K030780 · Newport Medical Instruments, Inc. · CBK · Jan 23, 2004 · Anesthesiology
Device Facts
Record ID
K030780
Device Name
NEWPORT E500 WAVE VENTILATOR
Applicant
Newport Medical Instruments, Inc.
Product Code
CBK · Anesthesiology
Decision Date
Jan 23, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Intended Use
The e500 Ventilator is intended to provide continuous (ET tube) or non-continuous (mask) ventilatory support and monitoring for infant, pediatric and adult patients with a tidal volume of ≥ 20 mL. The device is prescription use only. The intended environments include hospital, hospital-type and intra-hospital transport environments. Hospital use typically includes general care floors, operating rooms, special procedure areas, and intensive and critical care areas within the hospital. Hospital-type use includes facilities such as or similar to surgicenters, sub-acute centers, and special nursing facilities outside of the hospital. Intra-hospital transport includes patient transport within the hospital or hospital-type facility.
Device Story
The Newport e500 Wave Ventilator is a self-contained, electrically powered, microprocessor-controlled ventilator. It provides continuous or intermittent ventilatory support for infant, pediatric, and adult patients. The device features front-panel controls for selecting operational modes, including volume control, pressure control, and pressure support. It incorporates a comprehensive alarm system for both technical and patient-related parameters. The device is intended for use in hospital settings (ICU, OR, general care) and intra-hospital transport. It includes an internal battery providing approximately 1.5 hours of power. Healthcare providers use the device to manage patient ventilation; the output (ventilatory support and monitoring data) assists in clinical decision-making for respiratory care. The device benefits patients by providing reliable, controlled mechanical ventilation across a wide range of patient sizes.
Clinical Evidence
Bench testing only. The device was tested for performance against FDA Reviewer Guidance for Ventilators and various international standards including IEC601-1, CISPR 11, EN60601-1-2, EN 794-1, ISO9703-1&2, ASTM F 1100-90, MIL-STD461D, and MIL-STD810E. Biocompatibility analysis was performed to ensure safety for the intended environment.
Technological Characteristics
Self-contained, electrically powered, microprocessor-controlled ventilator. Supports infant/pediatric (≥ 20 mL) to adult patients. Features internal battery (1.5-hour capacity). Standards: IEC601-1, CISPR 11, EN60601-1-2, EN 794-1, ISO9703-1&2, ASTM F 1100-90, MIL-STD461D, MIL-STD810E. Connectivity: Standalone. Sterilization: Not specified.
Indications for Use
Indicated for infant, pediatric, and adult patients requiring mechanical ventilatory support (continuous via ET tube or non-continuous via mask) with tidal volumes ≥ 20 mL. Prescription use only.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
Predicate Devices
E200 Ventilator
Siemens Servo Ventilators
Related Devices
K061094 — NEWPORT E500 WAVE VENTILATOR · Newport Medical Instruments, Inc. · Jul 28, 2006
K053502 — NEWPORT E360 VENTILATOR · Newport Medical Instruments, Inc. · Jan 10, 2006
K992133 — NEWPORT HT50 VENTILATOR WITH HUMIDIFIER HT50-H, NEWPORT HT50 VENTILATOR WITHOUT HUMIDIFIER, HT50-H1, NEWPORT HT50 VENTIL · Newport Medical Instruments, Inc. · Aug 4, 2000
K101803 — NEWPORT E360 VENTILATOR · Newport Medical Instruments, Inc. · Sep 8, 2010
K022679 — RAPHAEL · Hamilton Medical AG · Oct 22, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
JAN 2 3 2004
Image /page/0/Picture/1 description: The image shows a handwritten text that appears to be an identifier or code. The text is "K039780" and is written in a slightly tilted orientation. The handwriting is simple and clear, with each character distinctly formed, though the overall appearance is somewhat rough due to the handwritten nature.
## Newport Medical Instruments Inc.
Traditional 510(k) Submission, e500 Wave Ventilator
Image /page/0/Picture/4 description: The image shows the word "NME" in bold, black letters. There are two horizontal lines on either side of the word, which are the same color and thickness as the letters. The letters are sans-serif and appear to be a custom font. The background is white.
