K243394 · Respironics, Inc. · CBK · Dec 30, 2024 · Anesthesiology
Device Facts
Record ID
K243394
Device Name
AF531 Oro-Nasal SE Face Mask
Applicant
Respironics, Inc.
Product Code
CBK · Anesthesiology
Decision Date
Dec 30, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
Medium and Large Size: The AF531 Oro-Nasal SE Face Mask is intended to provide a patient interface for application of noninvasive ventilation. The mask is to be used as an accessory to ventilators which have adequate alarms and safety systems for ventilation failure, and which are intended to administer CPAP or positive pressure ventilation for treatment of respiratory failure, respiratory insufficiency, or obstructive sleep apnea. The mask is for single use in the hospital/institutional environment only. The mask is to be used on patients (>30 kg) who are appropriate candidates for noninvasive ventilation. Small Size: The AF531 Oro-Nasal SE Face Mask is intended to provide a patient interface for application of noninvasive ventilation. The mask is to be used as an accessory to ventilators which have adequate alarms and safety systems for ventilation failure, and which are intended to administer CPAP or positive pressure ventilation for treatment of respiratory failure, respiratory insufficiency, or obstructive sleep apnea. The mask is for single use in the hospital/institutional environment only. The mask is to be used on patients 7 years or older (>20 kg) who are appropriate candidates for noninvasive ventilation.
Device Story
Oro-nasal face mask providing patient interface for noninvasive ventilation; covers nose and mouth. Used as accessory to ventilators administering CPAP or positive pressure ventilation. Features click-style elbow for circuit connection; requires separate exhalation device. Used in hospital/institutional environments by clinicians. Provides secure interface for respiratory support; benefits patients with respiratory failure, insufficiency, or obstructive sleep apnea. Available in small, medium, and large sizes with four-point or CapStrap headgear options.
Clinical Evidence
Bench testing only. Verification testing confirmed performance requirements including total mask leak, pressure drop, sound levels, and physical dead space. Biocompatibility evaluated per ISO 10993 and ISO 18562 standards.
Technological Characteristics
Materials: Polycarbonate/polysiloxane faceplate, silicone cushion, polypropylene/polycarbonate elbows, polyurethane foam forehead spacer. Connectivity: 22mm connector. Design: Oro-nasal mask with click-style elbow. Standards: ISO 17510, ISO 10993 series, ISO 18562 series, ISO 5356-1, ISO 14971. Non-sterile.
Indications for Use
Indicated for noninvasive ventilation as an accessory to ventilators for CPAP or positive pressure ventilation in patients with respiratory failure, respiratory insufficiency, or obstructive sleep apnea. Medium/Large sizes for patients >30 kg; Small size for patients ≥7 years (>20 kg). For single-patient use in hospital/institutional settings.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".
December 30, 2024
Respironics, Inc. Gary LeMere Principal Regulatory Affairs Specialist 1001 Murry Ridge Ln Murrysville, Pennsylvania 15668
Re: K243394
Trade/Device Name: AF531 Oro-Nasal SE Face Mask Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: Class II Product Code: CBK Dated: October 31, 2024 Received: October 31, 2024
Dear Gary Lemere:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Ethan L. Nyberg -S
Ethan Nyberg, Ph.D. Assistant Director DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
Submission Number (if known)
K243394
Device Name
AF531 Oro-Nasal SE Face Mask
#### Indications for Use (Describe)
Medium and Large Size:
The AF531 Oro-Nasal SE Face Mask is intended to provide a patient interface for application of noninvasive ventilation. The mask is to be used as an accessory to ventilators which have adequate alarms and safety systems for ventilation failure, and which are intended to administer CPAP or positive pressure ventilation for treatment of respiratory failure, respiratory insufficiency, or obstructive sleep apnea. The mask is for single use in the hospital/institutional environment only. The mask is to be used on patients (>30 kg) who are appropriate candidates for noninvasive ventilation.
