F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits – Size A (ONIV117A); F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits – Size B (ONIV117B); F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits – Size C (ONIV117C); F&P OptiNIV ONIV117-F Hospital Vented Full Face Mask with optional Expiratory Filter Compatible with Single-limb Circuits - Size A (ONIV117A-F); F&P OptiNIV ONIV117-F Hospital Vented Full
F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits – Size A (ONIV117A); F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits – Size B (ONIV117B); F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits – Size C (ONIV117C); F&P OptiNIV ONIV117-F Hospital Vented Full Face Mask with optional Expiratory Filter Compatible with Single-limb Circuits - Size A (ONIV117A-F); F&P OptiNIV ONIV117-F Hospital Vented Full
Applicant
Fisher & Paykel Healthcare, Ltd.
Product Code
CBK · Anesthesiology
Decision Date
Sep 10, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Fisher & Paykel Healthcare single-patient use masks are intended for use as an accessory to ventilators to enable non-invasive positive pressure ventilation (NPPV) therapy (CPAP or bi-level) to be delivered to spontaneously breathing adult patients (> 30 kg) with respiratory insufficiency or respiratory failure who have been prescribed NPPV. The masks are to be fitted and therapy maintained, by trained medical practitioners in a hospital/institutional environment with patient monitoring in place.
Device Story
Oro-nasal full face mask accessory for non-invasive positive pressure ventilation (NPPV) systems; delivers CPAP or bi-level therapy to spontaneously breathing adult patients. Device consists of plastic shell, soft seal for facial cushioning, and head straps. Connects to single-limb ventilator circuits via 22mm male ISO 5356-1 conical connectors. Used in hospital/institutional environments by trained medical practitioners. Provides interface for gas delivery; includes anti-asphyxiation valve. Benefits include respiratory support for patients with respiratory insufficiency or failure. Output is mechanical delivery of pressurized air/gas; healthcare providers monitor patient response and therapy parameters via the ventilator system.
Clinical Evidence
Bench testing only. Testing included ISO 5356-1 (connectors), ISO 17510 (masks), ISO 10993-1 (biocompatibility), and ISO 18562 (breathing gas pathway emissions). Performance testing confirmed pressure drop, resistance to flow, mask dead space, anti-asphyxiation valve function, venting leak (CO2 rebreathing), pressure waveform, humidity delivery, and product leak. Usability engineering performed per IEC 62366-1. Shelf-life and packaging integrity (ISTA 2A) also verified.
Technological Characteristics
Oro-nasal full face mask; plastic shell with soft seal; 22mm male ISO 5356-1 conical connector. Single-limb circuit compatible. Operating pressure range 4–30 cmH2O. Non-sterile. Materials evaluated per ISO 10993-1 and ISO 18562-1/2/3. Mechanical anti-asphyxiation valve. Available in three sizes (A, B, C).
Indications for Use
Indicated for spontaneously breathing adult patients (>30 kg) with respiratory insufficiency or respiratory failure requiring non-invasive positive pressure ventilation (CPAP or bi-level) in hospital/institutional settings.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 10, 2025
Fisher & Paykel Healthcare Ltd
Reena Daken
Regulatory Affairs Manager
15 Maurice Paykel Place, East Tamaki
Auckland, 2013
New Zealand
Re: K252173
Trade/Device Name: F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits - Size A (ONIV117A); F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits - Size B (ONIV117B); F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits - Size C (ONIV117C); F&P OptiNIV ONIV117-F Hospital Vented Full Face Mask with optional Expiratory Filter Compatible with Single-limb Circuits - Size A (ONIV117A-F); F&P OptiNIV ONIV117-F Hospital Vented Full Face Mask with optional Expiratory Filter Compatible with Single-limb Circuits - Size B (ONIV117B-F); F&P OptiNIV ONIV117-F Hospital Vented Full Face Mask with optional Expiratory Filter Compatible with Single-limb Circuits - Size C (ONIV117C-F)
Regulation Number: 21 CFR 868.5895
Regulation Name: Continuous ventilator
Regulatory Class: Class II
Product Code: CBK
Dated: July 11, 2025
Received: July 11, 2025
Dear Reena Daken:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252173 - Reena Daken
Page 2
adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K252173 - Reena Daken
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Ethan L. Nyberg -S
Ethan Nyberg, Ph.D.