Section 4 page 15
# SECTION 4. 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
- Submitter's Name and Address A. Newport Medical Instruments, Inc. 760 West 16th Street - Building N Costa Mesa, California 92627
Mailing Address:
P.O. Box 2600
Newport Beach, California 92658
- B. Submitter's Phone and FAX Number (949) 642-3910 Telephone
(949) 645-5026 FAX
- C. Name of Contact Person Richard Waters
Vice President, Quality Assurance & Regulatory Affairs
- D. Date the Summary was Prepared March 8, 2003
CONFIDENTIAL: Newport Medical Instruments Inc., considers the existence of this information to be confidential as defined in 21 CFR 814.9, and the data and information to be privileged or confidential as defined in 21 CFR Part 20.61. Therefore, Newport Medical Instruments requests that the confidentiality of this submission be afforded the protection provided by 21 CFR Part 814.9k, 21 CFR Part 20, 5 USC 551, 18 USC 1905, 21, 21 USC 331(j) and all other applicable regulations and laws.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the word "NME" in bold, black letters. There are two horizontal lines, one on top and one on the bottom, that flank the word. The lines are the same color and thickness as the letters. The background is white.
Traditional 510(k) Submission, e500 Wave Ventilator
#### E. Name of the Device
- 1. Trade or Proprietary Name
Newport e500 Wave Ventilator
- 2. Common or Usual Name
e500 Ventilator
- Classification Name 3.
868.5895 Continuous Ventilator 73CBK
- Device Class 4.
Class II
### Summary of Substantial Equivalence F.
The Newport e500 Wave Ventilator, the E200 Ventilator and the Siemens Servo Ventilators are intended to provide continuous or intermittent ventilatory support for the care of the individuals who require mechanical ventilation. The performance and clinical features of each ventilator supports infant/pediatric (= 20 mL) through adult applications.
The materials and design of this device are similar to those of the predicate devices. The technical characteristics of the e500 Ventilator do not introduce new questions regarding safety and effectiveness of critical care ventilators. Additionally, the labeling of the e500 provides similar information as the predicate devices.
The information provided in the 510(k) submission supports the determination of substantial equivalence. Biocompatibility analysis demonstrates that the product is safe for its intended environment.
Software design and development, including verification and validation testing, have been conducted using the FDA's Reviewer Guidance of Medical Device Software
CONFIDENTIAL: Newport Medical Instruments Inc., considers the existence of this information to be confidential as defined in 21 CFR 814.9, and the data and information to be privileged or confidential as defined in 21 CFR Part 20.61. Therefore, Newport Medical Instruments requests that the confidentiality of this submission be afforded the provided by 21 CFR Part 814.9k, 21 CFR Part 20, 5 USC 551, 18 USC 1905, 21, 21 USC 331(j) and all other applicable regulations and laws.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the word "NMI" in bold, black letters. There are two thick, horizontal lines above and below the word. The lines are the same color and thickness as the letters. The letters are sans-serif and appear to be a custom font.
## Traditional 510(k) Submission, e500 Wave Ventilator
Section 4 page 17
Submissions, December 15, 1995 draft as a guidance, and per company internal requirements.
### G. Description of the Device
The Newport e500 Wave Ventilator is a self-contained, electrically powered microprocessor controlled ventilator. Performance characteristics and clinical features support infant/pediatric (=20mL) through adult patients.
Front panel controls allow trained operators to select between a number of operational modes, pressure support and volume or pressure control. A comprehensive alarm system is built-in to alert the user to violations of preset safety limits. When fully charged, the internal battery provides an average of 1.5 hours of power.
The alarms associated with the e500 meet or exceed standards of critical care ventilators and have been developed in compliance with ISO9703-1, ISO9703-2 and EN475. The alarms of the e500 span both technical (ventilator related) alarms and non technical alarms (patient related alarms).
### H. Device Intended Use
The e500 Ventilator is intended to provide continuous (ET tube) or non-continuous (mask) ventilatory support and monitoring for infant, pediatric and adult patients with a tidal volume of = 20 mL. The device is prescription use only.
The intended environments include hospital, hospital-type and intra-hospital transport environments. Hospital use typically includes general care floors,
operating rooms, special procedure areas, and intensive and critical care areas within the hospital. Hospital-type use includes facilities such as or similar to surgicenters, sub-acute centers, and special nursing ficilities outside of the hospital. Intra-hospital transport includes patient transport within the hospital or hospital-type facility.
