NEWPORT AURA VENTILATOR

K121891 · Newport Medical Instruments, Inc. · CBK · Nov 9, 2012 · Anesthesiology

Device Facts

Record IDK121891
Device NameNEWPORT AURA VENTILATOR
ApplicantNewport Medical Instruments, Inc.
Product CodeCBK · Anesthesiology
Decision DateNov 9, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5895
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The AURA family of ventilators is intended to provide continuous or intermittent positive pressure mechanical ventilatory support through invasive or non-invasive interfaces to individuals that require such medical care. Specifically, the AURA family of ventilators is applicable for infant, pediatric and adult patients greater than or equal to 5 kg (11 lbs.). The AURA ventilator is a restricted medical device intended for use by qualified, trained, personnel under the direction of a physician. The AURA ventilator is suitable for use in hospital, sub-acute, emergency room, and home care environments, as well as for transport and emergency response applications.

Device Story

AURA ventilator provides mechanical ventilatory support (A/CMV, SIMV, SPONT, VTPC modes) for infant, pediatric, and adult patients. Device operates via AC/DC power or integrated battery; used in hospital, sub-acute, home, transport, and emergency settings by trained personnel. Front touch screen interface allows configuration in Hospital, Transport, or Basic domains; displays real-time ventilation data, waveforms, and trends. Includes QuickStart setup and weaning evaluation messaging. Optional SpO2 monitoring available. Clinicians use output to manage patient respiratory status; benefits include continuous ventilatory support across diverse clinical environments.

Clinical Evidence

Bench testing only. Verification and validation activities confirmed performance, safety, and electrical requirements. Testing included volume control, pressure support/control, trigger function, PEEP/bias flow, flow patterns, air/oxygen entrainment, waveforms, power performance, alarms, battery/O2 monitoring, and usability testing on less trained users.

Technological Characteristics

Continuous positive pressure ventilator; supports invasive/non-invasive interfaces. Features touch screen interface, integrated battery, and optional SpO2 monitoring. Modes: A/CMV, SIMV, SPONT, VTPC (Pressure/Volume). Power: AC/DC/Battery. Verification testing covered electrical safety, environmental safety, and mechanical performance.

Indications for Use

Indicated for infant, pediatric, and adult patients >= 5 kg (11 lbs.) requiring continuous or intermittent positive pressure mechanical ventilatory support via invasive or non-invasive interfaces.