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits – Size A
Page 8 of 30
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252173 | ? |
| Please provide the device trade name(s). | | ? |
| F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits – Size A (ONIV117A);
F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits – Size B (ONIV117B);
F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits – Size C (ONIV117C);
F&P OptiNIV ONIV117-F Hospital Vented Full Face Mask with optional Expiratory Filter Compatible with Single-limb Circuits - Size A (ONIV117A-F);
F&P OptiNIV ONIV117-F Hospital Vented Full Face Mask with optional Expiratory Filter Compatible with Single-limb Circuits - Size B (ONIV117B-F);
F&P OptiNIV ONIV117-F Hospital Vented Full Face Mask with optional Expiratory Filter Compatible with Single-limb Circuits - Size C (ONIV117C-F) | | |
| Please provide your Indications for Use below. | | ? |
| The Fisher & Paykel Healthcare single-patient use masks are intended for use as an accessory to ventilators to enable non-invasive positive pressure ventilation (NPPV) therapy (CPAP or bi-level) to be delivered to spontaneously breathing adult patients (> 30 kg) with respiratory insufficiency or respiratory failure who have been prescribed NPPV. The masks are to be fitted and therapy maintained, by trained medical practitioners in a hospital/institutional environment with patient monitoring in place. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
{4}
510(k) Summary
As Required by 21 CFR 807.92
# I. SUBMITTER
Company Name and Address
Fisher & Paykel Healthcare Limited
15 Maurice Paykel Place
East Tamaki
Auckland 2013, New Zealand
Telephone: +64 9 574 0100
Prepared and Submitted by
Jung Yun Lee
Regulatory Affairs Specialist
Contact Person
Reena Daken
Regulatory Affairs Manager
Telephone: +64 9 574 0100
Email: reena.daken@fphcare.co.nz
Date Prepared
05 September 2025
# II. DEVICE
Name of Device
F&P OptiNIV Vented Full Face Mask
Common/Usual Name
Full Face Mask
Classification Name
Ventilator, Continuous, Facility Use
Regulatory Class
Class II (21 CFR §868.5895)
Product Code
CBK
# III. PREDICATE DEVICE
Predicate device: F&P Visairo NIV Mask, K203449
Reference device: F&P Nivairo Mask, K191624
Page 1 of 5
{5}
Page 2 of 5
## IV. DEVICE DESCRIPTION
The F&P OptiNIV Vented Full Face Masks are oro-nasal full face masks that are intended for use as an accessory to deliver non-invasive positive pressure ventilation (NPPV) to a patient as part of a non-invasive ventilation system.
The OptiNIV Vented Full Face Masks are prescription-only, provided in a non-sterile state.
The F&P OptiNIV Vented Full Face Mask consists of a plastic shell with a seal to allow cushioning onto the patient's face, and straps to enable the mask to be secured over the head. The mask fits under the patient's nose and covers the nares of the nose and the mouth, with a soft seal to maintain pressure inside the mask whilst minimizing discomfort to the user.
The subject device is available in three sizes – A, B, and C to denote the seal sizes.
## V. INDICATIONS FOR USE
The Fisher & Paykel Healthcare single-patient use masks are intended for use as an accessory to ventilators to enable non-invasive positive pressure ventilation (NPPV) therapy (CPAP or bi-level) to be delivered to spontaneously breathing adult patients (> 30 kg) with respiratory insufficiency or respiratory failure who have been prescribed NPPV. The masks are to be fitted and therapy maintained, by trained medical practitioners in a hospital/institutional environment with patient monitoring in place.
