NEWPORT E500 WAVE VENTILATOR

K030780 · Newport Medical Instruments, Inc. · CBK · Jan 23, 2004 · Anesthesiology

Device Facts

Record IDK030780
Device NameNEWPORT E500 WAVE VENTILATOR
ApplicantNewport Medical Instruments, Inc.
Product CodeCBK · Anesthesiology
Decision DateJan 23, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5895
Device ClassClass 2
AttributesTherapeutic, Pediatric

Intended Use

The e500 Ventilator is intended to provide continuous (ET tube) or non-continuous (mask) ventilatory support and monitoring for infant, pediatric and adult patients with a tidal volume of ≥ 20 mL. The device is prescription use only. The intended environments include hospital, hospital-type and intra-hospital transport environments. Hospital use typically includes general care floors, operating rooms, special procedure areas, and intensive and critical care areas within the hospital. Hospital-type use includes facilities such as or similar to surgicenters, sub-acute centers, and special nursing facilities outside of the hospital. Intra-hospital transport includes patient transport within the hospital or hospital-type facility.

Device Story

The Newport e500 Wave Ventilator is a self-contained, electrically powered, microprocessor-controlled ventilator. It provides continuous or intermittent ventilatory support for infant, pediatric, and adult patients. The device features front-panel controls for selecting operational modes, including volume control, pressure control, and pressure support. It incorporates a comprehensive alarm system for both technical and patient-related parameters. The device is intended for use in hospital settings (ICU, OR, general care) and intra-hospital transport. It includes an internal battery providing approximately 1.5 hours of power. Healthcare providers use the device to manage patient ventilation; the output (ventilatory support and monitoring data) assists in clinical decision-making for respiratory care. The device benefits patients by providing reliable, controlled mechanical ventilation across a wide range of patient sizes.

Clinical Evidence

Bench testing only. The device was tested for performance against FDA Reviewer Guidance for Ventilators and various international standards including IEC601-1, CISPR 11, EN60601-1-2, EN 794-1, ISO9703-1&2, ASTM F 1100-90, MIL-STD461D, and MIL-STD810E. Biocompatibility analysis was performed to ensure safety for the intended environment.

Technological Characteristics

Self-contained, electrically powered, microprocessor-controlled ventilator. Supports infant/pediatric (≥ 20 mL) to adult patients. Features internal battery (1.5-hour capacity). Standards: IEC601-1, CISPR 11, EN60601-1-2, EN 794-1, ISO9703-1&2, ASTM F 1100-90, MIL-STD461D, MIL-STD810E. Connectivity: Standalone. Sterilization: Not specified.

Indications for Use

Indicated for infant, pediatric, and adult patients requiring mechanical ventilatory support (continuous via ET tube or non-continuous via mask) with tidal volumes ≥ 20 mL. Prescription use only.

