EXSALT T7 WOUND DRESSING

K113564 · Exciton Technologies, Inc. · FRO · Dec 15, 2011 · SU

Device Facts

Record IDK113564
Device NameEXSALT T7 WOUND DRESSING
ApplicantExciton Technologies, Inc.
Product CodeFRO · SU
Decision DateDec 15, 2011
DecisionSESE
Submission TypeSpecial
Device ClassClass U
AttributesTherapeutic

Intended Use

exsalt™ T7 Wound Dressing is indicated for use in partial and full thickness wounds, including decubitus ulcers, venous stasis ulcers, diabetic ulcers, first and second degree burns, grafts and donor sites, or other acute or chronic wounds. The dressing may be used over debrided and grafted wounds. The exsalt™ T7 Wound Dressing provides an antibacterial barrier that inhibits bacterial growth in the dressing for up to 7 days.

Device Story

exsalt™ T7 Wound Dressing is a silver-coated wound dressing designed to provide an antibacterial barrier. The device consists of a non-woven polyester inner layer sandwiched between two outer layers of gray Delnet® HDPE mesh, all treated with a proprietary chemical process to deposit oxidized silver species (0.4 mg/cm²; 4.5% w/w). The dressing is applied topically to wounds by clinicians to inhibit bacterial growth within the dressing for up to 7 days. It functions as a passive barrier to protect the wound environment. The device is intended for use on partial and full-thickness wounds, including ulcers and burns, and can be used over debrided or grafted sites. It benefits patients by reducing bacterial colonization in the dressing, potentially supporting the healing environment.

Clinical Evidence

No clinical data. Performance is supported by bench testing, including silver content analysis, moisture content, pH, absorbency, and in vitro antibacterial/bactericidal effectiveness against Staphylococcus aureus, Enterococcus faecalis, Staphylococcus epidermidis, Pseudomonas aeruginosa, and Acinetobacter baumannii. Biocompatibility and biological reactivity testing confirmed the device is nontoxic, non-irritant, and does not elicit sensitization.

Technological Characteristics

Non-woven polyester inner layer with two outer layers of HDPE mesh. Silver-coated using proprietary chemical process (0.4 mg/cm²; 4.5% w/w). Sterilized via gamma irradiation. Passive barrier dressing.

Indications for Use

Indicated for management of partial and full thickness wounds, including decubitus, venous stasis, and diabetic ulcers, first and second degree burns, grafts, donor sites, and other acute or chronic wounds in patients requiring an antibacterial barrier.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ DEC 1 5 2011 K113564 pag Exciton Technologies Inc. exsalt™ T7 Wound Dressing Special 510(k) # 510(k) Summary for exsalt™ T7 Wound Dressing ### 1. Trade (Proprietary) Name exsalt™ T7 Wound Dressing #### 2. Common Name Wound or Burn Dressing #### 3. Contact Information Melanie Ussyk Contact: Quality Assurance Manager - mussyk@excitontech.com Email: (780) 248-1281 Phone: Fax: (780) 248-5878 Address: Exciton Technologies Inc. Suite 4000-10230 Jasper Avenue Edmonton, Alberta T5J 4P6 Canada #### 4. Device Classification & Panel A final classification for wound/burn dressings has not been implemented; Class II has been proposed by the General & Plastic Surgery Devices Panel. 5. Predicate Device(s) exsalt™ SD7 Wound Dressing (K103067) exsalt™ SD7 Wound Dressing (K083870) #### 6. Device Description Exciton Technologies Inc. has developed the exsalt™ technology, a proprietary chemical process, which deposits oxidized silver species onto a non-woven polyester with gray Delnet® HDPE mesh layers thermally bonded on both sides. Silver in the exsalt™ T7 Wound Dressing inhibits bacterial growth in the dressing. The concentration of the silver and oxidized silver December, 2011 {1}------------------------------------------------ K113564 page 2/3 Exciton Technologies Inc. exsalt™ T7 Wound Dressing #### Special 510(k) species on the dressing is 0.4 mg/cm² (4.5% w/w). The exsalt™ T7 Wound Dressing is known to be effective in vitro against Staphylococcus aureus, Enterococcus faecalis, Staphylococcus epidermidis, Pseudomonas aeruginosa, and Acinetobacter baumannii for up to 7 days. #### 7. Intended Use exsalt™ T7 Wound Dressing is indicated for the management of partial and full thickness wounds, including decubitus ulcers, venous stasis ulcers, first and second degree burns, grafts and donor sites, or other acute or chronic wounds. The dressing may be used over debrided and grafted wounds. ## 8. Summary of Substantial Equivalence The labeled indications and directions for use of the exsalt™ T7 Wound Dressing are equivalent to those of the predicate device, exsalt™ SD7 Wound Dressing (K103067). The design, materials, and manufacturing process are the equivalent to those of the predicate device, exsalt™ SD7 Wound Dressing (K103067), and therefore do not raise any new issues concerning safety or effectiveness. #### a) Summary of Technological Characteristics The exsalt™ T7 Wound Dressing consists of 2 outer layers of HDPE with an inner layer of polyester which are all silver-coated. The skin-contacting materials in both the exsalt™ T7 Wound Dressing and the predicate are the same. The exsalt™ T7 Wound Dressing is sterilized by gamma irradiation. The change in the exsalt™ T7 Wound Dressing substrate configuration does not raise any concerns related to safety or effectiveness as compared to the predicate device (K103067). #### b) Summary of Performance Data The following performance tests were conducted on the exsalt™ SD7 Wound Dressing: - . Silver Content - Moisture Content . - рн . - . Absorbency - Anti-bacterial Effectiveness; against Staphylococcus aureus, Enterococcus faecalis, . Staphylococcus epidermidis, Pseudomonas aeruginosa, and Acinetobacter baumannii - . Bactericidal Effectiveness {2}------------------------------------------------ K113564 page 3/3 Special 510(k) - Biocompatibility . - Biological Reactivity . All performance characteristics of the exsalt™ T7 Wound Dressing are the same as the predicate(s). The exsalt™ T7 Wound Dressing raised no new safety concerns relative to biocompatibility. Testing performed on the exsalt™ SD7 Wound Dressing (K083870) showed that it was nontoxic, non-irritant, and did not elicit a sensitization response. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is often associated with medicine and healthcare. The caduceus is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002 DEC 1 5 2011 Exciton Technologies, Inc. % Ms. Melanie Ussyk 10230 Jasper Avenue, Suite 4000 Edmonton, Canada TSJ 4P6 Re: K113564 Trade/Device Name: Regulatory Class: Unclassified Product Code: FRO Dated: December 02, 2011 Received: December 02, 2011 Dear Ms. Ussyk: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {4}------------------------------------------------ Page 2 - Ms. Melanie Ussyk device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, for Pete Varm C.A. Vi. Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # exsalt™ T7 Wound Dressing Indications for Use 510(k) Number: Ktoso67 K | | 350- Device Name: exsalt™ T7 Wound Dressing Indications for Use: The exsalt™ T7 Wound Dressing is indicated for use in partial and full thickness wounds, including decubitus ulcers, venous stasis ulcers, diabetic ulcers, first and second degree burns, grafts and donor sites, or other acute or chronic wounds. The dressing may be used over debrided and grafted wounds. The exsalt™ T7 Wound Dressing provides an antibacterial barrier that inhibits bacterial growth in the dressing for up to 7 days. Prescription Use X (21 CFR Part 801 Subpart D) And/Or Over the Counter Use (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Daniel Krone for MXM Division of Surgical, Orthopedic, and Restorative Dev 510(k) Number K113564
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...