VASHE WOUND CLEANSING SYSTEM

K060901 · Sterilox Technologies, Inc. · FRO · Jun 29, 2006 · SU

Device Facts

Record IDK060901
Device NameVASHE WOUND CLEANSING SYSTEM
ApplicantSterilox Technologies, Inc.
Product CodeFRO · SU
Decision DateJun 29, 2006
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Intended Use

Vashe™ Wound Cleanser is intended for cleansing, irrigating, moistening, and debriding acute and chronic dermal lesions, such as Stage I-IV pressuring und stasis ulcers, diabetic ulcers, post-surgical wounds, first and second degree burns, abrasions and minor irritations of the skin in addition to moistening and lubricating absorbent wound dressings. The Vashe™ Wound Cleanser is intended for use by qualified health care personnel trained in its use.

Device Story

Vashe™ is a wound cleansing solution; used for cleansing, irrigating, and debriding dermal wounds; also used to moisten and lubricate absorbent wound dressings. Mechanism of action relies on mechanical fluid movement across the wound to remove foreign objects, dirt, and debris. Intended for use by qualified, trained healthcare personnel in clinical settings. Output is the physical cleansing/irrigation of the wound site; assists clinicians in wound management and preparation for healing; benefits patient by facilitating removal of contaminants from acute and chronic lesions.

Clinical Evidence

Bench testing only. Pre-clinical testing demonstrated biocompatibility of the solution.

Technological Characteristics

Wound cleansing solution. Mechanical action via fluid irrigation. No electronic components, software, or energy source. Biocompatible formulation.

