. The mask is to be used on<br>patients 7 years or older<br>(>40lbs/20kg) who are appropriate<br>candidates for noninvasive<br>ventilation. | only. The mask is to be used on<br>patients (>40lbs/20kg) who are<br>appropriate candidates for noninvasive<br>ventilation. |
| Patient<br>Population | 7 years or older (>40lbs/20kg) | Similar to K101129.<br>(>40lbs/20kg) |
| Patient<br>Usage Type /<br>Environment<br>of Use | Single use in the hospital/institutional | Unchanged from K101129. |
| Product Code | CBK | Unchanged from K101129. |
| Provided<br>Sterile or<br>Non-Sterile | Non-Sterile | Unchanged from K101129. |
| Anatomical<br>Sites | Nose and Mouth | Unchanged from K101129. |
| Device<br>Design | 1. Faceplate/frame<br>2. Cushion<br>3. Elbow<br>4. Headgear<br>5. Headgear clips | 1. Frame<br>2. Cushion<br>3. Elbow<br>4. Headgear<br>5. Headgear talon clips |
| Sizes | One size - Small | Over the Nose (OTN) version has four<br>sizes (S, M, L, XL)<br>Under the Nose (UTN) version has<br>three sizes (A, B, C) |
{6}------------------------------------------------
| Component | Primary Predicate: | Subject Device: |
|-------------------|---------------------------------------|-------------------------------------|
| | Device: | Device: |
| | AF531 SE Mask - Small Size<br>K101129 | AF541 SE Full Face Mask |
| | Manufacturer: | Manufacturer: |
| | Respironics, Inc. | Respironics, Inc. |
| Cushion | Silicone | Silicone |
| Frame | Polycarbonate | Polycarbonate |
| Elbow | Polycarbonate | Polypropylene |
| Headgear/Capstrap | Nylon/Spandex, Polyurethane<br>Foam | Nylon/Spandex, Polyurethane<br>Foam |
#### Table 2: Material comparison for the Respironics AF541 SE Full Face Mask and its predicate device
New materials used in this mask are classified as external communicating, tissue contact, with a contact duration C (> 30 days cumulative). The following biocompatibility tests were completed:
- Muscle Implantation Study in Rabbits 4 Weeks ●
- Muscle Implantation Study in Rabbits 12 Weeks ●
- Genotoxicity: Bacterial Reverse Mutation Study .
- . Genotoxicity: Mouse Lymphoma Assay
- . Genotoxicity: Mouse Peripheral Blood Micronucleus Study
- . Intracutaneous Injection Test
- Kligman Maximization Test
- Agar Diffusion Test (Direct Contact)
Design verification tests were performed on the AF541 SE Full Face Mask. All tests were verified to meet the required acceptance criteria. Respironics has determined that the modifications have not raised new safety and effectiveness concerns of the device. In summary, the device described in this submission is substantially equivalent to the predicate device.
## Clinical Tests
Clinical tests were not required to demonstrate the substantial equivalence of the AF541 SE Full Face Mask. Product functionality has been adequately assessed by non-clinical tests.
### Non-Clinical Tests
Performance testing was performed before and after cleaning treatments to verify that the device modifications did not raise new safety and effectiveness concerns of the subject device. Performance testing included:
- Total Mask Pressure Drop .
- Total Mask Leak ●
{7}------------------------------------------------
- . Deadspace
The AF541 SE Full Face Mask has been designed per the following standards:
- . ISO 17510-2 Sleep Apnoea Devices Part 2: Masks and Application Accessories
- . ISO 10993-1 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process
- . ISO 14971 Medical devices – Application of risk management to medical devices
### Conclusion
The performance and technological characteristics of the AF541 SE Full Face Mask are substantially equivalent to those of the AF531 SE Full Face Mask (K101129) and raise no new types of safety or effectiveness questions.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.