BCT ANTIMICROBIAL DRESSING & BCT SILVER BANDAGE

K140147 · Bio-Medical Carbon Technology Co., Ltd. · FRO · Apr 22, 2014 · SU

Device Facts

Record IDK140147
Device NameBCT ANTIMICROBIAL DRESSING & BCT SILVER BANDAGE
ApplicantBio-Medical Carbon Technology Co., Ltd.
Product CodeFRO · SU
Decision DateApr 22, 2014
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Intended Use

The BCT Antimicrobial Dressing & BCT Silver Bandage dressings are intended for the management of wounds and to provide an antimicrobial barrier. These dressings are applied topically and are in direct contact with the wound. Both are intended to be used for indications such as: - Partial and full thickness wounds; - Pressure ulcers; - Diabetic ulcers; - Surgical wounds; - Acute wounds (1st and 2nd degree burns).

Device Story

BCT Antimicrobial Dressing and BCT Silver Bandage are sterile, single-use wound dressings; composed of polyethylene terephthalate (PET) non-woven, silver-coated activated carbon fiber cloth, and polyethylene (PE) film; BCT Silver Bandage includes an adhesive layer. Device functions by absorbing wound fluid and exudates containing infectious organisms into the fabric; silver ions (100 µg/cm²) exert local antimicrobial action to eliminate bacteria and pathogens; activated carbon fiber absorbs bacterial toxins and offensive odors. Applied topically by healthcare providers in clinical settings; provides antimicrobial barrier; aids in wound management. Benefits include infection reduction and odor control.

Clinical Evidence

No clinical data provided. Substantial equivalence supported by bench testing and biocompatibility assessment. Antimicrobial efficacy demonstrated via AATCC Test Method 100-2004, showing >4 log reduction in microbial population. Biocompatibility testing performed in accordance with ISO 10993.

Technological Characteristics

Materials: PET non-woven, silver-coated activated carbon fiber, PE film, adhesive (bandage only). Silver content: 100 µg/cm². Principle: Passive absorption of exudate and toxins; silver ion antimicrobial release. Form factor: Various sizes. Sterilization: Sterile, single-use. No software or electronic components.

