BIODERM SCIENCES WOUND HEALING SPRAY

K042084 · Bioderm Sciences, Inc. · FRO · Jun 15, 2005 · SU

Device Facts

Record IDK042084
Device NameBIODERM SCIENCES WOUND HEALING SPRAY
ApplicantBioderm Sciences, Inc.
Product CodeFRO · SU
Decision DateJun 15, 2005
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Intended Use

BioDcrm Sciences Wound Spray is intended to cleanse, rinsc and externally manage dermal lesions such as lacerations, post-operative (surgical) wounds, partial and full-thickness wounds, burns and ulcers (diabetic, venous stasis, pressure). It may also be used in conjunction with a dressing that absorbs fluids (i.e. gauze, gel, alginate, foam, hydrocolloid). BioDerm Sciences Wound Spray is intended to clean, rinse and externally manage skin wounds such as minor lacerations, minor cuts, minor burns and abrasions.

Device Story

BioDerm Wound Spray is an acidic zinc-saline wet dressing for external wound management. It provides a moist, acidic environment to aid healing and removes foreign matter, bacteria, and tissue debris. The device is supplied in 100 ml pump-spray bottles for manual application by clinicians or patients. It is used as a standalone cleanser or in conjunction with absorbent dressings (gauze, gel, alginate, foam, hydrocolloid). The spray is non-irritating and suitable for use on skin, mucous membranes, mouth, and eyes. It benefits patients by facilitating wound cleansing and maintaining an environment conducive to healing.

Clinical Evidence

Bench testing and animal study provided. A study conducted at the University of Miami (Department of Dermatology & Cutaneous Surgery) using a porcine model demonstrated the effectiveness of the formulation as an aid to the healing of burns and partial-thickness wounds.

Technological Characteristics

Acidic zinc-saline wet dressing; pH <= 7. Contains 0.002% trace elements. Supplied non-sterile in 100 ml plunger-activated pump-spray bottles. No electronic components or software.