Indications for Use

Indicated for cleansing, irrigating, moistening, and debriding acute and chronic dermal lesions (Stage I-IV pressure ulcers, stasis ulcers, diabetic ulcers, post-surgical wounds, 1st/2nd degree burns, abrasions, minor skin irritations) and moistening/lubricating absorbent wound dressings. For use by trained healthcare personnel.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ KO60901 EXHIBIT C a marka mana mana mana mana mana mana mara mara mara mara marka marka mara mara mara mara mara mara mara mara mara mara marka marka marka marka marka marka marka marka marka # 510(k) SUMMARY OF VASHE™ WOUND CLEANSER - | 510(k) Summary | This summary of 510(k) safety and effectiveness information is being<br>submitted in accordance with the requirements of 21 C.F.R §807.92. | |--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter | PuriCore, Inc.<br>320 King of Prussia Road<br>Radnor, PA 19087 | | Contact Person | Howard Mann<br>320 King of Prussia Road<br>Radnor, PA 19087<br>484-321-2703<br>610-341-0503 fax | | Date Prepared | March 24th, 2006 | | Trade Name | Vashe™ | | Common Name | Wound Cleanser | | Classification Name | Solution, saline, (wound dressing) | | Predicate Device | Allklenz™ Cleanser; Healthpoint Medical K965120, March 21st 1997.<br>CarraKlenz Wound Cleanser; Carrington Laboratories, Inc. K022670,<br>Oct. 17th, 2002. Dermacyn™ Wound Cleanser, Oculus Innovative<br>Sciences, K042729, September 30th, 2004 and Dermacyn™ Wound<br>Dressing, Oculus Innovative Sciences, K041161, February 1st, 2005.<br>Biopure™ MTAD™ Root Canal Cleanser, Dentsply International,<br>K053167, December 8, 2005 | | Description | The subject device is a wound cleansing solution that is intended for<br>cleansing, irrigating, and debriding dermal wounds in addition to<br>moistening and lubricating absorbent wound dressings. The mechanical<br>action of fluid moving across the wound provides for the mechanism of<br>action and aids in the removal of foreign objects such as dirt and<br>debris. Vashe is offered in various volumes. | | Indications for Use | Vashe™ Wound Cleanser is intended for cleansing, irrigating,<br>moistening, and debriding acute and chronic dermal lesions, such as<br>Stage 1-IV pressure ulcers, stasis ulcers, diabetic ulcers, post-surgical<br>wounds, first and second degree burns, abrasions and minor irritations<br>of the skin in addition to moistening and lubricating absorbent wound<br>dressings. The Vashe™ Wound Cleanser is intended for use by<br>qualified health care personnel trained in its use. | | Substantial Equivalence | The product is similar in function and intended use to:<br>Allklenz™ and CarraKlenz Wound Cleansers, manufactured by<br>Healthpoint Medical and Carrington Laboratories, Inc. that include<br>among their labeled uses the cleansing and irrigation of dermal<br>wounds and removal of foreign material. Dermacyn™ Wound Cleanser and Wound Dressing manufactured<br>by Oculus Innovative Sciences, that include among it's labeled<br>uses the cleansing, moistening and debriding of dermal wounds in<br>addition to moistening and lubricating absorbent wound dressings. Biopure™ MTAD™ Root Canal Cleanser manufactured by<br>Dentsply International, that include among its labeled uses the<br>cleaning and disinfecting of the root canal system. | | Non-clinical Performance | Pre-clinical testing demonstrated biocompatibility of Vashe Wound<br>Cleanser. | | Conclusion | Vashe™ Wound Cleanser is substantially equivalent to the currently<br>cleared and marketed Allklenz™ Cleanser, CarraKlenz Wound<br>Cleansers, Dermacyn™ Wound Cleanser, Dermacyn™ Wound<br>Dressing and Biopure™ MTAD™ Root Canal Cleanser | .. ... .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ---- 17 --------------------------------------------------------------------------------------------------------------------------------------------------------------------------- - ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ {1}------------------------------------------------ Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three human profiles incorporated into its design. The eagle and profiles are encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAR 1 9 2007 PuriCore Inc. c/o Mr. Howard K. Mann Vice President Quality Assurance and Regulatory Affairs 320 King of Prussia Road, Suite 200 Radnor, Pennsylvania 19087 Re: K060901 Trade/Device Name: Vashe™ Wound Solution Regulatory Class: Unclassified Product Code: FRO Dated: May 25, 2006 Received: May 26, 2006 Dear Mr. Mann: This letter corrects our substantially equivalent letter of June 29, 2006. We have reviewed your Section 510(k) premarket notification of intent to market the device indication we nave reviewed your becally be device is substantially equivalent (for the indications felerenced above and have determined in marketed predicate devices marketed in interstate Tor use stated in the encrosure) to regally manatinent date of the Medical Device American Drug commerce pror to thing 20, 1978, in email and one with the provisions of the Federal Food, Drug, devices that have been roomssined in assee approval of a premarket approval application (PMA). and Cosment Act (Act) that do not require appto the general controls provisions of the Act. The r ou may, merciole, market the det result of the Act include requirements for annual registration, listing of general controls provisions of the rice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class II (PMA), it If your device is classified (secure) into existing major regulations affecting your device can may be subject to such additional controls. Existing major may be subject to such additional controller is the 21, Parts 800 to 898. In addition, FDA may be found in the Gode of I cases and Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please oe advised had IDA s issuation of a backers of a became of the Act that FDA has made a determination that your device complies with other must that FDA has made a decertifications administered by other Federal agencies. You must or any Federal statutes and regulations administer or registration and listing (21) comply with all the Act's requirements, including, but not limited to set CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CFR Part 807), labeling (21 CFR Part 800); good manager (20); and if applicable, the electronic forth in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {2}------------------------------------------------ Page 2 -- Mr. Howard K. Mann This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. **Sincerely yours,** Mark McMullen Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health {3}------------------------------------------------ ## Exhibit B #### Indications for Use Page __ 1___ of __ 1 __ ... . . . . . . . . . . . . . . . . . . . ## 510(k) Number: K060901 Device Name: Vashe™ Wound Cleansing System # Indications For Use: Vashe™ Wound Cleanser is intended for cleansing, irrigating, moistening, and debriding acute and chronic dermal lesions, such as Stage I-IV pressuring und stasis ulcers, diabetic ulcers, post-surgical wounds, first and second degree burns, abrasions and minor irritations of the skin in addition to moistening and lubricating absorbent wound dressings. The Vashe™ Wound Cleanser is intended for use by qualified health care personnel trained in its use. Prescription Use X (Per 21 CFR 801 Subpart D) OR Over-The Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) | Concurrence of CDRH, Office of Device Evaluation (ODE) | | |--------------------------------------------------------|--| | (Division Sign-Off) | | | Division of General, Restorative. | | | and Neurological Devices | | 16510(k) Number
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...