THE ADVANTAGE SERIES NON-VENTED (NV) FULL FACE MASK
K081670 · Cardinal Health 207, Inc. · CBK · Oct 9, 2008 · Anesthesiology
Device Facts
Record ID
K081670
Device Name
THE ADVANTAGE SERIES NON-VENTED (NV) FULL FACE MASK
Applicant
Cardinal Health 207, Inc.
Product Code
CBK · Anesthesiology
Decision Date
Oct 9, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic
Intended Use
THE ADVANTAGE SERIES® Non-vented (NV) Full Face Mask is intended to provide a patient interface for application of noninvasive ventilation. The mask is to be used as an accessory to ventilators which have adequate alarms and safety systems for ventilator failure, and which are intended to administer CPAP or positive pressure ventilation for treatment of respiratory failure, respiratory insufficiency, or obstructive sleep apnea. The mask is disposable and for single patient use. It is for use on adult patients (> 30 kg), who are appropriate candidates for noninvasive ventilation and use in a hospital/institutional environment.
Device Story
Patient interface for noninvasive ventilation; delivers positive pressure (3-40 cmH2O) from external ventilator to patient nose and mouth. Device consists of latex-free silicone cushion, molded polycarbonate frame, and 360° swiveling elbow with 22mm female connector (ISO 5356-1/EN1281-1 compliant). Held in place by adjustable headgear. Used in hospital/institutional environments by clinicians. Non-vented design requires separate exhalation/safety valve on ventilator circuit to remove exhaled gases and provide room air during pressure loss. Benefits patient by facilitating CPAP or positive pressure ventilation therapy.
Clinical Evidence
No clinical data. Bench testing only, including design specification conformance, interconnection compliance, environmental testing, dead space, pressure vs. flow, leak allowance, and cleaning validation.
Indicated for adult patients (>30 kg) requiring noninvasive ventilation for respiratory failure, respiratory insufficiency, or obstructive sleep apnea in hospital/institutional settings.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
Cardinal Health Advantage Series Full Face mask (K043382)
Related Devices
K060044 — ACUTE CARE FACE MASK, MODEL RT040 · Fisher & Paykel Healthcare, Ltd. · Apr 6, 2006
K140980 — TI NASAL MASK · Respironics, Inc. · Feb 10, 2015
K023135 — IMAGE3 SE DISPOSABLE FULL FACE MASK · Respironics, Inc. · Jan 3, 2003
K232836 — Family of DIMAR DMAX NIV masks · Dimar S.P.A. · Aug 14, 2024
K252173 — F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits Size A (ONIV117A); F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits Size B (ONIV117B); F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits Size C (ONIV117C); F&P OptiNIV ONIV117-F Hospital Vented Full Face Mask with optional Expiratory Filter Compatible with Single-limb Circuits - Size A (ONIV117A-F); F&P OptiNIV ONIV117-F Hospital Vented Full · Fisher & Paykel Healthcare, Ltd. · Sep 10, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
6081670
## 510(k) Summary
510(k) Summary as required by 21 CFR § 807.92
| 510(k) Submitter: | Cardinal Health 207, Inc.<br>Yorba Linda, CA 92887<br>Phone: (714) 283-8472<br>Fax: (714) 283-8472 | | |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|----------------|
| Contact Person | Andre von Muller<br>phone/fax: (714) 292-9464<br>email: andre.vonmuller@cardinalhealth.com | | |
| Establishment<br>Registration<br>Number | 2050001 | | |
| Date prepared | June 4, 2008 | | |
| Name of the device | THE ADVANTAGE SERIES® Non-vented (NV) Full Face Mask | | |
| Common/usual<br>name | Mask for use with ventilator (continuous, facility use) | | |
| Classification | THE ADVANTAGE SERIES® Non-vented (NV) Full Face Mask is classified as<br>a Class II device under the following classification code: | | |
| | Product<br>Code | CFR Section | Panel |
| | CBK | 21 CFR 868.5895 | Anesthesiology |
| Reason for the<br>submission | This is a new device to be marketed by Cardinal Health 207, Inc. | | |
| Substantially<br>equivalent device | THE ADVANTAGE SERIES® Non-vented (NV) Full Face Mask is substantially<br>equivalent to the following device:<br>➤ Image3 SE Disposable Full Face Mask (K023135). | | |
| | THE ADVANTAGE SERIES® Non-vented (NV) Full Face has the following<br>similarities to the predicate device:<br>➤ Same intended use<br>➤ Same technology | | |
| | Labeling for the predicate device to which substantial equivalence has been<br>claimed is included in Section 21 of this 510(k) submission. | | |
| | Device Description. THE ADVANTAGE SERIES® Non-vented (NV) Full Face Mask provides a<br>seal around the nose and mouth such that pressure from a positive pressure<br>source is directed into the patient's nose and mouth. It is held in place with an<br>adjustable headgear. The design consists of latex-free silicone cushion<br>connected to a molded polycarbonate frame with an elbow that can swivel<br>360°. The elbow shall have an interconnection mating with the tubing from the<br>positive pressure device. The interconnection shall conform to ISO 5356-1<br>(EN1281-1) with a 22mm female connector that can also swivel through 360°.<br>This mask can be used to deliver positive pressure therapy to patients<br>requiring pressures up from 3 to 40 cmH2O. This mask can be used with a<br>range of devices that provide a low level of pressure and that have a | | |