Small Size:
The AF531 Oro-Nasal SE Face Mask is intended to provide a patient interface for application of noninvasive ventilation. The mask is to be used as an accessory to ventilators which have adequate alarms and safety systems for ventilation failure, and which are intended to administer CPAP or positive pressure ventilation for treatment of respiratory failure, respiratory insufficiency, or obstructive sleep apnea. The mask is for single use in the hospital/institutional environment only. The mask is to be used on patients 7 years or older (>20 kq) who are appropriate candidates for noninvasive ventilation.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(K) Summary – K243394
| Date Prepared | December 27, 2024 |
|------------------------------------------------------|---------------------------------------------------------------------------------------|
| Company Name/ Owner | Respironics, Inc.<br>1001 Murry Ridge Lane<br>Murrysville, PA 15668<br>(724) 733-0200 |
| Official Contact | Gary LeMere<br>Principal Regulatory Affairs Specialist |
| Proprietary Name(s) | AF531 Oro-Nasal SE Face Mask |
| Common/Usual Name | Oro-Nasal Face Mask |
| Classification | Class II |
| Product Code | CBK - ventilator, continuous, facility use |
| Regulation Name | Continuous ventilator |
| Regulation Number | 21 CFR 868.5895 |
| Review Panel | Anesthesiology |
| Primary Predicate Device<br>510(k)<br>Product Code | AF541 SE Full Face Mask<br>K150638<br>CBK |
| Secondary Predicate Device<br>510(k)<br>Product Code | AF531 SE Full Face Mask<br>K101129<br>CBK |
{5}------------------------------------------------
#### Device Description
The AF531 Oro-Nasal SE Face Mask is designed for single-patient use in the hospital or institutional environment. AF531 Oro-Nasal SE Face Mask covers the nose and mouth while avoiding the eye region. The mask is available in three (3) sizes: Small, Medium, and Large.
The headgear options for the AF531 Oro-Nasal Masks include the four-point headgear and a CapStrap headgear.
The AF531 Oro-Nasal SE Face Mask requires use of a separate exhalation device. The mask utilizes a click-style elbow that secures to the mask hub yet can be removed by pressing the release tabs.
#### Indications for Use
#### AF531 Oro-Nasal SE Face Mask
Medium and Large Size: The AF531 Oro-Nasal SE Face Mask is intended to provide a patient interface for application of noninvasive ventilation. The mask is to be used as an accessory to ventilators which have adequate alarms and safety systems for ventilation failure, and which are intended to administer CPAP or positive pressure ventilation for treatment of respiratory failure, respiratory insufficiency, or obstructive sleep apnea. The mask is for single use in the hospital/institutional environment only. The mask is to be used on patients (>30 kg) who are appropriate candidates for noninvasive ventilation.
Small Size: The AF531 Oro-Nasal SE Face Mask is intended to provide a patient interface for application of noninvasive ventilation. The mask is to be used as an accessory to ventilators which have adequate alarms and safety systems for ventilation failure, and which are intended to administer CPAP or positive pressure ventilation for treatment of respiratory failure, respiratory insufficiency, or obstructive sleep apnea. The mask is for single use in the hospital/institutional environment only. The mask is to be used on patients 7 years or older (>20 kg) who are appropriate candidates for noninvasive ventilation.
| Design<br>Parameter or<br>Feature | Subject Device:<br>AF531 Oro-Nasal SE Face<br>Mask | Primary Predicate:<br>AF541 SE Full Face Mask<br>K150638<br>(September 18, 2015) | Secondary Predicate:<br>AF531 SE Full Face Mask<br>K101129<br>(September 16, 2010) | Comments |
|--------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|
| | | Intended Use | Intended Use | |
| Classification<br>Product Code | CBK | CBK | CBK | Identical |
| Regulation<br>Number | 21 CFR§ 868.5895 | 21 CFR§ 868.5895 | 21 CFR§ 868.5895 | Identical |
| Classification | II | II | II | Identical |