Regulatory Classification

Identification

A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ JAN 2 3 2004 Image /page/0/Picture/1 description: The image shows a handwritten text that appears to be an identifier or code. The text is "K039780" and is written in a slightly tilted orientation. The handwriting is simple and clear, with each character distinctly formed, though the overall appearance is somewhat rough due to the handwritten nature. ## Newport Medical Instruments Inc. Traditional 510(k) Submission, e500 Wave Ventilator Image /page/0/Picture/4 description: The image shows the word "NME" in bold, black letters. There are two horizontal lines on either side of the word, which are the same color and thickness as the letters. The letters are sans-serif and appear to be a custom font. The background is white. Section 4 page 15 # SECTION 4. 510(k) SUMMARY This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92. - Submitter's Name and Address A. Newport Medical Instruments, Inc. 760 West 16th Street - Building N Costa Mesa, California 92627 Mailing Address: P.O. Box 2600 Newport Beach, California 92658 - B. Submitter's Phone and FAX Number (949) 642-3910 Telephone (949) 645-5026 FAX - C. Name of Contact Person Richard Waters Vice President, Quality Assurance & Regulatory Affairs - D. Date the Summary was Prepared March 8, 2003 CONFIDENTIAL: Newport Medical Instruments Inc., considers the existence of this information to be confidential as defined in 21 CFR 814.9, and the data and information to be privileged or confidential as defined in 21 CFR Part 20.61. Therefore, Newport Medical Instruments requests that the confidentiality of this submission be afforded the protection provided by 21 CFR Part 814.9k, 21 CFR Part 20, 5 USC 551, 18 USC 1905, 21, 21 USC 331(j) and all other applicable regulations and laws. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the word "NME" in bold, black letters. There are two horizontal lines, one on top and one on the bottom, that flank the word. The lines are the same color and thickness as the letters. The background is white. Traditional 510(k) Submission, e500 Wave Ventilator #### E. Name of the Device - 1. Trade or Proprietary Name Newport e500 Wave Ventilator - 2. Common or Usual Name e500 Ventilator - Classification Name 3. 868.5895 Continuous Ventilator 73CBK - Device Class 4. Class II ### Summary of Substantial Equivalence F. The Newport e500 Wave Ventilator, the E200 Ventilator and the Siemens Servo Ventilators are intended to provide continuous or intermittent ventilatory support for the care of the individuals who require mechanical ventilation. The performance and clinical features of each ventilator supports infant/pediatric (= 20 mL) through adult applications. The materials and design of this device are similar to those of the predicate devices. The technical characteristics of the e500 Ventilator do not introduce new questions regarding safety and effectiveness of critical care ventilators. Additionally, the labeling of the e500 provides similar information as the predicate devices. The information provided in the 510(k) submission supports the determination of substantial equivalence. Biocompatibility analysis demonstrates that the product is safe for its intended environment. Software design and development, including verification and validation testing, have been conducted using the FDA's Reviewer Guidance of Medical Device Software CONFIDENTIAL: Newport Medical Instruments Inc., considers the existence of this information to be confidential as defined in 21 CFR 814.9, and the data and information to be privileged or confidential as defined in 21 CFR Part 20.61. Therefore, Newport Medical Instruments requests that the confidentiality of this submission be afforded the provided by 21 CFR Part 814.9k, 21 CFR Part 20, 5 USC 551, 18 USC 1905, 21, 21 USC 331(j) and all other applicable regulations and laws. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the word "NMI" in bold, black letters. There are two thick, horizontal lines above and below the word. The lines are the same color and thickness as the letters. The letters are sans-serif and appear to be a custom font. ## Traditional 510(k) Submission, e500 Wave Ventilator Section 4 page 17 Submissions, December 15, 1995 draft as a guidance, and per company internal requirements. ### G. Description of the Device The Newport e500 Wave Ventilator is a self-contained, electrically powered microprocessor controlled ventilator. Performance characteristics and clinical features support infant/pediatric (=20mL) through adult patients. Front panel controls allow trained operators to select between a number of operational modes, pressure support and volume or pressure control. A comprehensive alarm system is built-in to alert the user to violations of preset safety limits. When fully charged, the internal battery provides an average of 1.5 hours of power. The alarms associated with the e500 meet or exceed standards of critical care ventilators and have been developed in compliance with ISO9703-1, ISO9703-2 and EN475. The alarms of the e500 span both technical (ventilator related) alarms and non technical alarms (patient related alarms). ### H. Device Intended Use The e500 Ventilator is intended to provide continuous (ET tube) or non-continuous (mask) ventilatory support and monitoring for infant, pediatric and adult patients with a tidal volume of = 20 mL. The device is prescription use only. The intended environments include hospital, hospital-type and intra-hospital transport environments. Hospital use typically includes general care floors, operating rooms, special procedure areas, and intensive and critical care areas within the hospital. Hospital-type use includes facilities such as or similar to surgicenters, sub-acute centers, and special nursing ficilities outside of the hospital. Intra-hospital