K081670 Advantage Series Non-Vented (NV) Full Face Mask – Response to FDA
{1}------------------------------------------------
| Intended<br>Use/Indications for<br>Use | mechanism to adequately remove exhaled gases as well as a safety valve<br>that opens to atmosphere to provide room air in the event of loss of supply<br>pressure. THE ADVANTAGE SERIES® Non-vented (NV) Full Face Mask is<br>functionally similar to Cardinal Health Advantage Series Full Face mask<br>(K043382) except that it does not have the anti-asphyxia or an exhalation<br>port. Due to the fact that this is a non-vented mask, there is a need for a<br>separate mechanism to remove exhaled gases. Reference Section 22 of this<br>submission for a drawing of the mask. |
|-----------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended<br>Use/Indications for<br>Use | THE ADVANTAGE SERIES® Non-vented (NV) Full Face Mask is intended to<br>provide a patient interface for application of noninvasive ventilation. The mask<br>is to be used as an accessory to ventilators which have adequate alarms and<br>safety systems for ventilator failure, and which are intended to administer<br>CPAP or positive pressure ventilation for treatment of respiratory failure,<br>respiratory insufficiency, or obstructive sleep apnea. |
| | The mask is disposable and for single patient use. It is for use on adult<br>patients (> 30 kg), who are appropriate candidates for noninvasive ventilation<br>and use in a hospital/institutional environment. |
| Comparison of<br>technological<br>characteristics<br>between devices. | THE ADVANTAGE SERIES® Non-vented (NV) Full Face Mask is functionally<br>similar to the Cardinal Health Advantage Series Full Face mask (K043382)<br>except that it does not have the anti-asphyxia or an exhalation port. It is<br>constructed from the same materials as the Advantage Series Full Face<br>Mask. |
| | THE ADVANTAGE SERIES® Non-vented (NV) Full Face Mask is substantially<br>equivalent to the Image3 SE Disposable Full Face Mask (K023135). A<br>comparison of the Advantage Full-Face Non-Vented (NV) Mask to the Image3<br>SE Disposable Full Face Mask has been provided in Section 12 of this 510(k)<br>submission. |
| Summary of<br>non-clinical<br>performance<br>testing. | THE ADVANTAGE SERIES® Non-vented (NV) Full Face Mask was tested<br>under various conditions using the Cardinal Health design control process to<br>evaluate design parameters. Testing included:<br>• Conformance to design specifications<br>• Interconnection to patient circuit compliant with international standards<br>• Environmental testing<br>• Dead Space<br>• Pressure vs. Flow<br>• Leak Allowance<br>• Cleaning |
| | The mask passed the specified test criteria. |
| Summary of clinical<br>performance<br>experience. | THE ADVANTAGE SERIES® Non-vented (NV) Full Face Mask was not<br>subjected to human clinical studies to validate the performance of the device. |
| Conclusion | The bench performance data of the ADVANTAGE SERIES® Non-vented (NV)<br>Full Face Mask, when compared to the data and/or claims made on the<br>predicate devices, demonstrate that the device is as safe, as effective, and<br>performs as well as or better than the predicate devices. The intended use of<br>the ADVANTAGE SERIES® Non-vented (NV) Full Face Mask is the same as<br>the predicate device. No new questions of safety or effectiveness are raised. |
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines representing wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the symbol in a circular fashion.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 0 9 2008
Mr. Andre von Muller Senior Regulatory Affairs Engineer Cardinal Health 207, Incorporated 22745 Savi Ranch Parkway Yorba Linda, California 92887
Re: K081670
Trade/Device Name: ADVANTAGE SERIES® Non-vented (NV) Full Face Mask Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK Dated: September 26, 2008 Received: September 29, 2008
Dear Mr. von Muller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. von Muller
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
M. Samuels find mr forell
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known): ....................................................................................................................................................
Device Name: ADVANTAGE SERIES® Non-vented (NV) Full Face Mask
Indications for Use: The ADVANTAGE SERIES® Non-vented (NV) Full Face Mask is intended to provide a patient interface for application of noninvasive ventilation. The mask is to be used as an accessory to ventilators which have adequate alarms and safety systems for ventilation failure and which are intended to administer CPAP or positive pressure ventilation for treatment of respiratory failure, respiratory insufficiency, or obstructive sleep apnea.
> The mask is disposable and for single patient use. It is for use on adult patients (>30 kg) who are appropriate candidates for noninvasive ventilation and use in a hospital/institutional environment.
Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR
Over the Counter Use (21CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE IN ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Lan Aims
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K08/
Page *1* of *1*
K081670 Advantage Series Non-Vented (NV) Full Face Mask -- Response to FDA
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.