Indications for Use

Indicated for management of partial and full thickness wounds, pressure ulcers, diabetic ulcers, surgical wounds, and acute wounds (1st and 2nd degree burns) in patients requiring topical wound care and an antimicrobial barrier.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### 510(k) Summary #### For #### BCT Antimicrobial Dressing & BCT Silver Bandage #### 1. Submission Sponsor Bio-medical Carbon Technology Co., Ltd. No. 8-1. Gongyecyu 12th Rd. Siehe Village, Situn Dist. Taichung City 40755, Taiwan Phone: +886-4-2355-4588 Fax: +886-4-2355-4587 Contact: Jui-Hsing Lin, Manager, R&D Center #### 2. Submission Correspondent Emergo Group 816 Congress Avenue, Suite 1400 Austin, TX 78701 Office Phone: (512) 327.9997 Fax: (512) 327.9998 Contact: Dr. Diane Sudduth, Senior Consultant, QA Email: project.management@emergogroup.com #### 3. Date Prepared January 17, 2014 #### 4. Device Identification | Trade/Proprietary Name: | BCT Antimicrobial Dressing also marketed as KoCarbon Ag Antimicrobial Dressing<br>BCT Silver Bandage also marketed as KoCarbon Ag Silver Bandage | |----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Common/Usual Name: | Antimicrobial Dressing | | Classification Name: | Dressing, Wound, Drug | | Classification Regulation: | Unclassified; 21 CFR § 878.4020* | | Product Code: | FRO* | | Classification Panel: | General and Plastic Surgery | | Performance Standards: | No applicable performance standards have been established under<br>Section 514 of the FD&C Act. Biocompatibility tests were done in<br>conformance with relevant requirements of ISO 10993. | *A final classification for "Dressing" has not been implemented; however, a Class II classification has been proposed by the General and Plastic Surgery Devices Panel. At this time however, Classification Code FRO is unclassified. - Page 5 - 1 - {1}------------------------------------------------ #### 5. Legally Marketed Predicate Device(s) K022483 – ACTISORB Silver 220 Antimicrobial Binding Dressing (Johnson & Johnson Medical, Ltd.) #### 6. Device Description BCT Antimicrobial Dressing consists of polyethylene terphthalate (PET) non-woven, silver-coated activated carbon fiber cloth and polyethylene (PE) film. BCT Silver Bandage contains the same ingredients, but includes an adhesive layer. The dressings and bandages contain 100 µg/cm² of silver and are available in several sizes to accommodate different wound sizes. BCT Antimicrobial Dressing and BCT Silver Bandage absorbs wound fluid and exudates containing infectious organisms into the dressing fabric, where the silver exerts its antimicrobial action. The silver ions act locally within the dressing, eliminating the absorbed bacteria and pathogens. The activated carbon fiber cloth in BCT Dressings absorbs bacterial toxins and offensive odors. The dressings are individually packaged in a pouch. All dressings are sterile and are for single use only. #### 7. Indication for Use Statement The BCT Antimicrobial Dressing & BCT Silver Bandage dressings are intended for the management of wounds and to provide an antimicrobial barrier. These dressings are applied topically and are in direct contact with the wound. Both are intended to be used for indications such as: - Partial and full thickness wounds; - · Pressure ulcers; - . Diabetic ulcers; - . Surgical wounds; - Acute wounds (1st and 2nd degree burns) . #### 8. Substantial Equivalence Discussion The BCT Antimicrobial Dressing & BCT Silver Bandage have similar indications for use and technological characteristics as Johnson's ACTISORB Silver 220. Both dressings and bandage are manufactured with non-absorbable polymer materials and an activated carbon fiber layer and are available in a similar range of sizes. The biocompatibility testing showed comparable safety profile of the BCT Antimicrobial Dressing & BCT Silver Bandage and the predicate Bench testing demonstrated that the devices are substantially equivalent for the management of wounds and to provide an antimicrobial barrier. See Table 5A for a comparison of characteristics. | Manufacturer | Bio-medical Carbon Technology Co.,<br>Ltd. | Johnson & Johnson Medical, LTD. | | |--------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------| | Trade Name | BCT Antimicrobial Dressing & BCT<br>Silver Bandage | ACTISORB Silver 220 | | | 510(k) Number | Pending | K022483 | | | Product Code | FRO | FRO | | | Manufacturer | Bio-medical Carbon Technology Co.,<br>Ltd. | Johnson & Johnson Medical, LTD. | | | Trade Name | BCT Antimicrobial Dressing & BCT<br>Silver Bandage | ACTISORB Silver 220 | | | Regulation Number | None | None | | | Regulation Name | Dressing, Wound, Drug | Dressing, Wound, Drug | | | Intended Use/<br>Indications for Use | For the management of partial and full thickness dermal ulcers, pressure<br>ulcers, diabetic ulcers, acute wounds<br>such as 1st and 2nd degree burns and<br>surgical wounds. | ACTISORB Silver 220<br>Antimicrobial Binding. Dressing<br>provides an effective barrier to<br>bacterial penetration and for<br>adsorbing offending odor<br>resulting from wounds; the<br>binding properties of the<br>dressing trap bacteria, bacterial<br>toxins and odor. ACTISORB Silver<br>220 Antimicrobial Binding<br>Dressing may help reduce<br>infection in partial and