transport includes patient transport within the hospital or hospital-type facility. ### 1. Summary of Comparison of Technological Characteristics The e500 is substantially equivalent with the E200 Ventilator and the Siemens Servo Ventilators across the spectrum of the patient population (infant/pediatric to adult) for which each was designed. All of the devices share common modalities (volume control, CONFIDENTIAL: Newport Medical Instruments Inc., considers the existence of this information to be confidential as defined in 21 CFR 814.9, and the data and information to be privileged or confidential as defined in 21 CFR Part 20.61. Therefore, Newport Medical Instruments requests that the confidentiality of this submission be afforded the protection provided by 21 CFR Part 814.9k, 21 CFR Part 20, 5 USC 551, 18 USC 1905, 21, 21 USC 331(j) and all other applicable regulations and laws, {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the word "NME" in bold, black letters. The letters are stylized with horizontal lines above and below them. The font appears to be sans-serif and the overall design is simple and impactful. Traditional 510(k) Submission, e500 Wave Ventilator Section 4 page 18 pressure control, pressure support) and significantly overlap in the clinical range of function for their target population. Although some differences in design application are noted, the essential clinical function of each device is significantly similar and mimics each other in the typical frame of use by the health care provider. Each provide significant safety features in terms of alarms, back up ventilation, and fail safe mechanisms. Each provides internal battery, ease of use, simple circuit design and matching clinical function. CONFIDENTIAL: Newport Medical Instruments Inc., considers the existence of this information to be confidential as defined in 21 CFR 814.9, and the data and information to be privileged or confidential as defined in 21 CFR Part 20.61. Therefore, Newport Medical Instruments requests that the confidentiality of this submission be afforded the protection provided by 21 CFR Part 814.9k, 21 CFR Part 20, 5 USC 551, 18 USC 1905, 21, 21 USC 331(j) and all other applicable regulations and laws. {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the letters 'NME' in a bold, sans-serif font. The letters are black and appear to be part of a logo or title. There are two horizontal lines on either side of the letters, which are also black and bold. Section 4 page 19 Traditional 510(k) Submission, e500 Wave Ventilator #### J. Summary of Non-clinical Tests The Newport e500 Wave Ventilator meets all applicable device specification requirements for performance testing as identified in the FDA Reviewer Guidance for Ventilators as well as the Additionally, the e500 has been tested to and conforms to the following domestic and international standards. IEC601-1 CISPR 11 EN60601-1-2 EN 794-1 ISO9703-1&2 ASTM F 1100-90 MIL-STD461D MIL-STD810E #### K. Conclusion The Newport e500 Wave Ventilator as a self-contained, electrically powered, microprocessor controlled ventilator, has functionally proven to be safe. The device meets its stated performance specifications and the Reviewer Guidance for Premarket Computer Controlled Medical Devices Undergoing 510(k) Review. NMI concludes that the device will operate safely for its intended environments and will be effective in fulfilling its intended use. Given the intended use and patient criticality level, the e500 Wave Ventilator features are as safe and effective as the respective features provided on the predicate devices CONFIDENTIAL: Newport Medical Instruments Inc., considers the existence of this information to be confidential as defined in 21 CFR 814.9, and the data and information to be privileged or confidential as defined in 21 CFR Part 20.61. Therefore, Newport Medical Instruments requests that the confidentiality of this submission be afforded the protection provided by 21 CFR Part 814.9k, 21 CFR Part 20, 5 USC 551, 18 USC 1905, 21, 21 USC 331(i) and all other applicable regulations and laws. {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. Public Health Service JAN 2 3 2004 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Richard Waters Vice President, Regulatory Affairs/QA Newport Medical Instruments 760 West 16th Street, Bldg. N Costa Mesa, CA 92627 Re: K030780 Trade/Device Name: Newport e500 Wave Ventilator Regulation Number: 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK Dated: October 24, 2003 Received: October 28, 2003 Dear Mr. Waters: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Spccial Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {6}------------------------------------------------ Page 2 -- Mr. Richard Waters Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ # Indications for Use 510(k) Number (if known): K030780 Device Name: Newport e500 Wave Ventilator Indications for Use: The e500 Ventilator is intended to provide continuous (ET tube) or noncontinuous (mask) ventilatory support and monitoring for infant, peditric and adult patients with a tidal volume of ≥ 20 mL. The device is prescription use only. The intended environments include hospital, hospitaltype and intra- hospital transport environments. Hospital use typically includes general care floors, operating rooms, special procedure areas, and intensive and critical care areas within the hospital. Hospital-type use includes facilities suc h as or similar to surgicenters, sub-acute centers, and special nursing facilities outside of the hospital. Intra-hospital transport includes patient transport within the hospital or hospital-type facility. × Prescription Use _ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) foffloter/ vision Sinn-Off vision of Anesthesiology, G Infection Control. [ 510(k) Number: K030780
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...