### 1. Summary of Comparison of Technological Characteristics
The e500 is substantially equivalent with the E200 Ventilator and the Siemens Servo Ventilators across the spectrum of the patient population (infant/pediatric to adult) for which each was designed. All of the devices share common modalities (volume control,
CONFIDENTIAL: Newport Medical Instruments Inc., considers the existence of this information to be confidential as defined in 21 CFR 814.9, and the data and information to be privileged or confidential as defined in 21 CFR Part 20.61. Therefore, Newport Medical Instruments requests that the confidentiality of this submission be afforded the protection provided by 21 CFR Part 814.9k, 21 CFR Part 20, 5 USC 551, 18 USC 1905, 21, 21 USC 331(j) and all other applicable regulations and laws,
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "NME" in bold, black letters. The letters are stylized with horizontal lines above and below them. The font appears to be sans-serif and the overall design is simple and impactful.
Traditional 510(k) Submission, e500 Wave Ventilator
Section 4 page 18
pressure control, pressure support) and significantly overlap in the clinical range of function for their target population. Although some differences in design application are noted, the essential clinical function of each device is significantly similar and mimics each other in the typical frame of use by the health care provider. Each provide significant safety features in terms of alarms, back up ventilation, and fail safe mechanisms. Each provides internal battery, ease of use, simple circuit design and matching clinical function.
CONFIDENTIAL: Newport Medical Instruments Inc., considers the existence of this information to be confidential as defined in 21 CFR 814.9, and the data and information to be privileged or confidential as defined in 21 CFR Part 20.61. Therefore, Newport Medical Instruments requests that the confidentiality of this submission be afforded the protection provided by 21 CFR Part 814.9k, 21 CFR Part 20, 5 USC 551, 18 USC 1905, 21, 21 USC 331(j) and all other applicable regulations and laws.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the letters 'NME' in a bold, sans-serif font. The letters are black and appear to be part of a logo or title. There are two horizontal lines on either side of the letters, which are also black and bold.
Section 4 page 19
Traditional 510(k) Submission, e500 Wave Ventilator
#### J. Summary of Non-clinical Tests
The Newport e500 Wave Ventilator meets all applicable device specification requirements for performance testing as identified in the FDA Reviewer Guidance for Ventilators as well as the
Additionally, the e500 has been tested to and conforms to the following domestic and international standards.
IEC601-1 CISPR 11 EN60601-1-2 EN 794-1 ISO9703-1&2 ASTM F 1100-90 MIL-STD461D MIL-STD810E
#### K. Conclusion
The Newport e500 Wave Ventilator as a self-contained, electrically powered, microprocessor controlled ventilator, has functionally proven to be safe. The device meets its stated performance specifications and the Reviewer Guidance for Premarket Computer Controlled Medical Devices Undergoing 510(k) Review. NMI concludes that the device will operate safely for its intended environments and will be effective in fulfilling its intended use. Given the intended use and patient criticality level, the e500 Wave Ventilator features are as safe and effective as the respective features provided on the predicate devices
CONFIDENTIAL: Newport Medical Instruments Inc., considers the existence of this information to be confidential as defined in 21 CFR 814.9, and the data and information to be privileged or confidential as defined in 21 CFR Part 20.61. Therefore, Newport Medical Instruments requests that the confidentiality of this submission be afforded the protection provided by 21 CFR Part 814.9k, 21 CFR Part 20, 5 USC 551, 18 USC 1905, 21, 21 USC 331(i) and all other applicable regulations and laws.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Public Health Service
JAN 2 3 2004
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Richard Waters Vice President, Regulatory Affairs/QA Newport Medical Instruments 760 West 16th Street, Bldg. N Costa Mesa, CA 92627
Re: K030780 Trade/Device Name: Newport e500 Wave Ventilator Regulation Number: 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK Dated: October 24, 2003 Received: October 28, 2003
Dear Mr. Waters:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Spccial Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 -- Mr. Richard Waters
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known): K030780
Device Name: Newport e500 Wave Ventilator
Indications for Use:
The e500 Ventilator is intended to provide continuous (ET tube) or noncontinuous (mask) ventilatory support and monitoring for infant, peditric and adult patients with a tidal volume of ≥ 20 mL. The device is prescription use only.
The intended environments include hospital, hospitaltype and intra- hospital transport environments. Hospital use typically includes general care floors, operating rooms, special procedure areas, and intensive and critical care areas within the hospital. Hospital-type use includes facilities suc h as or similar to surgicenters, sub-acute centers, and special nursing facilities outside of the hospital. Intra-hospital transport includes patient transport within the hospital or hospital-type facility.
× Prescription Use _ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
foffloter/
vision Sinn-Off vision of Anesthesiology, G Infection Control. [
510(k) Number: K030780
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.