{6}
VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
| Design / Technological Characteristic | Subject Device
F&P OptiSIV Vented Full Face Mask | Predicate Device
F&P Visairo Mask (K203449)
RT077 Model | Comments |
| --- | --- | --- | --- |
| Intended Use / Indications for Use | | | |
| Indications for Use | The Fisher & Paykel Healthcare single patient use masks are intended for use as an accessory to ventilators to enable non-invasive positive pressure ventilation (NPPV) therapy (CPAP or bi-level) to be delivered to spontaneously breathing adult patients (>30 kg) with respiratory insufficiency or respiratory failure who have been prescribed NPPV. The masks are to be fitted and therapy maintained, by trained medical practitioners in a hospital/institutional environment with patient monitoring in place. | The Fisher & Paykel Healthcare single patient use masks are intended for use as an accessory to ventilators to enable non-invasive positive pressure ventilation (NPPV) therapy (CPAP or bi-level) to be delivered to spontaneously breathing adult patients (>30 kg) with respiratory insufficiency or respiratory failure who have been prescribed NPPV. The masks are to be fitted and therapy maintained, by trained medical practitioners in a hospital/institutional environment with patient monitoring in place. | Equivalent |
| Availability | Prescription use
(Part 21 CFR 801 Subpart D) | Prescription use
(Part 21 CFR 801 Subpart D) | Equivalent |
| Patient Population | Adult (>30 kg) | Adult (>30 kg) | Equivalent |
| Operating Environment | Hospital/institutional environments | Hospital/institutional environments | Equivalent |
| Reusability | Single use | Single use | Equivalent |
| Technical Specifications and Features | | | |
| Operating Pressure Range | 4 – 30 cmH₂O | 4 – 40 cmH₂O | Equivalent |
| Compatibility with F&P Systems | F&P 850 (K110019)
F&P 950 (K220703) | F&P 850™ (K110019)
F&P 950™ (K220703) | Equivalent |
| Interface Connections | ISO 5356-1 Conical Connectors
22mm male ISO medical taper | ISO 5356-1 Conical Connectors
22mm male ISO medical taper | Equivalent |
Page 3 of 5
{7}
| Design / Technological Characteristic | Subject Device
F&P OptiNIV Vented Full Face Mask | Predicate Device
F&P Visairo Mask (K203449)
RT077 Model | Comments |
| --- | --- | --- | --- |
| Mask Dead Space | < 200 cm³ | < 200 cm³ | Equivalent |
| Anti-Asphyxiation Valve Operation | • Open to Atmospheric Pressure
0.41 cmH₂O
• Closed to Atmospheric Pressure
0.92 cmH₂O | (F&P Visairo Mask RT075 model)
• Open to Atmospheric Pressure
0.41 cmH₂O
• Closed to Atmospheric Pressure
0.92 cmH₂O | Equivalent |
| Resistance to Flow through mask | • 0.23 cmH₂O @ 50 L/min
• 0.61 cmH₂O @ 100 L/min | • 0.26 cmH₂O @ 50 L/min
• 0.63 cmH₂O @ 100 L/min | Similar |
| Breathing Circuit | Single limb | Single limb | Equivalent |
| Sterility | Device not provided sterile | Device not provided sterile | Equivalent |
| Maximum duration of use | Mask 14 days
Filter 48 hours | 14 days | Similar |
| Shelf life | 3 years | 3 years | Equivalent |
| Sizes | Available in three sizes – A, B, and C | Available in three sizes – A, B, and C | Equivalent |
Page 4 of 5
{8}
Page 5 of 5
# VII. PERFORMANCE DATA
## Summary of non-clinical tests
The F&P OptiNIV Vented Full Face Mask range has been tested to applicable requirements and standards which have all passed:
- ISO 5356-1:2015 Anaesthetic and respiratory equipment — Conical connectors Part 1: Cones and sockets
- ISO 17510:2020 Medical devices — Sleep apnoea breathing therapy — Masks and application accessories
- Pressure drop/resistance to flow testing
- Mask deadspace testing
- Pressure-flow curve testing
- Anti-asphyxiation valve function testing
- ISTA 2A:2011 Procedure 2A: Packaged-products weighing 150 lb (68 kg) or less. Basic Requirements: atmospheric conditioning, compression, fixed displacement or random vibration, and shock vibration
- IEC 62366-1:2015/AMD 1:2020 Medical devices – Part 1: Application of Usability Engineering to Medical Devices – Amendment 1 and FDA Guidance “Applying Human Factors and Usability Engineering to Medical Devices”
- ISO 10993-1:2018 Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process
- ISO 18562-1:2024 Biocompatibility evaluation of breathing gas pathways in healthcare applications Part 1: Evaluation and testing within a risk management process
- ISO 18562-2:2024 Biocompatibility evaluation of breathing gas pathways in healthcare applications – Part 2: Tests for emissions of particular matter
- ISO 18562-3:2024 Biocompatibility evaluation of breathing gas pathways in healthcare applications – Part 3: Tests for emissions of volatile organic substances
Additional performance testing has also been completed to confirm the safety and effectiveness of the F&P OptiNIV Vented Full Face Mask range which have all passed:
- Shelf-life testing – all relevant functional testing was performed after simulated shelf-life of three years
- Venting leak testing to address CO2 rebreathing
- Pressure waveform testing
- Humidity delivery testing
- Product leak testing
# VIII. CONCLUSIONS
The F&P OptiNIV Vented Full Face Mask range is substantially equivalent to the predicate device, F&P Visario NIV Masks (K203449), based on patient population, intended uses, comparison of the general technological characteristics, and performance. In addition, the conclusions drawn from the non-clinical tests demonstrate that the subject device is substantially equivalent to the legally marketed predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.