full<br>thickness wounds, including:<br>• Pressure ulcers<br>• Venous ulcers<br>• Diabetic ulcers<br>• First and second-degree<br>burns<br>• Donor sites<br>• Surgical wounds<br><br>ACTISORB Silver 220 Antimicrobial<br>Binding Dressing is suitable for use<br>under compression bandaging. | | | Antimicrobial Activity<br>Microbes tested- | S. aureus<br>E. coli<br>C. albicans<br>P. aeruginose<br>S. epidermidis<br>S. saprophyticus<br>Multi-drug resistant P. aeruginose<br>E. faecalis<br>A. haemolyticus<br>M. furfur<br>P citrinum<br>S. marcescens<br>A. niger<br>T. rubrum | S. aureus<br>E. coli<br>C. albicans<br>P. aeruginose**<br>E. faecium**<br>E. faecalis** | | | Zone of Inhibition | No well-defined zone of inhibition | No well-defined zone of inhibition | | | Duration | Same | 7 days | | | Biocompatibility | Same | Biocompatibility established* | | | Key Factor | Same | Ag | | | Manufacturer | Bio-medical Carbon Technology, Co., Ltd. | | Johnson & Johnson Medical, LTD. | | Trade Name | BCT Antimicrobial Dressing & BCT Silver Bandage | | ACTISORB Silver 220 | | Silver process | Same | | Impregnation | | Silver (Ag) content | 100 µg/cm² | | 33 µg/cm² | | Silver (Ag) contact wound directly? | Same | | No | | Non-invasive device | Same | | Yes | | Structure and materials | Dressing:<br>A. Rayon and Polyester composite absorbent layer<br>B. Activated carbon fiber cloth with silver<br>C. Anti-adherent polyethylene film<br><br>Bandage:<br>A. Rayon and Polyester composite absorbent layer<br>B. Activated carbon fiber cloth with silver (ACF-Ag)<br>C. Anti-adherent polyethylene film<br>D. Adhesive layer | | Dressing:<br>A. Nylon non-woven<br>B. Silver-containing Plain cloth (Charcoal activated carbon with silver)<br>C. Nylon non-woven | | Sterilization | Same | | Gamma | #### Table 5A - Comparison of Characteristics {2}------------------------------------------------ . : : . . . . . . 100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000 {3}------------------------------------------------ * Stated in 510(k) Summary **Stated in J.R. Furr, et al J Hosp Infection (1994) 27, 201-208 #### 9. Non-Clinical Performance Data The biocompatibility of BCT Antimicrobial Dressing and BCT Silver Bandage has been demonstrated through appropriate in vivo and in vitro tests on the dressings. The products have been assessed in accordance with ISO 10993 and do not introduce any additional safety risk over the predicate device ACTISORB* Silver 220 (K022483). Antimicrobial testing demonstrated greater than a 4 log reduction in microbial population under AATCC Test Method 100-2004. Testing demonstrated comparable results between BCT Antimicrobial Dressing and the ACTISORB 220 predicate device. #### 10. Clinical Performance Data . . There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. These types of devices, including the predicate devices, have been on the market for many years with a proven safety and efficacy for the use of the device. The non-clinical testing detailed in this submission supports the substantial equivalence of the device. #### 11. Statement of Substantial Equivalence Based on the indications for use, technological characteristics and performance test results, the BCT Antimicrobial Dressing & BCT Silver Bandage is substantially equivalent to the referenced predicate device. {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with three wavy lines beneath it, possibly representing water or movement. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 22, 2014 Bio-medical Carbon Technology Company Ltd. % Ms. Diane Sudduth Emergo Group 816 Congress Avenue, Suite 1400 Austin, Texas 78701 Re: K140147 Trade/Device Name: BCT Antimicrobial Dressing & BCT Silver Bandage Regulatory Class: Unclassified Product Code: FRO . Dated: January 17, 2014 Received: January 22, 2014 Dear Ms. Sudduth: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be four of the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that 1 Dr has made a and regulations administered by other Federal agencies. You must of any I outh all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic ford in the quality of the (Sections 531-542 of the Act); 21 CFR 1000-1050. {5}------------------------------------------------ Page 2 - Ms. Diane Sudduth If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, ## David Krause -S for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration #### Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page. #### 510(k) Number (if known) K140147 #### Device Name BCT Antimicrobial Dressing & BCT Silver Bandage #### Indications for Use (Describe) The BCT Antimicrobial Dressing & BCT Silver Bandage Dressings are intended for the management of wounds and to provide an antimicrobial barrier. These dressings are applied topically and are in direct contact with the wound. Both are intended to be used for indications such as: - * Partial and full thickness wounds; - * Pressure ulcers; - * Diabetic ulcers; - * Surgical wounds; * Acute wounds ( 1st and 2nd degree burns ) Type of Use (Select one or both, as applicable) [x] Prescription Use (Part 21 CFR 801 Subpart D) [ ] Over-The-Counter Use (21 CFR 801 Subpart C) ### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) # David Krause
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...