Indications for Use

Indicated for cleansing, rinsing, and external management of dermal lesions including lacerations, surgical wounds, partial/full-thickness wounds, burns, and ulcers (diabetic, venous stasis, pressure) in clinical settings; also indicated for minor cuts, burns, and abrasions for OTC use. Contraindicated for patients with known allergies to components, third-degree burns, or severely damaged/missing dermis.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 042084 ## JUN 1 5 2005 ### 510(k) summary as described in § 807.92 #### 510(k) Summary | Submitter: | BioDerm Sciences, Inc. | |--------------------------|----------------------------------------------------------------------------| | | Enterprise Center | | | 9 Industrial Park Drive, Suite 1N | | | Oxford, Mississippi 38655 | | | Phone: (301) 216-3912 | | Contact Person: | Ed Gubish Ph. D. <egubish@biodermsciences.com><br>Chief Scientific Officer | | Preparation Date: | May 26, 2005 | | Proprietary Device Name: | BioDerm Wound Spray | | Classification Name: | Liquid Bandage | | Predicate Devices: | | BioDerm Wound Solution Dermagran wound cleanser with zinc Restore Wound Cleanser K040683 K945802 K040779 #### Description: BioDerm Wound Spray is an acidic zinc-saline wet dressing for external wound management. It aids the body in the healing process by providing a moist, acidic wound environment. The non-irritating components of BioDerm Wound Spray allow it to be used in the mouth, eyes and on mucous membrancs as well as on the skin. BioDerm Wound Spray is similar in chemical composition to the previously cleared BioDerm Scicnces Wound Solution with the spray also containing 0.002% of some trace elements. A difference between these products is that the predicate device BioDerm Scicnccs Wound Solution is supplied sterile in injection bottles, while the current device is supplied non-sterile in pump-spray bottles. The dressing is applied by spraying onto the affected arca. BioDerm Wound Spray is supplied in 100 ml spray bottles with a plunger activated pump applicator. #### Intended Use (Prescription): BioDerm Sciences Wound Spray is intended to cleanse, rinse and externally manage dermal lesions such as lacerations, post-operative (surgical) wounds, partial > Confidential BioDerm Sciences, Inc. {1}------------------------------------------------ and full-thickness wounds, burns and ulcers (diabetic, venous stasis, pressure). It is and thickness wounds, carrior with a sterile dressing that absorbs fluids (i.e. gauze, gel, alginate, foam, hydrocolloid). Intended Use (OTC): BioDerm Sciences Wound Spray is intended to clean, rinse mendou Ose (Oren Blow wounds such as minor lacerations, minor cuts, minor cuts, minor burns and abrasions. BioDerm Wound Spray can also be used as a wound cleanser to remove foreign matter, bacteria and tissue debris. BioDerm Wound Spray is contraindicated for use when patients are known to have had allergic reactions to this dressing or its components. It is not suitable for use on third degree aners, or for any wound for which the dermis has been severcly damaged or is missing. #### Comparison of Technological Characteristics: A number of wet dressings have been cleared for marketing by the FDA. All have in A numor of we we a caline solutions based on sodium chloride or metallic salts with a pH less than or equal to 7 (i.e. acidic). They may or may not contain additives such as nutrients. BioDerm Wound Spray is supplied non-sterile. A broad range of chemical compositions is currently available. The chemical composition of BioDerm Wound Spray is within the range of chemical compositions of these predicate devices and is substantially equivalent in terms of its safety and effectiveness. (Sec Table 1) #### Performance: A study of various formulations with varying strength levels of the metallic salts, performed at the University of Miami (Department of Dermatology & Cutaneous Surgery) using a porcine model demonstrated the effectiveness of BioDerm Wound Spray as an aid to the body's healing of burns, and partial wounds. #### Conclusions: When used as directed, BioDerm Wound Spray is safe and offective as a wound cleanser or wet dressing. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized symbol. The symbol consists of three abstract human profiles facing to the right, stacked on top of each other. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAY - 1 2007 Edward R. Gubish, Ph.D. Chief Scientific Officer BioDerm Sciences Inc. Enterprise Center 9 Industrial Park Drive, Suite 1N Oxford, Mississippi 38665 K042084 Re: Trade Name: BioDerm Sciences Wound Spray Regulatory Class: Unclassified Product Code: FRO Dated: May 17, 2005 Received: May 18, 2005 Dear Dr. Gubish: This letter corrects our substantially equivalent letter of June 15, 2005. We have reviewed your Section 510(k) premarket notification of intent to market the We have reviewed your Section 910(x) prematics is substantially equivalent (for the device relevenced above and nave decembly marketed predicate devices marketed indications for use stated in the enclosure) to the of the Modical Device indicalloris for use stated in the encreases of the enactment and of the Medical Device in in interstate commerce prior to may 20, 1770, and 11 accordance with the provisions of Amendments or to devices that have been reclassified in accordance corrected of a Amendinents of to devices that ha re over (Act) that do not require approval of a the Federal Food, Drug, and Cosmetic Act (Act) that do not requires website the the the Federal Food, Drug, and Cosment Act (100), market the device, subject to the general premarket approval (PMA). You may, therefore, market the Act include prematket approval (1 Mr.). "Totale general controls provisions of the Act include controls provisions of the Act. "The general of devices, good manufacturing practice, the requiremond for anitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III If your device is classified (sec above) into entist regulations affecting your (PMA), it may be subject to additional controls. Existing major regulations affecting your (PMA), if may be subject to additional controllar controlly of itle 21, Parts 800 to 898. In the Federal device can be found in the Code of Federal Regulations, Courter, Concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Edward R. Gubish, Ph.D. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health {4}------------------------------------------------ K042084 #### Indications for Use 510(k) Number: TBD Device Name: BioDcrm Sciences Wound Spray Indications for Use: BioDcrm Sciences Wound Spray is intended to cleanse, rinsc and externally manage dermal lesions such as lacerations, post-operative (surgical) wounds, partial and full-thickness wounds, burns and ulcers (diabetic, venous stasis, pressure). It may also be used in conjunction with a dressing that absorbs fluids (i.e. gauze, gel, alginate, foam, hydrocolloid). X Prescription Use (Part 21 CFR 801 Subpart D) Over-the-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) ![image.png](https://i.imgur.com/9499999.png) (Division Sign-Off) Division of General, Restorative and Neurological Devices 510(k) Number K042084 Confidential BioDerm Sciences, Inc. {5}------------------------------------------------ # Ko42084 #### Indications for Use 510(k) Number: TBD Device Name: BioDerm Sciences Wound Spray Indications for Use: BioDerm Sciences Wound Spray is intended to clean, rinse and externally manage skin wounds such as minor lacerations, minor cuts, minor burns and abrasions. Prescription Use (Part 21 CFR 801 Subpart D) Over-the-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | (Division Sign-Off) | | |--------------------------------|--| | Division of Gene., Restorative | | and Neurological Devices 510(k) Number _特殊 | 上一 Confidential BioDerm Sciences, Inc. 4
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...