Panther 5

K213098 · Origin Medical Devices · CBK · Jul 20, 2023 · Anesthesiology

Device Facts

Record IDK213098
Device NamePanther 5
ApplicantOrigin Medical Devices
Product CodeCBK · Anesthesiology
Decision DateJul 20, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5895
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The Panther 5 is intended to provide continuous ventilation to patients requiring respiratory support. The device is intended for patients from neonates with Veight (VLBW) <1500 gm to adult and for a wide variety of clinical conditions. The device is intended to be used in hospital type facilities and during hospital and hospital-type transport provided that electrical power and compressed Oxygen are supplied.

Device Story

Panther 5 is a continuous ventilator providing respiratory support to patients ranging from neonates (VLBW <1500g) to adults. Device operates in hospital facilities and during hospital-type transport; requires external electrical power and compressed oxygen. Provides mechanical ventilation to assist or control patient breathing. Clinicians use device to manage respiratory function for diverse clinical conditions. Benefits include life-sustaining respiratory support for patients unable to breathe adequately on their own.

Indications for Use

Indicated for patients requiring respiratory support, ranging from neonates (VLBW <1500g) to adults, across various clinical conditions. For prescription use in hospital settings and hospital-type transport.

Regulatory Classification

Identification

A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. July 20, 2023 Origin Medical Devices % Ray Kelly Consultant Arazy Group Consultants Inc. 3422 Leonardo Lane New Smyrna Beach, Florida 32168 Re: K213098 Trade/Device Name: Panther 5 Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: Class II Product Code: CBK Dated: June 13, 2023 Received: June 20, 2023 Dear Ray Kelly: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, | Ethan L.<br>Nyberg -S | 2023.07.20<br>16:53:50 -04'00' | |--------------------------------------------|--------------------------------| | Ethan Nyberg, Ph.D. | | | Assistant Director | | | DHT1C: Division of Sleep Disordered | | | Breathing, Respiratory and | | | Anesthesia Devices | | | OHT1: Office of Ophthalmic, Anesthesia, | | | Respiratory, ENT and Dental Devices | | | Office of Product Evaluation and Quality | | | Center for Devices and Radiological Health | | Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K213098 Device Name Panther 5 ## Indications for Use (Describe) The Panther 5 is intended to provide continuous ventilation to patients requiring respiratory support. The device is intended for patients from neonates with Veight (VLBW) <1500 gm to adult and for a wide variety of clinical conditions. The device is intended to be used in hospital type facilities and during hospital and hospital-type transport provided that electrical power and compressed Oxygen are supplied. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| | <span> ☒ Prescription Use (Part 21 CFR 801 Subpart D) </span> | |-----------------------------------------------------------------------------| | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...