| Indications for<br>Use | Medium and Large:<br>The AF531 Oro-Nasal SE<br>Face Mask is intended to<br>provide a patient interface<br>for application of<br>noninvasive ventilation. The<br>mask is to be used as an<br>accessory to ventilators<br>which have adequate alarms<br>and safety systems for<br>ventilation failure, and which | The AF541 SE Full Face<br>Mask is intended to provide<br>a patient interface for<br>application of noninvasive<br>ventilation. The mask is to<br>be used as an accessory to<br>ventilators which have<br>adequate alarms and safety<br>systems for ventilation<br>failure, and which are | The AF531 SE Full Face<br>Mask is intended to<br>provide a patient interface<br>for application of<br>noninvasive ventilation.<br>The mask is to be used as<br>an accessory to ventilators<br>which have adequate<br>alarms and safety systems<br>for ventilation failure, and | Equivalent |
| Design<br>Parameter or<br>Feature | Subject Device:<br>AF531 Oro-Nasal SE Face<br>Mask with click-style elbow | Primary Predicate:<br>AF541 SE Full Face Mask<br>K150638<br>(September 18, 2015) | Secondary Predicate:<br>AF531 SE Full Face Mask<br>K101129<br>(September 16, 2010) | Comments |
| | are intended to administer<br>CPAP or positive pressure<br>ventilation for treatment of<br>respiratory failure,<br>respiratory insufficiency or<br>obstructive sleep apnea.<br>The mask is for single use in<br>the hospital/institutional<br>environment only. The mask<br>is to be used on patients<br>(>30 kg) who are<br>appropriate candidates for<br>noninvasive ventilation.<br><br>Small:<br>The AF531 Oro-Nasal SE<br>Face Mask is intended to<br>provide a patient interface<br>for application of<br>noninvasive ventilation. The<br>mask is to be used as an<br>accessory to ventilators<br>which have adequate alarms<br>and safety systems for<br>ventilation failure, and which<br>are intended to administer<br>CPAP or positive pressure<br>ventilation for treatment of<br>respiratory failure,<br>respiratory insufficiency or<br>obstructive sleep apnea.<br>The mask is for single use in<br>the hospital/institutional<br>environment only. The mask<br>is to be used on patients 7<br>years or older (>20 kg) who<br>are appropriate candidates<br>for noninvasive ventilation. | intended to administer<br>CPAP or positive pressure<br>ventilation for treatment of<br>respiratory failure,<br>respiratory insufficiency or<br>obstructive sleep apnea.<br>The mask is for single use<br>in the hospital/ institutional<br>environment only. The<br>mask is to be used on<br>patients (>40lbs/20kg) who<br>are appropriate candidates<br>for noninvasive<br>ventilation. | which are intended to<br>administer CPAP or<br>positive pressure<br>ventilation for treatment of<br>respiratory failure,<br>respiratory insufficiency, or<br>obstructive sleep apnea.<br>The mask is for single use<br>in the hospital/institutional<br>environment only. The<br>mask is to be used on<br>patients 7 years or older<br>(>40lbs/20kg) who are<br>appropriate candidates for<br>noninvasive<br>ventilation | |
| Patient<br>Population | Small Size: Patients >7<br>years (>20kg)<br>Medium/Large Size: >30kg | All sizes: Patients<br>>40lbs/20kg | Small Size: Patients >7<br>years (>40lbs/20kg) | Equivalent |
| Functional<br>Indication | Interface for application of<br>noninvasive ventilation | Interface for application of<br>noninvasive ventilation | Interface for application of<br>noninvasive ventilation | Identical |
| Environment<br>of Use | Hospital/<br>Institutional Environment | Hospital/<br>Institutional Environment | Hospital/<br>Institutional Environment | Identical |
| Reprocessing<br>Claims | Single use | Single use | Single use | Identical |
| Anatomical<br>Sites | Nose and mouth | Nose and mouth | Nose and mouth | Identical |
| Design<br>Parameter or<br>Feature | Subject Device:<br>AF531 Oro-Nasal SE Face<br>Mask with click-style elbow | Primary Predicate<br>AF541 SE Full Face Mask<br>K150638<br>(September 18, 2015) | Secondary Predicate:<br>AF531 SE Full Face Mask<br>K101129<br>(September 16, 2010) | Comments |
| Sterile State as<br>Provided | Non-Sterile | Non-Sterile | Non-Sterile | Identical |
| Useful Life | ≤ 7 days | Not Defined | Not Defined | Equivalent |
| Device Design | Polycarbonate faceplate<br>Silicone cushion<br>Click-Style Elbows<br>SE option without<br>entrainment valve<br>Elbow retention hub to<br>retain click-style elbow<br>Headgear<br>Accessory Port Cap | Nylon Faceplate<br>Silicone cushion<br>Click-Style Elbows<br>SE option without<br>entrainment valve<br>No elbow retention hub<br>integral to AF541<br>design; elbows attach<br>via faceplate<br>Headgear<br>Accessory Port Cap | Polycarbonate<br>faceplate<br>Silicone cushion<br>Press-fit Elbows<br>SE option without<br>entrainment valve<br>Elbow retention hub to<br>retain press-fit elbow<br>Headgear<br>Pick-off Port Cap | Equivalent |