Regulatory Classification

Identification

A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K121891 9 2012 NOV Image /page/0/Picture/1 description: The image shows the logo for Newport Medical. The word "Newport" is written in a cursive font, with the word "MEDICAL" written in block letters below it. The logo is black and white. Breathing care into all we do. ## 510(k) Summary AURA Ventilator Tom Colonna 10 October 2012 - l. Submitter Information: Newport Medical Instruments, Inc. 1620 Sunflower Avenue Costa Mesa, CA 92626 Contact Person: Senior Director, Regulatory Affairs Summary Date: ### Device Name | Proprietary: | AURA | |-----------------|---------------------------------------------------------------------------| | Common: | Ventilator, Continuous, Facility Use;<br>Ventilator, Continuous, Home Use | | Classification: | II | Product Code: CBK; NOU CFR Section: 868.5895 #### III. Predicate Devices The AURA is substantially equivalent to the following legally marketed predicate devices: - NEWPORT HT70® Ventilator cleared under K090888, K111146 ● - NEWPORT e360 cleared under K053502, K101803 . #### IV. Device Description The AURA Ventilator provides ventilatory support for infant, pediatric and adult patients. The AURA ventilator is a prescription device that can be used in hospital, sub-acute, emergency response, transport and homecare environments under the direction of a physician. The AURA ventilator may be operated from AC or DC external power sources or from the integrated battery system. Any time the ventilator is connected to external power, the integrated battery system is charged, including while the ventilator is in use. The AURA ventilator has A/CMV, SIMV, SPONT and VTPC ventilatory modes (Pressure or Volume). ll. {1}------------------------------------------------ The controls of the AURA ventilator are easily found on the front touch screen and panel. The touch screen can be set up in Hospital, Transport or Basic functional Domains to provide the user with full or limited access. The AURA has a Trends screen for displaying data graphically. User Help screens present information for all controls and features. The AURA family of ventilators has models that can be differentiated based on whether or not the device has SpO2 monitoring capability. #### V. Intended Use The AURA family of ventilators is intended to provide continuous or intermittent positive pressure mechanical ventilatory support through invasive or non-invasive interfaces to individuals that require such medical care. Specifically, the AURA family of ventilators is applicable for infant, pediatric and adult patients greater than or equal to 5 kg (11 lbs.). The AURA ventilator is a restricted medical device intended for use by qualified, trained, personnel under the direction of a physician. The AURA ventilator is suitable for use in hospital, sub-acute, emergency room, and home care environments, as well as for transport and emergency response applications. #### VI. Reason for the Submission The purpose of this traditional 510(k) is to request authorization to market the AURA ventilator. The AURA ventilator includes the following features: - . A QuickStart feature for users to set-up the ventilator - Message for the evaluation of possible weaning . The AURA ventilator has been determined to be substantially equivalent to other legally marketed predicate ventilators. The design and technology of used for the AURA ventilator are common to the HT70 ventilator. #### VII. Nonclinical Data The AURA ventilator has undergone extensive verification, validation and design performance safety testing, all of which confirms that the device meets its design, performance, and electrical safety requirements. The following testing was completed to verify the performance of the AURA ventilator: | Volume Control | |------------------------------------| | Pressure Support | | Pressure Control | | Trigger Function | | PEEP/Bias Flow | | Flow Patterns | | Air/Oxygen Entrainment/Monitoring | | Screens | | Waveforms | | Power Performance | | Intake Valve/Exhalation Resistance | 2 {2}------------------------------------------------ | Respiration Rate, Inspiration time & I/E Ratio | |------------------------------------------------| | Functional Safety | | Alarms | | Endurance | | O2 Cylinder Time Monitoring | | SpO2 Monitoring | | Battery Time Monitoring | | Volume Target Pressure Control (VTPC) | | SIMV | | Export Data | | Maintenance Due | | Usability Test on Less Trained Users | | Emergency/Transport and Environmental Safety | | Shipping Container | #### ·VIII. Conclusions Verification and validation activities were conducted to establish the performance and safety characteristics of the Aura ventilator. All testing demonstrated that the AURA ventilator performs as intended and has acceptable mechanical properties when used in accordance with its labeling; the device is therefore suitable for its intended use. As the device's intended use, operating principles, ventilation modes and performance parameters are comparable to the referenced predicate devices. Therefore, the AURA ventilator is substantially equivalent to the predicate devices. {3}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized abstract symbol resembling an eagle or bird in flight, with three curved lines representing its wings. The symbol is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA", which is arranged in a circular fashion around the symbol. November 9, 2012 Mr. Tom Colonna Senior Director, Regulatory Affairs Newport Medical Instrument, Incorporated 1620 Sunflower Avenue Costa Mesa, California 92626 Re: K121891 Trade/Device Name: AURA Family of Ventilators Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK; NOU Dated: October 10, 2012 Received: October 12, 2012 Dear Mr. Colonna: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002 {4}------------------------------------------------ ### Page 2 - Mr. Colonna Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. Kwame O. Ulmer Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use 510(k) Number: K121891 AURA Family of Ventilators Device Name: ### Indications for Use: The AURA Family of Ventilators is intended to provide continuous or intermittent positive pressure mechanical ventilatory support through invasive or non-invasive interfaces to individuals that require such medical care. Specifically, the AURA Family of Ventilators is applicable for infant, pediatric and adult patients greater than or equal to 5 kg (11 lbs.). The AURA ventilator is a restricted medical device intended for use by qualified, trained, personnel under the direction of a physician. The AURA ventilator is suitable for use in hospital, sub-acute, emergency room, and home care environments, as well as for transport and emergency response applications. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) # (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) \$\qquad\$ (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices Page 1 of 1 510(k) Number: K|2.189|
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...