| Faceplate/<br>Cushion Type | Covers nose and mouth | Covers nose and mouth | Covers nose and mouth | Identical |
| Exhalation<br>Device Design | Separate exhalation device<br>required | Separate exhalation device<br>required | Separate exhalation device<br>required | Identical |
| Headgear Type | CapStrap or four-point<br>headgear | CapStrap or four-point<br>headgear | CapStrap or four-point<br>headgear | Identical |
| Patient Circuit<br>Connection | 22mm click-style elbow with<br>22mm connector | 22mm click-style elbow with<br>22mm connector | 22mm press-fit<br>elbow with 22mm<br>connector | Equivalent |
| Mask Sizes | Small, Medium, and Large | Small, Medium, Large, and<br>Extra-Large | Small | Equivalent |
| Materials -<br>Elbows | SE Click-Style Elbow:<br>Polypropylene (Blue)<br>Accessory Port Cap:<br>Silicone | SE Click-Style Elbow:<br>Polypropylene (Blue)<br>Accessory Port Cap:<br>Silicone | SE Press-Fit Elbow:<br>Polycarbonate (Blue)<br>Pick-off Port Cap:<br>Silicone | Identical to<br>Primary<br>predicate |
| Materials -<br>Mask and<br>Headgear<br>Component | Mask Materials:<br>Faceplate:<br>Polycarbonate/polysiloxane<br>Cushion: Polysiloxane<br>Click-Style Elbow Hub:<br>Polycarbonate<br>Split Washer: Polycarbonate<br>Forehead Arm:<br>Polycarbonate/ polysiloxane<br>Support Retaining Clip:<br>Polycarbonate<br>Forehead Bracket:<br>Polycarbonate/<br>polysiloxane<br>Disposable Foam Forehead<br>Spacer: Polyurethane foam | Mask Materials:<br>Faceplate: Nylon<br>Over the nose / under the<br>nose Cushion: Silicone<br>Over the Nose / under the<br>nose Hub: Nylon<br>Headgear Materials:<br>4 Point Crown Headgear<br>Materials:<br>Forehead Pad Bracket:<br>Nylon<br>Forehead Adjuster Ball<br>Post:<br>Nylon<br>Forehead Adjuster Button:<br>Nylon<br>Disposable Foam Forehead<br>Spacer:<br>Polyurethane | Mask Materials:<br>Faceplate: Polycarbonate/<br>polysiloxane<br>Cushion: Polysiloxane<br>Press-Fit Elbow Hub:<br>Polycarbonate<br>Split Washer:<br>Polycarbonate<br>Forehead Arm:<br>Polycarbonate/<br>polysiloxane<br>Support Retaining Clip:<br>Polycarbonate<br>Forehead Bracket:<br>Polycarbonate/<br>polysiloxane<br>Disposable Foam<br>Forehead Spacer:<br>Polyurethane foam | Equivalent |
| Design<br>Parameter or | Subject Device:<br>AF531 Oro-Nasal SE Face | Primary Predicate<br>AF541 SE Full Face Mask | Secondary Predicate:<br>AF531 SE Full Face Mask | Comments |
| Feature | Mask with click-style elbow | K150638<br>(September 18, 2015) | K101129<br>(September 16, 2010) | |
| | | Headgear Clip:<br>Polypropylene,<br>Tension Clip:<br>Polyoxymethylene<br>Velcro Hook: Nylon<br>Headgear Material: Nylon<br>Polyester Polyurethane,<br>Nylon | | |
| | CapStrap Materials:<br>CapStrap Halo: Nylon<br>CapStrap Soft Goods:<br>Nylon, Polyester/<br>Polyurethane,<br>Polyester/Lycra Knit Elastic<br>CapStrap Clip:<br>Acetal copolymer | CapStrap Headgear:<br>Cap: Nylon Fabric, Foam,<br>Pad Bracket: Nylon<br>Adhesive: Loctite<br>Headgear Clip:<br>Polypropylene<br>Velcro Hook: Nylon | CapStrap Materials:<br>CapStrap Halo: Nylon<br>CapStrap Soft Goods:<br>Nylon,<br>Polyester/ Polyurethane,<br>Polyester/Lycra Knit<br>Elastic<br>CapStrap Clip:<br>Acetal copolymer | |
| | Headgear Materials:<br>Headgear Clip:<br>Acetal copolymer<br>Headgear Material: Nylon,<br>Lycra, Polyester/<br>polyurethane,<br>Polyoxymethylene | Headgear Material: Nylon<br>Polyester Polyurethane | Headgear Materials:<br>Headgear Clip:<br>Acetal copolymer.<br>Headgear Material: Nylon,<br>Lycra Polyester/<br>polyurethane,<br>Polyoxymethylene | |
| Performance Specifications | | | | |
| Pressure<br>Range | 4 to 40cmH2O | 4 to 40cmH2O | 4 to 40cmH2O | Identical |
| Total Mask<br>Leak | Specification:<br>≤15 SLPM@4.0cmH2O<br>≤25 SLPM@40.0cmH2O | SE Specification, all sizes:<br>≤15 SLPM@4.0cmH2O<br>≤25 SLPM@40.0cmH2O | Specification:<br>≤15 SLPM@4.0cmH2O<br>≤25 SLPM@40.0cmH2O | Identical |
| Pressure Drop | Specifications:<br>≤ 1.0cmH2O@50SLPM<br>≤ 4.0cmH2O@100SLPM | Specifications:<br>≤ 1.0cmH2O@50SLPM<br>≤ 4.0cmH2O@100SLPM | Specifications:<br>≤ 1.0cmH2O@50SLPM<br>≤ 4.0cmH2O@100SLPM | Identical |
| Sound Power<br>and Pressure<br>Level | Test results at 10cmH2O:<br>Power Level:18.2dBA@1m<br>Pressure<br>Level:10.9dBA@1m…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.